OTRIEVA TAPERED OVUM ASPIRATION NEEDLE SINGLE LUMEN
K141365 · William A. Cook Australia Pty. , Ltd. · MQE · Aug 7, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141365
Device Name
OTRIEVA TAPERED OVUM ASPIRATION NEEDLE SINGLE LUMEN
Applicant
William A. Cook Australia Pty. , Ltd.
Product Code
MQE · Obstetrics/Gynecology
Decision Date
Aug 7, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.6100
Device Class
Class 2
Indications for Use
The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for laproscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles.
Device Story
Otrieva™ Tapered Ovum Aspiration Needle is a sterile, single-use device for oocyte retrieval. It consists of a tapered cannula (17G proximal to 20G distal) with a constant 0.60 mm internal diameter, aspiration tubing, and a vacuum line. The needle is inserted transvaginally under ultrasound guidance or via cannula during laparoscopy to access ovarian follicles. The device connects to an external vacuum source to aspirate and flush oocytes. Used by clinicians in clinical settings for in vitro fertilization procedures. The tapered design provides the rigidity of a larger needle with the reduced puncture trauma of a smaller tip. The device facilitates oocyte collection, which is essential for subsequent laboratory fertilization and embryo development.
Clinical Evidence
Bench testing only. Testing included resistance to breakage, air tightness, tensile strength, and leakage testing. Performance specifications were verified, including a Mouse Embryo Assay (MEA) change from 1-cell (75% blastocyst rate at 96h) to 2-cell (80% blastocyst rate at 72h), confirming no adverse impact on embryo toxicity.
Indicated for laparoscopic or ultrasound-guided transvaginal aspiration and flushing of oocytes from ovarian follicles. Contraindicated in patients with active vaginal or intrauterine infection, sexually transmitted disease, recent uterine perforation, recent caesarean section, or pregnancy.
Regulatory Classification
Identification
Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2014
William A. Cook Australia Pty, Ltd. Katie-Lou Buscher Project Manager, Regulatory Affairs 95 Brandl Street Brisbane Technology Park, Eight Mile Plains Brisbane, QLD 4113 AUSTRALIA
Re: K141365 Trade/Device Name: Otrieva™ Tapered Ovum Aspiration Needle Regulation Number: 21 CFR& 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MQE Dated: July 14, 2014 Received: July 15, 2014
Dear Katie-Lou Buscher,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K141365
Device Name
Otrieva™ Tapered Ovum Aspiration Needle
Indications for Use (Describe)
The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for laproscopic or ultrasound guided transvaginal aspiration and flushing of oocytes from ovarian follicles.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters, and the word "MEDICAL" is written in smaller, white, sans-serif letters below the word "COOK".
## 510(k) Summary - K141365 Otrieva™ Tapered Ovum Aspiration Needle - 14 July 2014
| Submitted by: | William A. Cook Australia<br>95 Brandl Street<br>Eight Mile Plains<br>Brisbane<br>Queensland<br>4113<br>Australia |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:<br>Email:<br>Telephone:<br>Fax: | Katie-Lou Buscher<br>Katie-Lou.Buscher@CookMedical.com<br>+61 7 3841 1188<br>+61 7 3841 1288 |
| Trade name | Otrieva™ Tapered Ovum Aspiration Needle |
| 510(k) Number | K141365 |
| Classification name | Assisted Reproduction Needles (21 CFR 884.6100, Product Code<br>MQE) |
| Predicate Device | Ovum Pick-Up Aspiration Needle, cleared under K983593 |
| Device Description | This device is intended for laparoscopic or ultrasound guided<br>transvaginal aspiration and flushing of oocytes from ovarian follicles<br><br>The needle is passed through a transvaginal ultrasound transducer or<br>placed through a cannula for a laparoscopic procedure to advance into<br>the ovarian follicle.<br><br>The main body of the cannula is 17 gage, which tapers to 20 gage at<br>the distal tip. Although the outer diameter of the needle tapers, a<br>constant 0.60 mm internal diameter is maintained.<br><br>The Otrieva™ needles are available in two lengths: 30cm and 35cm.<br>The needles are provided with an aspiration tube of 90 cm length and a<br>vacuum tube with a length of 50 cm. The tubing is connected to a<br>silicone bung.<br><br>The materials used in the device include: stainless steel, polycarbonate,<br>PTFE, silicone, FEP and copolyester. |
| Indication for Use | The Cook Otrieva™ Tapered Ovum Aspiration Needles are used for<br>laparoscopic or ultrasound guided transvaginal aspiration and flushing<br>of oocytes from ovarian follicles. |
| Comparison to | The Otrieva™ Tapered Ovum Aspiration Needle is a modification of the |
| Predicate Device | Ovum Pick-up Aspiration Needle (K983593). The modifications are:<br>• Incorporation of a tapered cannula instead of fixed diameter cannula - the tapered cannula is designed to allow the clinician the rigidity of a larger diameter cannula while providing a smaller diameter tip for puncture. As the inner and outer diameter of the cannula at both ends falls within the range of the predicate device, and the internal lumen of the cannula is the same as the predicate device, this change does not represent a new technology that could raise new types of safety or effectiveness questions. |
| | • Addition of a 50 cm vacuum line - the addition of the vacuum line simply provides an extension of the connection to the vacuum source and does not change the use of the device. Attachment of an Ovum Aspiration Needle to the vacuum source using tubing is not a new technology and therefore there are no new safety or effectiveness questions raised. |
| | • Incorporation of a smaller (1.32 mm) diameter aspiration tube – The smaller diameter aspiration tube is applied to better match the inner lumen of the needle cannula. This modification is not a new technology and does not raise new types of safety or effectiveness questions. |
| | In addition to the modifications described above, the Indication for Use statement has been modified from the predicate device to include transvaginal ultrasound and laparoscopic approaches, both of which were specifically stated within the instructions for the predicate device, just not stated directly within the Indications for Use statement. This modification to the Indication for Use statement does not impact safety and efficacy of the device. |
| | The Mouse Embryo Assay (MEA) specification has changed from one-cell (75% or greater blastocyst rate at 96 h) to 2-cell (80% or greater blastocyst rate at 72 h). This is a change in performance specification, and does not represent a new technology, therefore the safety and effectiveness is not altered. |
| | A full comparison is provided in Table 1 over the page. |
{4}------------------------------------------------
{5}------------------------------------------------
## Table 1 - Comparison of Otrieva and predicate device
| Device Name | Otrieva™ Tapered<br>Ovum Aspiration<br>Needle | Ovum Pick-Up<br>Aspiration Needle | Comparison |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Cook Otrieva™<br>Tapered Ovum<br>Aspiration Needles are<br>used for laparoscopic<br>or ultrasound guided<br>transvaginal aspiration<br>and flushing of oocytes<br>from ovarian follicles. | The Ovum Pick-Up<br>Aspiration Needles are<br>used for aspiration and<br>flushing of oocytes from<br>ovarian follicles. | The Otrieva Indication for<br>Use has the addition of<br>ultrasound and<br>laparoscopic approaches<br>explicitly stated. The<br>predicate device is<br>intended to be used<br>laparoscopically and with<br>transvaginal ultrasound,<br>as described in the<br>predicate 510(k)<br>summary, and these<br>approaches were also<br>included in the cleared<br>labelling for the predicate<br>device.<br>This is not a new<br>intended use, as the<br>clinical practise has<br>always included<br>transvaginal ultrasound<br>and laparoscopy. |
| Contraindications | This device should not<br>be used on a patient<br>with an active vaginal<br>or intrauterine infection,<br>a sexually transmitted<br>disease, a recent<br>uterine perforation, a<br>recent caesarean<br>section, or who is<br>currently pregnant. | This device should not<br>be used on a patient<br>with an active vaginal or<br>intrauterine infection, a<br>sexually transmitted<br>disease, or who is<br>currently pregnant. | These patient populations<br>were added to ensure<br>consistency with clinical<br>practices for in vitro<br>fertilization. |
| Product Description | Needle passed through<br>transvaginal ultrasound<br>transducer or placed<br>through a cannula for a<br>laparoscopic procedure<br>to advance into the<br>ovarian follicle. | Needle passed through<br>transvaginal ultrasound<br>transducer or placed<br>through a cannula for a<br>laparoscopic procedure<br>to advance into the<br>ovarian follicle. | The description of the<br>Otrieva™ is substantially<br>equivalent to the<br>predicate device. |
| Dimensions | | | |
| Gage | 17 G tapering to 20 G | 12 G - 20 G | The Otrieva™ Tapered<br>Ovum Aspiration Needle |
| Cannula Inner<br>Diameter | 0.60 mm | 2.10 mm - 0.60 mm | has an increased wall<br>thickness at the 17 G<br>proximal portion of the |
| Needle Length | 30 & 35 cm | 20 cm - 40 cm | |
| Aspiration Tubing<br>Length | 90 cm | 25cm - 90 cm | needle cannula which is<br>necessary to ensure an<br>appropriate wall<br>thickness at the distal,<br>tapered end of the needle<br>while maintaining a 0.60 mm<br>inner diameter. |
| Device Name | Otrieva™ Tapered<br>Ovum Aspiration<br>Needle | Ovum Pick-Up<br>Aspiration Needle | Comparison |
| Aspiration Tubing<br>External Diameter | 1.32 mm | 1.67 mm - 2.7 mm<br>(5 Fr - 8 Fr) | The diameter of the<br>Otrieva aspiration tubing<br>is smaller than the<br>predicate. |
| Vacuum Tubing<br>Length | 50 cm | N/A | Vacuum tubing has been<br>added to the design of the |
| Vacuum Tubing<br>External Diameter | 2.06 mm | N/A | Otrieva™ Tapered Ovum<br>Aspiration Needle |
| Materials | | | |
| Needle Cannula | Stainless Steel #304 | Stainless Steel #304 | Same |
| Needle Hub | Polycarbonate | Polycarbonate | Same |
| Aspiration Tubing | PTFE | PTFE | Same |
| Vacuum Tubing | FEP | N/A | Vacuum tubing has been<br>added to the design of the<br>Otrieva™ Tapered Ovum<br>Aspiration Needle |
| Test Tube Stopper | Silicone | Silicone | Same |
| Vacuum Line Female<br>Conical Fitting | Copolyester or<br>Polycarbonate | N/A | Vacuum tubing has been<br>added to the design of the<br>Otrieva™ Tapered Ovum<br>Aspiration Needle |
| Vacuum Line Cannula | Stainless steel #304 | Stainless steel #304 | Same |
| Other characteristics | | | |
| Lumens | Single | Single or Double | Otrieva™ Tapered Ovum<br>Aspiration Needle lumen<br>type falls within the range<br>of the predicate device |
| Echo-tip | Echo-Tip only | With or without Echo-Tip | Otrieva™ Tapered Ovum<br>Aspiration Needle lumen<br>type falls within the range<br>of the predicate device |
| Shelf Life | 3 years | 3 years | Same |
| Single Use/Reusable | Single Use | Single Use | Same |
| Sterility | Sterile (SAL 10-6) | Sterile (SAL 10-6) | Same |
| Endotoxin<br>Specification | USP endotoxin (LAL)<br>tested and passed with<br>20EU or less per<br>device | USP Endotoxin (LAL)<br>tested and passed with<br>20EU or less per device | Same |
| MEA specification | Two-cell MEA tested<br>and passed with 80%<br>or greater Blastocyst<br>rate at 72h. | Once cell MEA tested<br>and passed with 75% or<br>greater Blastocyst at 96<br>h. | MEA test has changed<br>from 1-cell MEA (75% or<br>greater Blastocyst at 96<br>h) to 2-cell MEA (80% or<br>greater Blastocyst at 72<br>h). Both techniques are<br>acceptable and provide<br>indication of embryo<br>toxicity. |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Packaging | Polyester/low density<br>polyethylene film with<br>Tyvek-Poly pouch | Tyvek-Poly pouch | Same |
{6}------------------------------------------------
{7}------------------------------------------------
| Non-clinical testing<br>summary | The Otrieva™ Tapered Ovum Aspiration Needle has undergone non-clinical testing to verify the design modifications did not affect the performance of the needle. The testing included Resistance to breakage testing Air tightness testing Tensile strength testing between the tubing and handle junction Leakage testing |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The design modifications were shown to have no adverse effect on the needle performance in any of these tests. |
| Safety and Efficacy | The results of the testing provide reasonable assurance that the Otrieva™ Tapered Ovum Aspiration Needle is as safe and effective as the predicate device and support a determination of substantial equivalence. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.