PICSI is intended for the selection of mature sperm for ICSI.
Device Story
PICSI Sperm Selection Device is a sterile plastic culture dish featuring three microdots of hyaluronan hydrogel and locating lines. Used in clinical laboratory settings by embryologists/technicians during ICSI procedures. Mature sperm bind to hyaluronan microdots via specific membrane receptors; binding stops progressive movement while tails continue beating, allowing identification of mature sperm. Operator uses micropipette to capture bound sperm for injection into oocytes. Device facilitates selection of mature, viable sperm compared to standard methods. Benefits include improved selection of mature sperm for fertilization.
Clinical Evidence
Clinical trial compared PICSI to Sperm Catch using side-by-side fertilization of oocytes via ICSI. Endpoints included fertilization rates, 3-day embryo cell counts, percentage of embryos with ≥8 cells, and morphological characterization of 3-, 5-, and 6-day embryos. Results showed no statistically significant differences in fertilization rates or embryo outcomes between the two groups.
Technological Characteristics
Sterile plastic culture dish with hyaluronan hydrogel microdots. Operates at 18-28°C. Radiation sterilized. Biocompatibility verified via 1-cell Mouse Embryo Assay (≥75% conversion in 96 h). Endotoxin levels <1 EU/device.
Indications for Use
Indicated for selection of mature sperm for intracytoplasmic sperm injection (ICSI) in the treatment of infertile couples.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
## APR 1 3 2006
## 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K052457.
| Trade/proprietary name: | PICSI™ Sperm Selection Device |
|-------------------------|---------------------------------------------------------------|
| Common or usual name: | Assisted Reproduction Media,<br>Assisted Reproduction Labware |
Classification: Assisted Reproduction Media, CFR 884.6180, Class II, Procode 85 MQL; Assisted Reproduction Labware, CFR 884.6160, Class II, Procode 85 MQK
PICSI is substantially equivalent to Sperm Catch™ Assisted Reproduction Media, (510(k) K011607, decision date: 8/10/2001) and to FALCON® IVF Round Dish, (510(k) K991253, decision date 5/5/1999)
The PICSI™ Sperm Selection Device is a sterile, plastic culture dish with Description: three microdots of hyaluronan hydrogel attached to the dish and three locating lines three microuots of the bottom exterior of the dish to aid the operator to find the microdots. emoussed on the boundtiach themselves to hyaluronan through specific receptors on the Mature hema membrane. At temperatures from 18-28° C, mature sperm attach sperin prasmamy to the hyaluronan microdot and no longer exhibit progressive movement, although their tails beat and they are capable of motility. Such sperm are movement, atthough moved from the hyaluronan by micropipet for intracytoplasmic sperm injection (ICSI).
Intended Use: PICSI is intended for the selection of mature sperm for ICSI.
| Factor/attribute | PICSI dish | Sperm Catch | FALCON dish |
|-----------------------------------------------------------------------------|-------------------------------------------------------|----------------------------------|-------------|
| Method of facilitating<br>sperm selection and<br>capture by<br>micropipette | Stops progressive<br>movement | Slows<br>progressive<br>movement | N/A |
| Active Ingredient | Hyaluronan<br>(hydrogel<br>attached to Petri<br>dish) | Hyaluronan<br>(solution) | N/A |
| Sperm fraction subject | Mature motile | Total motile | N/A |
Technological Characteristics Compared to Sperm Catch and to FALCON IVF dish:
{1}------------------------------------------------
| to selection | fraction | fraction | fraction |
|------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------|------------------------------------------------------------------------------|
| Useful Operating<br>Temperature | 18-28° C | 30-37° C | 0-37°C |
| Minimum time for set<br>up and selection of<br>sperm | 15 min | 10 min | N/A |
| Storage conditions | Dry, room<br>temperature | (?) Refrigerated | (?) Dry, room<br>temperature |
| Shelf Life | 1 y (anticipated to<br>be extended as<br>aging studies are<br>completed) | Unknown | Unknown |
| Sterility | Radiation<br>sterilized | Filtration<br>sterilized | Radiation<br>sterilized |
| Endotoxin | <1 EU/device | <1 EU/mL | < 20 EU/device |
| Biocompatibility | Compatible (1-<br>cell Mouse<br>Embryo Assay,<br>≥75% conversion<br>in 96 h) | Compatible<br>(> 24 h survival<br>in human sperm<br>survival assay) | Compatible (2-cell<br>Mouse Embryo<br>Assay, ≥ 70%<br>conversion<br>in 72 h) |
Performance: In a clinical trial involving side-by-side fertilization of oocytes by ICSI, using sperm selected on PICSI and in Sperm Catch, the fertilization rates, quality of the resulting embryos and the disposition of embryos in the PICSI and the Sperm Catch groups were compared. No statistically significant differences were found in fertilization rates, the average number of cells in the 3-day embryos, the number of 3-day embryos with 8 or more cells, the morphological characterization of the 3-, 5- and 6-day embryos and the disposition (outcomes) of the 3- and 6-day embryos.
Signature
James B. Johnston
22 February 2006
Date
K 052457 Premarket Notification [510(k)] Number
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a cross. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
APR 1 8 2006
James B. Johnston, Ph.D. Principal Investigator BIOCOAT Incorporated 211 Witmer Road HORSHAM PA 19044
Re: K052457
Trade/Device Name: PICSI™ Sperm Selection Device Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted reproduction labware Product Code: MQK Regulation Number: 21 CFR §884.6180 Regulation Name: Reproductive media and supplements Product Code: MQL Regulatory Class: II Dated: February 24, 2006 Received: February 27, 2006
## Dear Dr. Johnston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows the FDA Centennial logo. The logo is circular and contains the text "FDA Centennial" in the center. The years "1906 - 2006" are at the top of the circle. The text "Protecting and Promoting Public Health" is below the logo.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K052457
Device Name: PICSI Sperm Selection Device
Indications For Use: In the treatment of infertile couples by ICSI, PICSI is indicated for the selection of mature sperm for injection.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymann
(Division Sign-Off)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devi 510(k) Number.
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.