ISOCARE ONE STEP SPERM PROCESSING MEDIUM CAT# 2207-12
K023222 · Invitrocare, Inc. · MQL · Dec 26, 2002 · Obstetrics/Gynecology
Device Facts
Record ID
K023222
Device Name
ISOCARE ONE STEP SPERM PROCESSING MEDIUM CAT# 2207-12
Applicant
Invitrocare, Inc.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Dec 26, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
ISOCARE ONE-STEP Sperm Processing Medium is intended for use in assisted reproductive technology procedures that involve the isolation, processing and transfer of sperm. Specifically, ISOCARE ONE-STEP Sperm Processing Medium is intended for use as a sperm isolation and processing medium in the isolation and processing of sperm during intrauterine and other reproductive technology procedures.
Device Story
ISOCARE ONE-STEP Sperm Processing Medium is a synthetic, defined culture medium; formulated to mimic human fallopian tube fluid; contains salts, physiologically compatible substances, and 5mg/ml human serum albumin. Used in clinical settings by laboratory personnel during sperm isolation and processing; semen sample suspended in medium; centrifuged to separate viable sperm from seminal fluid constituents; spent semen decanted; concentrated sperm layer used for intrauterine insemination or further purified via dilution and additional centrifugation. Buffering system (sodium bicarbonate/HEPES) allows use in procedures without carbon dioxide atmosphere. Benefits include concentration of viable sperm for fertilization procedures.
Clinical Evidence
Bench testing only. Performance validated via cytotoxicity testing, sperm motility/hyperactivation analysis, and mouse embryo assays to ensure functionality and absence of toxic components. Sterility and endotoxin testing performed as release criteria.
Technological Characteristics
Synthetic culture medium; composition: salts, physiologically compatible substances, 5mg/ml human serum albumin. Buffering: sodium bicarbonate/HEPES. Form factor: liquid medium. Sterilization: not specified. No software or electronic components.
Indications for Use
Indicated for use in assisted reproductive technology (ART) procedures, including intrauterine insemination, for the isolation, processing, and transfer of sperm.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335 (K983588)
Submission Summary (Full Text)
{0}------------------------------------------------
Page -1 -
### 510(k) SUMMARY
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
The assigned 510(k) number is: k023222.
Submitted by:
InVitroCare, Inc. 11760-H Sorrento Valley Rd. San Diego, CA 92121 (858) 792-1311 Telephone: Facsimile: (858) 792-1318 info@invitrocare.com E-mail: Robert E. Lovins, PhD Contact: Date Submitted: September 27,2002
#### Device Identification:
| Trade Name: | ISOCARE ONE-STEP Sperm Processing Medium |
|----------------------|--------------------------------------------------------------------|
| Common Name: | Sperm isolation and processing medium,<br>Human Tubal Fluid medium |
| Classification Name: | Reproductive Media (21CFR, 886.6180) |
#### Predicate Device:
Notice of Final Rule, 63 FR 48428, Docket number 97N-0335 and 510(k) Reference Number K983588
### Description:
ISOCARE ONE-STEP Sperm Processing Medium is a synthetic, defined culture medium composed of a mixture of salts and other physiologically compatible substances supplemented with 5mg/ml human serum albumin. ISOCARE ONE-STEP Sperm Processing Medium is intended for use in sperm processing procedures prior to intrauterine or other assisted reproductive technology procedures. It has been formulated to mimic the composition of the fluid found in fallopian tubes as defined by Quinn etal (Quinn P, Kerin JF, Warnes GM: Fertil Steril 1985;44:493-498). ISOCARE ONE-STEP Medium uses a combined sodium bicarbonate/HEPES (J4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid () buffering system and is appropriate for those procedures that do not use a carbon dioxide atmosphere.
{1}------------------------------------------------
#### Intended Use:
ISOCARE ONE-STEP Sperm Processing Medium is intended for use in assisted reproductive technology procedures that involve the isolation, processing and transfer of sperm. Specifically, ISOCARE ONE-STEP Sperm Processing Medium is intended for use as a sperm isolation and processing medium in the isolation and processing of sperm during intrauterine and other reproductive technology procedures.
## Design Characteristics:
ISOCARE ONE-STEP Sperm Processing Medium is intended for use as a culture medium for a variety of assisted reproductive procedures. During sperm isolation and processing procedures, sperm cells are separated from the other constituents of seminal fluid in an effort to concentrate the viable sperm and increase the number of sperm available for fertilization. A culture medium such as ISOCARE ONE-STEP Sperm Processing Medium is used to suspend the semen, the sample is centrifuged, and, after the spent semen layer is decanted, the lower ISOCARE Layer, containing the concentrated sperm sample can be used directly for intrauterine insemination. Alternatively, the sperm containing ISOCARE layer can be diluted with sperm wash medium and centrifuged to obtain a purified sperm pellet which may be used used for the desired fertilization procedure. ISOCARE ONE-STEP Medium is therefore intended for use as a sperm isolation and processing medium in assisted reproductive technology procedures.
## Performance Data:
ISOCARE ONE-STEP Sperm Processing Medium is subjected to cytotoxicity testing and sperm motility/hyperactivation analysis. Each lot of ISOCARE ONE-STEP Medium is also assayed by a mouse embryo assay prior to its release to market. These assays assure that the product is both functional for its intended use, and that no toxic components are present in the formulation. Sperm isolation and processing media have been used in a variety of clinical settings, for the intended use for a number of years. In that time the product has become the standard media used for sperm isolation, washing and transport.
## Additional Information:
Mouse embryo testing will be performed as a condition of release for ISOCARE ONE-STEP Sperm Processing Medium as well as endotoxin and sterility testing. Results of all release assays will be reported on a lot-specific certificate of analysis and will be indicated on the labeling.
{2}------------------------------------------------
#### Conclusion:
The conclusion from performance testing, as well as a review of published historical information contained in the professional literature shows that ISOCARE ONE-STEP Sperm Processing Medium is suitable for its intended use and meets the criteria outlined in the Final Rule, 63 FR48428, Docket number 97N-0335.
### PROPOSED LABELING
Three copies of the proposed labeling for ISOCARE ONE-STEP Sperm Processing Medium are enclosed with is submission, beginning on the following page. These labels include vial and box labels, and proposed product brochure. Some information such as results of endotoxin tests and mouse embryo assays, will be included in the lot-specific certificate of analysis provided with the product. An example of the format for these certificate of analysis, and the information supplied in them may be found in labeling section of this submission.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEPARTMENT OF HEALTH & HUMAN SERVICES
**DEC 26 2002**
Robert E. Lovins, Ph.D. President, InVitroCare, Inc. . 11760-H Sorrento Valley Rd. SAN DIEGO CA 92121
### Re: K023222
Trade/Device Name: ISOCARE ONE-STEP Sperm Processing Medium Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements
Regulatory Class: II Product Code: 85 MQL Dated: December 3, 2002 Received: December 5, 2002
Dear Dr. Lovins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed . predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Henry C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
InVitroCare, Inc.
September 27, 2002 Page - 1
# INDICATIONS FOR USE STATEMENT (Page 1 of 1
510(k) number:
Device Names: ISOCARE ONE-STEP Sperm Processing Medium
Indications for Use:
ISOCARE ONE-STEP Sperm Processing Medium is intended for use in assisted reproductive technology procedures that involve the isolation, processing and transfer of sperm. Specifically, ISOCARE ONE-STEP Sperm Processing Medium is intended for use as a sperm isolation and processing medium in the isolation and processing of sperm during intrauterine and reproductive technology procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K023377
Prescription Use (per 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.