Soft-Pass Embryo Mock Transfer Catheter Sets (Soft-Pass Embryo Mock Transfer Catheter Set and Soft-Pass Embryo Mock Transfer Catheter Set with Echogenic Tip)
Soft-Pass Embryo Mock Transfer Catheter Sets (Soft-Pass Embryo Mock Transfer Catheter Set and Soft-Pass Embryo Mock Transfer Catheter Set with Echogenic Tip)
Applicant
Cook Incorporated
Product Code
MQF · Obstetrics/Gynecology
Decision Date
May 17, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.6110
Device Class
Class 2
Indications for Use
The Soft-Pass™ Embryo Mock Transfer Catheter Sets are used for mock (simulated) embryo transfer procedures to aid in verifying catheter access and positioning in advance of an embryo transfer procedure.
Device Story
Device consists of polyethylene guide catheter (6.8 Fr) and polyethylene mock transfer catheter (4.4 Fr) with closed tip; echogenic tip versions include stainless steel band for ultrasound visualization and stainless steel cannula for support. Both catheters feature depth markers for placement guidance. Used by clinicians during simulated embryo transfer procedures to verify uterine access and positioning before actual embryo transfer. Single-use; ethylene oxide sterilized.
Clinical Evidence
No clinical data. Bench testing from predicate K173103 supports the subject device as it is identical in design and materials.
Indicated for use in mock (simulated) embryo transfer procedures to verify catheter access and positioning prior to actual embryo transfer. For prescription use only.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Predicate Devices
Pivet and Soft-Pass Embryo Transfer Catheter Sets (K173103)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2019
Cook Incorporated Ian Herrman Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
Re: K191015
Trade/Device Name: Soft-Pass™ Embryo Mock Transfer Catheter Sets (Soft-Pass™ Embryo Mock Transfer Catheter Set and Soft-Pass™ Embryo Mock Transfer Catheter Set with Echogenic Tip) Regulation Number: 21 CFR 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: Class II Product Code: MQF Dated: April 16, 2019 Received: April 17, 2019
Dear Ian Herrman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191015
#### Device Name
Soft-Pass™ Embryo Mock Transfer Catheter Sets (Soft-Pass™ Embryo Mock Transfer Catheter Set and Soft-Pass™ Embryo Mock Transfer Catheter Set with Echogenic Tip)
#### Indications for Use (Describe)
The Soft-Pass™ Embryo Mock Transfer Catheter Sets are used for mock (simulated) embryo transfer procedures to and in verifying catheter access and positioning in advance of an embryo transfer procedure.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div></td><td><div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div></td></tr></table> | <div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| <div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="16" width="16"> <rect fill="none" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K191015 Soft-Pass™ Embryo Mock Transfer Catheter Sets
| Submitted By: | Cook Incorporated |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Ian Herrman |
| Contact: | 750 Daniels Way |
| Applicant Address: | P.O. Box 489 |
| | Bloomington, IN 47402 |
| Contact Phone Number: | (812) 339-2235 x104034 |
| Contact Fax Number: | (812) 332-0281 |
| Date Prepared: | May 15, 2019 |
| Device Information: | |
| Trade Names: | Soft-Pass™ Embryo Mock Transfer Catheter Sets<br>(Soft-Pass™ Embryo Mock Transfer Catheter Set and<br>Soft-Pass™ Embryo Mock Transfer Catheter Set with<br>Echogenic Tip) |
| Common Name: | Assisted Reproduction Catheter |
| Regulation Number: | 21 CFR §884.6110 |
| Regulation Name: | Assisted Reproduction Catheters |
| Product Code: | MQF - Catheter, Assisted Reproduction |
| Device Class: | II |
### Predicate Device:
K173103 - Pivet and Soft-Pass Embryo Transfer Catheter Sets. Specifically, the Soft-Pass Embryo Mock Transfer Catheter Sets cleared in the predicate submission.
The predicate devices have not been subject to a design-related recall.
### Device Description
The Soft-Pass Embryo Mock Transfer Catheter Sets are used to aid in verifying access and positioning of embryo transfer catheters prior to placing transfer catheters containing embryos into the uterine cavity. These devices consist of a polyethylene guide catheter that is 6.8 Fr in diameter and either 12 or 17.3 cm long, and a polyethylene mock transfer catheter that is 4.4 Fr in diameter and 19.7 or 24.7 cm long. The tip of the mock transfer catheter is closed. The echogenic tip versions of the device also include an echogenic stainless steel band on the distal tip of the mock transfer catheter to aid in tip visualization when using ultrasound. The echogenic tip versions also include a supporting stainless steel cannula in the proximal portion of the mock transfer catheter. Depth markers are also included on both the guide and mock transfer catheters to aid in placement of the device during a mock procedure. These devices are sterilized by
{4}------------------------------------------------
ethylene oxide, single-use only, and have a three-year shelf-life. The product number and trade name for the subject device sets are presented in the table below.
| Model Number | Trade Name |
|---------------------|-----------------------------------------------------------------|
| J-SPPE-681200-MC | Soft-Pass™ Embryo Mock Transfer Catheter Set |
| J-SPPE-681700-MC | Soft-Pass™ Embryo Mock Transfer Catheter Set |
| J-SPPE-681210-ET-MC | Soft-Pass™ Embryo Mock Transfer Catheter Set with Echogenic Tip |
| J-SPPE-681710-ET-MC | Soft-Pass™ Embryo Mock Transfer Catheter Set with Echogenic Tip |
### Indications for Use
The Soft-Pass™ Embryo Mock Transfer Catheter Sets are used for mock (simulated) embryo transfer procedures to aid in verifying catheter access and positioning in advance of an embryo transfer procedure.
### Comparison of Intended Use and Technological Characteristics with the Predicate Device:
| | | Predicate Device | Subject Device |
|--------------------------|--------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Pivet and Soft-Pass Embryo<br>Transfer Catheter Sets | Soft-Pass Embryo Mock<br>Transfer Catheter Sets |
| 510(k) Number | | K173103 (Specifically, the Soft-Pass<br>Embryo Mock Transfer Catheter Sets<br>with a polyethylene guide catheter, J-<br>SPPE-68XXXX-MC[-ET]) | Subject of this submission |
| Indications for Use | | The Embryo Transfer Catheters/Sets are<br>used for transferring IVF embryos into<br>the uterine cavity. | The Soft-PassTM Embryo<br>Mock Transfer Catheter Sets<br>are used for mock (simulated)<br>embryo transfer procedures to<br>aid in verifying catheter<br>access and positioning in<br>advance of an embryo<br>transfer procedure. |
| Material Specification | | | |
| Transfer Catheter | | Polyethylene, stainless steel (echogenic<br>tip versions only), marker ink | Identical |
| Guide Catheter | | Polyethylene, marker ink | Identical |
| Physical Characteristics | | | |
| Hub | | Flared hub on the guide catheter | Identical |
| Dimensional Data | | | |
| Transfer | OD | 4.4 Fr | Identical |
| Catheter | Length | 19.7 cm to 24.7 cm | Identical |
| Guide | OD | 6.8 Fr | Identical |
| Catheter | Length | 12 cm to 17.3 cm | Identical |
| Physical Features | | | |
| Transfer | | With graduation marks | Identical |
| | | No end-hole | Identical |
{5}------------------------------------------------
| | Predicate Device | Subject Device |
|------------------------|------------------------------------------------------|-------------------------------------------------|
| | Pivet and Soft-Pass Embryo<br>Transfer Catheter Sets | Soft-Pass Embryo Mock<br>Transfer Catheter Sets |
| | | |
| Guide | With or without Echotip | Identical |
| | With graduation marks | Identical |
| | Without positioner | Identical |
| | Without Echotip | Identical |
| | Straight | Identical |
| Manufacturing Features | | |
| Sterilization | ETO | Identical |
| Packaging | Tyvek Sterile Barrier | Identical |
| Shelf Life | 3 years | Identical |
The purpose of this submission is to modify the indications statement for the Soft-Pass Embryo Mock Transfer Catheter Sets to clarify the cleared uses of these devices. As shown in the table above, the subject and predicate devices have different indications for use statements; however, the intended uses of the predicate and subject devices are the same (i.e., catheter used to aid in embryo transfer procedures).
In regards to technological characteristics, the subject Soft-Pass Embryo Mock Transfer Catheter Sets are identical to those cleared under K173103. There are no differences raising different questions of safety or effectiveness.
# Performance Data:
The subject device, Soft-Pass Embryo Mock Transfer Catheter Sets, are identical to those cleared under K173103 and have not undergone any design changes since clearance. The changes in indications for use included in this submission did not require additional performance testing. Therefore, the performance testing included in K173103 supports the Soft-Pass Embryo Mock Transfer Catheter Sets included in this submission.
# Conclusion:
The Soft-Pass Embryo Mock Transfer Catheter Sets are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.