Intelligent Wireless Fitness Apparatus is indicated to be used for: Improvement of abdominal tone, strengthening of the abdominal muscles development of firmer abdomen. Strengthening, Toning and Firming of buttocks and thighs.
Device Story
Intelligent Wireless Fitness Apparatus (Models AST-301, AST-302, AST-303) is a powered muscle stimulator for OTC use. Device consists of a main unit and electrode pads. User operates device via three buttons: On/Off/Mode selection, and intensity adjustment (+/-). Device delivers electrical stimulation to nerves to induce muscle contraction. Used for abdominal, buttock, and thigh toning. Provides benefits of muscle strengthening and firming. Healthcare providers do not operate the device; it is intended for patient self-use. Output is visual (LED indicators for status, battery, and intensity).
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 62366), software verification/validation, and battery safety (IEC 62133). Biocompatibility of electrode pads confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/hypersensitivity). No clinical data provided.
Technological Characteristics
Powered muscle stimulator; Class II; Product Code NGX. Materials: ABS plastic housing, transparent PET silicone/conductive silicone electrode pads. Energy: Rechargeable 3.7Vdc Li-ion battery or 5V/1A adapter. Output: 2 channels, voltage-controlled, pulsed symmetric biphasic waveform. Connectivity: Standalone. Sterilization: Not specified. Software: Microprocessor-controlled with automatic shut-off and patient override.
Indications for Use
Indicated for individuals seeking improvement of abdominal tone, strengthening of abdominal muscles, development of a firmer abdomen, and strengthening, toning, and firming of buttocks and thighs.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
K181728 — Muscle Trainer · Shenzhen Kentro Medical Electronics Co., Ltd. · Oct 17, 2018
K102295 — X2ABS DUAL CHANNEL FITNESS BELT · Leto Enterprises Incorporation · Feb 18, 2011
K122414 — AB COMMAND DUAL CHANNEL FITNESS BELT · Leto Enterprises Incorporation · Dec 21, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2019
Shenzhen OSTO Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou. 510663 CN
Re: K182136
Trade/Device Name: Intelligent Wireless Fitness Apparatus (Model: AST-301, AST-303) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: September 10, 2019 Received: September 13, 2019
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Vivek Pinto. PhD Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182136
Device Name Intelligent Wireless Fitness Apparatus (Model: AST-301, AST-302, AST-303)
Indications for Use (Describe)
Intelligent Wireless Fitness Apparatus is indicated to be used for:
Improvement of abdominal tone, strengthening of the abdominal muscles development of firmer abdomen. Strengthening, Toning and Firming of buttocks and thighs.
Type of Use (Select one or both, as applicable)
|--|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
510(k) Owner's Name: Shenzhen OSTO Technology Co., Ltd. Establishment Registration Number: Applying Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street Longgang District, Shenzhen City Guangdong Province, CHINA Tel: +86-755-29769546 Fax: +86-755-29769540 Contact Person: Li Yang (General Manager) Email: annaosto@163.com
## Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Tel: +86-20-61099984 Email: regulatory@glomed-info.com
#### 2. Subject Device Information
Company Name: Shenzhen OSTO Technology Co., Ltd. Trade/Device Name: Intelligent Wireless Fitness Apparatus Model Name: AST-301, AST-302, AST-303 Common Name: Powered Muscle Stimulator Product Code: NGX Regulation Number: 21 CFR 890.5850 Regulatory Class: 2
#### 3. Predicate Device Information
#### Predicate Device 1:
510(K) Number: K131291 Company Name: Johari Digital HealthCare Ltd. Trade/Device Name: Torc Body Common Name: Powered Muscle Stimulator Product Code: NGX Regulation Number: 21 CFR 890.5850 Regulatory Class: 2
#### Predicate Device 2:
510(K) Number: K133929 Company Name: Shenzhen OSTO Technology Co., Ltd. Trade Device Name: Health Expert Electronic Stimulator Common Name: Electronic Stimulator Product Code: NUH, NGX Regulation Number: 882.5890, 890.5850 Regulatory Class: 2
#### 4. Device Description
{4}------------------------------------------------
Intelligent Wireless Fitness Apparatus is indicated to be used for: Improvement of abdominal tone, strengthening of the abdominal muscles development of firmer abdomen. By muscle stimulation, the device helps to strengthen and firm the abdomen, arms and thighs.
The device is including a main unit and 3 kinds of electrode pads. The therapeutic parameters are easily set using the buttons on the device.
There are three buttons on the main unit, On/off button for on/off and mode selection, "+" button and "-" button for level adjustment.
The accessories is electrode pads. The electrode pads connect to the main unit by buttons (which is permanent secured on the main unit and the electrode pad). All the accessories can only be changed by local distributor.
## 5. Intended Use / Indications for Use
Intelligent Wireless Fitness Apparatus is indicated to be used for:
lmprovement of abdominal tone, strengthening of the abdominal muscles development of firmer abdomen. Strengthening, toning and firming of buttocks and thighs.
## 6. Test Summary
Intelligent Wireless Fitness Apparatus, Model: AST-301, AST-302, AST-303 had been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards �
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Usability test according to IEC 62366 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- � IEC 62133 Edition 2.0 2012-12 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications [Including: Corrigendum 1 (2013)]
Note:
The materials and manufacturing used for the Intelligent Wireless Fitness Apparatus are identical to those of the predicate device K133929, and the Electrode Pads have been demonstrated to conform with the following biocompatibility standards:
• ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity
• ISO 10993-10: 2009, Biological evaluation of medical devices - Part 10: Tests for irritation and delayedtype hypersensitivity
## 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Intelligent Wireless Fitness Apparatus is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Predicate Device | Remark |
|---------------------------|-------------------------------------------|------------------|----------------------------------------|--------|
| 510(k) number | Applying | K131291 | K133929 | -- |
| Device name | Intelligent Wireless<br>Fitness Apparatus | Torc Body | Health Expert<br>Electronic Stimulator | -- |
| Model Name: | AST-301, AST-302, | -- | AST-300C and AST- | -- |
{5}------------------------------------------------
| | | AST-303 | | 300D | | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------|
| Company name | | Shenzhen OSTO<br>Technology Co., Ltd. | Johari Digital<br>HealthCare Ltd | Shenzhen OSTO<br>Technology Co., Ltd. | | |
| Product Code | | NGX | NGX | NUH, NGX | SE | |
| Regulation Number | | 21 CFR 890.5850 | 21 CFR 890.5850 | 882.5890, 890.5850 | SE | |
| Indications for<br>Use | | Intelligent Wireless<br>Fitness Apparatus is<br>indicated to be used<br>for:<br>Improvement of<br>abdominal tone,<br>strengthening of the<br>abdominal muscles<br>development of<br>firmer abdomen.<br>Strengthening, toning<br>and firming of<br>buttocks and thighs. | Torc Body is<br>indicated to be used<br>for:<br>Improvement of<br>abdominal tone, for<br>strengthening of the<br>abdominal muscles,<br>for development of<br>firmer abdomen.<br>Strengthening,<br>Toning and Firming<br>of buttocks & thighs. | PMS (Mode 1~8) It is<br>intended to stimulate<br>healthy muscles in<br>order to improve and<br>facilitate muscle<br>performance. TENS<br>(Mode 9~25) To be<br>used for temporary<br>relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back, back of the<br>neck, arm, leg, and<br>foot due to strain from<br>exercise or normal<br>household work<br>activities by applying<br>current to stimulate<br>nerve. | SE | |
| Primary Function | | Muscle stimulation | Muscle stimulation | Muscle stimulation | SE | |
| Principle of<br>Action | | Initiating action<br>potential of nerves<br>results in muscle<br>contraction | Initiating action<br>potential of nerves<br>results in muscle<br>contraction | Initiating action<br>potential of nerves<br>results in muscle<br>contraction | SE | |
| Clinical Use | | OTC | OTC | OTC | SE | |
| Target area | | abdominal muscles,<br>buttocks and thighs | abdominal muscles,<br>buttocks and thighs | abdominal muscles,<br>buttocks thighs and<br>foot | SE | |
| Shape of electrode | | Butterfly-shaped,<br>irregular | Butterfly-shaped,<br>irregular | Rectangle, irregular | SE | |
| Method of electrode<br>application | | Self-adhesive with<br>belt versus adhesive<br>silicone | Self-adhesive with<br>belt versus adhesive<br>silicone | Self-adhesive with<br>belt versus adhesive<br>silicone, non-adhesive | SE | |
| Power Source(s) | | Adapter (Model<br>HDMU05E-050100,<br>HDMU05B-050100,<br>HDMU05U-050100 )<br>Input:100-240 Vac;<br>50/60 Hz; 0,3A;<br>Output: 5 V ; 1A<br>Rechargeable<br>Lithium-ion Battery:<br>3.7Vdc | 24 V DC Battery<br>pack | 100 - 240 V AC, 50-<br>60 Hz | SE<br>Note 2 | |
| Electrical Protection | | Class II, BF | Class II, BF | Class II, BF | SE | |
| -Method of Line<br>Current Isolation | | Type BF Applied Part | Type BF Applied Part | Type BF Applied Part | SE | |
| Patient<br>Leakage | NC | $\leq$ 0.1 $\mu$ A | -- | -- | SE | |
| Current | | | | | | |
| Number of Output<br>Modes | | 8 | 2 | 25 | SE<br>Note 1 | |
| Number of Output<br>Channels: | | 2 | -- | 2 | SE | |
| Synchronous or<br>Alternating | | Alternating | -- | Alternating | SE | |
| Method of Channel<br>Isolation | | Voltage Transform<br>Isolation | -- | Voltage Transform<br>Isolation | SE | |
| Regulated Current or<br>Regulated Voltage | | Voltage Control | Voltage Control | Voltage Control | SE | |
| Software/Firmware/M<br>icroprocessor<br>Control | | Yes | Yes | Yes | SE | |
| Automatic Overload<br>Trip | | No | No | No | SE | |
| Automatic No-Load<br>Trip | | No | No | No | SE | |
| Automatic Shut Off | | Yes | -- | Yes | SE | |
| Patient Override<br>Control | | Yes | Yes | Yes | SE | |
| | On/Off<br>Status | Yes | -- | Yes | SE | |
| Indicator<br>Display | Low<br>Battery | Yes | -- | No | SE | |
| | Voltage/<br>Current<br>Level | Yes | | Yes | SE | |
| Timer Range<br>(minutes) (clarify for<br>each mode if<br>different) | | 15min | 1- 60 min | 25min | SE<br>Note1 | |
| Compliance with<br>Voluntary Standards | | Complied with<br>ISO 10993-5,<br>ISO 10993-10,<br>IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10 | Complied with<br>ISO 10993-5,<br>ISO 10993-10,<br>IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10 | Complied with<br>ISO 10993-5,<br>ISO 10993-10,<br>IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10 | SE | |
| | Weight | | 80g (Without<br>accessories) | -- | 2Kg (Without<br>accessories) | SE<br>Note1 |
| | Housing Materials<br>and Construction | | Main unit: ABS<br>plastic | -- | Main unit: ABS plastic | SE |
| | Waveform (e.g.,<br>pulsed monophasic,<br>biphasic) | | Pulsed, symmetric,<br>biphasic | -- | Pulsed, symmetric,<br>biphasic | SE |
| | Maximum Output<br>Voltage | | 44V±10% @ 500Ω<br>80V±10% @ 2ΚΩ<br>112V±10% @ 10ΚΩ | -- | 44V±10% @ 500Ω<br>80V±10% @ 2ΚΩ<br>112V±10% @ 10ΚΩ | SE |
| Maximum Output<br>Current | | 88mA±10% @ 500Ω<br>40mA±10% @ 2ΚΩ<br>11.2mA±10%<br>@10ΚΩ | -- | 88mA±10% @ 500Ω<br>40mA±10% @ 2ΚΩ<br>11.2mA±10% @10ΚΩ | SE | |
| | For multiphasic<br>waveforms only: - | | Symmetrical phases | -- | Symmetrical phases | SE |
| Symmetrical phases | | | | | | |
| Phase Duration | | $120 \mu s$ | $290 \mu s$ | $120 \mu s$ | SE | |
| Net Charge (mC per pulse) | | $10.56 \mu C @ 500\Omega$ | -- | $10.56 \mu C @ 500\Omega$ | SE | |
| Maximum Phase Charge, (mC) | | $12.78 \mu C @ 500\Omega$ | $35.7 \mu C @ 500\Omega$ | $12.78 \mu C @ 500\Omega$ | SE | |
| Maximum Average Current | | $1.69mA @ 500\Omega$ | -- | $1.69mA @ 500\Omega$ | SE | |
| Maximum Current Density, (mA/cm²) | | $0.026mA/cm^2 @ 500\Omega$ | $0.176mA/cm^2$ | $0.026mA/cm^2 @ 500\Omega$ | SE | |
| Maximum Power Density | | $16.37 \mu W/cm^2 @ 500\Omega$ | $0.0089 W/cm^2$ | $16.37 \mu W/cm^2 @ 500\Omega$ | SE | |
| Burst Mode (i.e., pulse trains) | Pulses per burst | 1 | -- | 1 | SE | |
| | Bursts per second | 77 | -- | 77 | SE | |
| | Burst duration (seconds) | 12.94ms | -- | 12.94ms | SE | |
| | Duty Cycle [Line (b) x Line (c) | $240 \mu s/12.94ms$ | -- | $240 \mu s/12.94ms$ | SE | |
| ON Time (seconds) | | 3s | -- | 3s | SE | |
| OFF Time (seconds) | | 3s | -- | 3s | SE | |
| Type of Energy | | Electrical | Electrical | Electrical | SE | |
| Type of Operation | | Continuous | Continuous…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.