The micropipet family's intended use is to manipulate zygotes during the ICSI procedure. The micropipets are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to reimplantation. The micropipets are tools used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male fertility is impaired.
Device Story
Micropipets are manual tools fabricated from borosilicate glass capillary tubes; used by embryologists in clinical IVF laboratories. Devices are pulled, ground, and polished to specific dimensions (2-200 um) and bevel angles (15-50 degrees) to facilitate micromanipulation of zygotes and gametes. Specific variants include ICSI, Spermatid ICSI, Holding, Assisted Hatching, SUZI, PZD, and Denuding micropipets. Tools are used under a microscope to physically inject sperm, hold eggs, or assist in hatching. These manual interventions support fertilization and embryo development for infertile couples. Devices are provided sterile for single-use.
Clinical Evidence
Bench testing only. Each lot is validated for toxicity using a two-cell mouse embryo bioassay; embryos are considered non-toxic if expanded/hatching blastocyst development is within 10% of control group. Endotoxin levels are verified via Limulus Amebocyte Lysate assay (must be < 20 EU/device).
Technological Characteristics
Materials: Borosilicate glass. Dimensions: 1 mm diameter capillary tubes pulled to 2-200 um tips. Configuration: Straight or angled (15-45 degrees). Sterilization: Gamma radiation (21.8-40 kGy) per ANSI/AAMI/ISO 11137-1995. Connectivity: None (manual tool).
Indications for Use
Indicated for use by embryologists to manipulate zygotes during ICSI procedures, including sperm injection into eggs and assisted embryo hatching, to aid infertile couples, particularly those with male factor infertility, in achieving pregnancy.
Regulatory Classification
Identification
Assisted reproduction microtools are pipettes or other devices used in the laboratory to denude, micromanipulate, hold, or transfer human gametes or embryos for assisted hatching, intracytoplasmic sperm injection (ICSI), or other assisted reproduction methods.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when the assisted reproduction microtools (pipettes) are manufactured from glass, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 14 1999
# K990847
## 510(k) SUMMARY
Submitter's Name: Humagen Fertility Diagnostics, Inc.
Address: 2400 Hunter's Way Charlottesville, VA 22911
Telephone #: (804) 979-4000
FAX #: (804) 295-5912
Contact person: Cindy Showalter
Date summary prepared: May 28 , 1999
{1}------------------------------------------------
### Device name:
Classification name: Assisted reproduction microtools (per CFR# 884.6130)
Common/Usual name: Micropipets used for in vitro fertilization
Proprietary names: Intracytoplasmic Sperm Injection Micropipets (ICSI) Spermatid ICSI Micropipets Holding Micropipets をした。 Assisted Hatching Micropipets Subzonal Injection Micropipets (SUZI) Partial Zona Dissection Micropipets (PZD) Denuding Micropipets
### Substantial Equivalence:
Substantial equivalence is being supported by the Federal Register Notice Final Rule entitled "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In Vitro Fertilization and Related Assisted Reproduction Procedures. This equivalence is supported by the SUMMARY statement:
Upon the effective date, the Federal Register document may be cited in the absence of an existing predicate device which would be used to support substantial equivalence.
### Description of Device:
Micropipets are fabricated from 1 mm. diameter borosilicate glass capillary tubes. One end of the glass tube is pulled to a much smaller diameter using a glass puller with a heated coil. Diameters range from 2 - 3 um to 200 um depending upon the type of micropipet. Most micropipets are ground on a microgrinder to produce a beveled or flat opening. Generally, a sharp spike is pulled on beveled micropipets by touching the bevel to a heated glass bead and withdrawing quickly. Angles may be added along the shaft of the micropipet per request of the end user to best fit their micromanipulation equipment. Specifications for each variation of micropipet is as follows:
Intracytoplasmic Sperm Injection Micropipets (ICSI) - pulled to a 6 to 7 um outer diameter, 4 to 5 um inner diameter with a 35 or 50 degree bevel and a short, sharp point or may be made with no point. May be straight or with an angle (15 - 45 degrees as specified), 0.5mm from the beveled end. Any of these parameters may be modified to meet customer specifications.
Spermatid ICSI Micropipets - pulled to a 9 um outer diameter, 7 to 8 um inner diameter with a 35 or 50 degree bevel and a short, sharp point at the tip. May be
{2}------------------------------------------------
#### 510K SUMMARY
straight or with an angle (15 to 45 degrees as specified), 0.5 mm from the beveled end. Again, parameters may vary somewhat depending on customer specifications.
Holding Micropipets - one end is pulled to the specified outer diameter, which can range from 65 um to 180 um, scored and cut to yield a blunt opening of the appropriate size, and polished over a hot glass bead to a 20 to 25 um inner diameter opening, unless otherwise specified. Small holding micropipets are pulled to an outer diameter of 65 to 95 um, medium holdings are pulled to an outer diameter of 100 to 120 um, large holdings are pulled to an outer diameter of 125 to 150 um, and extra large holding micropipets are pulled to an outer diameter of 155 to 180 um. Holding micropipets may be angled or straight.
Assisted Hatching Micropipets - one end is pulled to a standard outer diameter of 8 to 10 um unless otherwise specified, with a blunt opening. Other outer diameter sizes are available upon request. Assisted hatching micropipets may be straight or angled (15 to 45 degrees as specified), 0.5 mm from the tapered end.
Subzonal Insertion Micropipets (SUZI) - one end is pulled to a 10 to 12 um outer diameter with a 40 degree bevel, and a short, sharp point at the tip. SUZI micropipets may be straight or angled (15 to 45 degrees as specified), 0.5 mm from the tapered end.
Partial Zona Dissection Micropipets (PZD) - one end is pulled to a long, thin taper with a closed, sharp point. PZD micropipets may be straight or angled (15 to 45 degrees as specified), 0.5 mm from the tapered end.
Denuding Micropipets - one end is pulled and then cut with a diamond knife to an opening of 150 or 190 um inner diameter with a blunt opening.
### Testing Procedures:
Each lot of micropipets is mouse embryo tested for toxicity. We use a two-cell mouse embryo bioassay. A two-cell mouse embryo test was performed as follows: 25 ul tissue culture medium containing serum albumin and penicillin/streptomycin was overlaid with 1 ml mineral oil and equilibrated overnight in 5.5% CO2. On the morning of the embryo harvest, a droplet of medium was exposed to product and embryos were cultured in the media beginning at the two-cell stage. Control droplets of medium were not exposed to product before receiving embryos. The combined percent of embryos developing to expanded and/or hatching blastocysts were assessed at 72 hours of culture. Embryos exposed to the pipets are considered non-toxic if the number of expanded and/or hatching blastocysts in the treated group is within 10% of the control group, and no contamination of the culture has occurred. Greater than 80% hatching/expanded blastyocysts in the control group indicates a valid assay.
{3}------------------------------------------------
Sterilization is performed by gamma radiation. The sterilization validation program follows ANSI/AAMI/ISO 11137-1995 procedures. Quarterly dose audits are conducted by an outside testing laboratory by AAMI Method 1.
The sterility assurance level for the micropipets is 10°.
The minimum established radiation dose is 21.8 kGy established by AAMI Method 1, SIP. A maximum limit of 40 kGy has been established based upon the physical characteristics of the materials.
Each lot of micropipets is also tested for endotoxin levels using the Limulus Amebocyte Lysate assay. The level of endotoxin units per device must be less than 20 to be considered acceptable.
#### Intended use statement:
The micropipet family's intended use is to manipulate zygotes during the ICSI procedure.
The micropipets are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to reimplantation.
The micropipets are tools used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male fertility is impaired.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 14 1999
Ms. Cindy Showalter Quality Assurance Manager Humagen Fertility Diagnostics, Inc. 2400 Hunter's Way Charlottesville, VA 22911
Re: K990847 Intracytoplasmic Sperm Injection Micropinets (ICSI) Dated: May 31, 1999 Received: June 2, 1999 Requiatory Class: Il 21 CFR §884.6130/Procode: 85 MQH
Dear Ms. Showalter:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiiance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K990847 510(k) NUMBER (IF KNOWN):
Intracytoplasmic Sperm Injection Micropipets (ICSI) DEVICE NAME:
INDICATIONS FOR USE:
The micropipet family's indication for use is to manipulate zygotes during the ICSI procedure.
Micropipets are used in the tissue culture techniques performed by embryologists when injecting a single sperm into an egg, or assisting an embryo in hatching prior to re-implantation.
The micropipets are tools used in procedures that have been developed to aid infertile couples achieve pregnancy. Specifically, the ICSI procedure is beneficial in cases where the male fertility is impaired.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter-Use (Optional Format 1-2-96
Ra-C. Paez
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological I 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.