FreezeKit Cleave , ThawKit Cleave

K150950 · Vitrolife Sweden AB · MQL · Dec 21, 2015 · Obstetrics/Gynecology

Device Facts

Record IDK150950
Device NameFreezeKit Cleave , ThawKit Cleave
ApplicantVitrolife Sweden AB
Product CodeMQL · Obstetrics/Gynecology
Decision DateDec 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6180
Device ClassClass 2

Indications for Use

FreezeKit™ Cleave is intended for freezing of pronuclear (2PN) and cleavage-stage embryos. ThawKit™ Cleave is intended for thawing of frozen pronuclear (2PN) and cleavage-stage embryos.

Device Story

FreezeKit™ Cleave and ThawKit™ Cleave are reproductive media kits used in IVF laboratories for cryopreservation and thawing of human embryos. FreezeKit™ Cleave (two solutions) facilitates freezing of 2PN and cleavage-stage embryos; ThawKit™ Cleave (three solutions) facilitates thawing. Solutions contain MOPS buffer, gentamicin, human serum albumin, and hyaluronan; cryoprotectants include 1,2-propanediol and sucrose (freezing) or sucrose (thawing). Embryologists perform procedures in clinical laboratory settings. Media protects embryo viability during temperature-controlled storage. Clinical benefit includes successful preservation of embryos for future transfer, supporting IVF treatment cycles.

Clinical Evidence

Clinical evaluation assessed survival and development of human 2PN and cleavage-stage embryos cryopreserved with subject devices versus predicates. Results demonstrated satisfactory survival and development rates for embryos treated with subject devices.

Technological Characteristics

Reproductive media containing MOPS buffer, gentamicin, human serum albumin, and hyaluronan. Cryoprotectants: 1,2-propanediol and sucrose. Sterilization: aseptic filtration. Specifications include pH, osmolality, endotoxin levels, and Mouse Embryo Assay (MEA) performance (1-cell MEA blastocyst expansion and cell count).

Indications for Use

Indicated for freezing and thawing of pronuclear (2PN) and cleavage-stage embryos in patients undergoing IVF procedures.

Regulatory Classification

Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 21, 2015 Vitrolife Sweden AB % Brett Sonet Glazar Vitrolife, Inc. Embryologist & MEA Laboratory Manager 3601 South Inca Street Englewood, CO 80110 Re: K150950 > Trade/Device Name: Freezekit™ Cleave, Thawkit™ Cleave Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: Class II Product Code: MOL Dated: November 9, 2015 Received: November 10, 2015 Dear Brett Sonet Glazar, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a large, bold font. The text is arranged horizontally, with the name "Herbert P. Lerner" appearing prominently. The "-S" is slightly smaller and set off to the right. The background is plain and does not distract from the text. Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150950 Device Name FreezeKit™ Cleave and ThawKit™ Cleave Indications for Use (Describe) FreezeKit™ Cleave is intended for freezing of pronuclear (2PN) and cleavage-stage embryos. ThawKit™ Cleave is intended for thawing of frozen pronuclear (2PN) and cleavage-stage embryos. Type of Use (Select one or both, as applicable) | <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |------------------------------------------------------------------------------------| | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Vitrolife" in blue font, with a blue curved line to the right of the word. The word "Vitrolife" is written in a serif font, and the curved line is thick and smooth. The image is simple and clean, with a white background. # 510(k) Summary | Submitted by: | Vitrolife Sweden AB<br>Box 9080<br>SE-40092 Göteborg<br>SWEDEN | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Brett Sonet Glazar, M.S.<br>Embryologist & MEA Laboratory Manager<br>Vitrolife Inc.<br>3601 South Inca Street<br>Englewood, Co. 80110<br>USA<br>Tel: +1 720 473 5577<br>Fax: +1 303 781 5615<br>E-mail: bglazar@vitrolife.com | | Date Prepared: | November 21, 2014 | | Trade Name: | FreezeKit™ Cleave and ThawKit™ Cleave | | Common Name: | Assisted Reproduction Media for<br>cryopreservation | | Classification Name: | Reproductive Media and Supplements<br>(21 C.F.R. § 884.6180) | | Predicate Devices: | Cook IVF Cryopreservation Kit and Cook IV<br>Thawing Kit (K011157) | ## Devices Description: FreezeKit™ Cleave and ThawKit™ Cleave are devices used for freezing of pronuclear (2PN) and cleavage-stage embryos, and thawing of frozen pronuclear (2PN) and cleavage-stage embryos during in vitro fertilization (IVF) procedures. The proposed devices are two separate kits, containing two and three solutions respectively. The two kits are marketed separately, yet are recommended to be used together to perform freezing and thawing. All solutions in both kits contain MOPS buffered solution, gentamicin as an antibacterial agent, human serum albumin and hyaluronan. The {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "Vitrolife" in blue font. To the right of the word is a blue swoosh design. The word and design are on a white background. cryoprotectants 1,2 - propanediol and sucrose are included in the freezing solution, and only sucrose is included as a cryoprotectant in the thawing solutions. #### Indications for Use: FreezeKit™ Cleave is intended for freezing of pronuclear (2PN) and cleavage stage embryos. ThawKit™ Cleave is intended for thawing of frozen pronuclear (2PN) and cleavage-stage embryos. The subject device indications for use are slightly different from the predicate device, which is indicated more generally for cryopreservation and thawing of zygotes and embryos. However, this difference does not alter the intended use of the subject device from that of the predicate. The intended use of the subject and predicate devices are comparable. ## Technological Characteristics Comparison The formulation of the subject devices are similar to that of the predicates. The compositions of the proposed devices and predicate devices contain antibiotics (gentamicin) and human serum albumin (HSA), as well as the same cryoprotectants (sucrose and 1,2 -propanediol). However, hyaluronan is included only in the proposed devices. In addition, the subject devices have a different buffer base compared to the predicate devices. The addition of hyaluronan in the subject device and the different buffer base do not raise any different types of safety and effectiveness questions. The subject and predicate devices have similar specifications, including pH, osmolality, endotoxin levels and MEA performance. In addition, both the subject and predicate devices are sterilized by aseptic filtration. The subject device includes one additional specification for the MEA analysis is evaluated for the proposed devices. However, this does not represent a difference in technological characterisicts and does not raise any different types of safety and effectiveness questions. ## Non-Clinical Bench Testing The following parameters were measured and evaluated to assure satisfactory operating performance of FreezeKit™ Cleave and ThawKit™ Cleave: {5}------------------------------------------------ - pH at +20°C and ambient atmosphere - Osmolality in mOsm/kg ● - . Sterility - Bacterial Endotoxins (LAL assay) in EU/mL - Mouse embryo assay (1-cell MEA) [% expanded blastocyst within 96h| - Mouse embryo assay (1-cell MEA) [blastocyst cell number within 96h] All bench tests performed met the predefined acceptance criteria. ## Clinical Testing A clinical evaluation was performed to assure satisfactory performance of FreezeKit™ Cleave and ThawKit™ Cleave. The testing assessed survival and development of frozen and thawed human 2PN embryos and cleavage-stage embryos after being cryopreserved with the subject devices as compared to the predicates. Results for the subject device demonstrated satisfactory survival and development of treated 2PN embryos. ## Conclusions The comparison of indications for use / intended purposes, product characteristics, specifications, composition and performance between the predicate devices (Cook IVF Cryopreservation Kit and Cook IVF Thawing Kit) and the proposed devices (FreezeKit™ Cleave and ThawKit™ Cleave) demonstrates that they are substantially equivalent. Although there are differences in the terminology of the indications for use, as well as minor differences in the composition, none of these differences render the subject device not substantially equivalent. Pre-clinical and clinical evaluation studies demonstrate that the subject device performs comparably to the predicate device and support a determination of substantial equivalence. Therefore, FreezeKit™ Cleave and ThawKit™ Cleave are as safe and effective as Cook IVF Cryopreservation Kit and Cook IVF Thawing Kit. FreezeKit™ Cleave and ThawKit™ Cleave are substantially equivalent to Cook IVF Cryopreservation Kit and Cook IVF Thawing Kit.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...