Guardia™ Access Malleable Obturator, Guardia™ Obturator, Soft-Pass™ Obturator, and Soft -Trans Malleable Obturator: Used to supplement and assist uterine access of a cleared, dimensionally compatible embryo transfer device for placement of in vitro fertilized (IVF) embryos into the uterine cavity. GuardiaTM ETS Embryo-Tested Syringe: For use with catheters indicated for intrauterine insemination and embryo transfer. Flushing Catheter: The Flushing Catheter is intended for instillation of sterile medium into the cervical canal prior to embryo transfer.
Device Story
Device family includes malleable obturators, embryo-tested syringe, and flushing catheter; used in clinical settings by physicians during IVF/IUI procedures. Obturators assist uterine access for embryo transfer catheters; flushing catheter instills sterile medium into cervical canal; syringe facilitates fluid delivery. Obturators feature adjustable handles and malleable wires to navigate uterine anatomy. Syringe is two-part (polypropylene barrel, high-density polyethylene plunger). Flushing catheter includes depth indicator and side port. Devices are sterile, single-use. Clinical benefit: facilitates precise placement of embryos and preparation of cervical canal, improving procedural success. Healthcare providers use these tools to guide catheters and manage fluids, directly impacting the accuracy of embryo transfer.
Clinical Evidence
Bench testing only. Studies included: sterilization validation (ISO 11135-1:2007), biocompatibility (ISO 10993-5, -10), endotoxin testing (AAMI/ANSI ST72:2011), package integrity (ASTM F2096-04, F88-09), and transportation simulation (ASTM D4169-05). Mouse Embryo Assay (MEA) confirmed >80% embryo development to blastocyst stage. Bench performance verified dimensions, leak resistance, tensile strength, handle securement, and fracture resistance (ISO 11070:1999).
Technological Characteristics
Materials: Polyurethane, stainless steel, polyethylene, nylon, silicone. Obturators: Malleable wire/stylet, adjustable handles. Flushing catheter: 20 cm tube, 14.9 Fr to 3.5 Fr taper, side port. Syringe: Polypropylene barrel, HDPE plunger. Sterilization: Ethylene oxide (ISO 11135-1). Connectivity: None (mechanical).
Indications for Use
Indicated for patients undergoing in vitro fertilization (IVF) or intrauterine insemination (IUI) procedures requiring uterine access or cervical canal preparation.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 24, 2018
Cook Incorporated Naomi Funkhouser Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, Indiana 47402
Re: K180302
> Trade/Device Name: Guardia™ Access Malleable Obturator, Guardia™ Obturator, Soft-Pass™ Obturator, Soft-Trans Malleable Obturator, Guardia™ ETS Embryo-Tested Syringe, Flushing Catheter Regulation Number: 21 CFR 884.6110 Regulation Name: Assisted Reproduction Catheters Regulatory Class: Class II Product Code: MOF Dated: September 26, 2018 Received: September 27, 2018
Dear Naomi Funkhouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K180302
#### Device Name
Guardia™ Access Malleable Obturator, Guardia™ Obturator, Soft-Pass™ Obturator, Soft -Trans Malleable Obturator, GuardiaTM ETS Embryo-Tested Syringe, Flushing Catheter
#### Indications for Use (Describe)
Guardia™ Access Malleable Obturator, Guardia™ Obturator, Soft-Pass™ Obturator, and Soft -Trans Malleable Obturator:
Used to supplement and assist uterine access of a cleared, dimensionally compatible embryo transfer device for placement of in vitro fertilized (IVF) embryos into the uterine cavity.
GuardiaTM ETS Embryo-Tested Syringe:
For use with catheters indicated for intrauterine insemination and embryo transfer.
#### Flushing Catheter:
The Flushing Catheter is intended for instillation of sterile medium into the cervical canal prior to embryo transfer.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white letters on a red background. The red background is shaped like a rectangle with a notch cut out of the upper left corner.
# 510(k) Summary - K180302
### 1. Submitter Information:
| Applicant: | Cook Incorporated |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contacts: | Naomi Funkhouser<br>Rohini Patel |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone: | (812) 339-2235 x104371 |
| Contact Fax: | (812) 332-0281 |
| 2. Date Prepared: | October 24, 2018 |
| 3. Device Information: | |
| Trade Names: | Guardia™ Access Malleable Obturator, Guardia™<br>Obturator, Soft-Pass™ Obturator, Soft-Trans Malleable<br>Obturator, Guardia™ ETS Embryo-Tested Syringe,<br>Flushing Catheter |
| Common Name: | Embryo Transfer Obturators and Accessories |
| Classification Name: | Assisted Reproduction Catheters (21 CFR 884.6110) |
| Classification Regulation: | MQF (Catheter, Assisted Reproduction) |
| Regulatory Class | II |
### 4. Predicate Device:
Embryo Transfer Catheter/Sets (K983594) manufactured by Cook Ob/Gyn. This predicate device has not been subject to any design related recalls.
# 5. Device Description:
This 510(k) covers eight devices from five of COOK's device families. These devices, as outlined in the table below, are sold separately.
| Product family | Product name | RPN | Diameter/length or volume |
|----------------|--------------------------------------------------------|------------|---------------------------|
| Guardia Access | Guardia™ Access Malleable<br>Obturator | K-JET-7002 | 4.0 Fr/19.9 cm |
| Guardia | Guardia™ Obturator (Malleable<br>Adjustable Obturator) | J-UOB-2828 | 2.8 Fr/28 cm |
| | Guardia™ Obturator (Malleable<br>Adjustable Obturator) | J-UOB-4028 | 4.0 Fr/28 cm |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, sans-serif font, placed on a red square. Below the red square is a red banner with the word "MEDICAL" in white, sans-serif font. The banner is slightly offset from the square above it.
| | Guardia™ Obturator (Stiff Adjustable<br>Obturator) | J-UOB-4028-ST | 4.0 Fr/28 cm |
|-------------------|----------------------------------------------------|------------------|----------------|
| | Guardia™ ETS Embryo-Tested<br>Syringe | K-ETS-1000 | 1 ml |
| Soft-Pass | Soft-Pass™ Obturator | J-SP-4420 | 4.4 Fr/20.2 cm |
| Soft-Trans | Soft-Trans Malleable Obturator | K-SOFT-4018 | 4.0 Fr/18 cm |
| Flushing Catheter | Flushing Catheter | J-IUIC-352000-CE | 3.5 Fr/20 cm |
The polyurethane obturators are designed to fit dimensionally compatible embryo transfer guide catheters. Some of these devices are designed to be adjustable in length.
The flushing catheter is a 20 cm tube measuring 14.9 Fr at the proximal end and tapering to 3.5 Fr at the distal end. It features a depth indicator and directional side port with an ink mark aligned with the position of the side port. This device is identical to the Insemi-Cath IUI catheter cleared under K172321.
The Guardia Embryo-Tested Syringe is two-part syringe with a polypropylene barrel and a highdensity polyethylene plunger. This device undergoes Mouse Embryo Assay (MEA) and endotoxin testing (LAL) before lot release.
All subject devices are supplied sterile and intended for one-time use. The Guardia Embryo-Tested Syringe has a shelf-life of five years, whereas the remaining subject devices have a shelflife of three years.
# 6. Indications for Use:
Guardia Malleable Obturator, Guardia TM Obturator, Soft-Pass™ Obturator, and Soft-Trans Malleable Obturator:
Used to supplement and assist uterine access of a cleared, dimensionally compatible embryo transfer device for placement of in vitro fertilized (IVF) embryos into the uterine cavity.
# GuardiaTM ETS Embryo-Tested Syringe
For use with catheters indicated for intrauterine insemination and embryo transfer.
### Flushing Catheter
The Flushing Catheter is intended for instillation of sterile medium into the cervical canal prior to embryo transfer.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom left.
# 7. Comparison of Intended Use and Technological Characteristics with the Predicate Device:
# Comparison of obturators with the predicate device
| Device | Subject device (K180302) | Predicate device (K983594) |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same as the predicate device | Used to supplement and assist uterine access of<br>cleared, dimensionally compatible embryo transfer<br>devices for placement of in vitro fertilized (IVF)<br>embryos into the uterine cavity. |
| Design | • Adjustable handles to be sized<br>to fit inside guide catheters<br>• Polyethylene over soft<br>stainless or stylet wire for<br>malleability<br>• Smooth and rounded distal tip | • Sized to fit inside guide catheters<br>• Malleability not specified<br>• Tip characteristic not specified |
| Dimension | • OD: 2.8-4.4 Fr<br>• Length: 18-28 cm | • OD: Matched to fit within guide catheter<br>• Length: matched with guide catheter at tip |
| Materials | Same as the predicate device | Polyethylene, stainless steel |
Both subject and predicate devices are used to supplement and assist uterine access of cleared, dimensionally compatible embryo transfer devices for placement of IVF embryos into the uterine cavity. Therefore, they have the same intended use.
The subject and predicate devices use the same materials but have different design and dimensions. However, the differences are commonly seen in cleared devices and therefore, do not raise different questions of safety and effectiveness. The differences can be assessed by biocompatibility and bench performance testing.
### Comparison of Guardia™ ETS Embryo-Tested Syringe with the predicate device
The predicate device (K983594) includes a syringe intended for aiding in embryo transfer procedures. Therefore, the subject device and predicate devices have the same intended use. In addition, these two devices have comparable technological characteristics. The Guardia™ ETS Embryo-Tested Syringe is identical to a cleared device intended for general injection use. To ensure biological safety, the subject syringe is evaluated for embryotoxicity and endotoxin.
| Device | Subject device (K180302) | Predicate device (K983594) |
|-----------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Design | • Without graduation marks<br>• Close end with side port<br>• Without hubs<br>• With positioner | • With or without graduation marks<br>• Open end with side port<br>• With or without hubs<br>• Without positioner |
| Dimension | OD 3.5 Fr / Length 20 cm | OD 2-8 Fr / Length 12-30 cm |
| Materials | Nylon, silicone | Polyethylene, Teflon, stainless steel |
### Comparison of Flushing Catheter with the predicate device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, sans-serif font, stacked on top of the word "MEDICAL" in a smaller, white, sans-serif font. Both words are set against a red background. The "COOK" has a trademark symbol next to it.
The subject and predicate devices have similarity in design (e.g., side port, hubless, no graduation marks and comparable dimensions). There are differences in design, materials, and dimensions, but these differences do not raise different questions of safety or effectiveness as compared to the predicate device and are commonly seen in other comparable cleared Assisted Reproduction Technology devices. The differences can be assessed by biocompatibility and bench performance testing.
# 8. Summary of Non-Clinical Performance Testing:
The following studies have been performed to support substantial equivalence to the predicate device:
- . Sterilization Validation testing per ISO 11135-1:2007
- . Biocompatibility studies, as follows:
- Cytotoxicity testing per 10993-5:2009 o
- Guinea Pig Maximization Sensitization testing per ISO 10993-10:2002/2010 o
- Intracutaneous Irritation testing per ISO 10993-10:2002/2010 o
- Endotoxin testing per AAMI/ANSI ST72:2011 (<20 EU/device) .
- Transportation Simulation study per ASTM D4169-05 .
- Package Integrity testing after real-time aging: ●
- Bubble Leak test per ASTM F2096-04 o
- Seal Strength testing per ASTM F88-09 o
- Visual Inspection: No package displayed damage (tears, folds, puncture holes, etc.) o
- Mouse Embryo Assay (MEA) before and after aging:
One-cell mouse embryos were exposed to subject devices and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed. The testing demonstrated that the devices met acceptance criterion of "1-cell MEA >80% embryos developed to blastocyst in 96 hours."
- Bench Performance studies before and after aging demonstrated that all predetermined acceptance criteria were met in the following tests:
- Obturator/Flushing Catheter dimensional verification Devices are measured and o verified against device input requirements.
- Flushing Catheter Leak Test Testing ensures that fluid path does not leak under a o predetermined injection pressure.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, block letters on a red background. The red background extends further to the left than the word "COOK".
- Obturator/Flushing Catheter Tensile Test Testing demonstrates that the tensile o strength value is greater than the predetermined acceptance criterion.
- Obturator Handle Securement Test Testing demonstrates the adjustment o mechanism does not allow the obturator to move from its set position under worst case handling under the predetermined acceptance criterion.
- Obturator Fracture Test This testing demonstrated that the test articles do not o facture, loosen, or fail when tested in accordance with ISO 11070:1999
In addition, performance data in K172321 and K101547 are leveraged to support substantial equivalence of Flushing Catheter and Guardia™ ETS Embryo-Tested Syringe, respectively.
### 9. Conclusion:
The subject and predicate devices have the same intended use. Although there are differences in technological characteristics between the subject and predicate devices, these differences do not raise different questions of safety or effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.