The EmbryoSlide+ ic8 dish is intended for culturing, handling and preparation for transfer of human embryos. The EmbryoSlide+ ic8 dish must be used together with the EmbryoScope+ incubator.
Device Story
EmbryoSlide+ ic8 dish is a single-use, sterile culture dish for human embryos; used in assisted reproduction technology. Device features eight culture wells (20µL each) and four rinsing wells (25-30µL each); requires 1.6mL culture oil overlay. Used in conjunction with EmbryoScope+ incubator. Provides physical containment for embryos during culture; includes barcode area for identification. Clinicians/embryologists use the device to maintain embryo environment; supports embryo development prior to transfer. Benefits include standardized culture environment and traceability.
Clinical Evidence
Bench testing only. Performance validated via Mouse Embryo Assay (MEA) showing ≥80% development to blastocyst at 96 hours. Sterility validated to SAL 10^-6 per ISO 11137. Endotoxin levels ≤20 EU/device per USP <85>. Package integrity and shelf-life validated via accelerated aging (ASTM F1980) and transportation simulation (ASTM D4169).
Technological Characteristics
Polystyrene construction; 8 culture wells, 4 rinsing wells; sterile (SAL 10^-6); non-cytotoxic (ISO 10993-5, USP <87>). Dimensions/form factor: culture dish with lid. No electronic components or software; standalone labware.
Indications for Use
Indicated for culturing, handling, and preparation for transfer of human embryos in assisted reproduction technology procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 11, 2022
Vitrolife A/S Belinda Dueholm Regulatory Affairs Specialist Jens Juuls Vei 20 Viby J, 8260 Denmark
K213869 Re: Trade/Device Name: EmbryoSlide+ ic8 dish Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MQK Dated: December 7, 2021 Received: December 13, 2021
Dear Belinda Dueholm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213869
Device Name EmbryoSlide+ ic8 dish
Indications for Use (Describe)
The EmbryoSlide+ ic8 dish is intended for culturing, handling and preparation for transfer of human embryos. The EmbryoSlide+ ic8 dish must be used together with the EmbryoScope+ incubator.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K213869 EmbryoSlide+ ic8 dish
#### 1. SUBMITTER INFORMATION
| Submitter: | Vitrolife A/S |
|------------|-------------------|
| | Jens Juuls Vej 20 |
| | 8260 Viby J |
| | Denmark |
| Contact Person: | Ms. Belinda Dueholm |
|-----------------|------------------------|
| Telephone: | +45 7221 7900 (main) |
| | +45 2076 3707 (direct) |
#### 2. Date Prepared
March 10, 2021
#### 3. Device Information
| Device Name | EmbryoSlide+ ic8 dish |
|-------------------|--------------------------------------|
| Common Name | Culture dish |
| Regulation Number | 21 CFR 884.6160 |
| Regulation Name | Assisted Reproduction Labware |
| Product Code | MQK (Labware, Assisted Reproduction) |
| Regulatory Class | II |
#### 4. Predicate Device
EmbryoSlide+ culture dish, K173264, (Vitrolife A/S).
The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
#### 5. DEVICE DESCRIPTION
The EmbryoSlide+ ic8 dish is single use, sterile for embryo culture, handling, and preparation for transfer during assisted reproduction technology procedures. The EmbryoSlide+ ic8 dish is equipped with barcode label area for unique identification of the culture dish, and a small area which can be used for manual writing.
The EmbryoSlide+ ic8 dish has eight culture wells. Each well holds a single embryo in 20µL of culture medium.
In addition, four rinsing wells are available, each rinsing well holding 25-30μL medium. The culture area of the dish, comprising both the culture wells, must be covered with 1.6mL culture oil overlay to minimize osmolarity changes during culture in a dry environment.
| Parameter | Specification |
|--------------------------------------|-------------------------------------------------------------------------------|
| Appearance | Good clarity with no apparent flaw |
| Sterility | SAL 10-6 (based on process validation data and package<br>integrity testing). |
| Endotoxin Testing (USP <85>) | ≤ 20 EU/device |
| Mouse Embryo Assay (MEA) | 1-cell MEA: ≥80% embryos developed to blastocyst at 96<br>hours |
| Cytotoxicity (ISO 10993-5, USP <87>) | Non-cytotoxic |
The EmbryoSlide+ ic8 dish has the following device specifications:
#### 6. Indications for Use
The EmbryoSlide+ ic8 dish is intended for culturing, handling and preparation for transfer of human embryos. The EmbryoSlide+ ic8 dish must be used together with the EmbryoScope+ incubator.
### 7. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
{5}------------------------------------------------
| Attribute | Subject Device<br>EmbryoSlide+ ic8 dish<br>K213869 | Predicate Device<br>EmbryoSlide Culture Dish<br>173264 | Differences |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The EmbryoSlide+ ic8 dish is intended<br>for culturing, handling and preparation<br>for transfer of human embryos. The<br>EmbryoSlide+ ic8 dish must be used<br>together with the EmbryoScope+<br>incubator. | The EmbryoSlide+ culture dish is<br>intended for preparing, storing,<br>and transferring human embryos.<br>The EmbryoSlide+ culture dish<br>must be used together with the<br>EmbryoScope+ incubator. | The indications for the<br>subject and predicate<br>devices are similar. Both the<br>predicate and subject<br>devices are indicated for<br>culturing of embryos for<br>transfer. |
| General design | Optically clear culture dish with a lid | Optically clear culture dish with a<br>lid | Same |
| Material | Polystyrene | Polystyrene | Same |
| Number of culture<br>wells for individual<br>embryo incubation | 8 | 16 | Different: Different number<br>of wells are present in the<br>subject and predicate<br>devices. Differences in well<br>numbers do not raise<br>different questions of safety<br>and effectiveness. |
| Traceability/ID | Individually numbered wells | Individually numbered wells | Same |
| Number of rinsing<br>wells | 4 | 4 | Same |
| Barcode label | Area for label (barcode) | Area for label (barcode) | Same |
As stated in the table, the subject and predicate devices have similar indications for use statements and have the same intended use (i.e., culturing embryos for transfer). The subject and predicate devices have differences in their technological characteristics (i.e., different number of culture wells). The differences in technological characteristics do not raise different questions of safety and effectiveness.
#### 8. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
The following non-clinical performance testing has been conducted on the EmbryoSlide+ ic8 dish to support substantial equivalence to the predicate device:
- . Radiation sterilization and validation testing per ISO 11137-1:2006/A2013, ISO 11137-2:2013, ISO 11737-1:2018, and ISO 11737-2:2019.
- Transportation simulation testing per ASTM D4169-16
- Package integrity testing after accelerated aging per ASTM F1908-16:
- Bubble test per ASTM F2096-11 O
- Peel strength testing ASTM F88/F88M-09 O
- . Endotoxin testing per USP <85>. The testing demonstrated that the device met the specification of ≤20 EU/device.
- Mouse embryo assay (MEA) per the 2021 FDA guidance document "Mouse Embryo
{6}------------------------------------------------
Assay for Assisted Reproduction Technology Devices:"
One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group. The testing demonstrated that the device met acceptance criterion of "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours."
- Shelf-life testing (accelerated aging) per ASTM F1980:2016. The following testing was . conducted:
- Endotoxin testing per USP <85> O
- MEA per the 2021 FDA guidance document ""Mouse Embryo Assay for Assisted o Reproduction Technology Devices"
#### 9. Conclusions
The results of the performance testing described above demonstrate that the subject media products are as safe and effective as the predicate devices and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.