The Cryotop®US is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Device Story
Cryotop®US is a single-use, sterile cryopreservation storage device for human embryos. It consists of an ABS handle shaft with a PET film tip for embryo loading and a polypropylene straw enclosure. During vitrification, embryos are loaded onto the film tip; the tip is then inserted into the straw enclosure, creating a hermetic seal that isolates the embryos from liquid nitrogen. The device is used in clinical laboratory settings by embryologists. The straw includes a weighted distal end for proper orientation in storage containers. By maintaining embryo integrity during freezing and thawing, the device facilitates successful cryopreservation, supporting assisted reproduction procedures.
Clinical Evidence
Bench testing only. Testing included cooling/warming rate verification, dimensional analysis, durability (liquid nitrogen immersion), mechanical tensile strength (5N), endotoxin levels (≤0.5 EU/device), sterility (no microbial growth), and mouse embryo assay (≥80% development at 96 hours). Shelf life and package integrity were also validated.
Technological Characteristics
Materials: ABS handle, PET film tip, polypropylene straw. Dimensions: specified outer diameter and length. Energy: passive storage in liquid nitrogen. Sterilization: radiation (SAL 10^-6). Single-use. Closed system design with hermetic seal.
Indications for Use
Indicated for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 14, 2016
KITAZATO BioPharma Co., Ltd. % Richard A. Vincins, CBA, CQA, RAC (US,EU) Vice President, Quality Assurance Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K153027
> Trade/Device Name: Cryotop® US Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MOK Dated: February 9, 2016 Received: February 12, 2016
Dear Richard A. Vincins.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
# Page 2 - Richard A. Vincins, CBA, CQA, RAC (US,EU)
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153027
Device Name Cryotop®US
Indications for Use (Describe)
The Cryotop®US is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Cryotop®US
# K153027
#### 1. Submission Sponsor
KITAZATO BioPharma Co., Ltd.
81 Nakajima, Fuji-city
Shizuoka 416-0907
JAPAN
Phone number: +(81) 545-66-2202
Contact: Futoshi Inoue
Title: President
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
816 Congress Avenue, Suite 1400
Austin, TX 78701
Office Phone: (512) 327 - 9997
Contact: Richard A. Vincins, Vice President, Quality Assurance/Regulatory Affairs
Email: project.management@emergogroup.com
#### 3. Date Prepared
March 9, 2016
#### 4. Device Identification
| Trade/Proprietary Name: | Cryotop®US |
|-------------------------|---------------------------------|
| Common/Usual Name: | Cryopreservation Storage Device |
| Classification Name: | Assisted Reproduction Labware |
| Regulation Number: | 884.6160 |
{5}------------------------------------------------
Product Code:
Device Class: Class II
Classification Panel: Obstetrics/Gynecology
MOK
### 5. Legally Marketed Predicate Device
K122982, CRYOLOCK™, BioTech Inc.
#### 6. Device Description
CryotopUS is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos. As part of the vitrification procedure, the cells to be stored are loaded on the tip of the CryotopUS device for subsequent storage.
The CryotopUS device is composed of an acrylonitrile butadiene styrene (ABS) handle shaft with a polyethylene terephthalate (PET) film tip and a polypropylene straw enclosure. The fine tip has a flat film area for loading embryos. The CryotopUS device has a "straw" enclosure system for when the unit is placed in the liquid nitrogen. The handle shaft is designed with a "stop" for inserting the film tip/shaft handle into the straw enclosure. A hermetic seal is created via a tapered shaft handle with a stop location integrated into the handle is placed into the straw enclosure this creates a closed system keeping the film tip isolated from the liquid nitrogen. The straw enclosure system has a weight at the distal end to place the straw and the shaft inside in a correct position in the liquid nitrogen. The CryotopUS device is provided sterile and is for single use only. The CryotopUS device has been designed to maintain the integrity of the human embryos through the freezing and thawing process.
#### 7. Indication for Use Statement
The Cryotop®US is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human 4-8 cell and blastocyst stage embryos.
#### 8. Substantial Equivalence Discussion
The following Table 1 compares the CryotopUS to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
{6}------------------------------------------------
| Manufacturer | Kitazato BioPharma Co.,<br>Ltd. | BioTech Inc. | Significant<br>Differences |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Cryotop®US | CRYOLOCK™ | |
| 510(k) Number | Not assigned | K122982 | |
| Product Code | MQK | MQK | Same |
| Regulation Number | 884.6160 | 884.6160 | Same |
| Regulation Name | Assisted Reproduction<br>Labware | Assisted Reproduction<br>Labware | Same |
| Indications for Use | The Cryotop®US is a<br>cryopreservation storage<br>device that is intended<br>for use in vitrification<br>procedures to contain<br>and maintain human 4-8<br>cell and blastocyst stage<br>embryos | The CRYOLOCK™ is a<br>cryopreservation storage<br>device that is intended<br>for use in Vitrification<br>procedures to contain<br>and maintain human 1-<br>cell stage embryos | Similar; the<br>indications for use are<br>both for the<br>vitrification and<br>storage of human<br>embryos. |
| Mechanism of Action | Vitrification Method | Vitrification Method | Same |
| Technology<br>Overview | The CryotopUS is<br>designed to contain,<br>freeze and maintain<br>embryos. The device<br>consists of a two piece<br>assembly comprised of<br>the main part containing<br>the fine tip film area and<br>the "straw." The handle<br>shaft and straw are<br>designed to be closed<br>system. The straw is<br>weighted to allow proper<br>alignment in the storage<br>container. The<br>CryotopUS device is<br>packaged in a single<br>barrier sterilization<br>pouch. | The Cryolock is designed<br>to contain, freeze and<br>maintain embryos. The<br>device is a square shape<br>stick with four flat<br>surfaces. The device is<br>composed of two piece<br>assembly with the main<br>part for the embryo<br>placement and the<br>handle shaft and cap<br>designed to be a closed<br>system. The CryoLock<br>device is packaged in a<br>single barrier sterilization<br>pouch. | Same |
| Manufacturer | Kitazato BioPharma Co.,<br>Ltd. | BioTech Inc. | Significant<br>Differences |
| Trade Name | Cryotop®US | CRYOLOCK™ | |
| Material<br>Composition | PET, ABS, Polypropylene | PET, Polypropylene | Similar; both devices are made of medical grade plastics that are commonly used. |
| Sterile | Radiation, SAL 10-6 | Radiation, SAL 10-6 | Same |
| Single-Use | Yes | Yes | Same |
| Shelf Life | 3 years | 3 years | Same |
| Cooling Rate | 3,000°C/min | 1,494°C/min | Different; though the cooling rate is different, non-clinical performance testing from both devices support intended use. |
| Rewarming Rate | 44,000°C/min | 21,000°C/min | Different; though the warming rate is different, non-clinical performance testing from both devices support intended use. |
| Contact with<br>Warming Medium | The tip (film) and the shaft of CryotopUS are taken out from the straw. Directly immerse the tip (film) into thawing solution | The tip (film) and the shaft of Cryolock are taken out from the body. Directly immerse the tip (film) into thawing solution. | Same |
| Performance Testing | Passed | Passed | Same |
| Mouse Embryo<br>Testing | Passed | Passed | Same |
| Sterility Validation<br>Testing | Passed | Passed | Same |
## Table 1– Comparison of Characteristics
{7}------------------------------------------------
The technological characteristics of the CryotopUS are comparable to the predicate device. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{8}------------------------------------------------
#### 9. Non-Clinical Performance Data
As part of demonstrating substantial equivalence to the predicate device, Kitazato BioPharma completed a number of non-clinical performance tests. The CryotopUS device meets all the requirements for overall design, sterilization, and performance testing results, confirming that the design output meets the design inputs and specifications for the device.
The CryotopUS device passed all the testing in accordance with internal requirements and applicable standards that is shown below to support substantial equivalence of the subject device:
- . Cooling Rate Testing: Cooling rate of 3,000 °C/min passed
- . Warming Rate Testing: Warming rate of 44,000 °C/min passed
- Dimensional Testing: Passes outer diameter and length according to specifications
- . Durability Testing: No burst or liguid nitrogen inside the straw after 30 second immersion
- . Mechanical Tensile Testing: Tensile strength to withstand 5N
- . Endotoxin Testing: Endotoxin values conform to the value ≤0.5 EU/device
- . Sterility Testing: No microbial growth from sterility testing
- . Mouse Embryo Assay: ≥80% of 1-cell control embryos develop at 96 hours
- . Shelf life testing
- . Package integrity testing
Note: The performance testing, Mouse Embryo Assay (MEA), and sterility test are all performed on samples from routine manufactured lots; a Certificate of Analysis is provided with each lot of CryotopUS device.
#### 10. Conclusion
The results of the testing described above provide reasonable assurance that the CryotopUS is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.