K151367 · William A. Cook Australia Pty. , Ltd. · MQL · Jan 27, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K151367
Device Name
Sydney IVF Hyaluronidase
Applicant
William A. Cook Australia Pty. , Ltd.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Jan 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Sydney IVF Hyaluronidase is intended for use to facilitate removal of the cumulus cells surrounding oocytes in assisted reproduction technology (ART) procedures.
Device Story
Sydney IVF Hyaluronidase is an enzymatic preparation containing 80 IU/mL Hyaluronidase of ovine origin; supplemented with bicarbonate buffered media, 5 mg/mL Human Serum Albumin, and 10 ug/mL Gentamicin. Used in clinical ART laboratories by embryologists to enzymatically digest and remove cumulus cells from oocytes prior to fertilization. Provided as a sterile, aseptically filtered solution in glass vials. Ovine source material is certified free from BSE/TSE risks. Product facilitates oocyte preparation, improving workflow efficiency in ART procedures.
Clinical Evidence
Bench testing only. Evidence includes Mouse Embryo Assay (MEA) to screen for embryo toxicity, comparing 2-cell MEA (72hrs, ≥80% blastocyst development) against predicate 1-cell MEA (96hrs, ≥75% blastocyst development). Stability studies validated shelf-life of 20 weeks at 2-8°C, assessing pH, osmolality, endotoxin, sterility, amino acid/pyruvate concentrations, ammonia levels, and enzymatic activity. No clinical data presented.
Technological Characteristics
Enzymatic solution containing 80 IU/mL ovine-derived Hyaluronidase, bicarbonate buffer, 5 mg/mL Human Serum Albumin, and 10 ug/mL Gentamicin. Aseptically filtered. Packaged in borosilicate type 1 glass vials with FluroTec coated rubber stoppers. Performance specs: osmolality 285-295 mOsm/kg; endotoxin < 0.40 EU/mL. Shelf-life: 20 weeks at 2-8°C.
Indications for Use
Indicated for use in assisted reproduction technology (ART) procedures to facilitate the removal of cumulus cells surrounding oocytes.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
November 19, 2019
William A. Cook Australia Pty Ltd Gordana Pozvek Senior Regulatory Affairs Specialist 95 Brandl Street Brisbane Technology Park, Eight Mile Plains Brisbane, OLD 4113, AUSTRALIA
Re: K151367
Trade/Device Name: Sydney IVF Hyaluronidase Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: Class II Product Code: MOL Dated: December 15, 2015 Received: December 18, 2015
Dear Gordana Pozvek:
This letter corrects our substantially equivalent letter of January 27, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name Sydney IVF Hyaluronidase
Indications for Use (Describe)
Sydney IVF Hyaluronidase is intended for use to facilitate removal of the cumulus cells surrounding oocytes in assisted reproduction technology (ART) procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif font, with the registered trademark symbol next to it. Below "COOK", the word "MEDICAL" is written in a smaller, white, sans-serif font.
WILLIAM A. COOK AUSTRALIA PTY. LTD. 95 BRANDL STREET BRISBANE TECHNOLOGY PARK, EIGHT MILE PLAINS BRISBANE, QLD 4113, AUSTRALIA PHONE: 1800.777.222 FAX: +61.7.3841.1288 www.cookmedical.com
## 510(k) Summary
#### SUBMITTED BY:
William A. Cook Australia Pty Ltd 95 Brandl Street Eight Mile Plains OLD 4113 Australia
| Contact Person: | Gordana Pozvek Ph.D. |
|-----------------|--------------------------------|
| Tel: | +61 (7) 3841 1188 |
| Fax: | +61 (7) 3841 3905 |
| E-mail: | Gordana.Pozvek@CookMedical.com |
Date Prepared: January 26, 2016
### DEVICE IDENTIFICATION:
| Trade Name: | Sydney IVF Hyaluronidase (product code: K-SIHY-1-5) |
|-------------------|-----------------------------------------------------|
| Common Name: | Hyaluronidase Solution |
| Regulation No: | 21 CFR 884.6180, Reproductive Media & Supplements |
| Regulatory Class: | II |
| Product Code: | MQL - Media, Reproductive |
#### PREDICATE DEVICE:
K023353 - Sydney IVF Hyaluronidase (product code: K-SIHY-1-5) Date of clearance - December 20, 2002.
#### DEVICE DESCRIPTION:
Sydney IVF Hyaluronidase is an enzymatic preparation containing 80 IU/mL Hyaluronidase (of ovine origin) in a Bicarbonate buffered media supplemented with 5 mg/mL Human Serum Albumin (HSA) and 10 ug/mL Gentamicin.
Sydney IVF Hyaluronidase is provided sterile (aseptically filtered) and is packaged in glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. It is available in a 5 pack carton box with each vial containing 1 mL of Sydney IVF Hvaluronidase.
{4}------------------------------------------------
#### INDICATIONS FOR USE:
Sydney IVF Hyaluronidase is intended for use to facilitate removal of the cumulus cells surrounding oocytes in assisted reproduction technology (ART) procedures.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The Sydney IVF Hyaluronidase and the predicate device (K023353) have the same fundamental technology and similar technological characteristics including the following:
- Similar chemical formulation
- . Similar performance specifications:
- Osmolality 285 295 mOsm/kg —
- -Endotoxin < 0.40 EU/mL
- । A Mouse Embryo Assay (MEA) is used to screen the product for embryo toxicity.
- Same method of sterilisation - Aseptically filtered.
- Same packaging - borosilicate type 1 vials with FluroTec coated stopper and tamper evident seals
The modification that was made to the predicate device was a change in shelf-life from 6 months at -20℃ (for predicate device) to 20 weeks at 2-8℃.
The (ovine derived) Sydney IVF Hyaluronidase is functionally comparable to the (bovine derived) predicate (K023353) and shares similarity in intended use, method of use, formulation, and performance specifications.
Sydney IVF Hyaluronidase, sourced from ovine raw material certified free from the risk of BSE or other TSE forms, has been licensed and sold globally since 2007. During this time there has been no evidence that the product has been the cause of any adverse events in connection with its intended use.
Based on MEA data, and our experience with the (ovine derived) Sydney IVF Hyaluronidase marketed post-2007, this supports the safety and effectiveness of the product for its intended use. Thus the difference in technology between bovine and ovine derived material does not raise concerns regarding safety and efficacy and demonstrates that the (ovine derived) Sydney IVF Hyaluronidase is substantially equivalent to the predicate device (K023353).
#### PERFORMANCE DATA:
The product specifications for the Sydney IVF Hyaluronidase and the predicate device are the same regarding osmolality, sterility and endotoxin.
Bovine spongiform encephalopathy (BSE)
<sup>2</sup> Transmissible Spongiform Encephalopathies (TSE)
{5}------------------------------------------------
#### Mouse Embryo Assay (MEA)
The MEA is used for both the predicate and the proposed device, but the test assay and specification has changed from 1-cell MEA (96hrs) with ≥75% of control that develop to blastocyst (predicate) to 2-cell MEA (72hrs) with ≥80% of control that develop to blastocyst. Both 1-cell and 2-cell Mouse Embryo Assays detect toxicity and test the functionality of IVF Media.
#### Stability & Shelf Life
The shelf-life of Sydney IVF Hyaluronidase has been validated in stability studies to 20 weeks at 2 - 8°C. Stability tests included pH, osmolality, endotoxin, MEA, sterility, the concentrations of proline (amino acid), pyruvate and the HSA by-product ammonia, and the enzymatic activity of Hyaluronidase.
#### CONCLUSION:
The results of the testing provide reasonable assurance that the Sydney IVF Hyaluronidase is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.