K132870 · Origio, Inc. · MQG · May 7, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K132870
Device Name
Origio Gas Line Filter
Applicant
Origio, Inc.
Product Code
MQG · Obstetrics/Gynecology
Decision Date
May 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6120
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K132870 · May 7, 2014
Origio Gas Line Filter
Origio, Inc.
Filters retrieved from routine clinical service in IVF laboratories
The sponsor tested filters that had been in service in IVF laboratories for 6 months to confirm their continued effectiveness in removing VOCs at the end of their claimed 'Use Life'.
VOC removal efficiency; Real-world clinical service; Filter use life
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
VOC removal efficiency testing of in-service filters; Retrospective/Observational assessment of filters used in clinical practice; Follow-up/Duration: 6 months of clinical use
IVF laboratory gas supply lines
Not applicable for this study
VOC removal efficiency
Indications for Use
The Origio Gas Line Filter is intended for the reduction of volatile organic compounds (VOC's), and particulate contaminants from supply gas lines to incubators and work stations that are used in assisted reproduction technologies and/or In vitro fertilization procedures.
Device Story
The Origio Gas Line Filter is an inline filtration accessory for IVF incubator gas supply lines (N2, CO2, or TriGas). It consists of a polypropylene capsule containing a 0.5um glass fiber HEPA filter and a rolled activated carbon fiber felt media impregnated with potassium permanganate (KMnO4). The device is installed between the gas supply and the incubator/workstation using integrated 1/4" quick-connect fittings. It functions by physically trapping particulates via the HEPA media and chemically adsorbing VOCs via the impregnated activated carbon. The filter is non-sterile, vacuum-packed, and intended for a 6-month use life. It is operated by laboratory personnel (embryologists) in clinical IVF settings. By maintaining gas purity, the device protects cultured embryos from environmental contaminants, potentially improving outcomes in assisted reproduction procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing and Mouse Embryo Assay (MEA). Bench testing confirmed structural integrity (flow rate, pressure drop, burst pressure), HEPA/ULPA filtration efficiency (≥99.999% for 0.1-0.3 micron particles), and VOC removal efficiency (>99% for 17 VOCs) over a 6-month simulated use life. MEA testing verified that the filter does not negatively impact mouse embryo development.
Technological Characteristics
Capsule: Polypropylene. Filtration: 0.5um glass fiber HEPA media and potassium permanganate (KMnO4) impregnated activated carbon fiber felt. Connectivity: 1/4" male/female quick-connect fittings. Form factor: 5.5-inch capsule. Non-sterile. Standalone accessory.
Indications for Use
Indicated for use in IVF and ART laboratories to reduce VOCs and particulate contaminants in supply gas lines (N2, CO2, or TriGas) for incubators and workstations. Intended for use by embryologists.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
# MAY 0 7 2014
## 510(k) Summary: K132870
## 510(k) Summary in accordance with the requirements of 21 CFR 807.92
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c).
| Submitter: | Origio Inc.<br>2400 Hunters Way<br>Charlottesville, VA 22911 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John Clay<br>Director, Regulatory Affairs<br>Origio Inc.<br>77 Elbo Lane<br>Mount Laurel, NJ 08054<br>Phone – (800) 648-1151 or (856) 762-2000 |
Fax - (856) 762-2009
Date Prepared: April, 24 2014
### Device Information
| Trade Name: | Origio Gas Line Filter |
|-----------------------|------------------------------------|
| Common/Usual Name: | Gas Line Filter |
| Classification: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
| Regulation Number: | 21 CFR 884.6120 |
| Regulation Name: | Accessories, Assisted Reproduction |
| Product Code: | MQG |
### Predicate Device
Origio has compared the Origio Gas Line Filter to the CODA Inline Filter, which was included as part of 510(k) K001486 for the <genX> CODA Air Purification System, submitted by <genX> International Company.
Origio Gas Line Filter - Traditional 510(K)
{1}------------------------------------------------
### Device Description
The Origio Gas Line Filter combines a High Efficiency Particulate Air (HEPA) Filter Media and an Activated Carbon Media Filter which is impregnated with Potassium Permanganate (KMNO4). The 1.5", 0.5um Glass Fiber HEPA Media is designed to remove particulates from the supply gas, №2, СО₂ or TriGas (N₂, CO₂ and O₂ ) blend which is commonly used in IVF Incubators and other IVF work stations. The 3.5" Potassium Permanganate – Impregnated Activated Carbon Fiber Felt Media is designed to remove Volatile Organic Compounds (VOC's) from the supply gas provided to the IVF Incubator/Work Station.
The HEPA Filter Media and the Activated Carbon Filter Media are encased in a polypropylene outer shell creating a capsule for the filtration of the medical supply gas. The Origio Gas Line Filter is designed to be installed between the supply gas and the IVF Incubator/Work station utilizing the integrated ¼" male and ¼" female quick- connections. Quick disconnect "intake and output fittings" and clamps are provided along with a magnetic holding bracket for easy installation on to the supply gas line. The Origio Gas Line Filter is provided non-sterile and is individually vacuum packed in a double poly bag.
### Indication for Use
The Origio Gas Line Filter is intended for the reduction of volatile organic compounds (VOC's), and particulate contaminants from supply gas lines to incubators and work stations that are used in assisted reproduction technologies and/or In vitro fertilization procedures.
### Technological Characteristics Comparison
The materials of construction and design features of the Origio Gas Line Filter are very similar to the predicate Coda InLine filter. The capsule size for both filters are also comparable in that the Origio Filter has a 5 ½ inch long capsule and the Coda InLine Filter capsule is approximately 6½ inches long. Each filter is designed with integral ¼ inch male and female quick disconnect features to enable easy installation on to the supply gas line.
Each filter has an activated carbon filter component and a HEPA filter media component contained in a capsule that is placed on the gas line supplying N2 CO2 or TriGas (N2, CO2 and O2) blend. The main difference between the Origio filter and the predicate CODA filter is the addition of potassium permanganate (KMNQ4), which is impregnated into the rolled activated carbon fiber felt component contained in the Origio Gas Line Filter, as an additional measure to reduce VOC's which can be found in the supply gas lines.
Origio Gas Line Filter - Traditional 510(K)
{2}------------------------------------------------
# origin
The activated carbon and the potassium permanganate contained in the Origio Gas Line Filter are not loose or granulated and do not pass through the secondary particle filter. The particle filtration testing conducted has demonstrated that this secondary filter meets ULPA efficiency requirements. Additionally, there were no toxic effects observed with cultured embryos while using the Origio Gas Line Filters on the incubation system utilized for Mouse Embryo Assay testing. Therefore, we believe the use of the potassium permanganate within the Activated Carbon filter component of the Origio Gas Line Filter does not pose additional safety or effectiveness questions as compared to the predicate device.
While there are differences in the construction and materials of the capsule filter with respect to the HEPA/ULPA filter and the activated carbon (Granulated Activated Carbon for the CODA Filter vs. Rolled Carbon fiber felt/with impregnated KMNO4 for the Origio Filter), the intended purpose of these filtering components are the same, which is to remove VOC's, and particulate contaminants from the supply gas. Both the Origio Gas Line Filter and the predicate Coda InLine Filter are used in the same manner and for the same intended purpose as an accessory to the IVF incubator system and/or workstation by filtering VOC's and particulate contaminants from the supply gas lines, which are used in assisted reproduction technologies and/or In vitro fertilization procedures.
| Device Name | Origio Gas Line Filter | (CODA InLine Filter) |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject 510(k) | K001486<br><genx> CODA Air Purification System</genx> |
| Indication for Use | The Origio Gas Line Filter is intended for the<br>reduction of volatile organic compounds (VOC's),<br>and particulate contaminants from supply gas<br>lines to incubators and work stations that are<br>used in assisted reproduction technologies and/or<br>In vitro fertilization procedures. | The CODA Air Purification System will be used as<br>an accessory during assisted reproductive<br>procedures. |
| Target Population | Embryologists in IVF and ART Laboratories | Embryologists in IVF and ART Laboratories |
| | Device Characteristics and Design Features | |
| Gas Line | Includes ¼" male and female quick disconnect | Includes ¼" male and female quick disconnect |
| Connection<br>Features | features on proximal and distal end of capsule. | features on proximal and distal end of capsule. |
| Installation | Filter placement between supply gas and<br>incubator using supplied standard ¼" male and<br>female quick connect fittings and holding bracket. | Filter placement between supply gas and<br>incubator using supplied standard ¼" male and<br>female quick connect fittings and holding bracket. |
| "Use Life" -<br>Frequency of<br>filter change | Up to 6 months | 2 months |
| Shelf Life | 3 years | No Information Available |
| Filtration<br>Efficiency | HEPA filter must capture a minimum of 99.97% of<br>particulate contaminants at 0.3 microns in size.<br><br>Actual Efficiency - The Origio Filter meets<br>requirements for both ULPA (Ultra Low<br>Penetration Air) and HEPA (High Efficiency<br>Particulate Air) with $\geq$ 99.999% efficiency for<br>particle sizes between 0.1-0.2 microns and 0.2-0.3<br>microns throughout the 3 year shelf life.<br><br>Use Life of up to 6 months was demonstrated<br>with filtration efficiencies meeting both ULPA and<br>HEPA requirements for filters challenged with a<br>worse case particle load simulating 6 months of<br>usage in an IVF Lab. | HEPA filter must capture a minimum of 99.97% of<br>particulate contaminants at 0.3 microns in size.<br><br>Actual Efficiency - Unknown |
| VOC Removal<br>Efficiency | Testing demonstrates >99% efficiency up to a 6<br>month "use life". | Actual VOC Efficiency - Unknown |
| Materials of Construction | | |
| Capsule<br>Material/Size | Polypropylene (Shell) ~5 ½ inches | Molded Plastic (Shell) ~6½ inches<br>(Material Unknown) |
| HEPA/ULPA Filter | Microglass Fiber Media | HEPA Filter Media (Unknown) |
| Activated Carbon<br>Filter Media<br>Component | Activated Carbon Fiber Felt Media (rolled),<br>impregnated with potassium permanganate<br>(KMNO₄) | Activated Carbon Media (Loose/Granulated) |
| Sterilization and Packaging | | |
| Sterilization | Filter is provided non-sterile | Filter is not labeled as sterile. |
| Packaging<br>Materials | Vacuum sealed in a double layer Polyethylene bag<br>with Corrugated outer box. | Vacuum sealed in a single layer plastic bag with<br>corrugated outer box. |
| Labeling | Includes flow direction markings molded into<br>capsule as well as included on labelling. | Includes flow direction markings on labelling. |
### Comparison of Key Design Features and Characteristics
Origio Gas Line Filter - Traditional 510(K)
{3}------------------------------------------------
:
:
# ouldlo
:
{4}------------------------------------------------
### Performance Testing and Safety
Performance testing was conducted on the Origio Gas Line Filter demonstrating the overall structural integrity, filtration efficiency and the effectiveness in the reduction/removal of Volatile Organic Compounds. Structural integrity testing included the following: Flow Rate and Pressure Drop, Diffusion and Bubble Point and Burst Pressure.
Particle filtration efficiency testing was conducted using an industry standard test method to demonstrate the filters met HEPA and ULPA requirements. Particle filtration efficiency was conducted with new filters, aged filters and filters that were challenged with a simulated particle load representing 6 months of use life based upon reported "source air" particle levels found in typical IVF Laboratories.
VOC removal efficiency testing was performed using a challenge supply gas concentration containing 17 different VOCs based upon the published literature for the reported VOC levels found in the typical IVF Laboratory. VOC removal efficiency values for the Origio Gas Line Filter was demonstrated up to the 6-month use life for new and aged filters. Additional VOC efficiency testing was conducted confirming that filters which had already been in service in IVF laboratories for 6 months were still effective in removing VOC's at the end of their actual "Use Life".
Mouse Embryo Assay testing of the Origio Gas Line Filter was conducted with the Origio Gas Line Filter installed on the incubator gas supply line to provide verification that the use of Origio Gas Line Filter does not negatively impact or effect mouse embryo development.
### Conclusion
Based on our performance testing and the assessment of the similarities and differences between the Origio Gas Line Filter and the predicate Coda Inline Filter, Origio believes the Origio Gas Line Filter is substantially equivalent with respect to the fundamental scientific technology, materials of construction, design features, filtration efficiency, VOC removal capabilities and intended use as compared to the predicate device "Coda InLine Filter", that was previously cleared under K001486, <genX> CODA Air Purification System.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three wavy lines that resemble a caduceus, a symbol often associated with healthcare.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 7, 2014
Origio, Inc. John Clay Director of Regulatory Affairs 2400 Hunters Way Charlottesville, VA 22911
K132870 Re:
> Trade/Device Name: ORIGIO® Gas Line Filter Regulation Number: 21 CFR§ 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: II Product Code: MQG Dated: March 27, 2014 Received: March 31, 2014
Dear John Clay,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - John Clay
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/7 description: The image shows the name "Herbert P. Lerner - S" in a bold, sans-serif font. The letters are black against a white background, creating a high contrast. The text appears to be a signature or a title, with the "P." possibly indicating a middle initial.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K132870
Device Name ORIGIO® Gas Line Filter
Indications for Use (Describe)
The ORIGIO® Gas Line Filter is intended for the reduction of volatile organic compounds (VOC's) and particulate contaminants from supply gas lines to incubators and work stations that are used in assisted reproduction technologies and/or In vitro fertilization procedures.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY FOR : :::
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Herbert P. Lerner -S
2014.05.07 13:34:25 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Explration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.