ALEVE Direct Therapy (ALEVE Direct Therapy TENS device)

K152852 · Bayer Healthcare, LLC · NUH · Dec 22, 2015 · Neurology

Device Facts

Record IDK152852
Device NameALEVE Direct Therapy (ALEVE Direct Therapy TENS device)
ApplicantBayer Healthcare, LLC
Product CodeNUH · Neurology
Decision DateDec 22, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities.

Device Story

Battery-powered TENS device for OTC pain relief; applies electrical current to peripheral nerves via skin-contact electrodes. Comprised of TENS unit with integral silicone electrodes, replaceable hydrogel pads, and wireless remote control. User adheres device to lower back; uses remote to adjust stimulation intensity. Device reduces pain perception through electrical stimulation. Operates on fixed 30-minute 42-second cycle. Intended for home use by patients.

Clinical Evidence

Bench testing only. Provided verification of output characteristics, waveform data, and software verification/validation (Moderate level of concern). Biocompatibility testing (ISO 10993-1, -5, -10) and electrical safety/EMC testing (IEC 60601 series) performed. Usability supported by right of reference to 15-subject study from predicate K120500.

Technological Characteristics

Battery-powered (2 AAA, 1 Lithium coin). ABS plastic housing. Integral silicone electrodes with carbon rubber conductive area. Hydrogel pads. Dimensions: 7.5" x 3.5" x 0.7". Weight: 4.8 oz. Voltage-regulated output. Software-controlled. Complies with AAMI ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014, IEC 60601-1-11:2010, IEC 60601-2-10:2012, ISO 10993 series.

Indications for Use

Indicated for temporary relief of lower back muscle pain due to strain from exercise or normal household/work activities in adults.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 22, 2015 Bayer Healthcare, LLC William Walsh Director, Us Regulatory Affairs 100 Bayer Boulevard Whippany, New Jersey 07981-0915 Re: K152852 Trade/Device Name: Aleve Direct Therapy (Aleve Direct Therapy Tens Device) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: August 25, 2015 Received: September 29, 2015 Dear Mr. Walsh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # William J. Heetderks -S for Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152852 Device Name ALEVE Direct Therapy TENS Device Indications for Use (Describe) Temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(K) SUMMARY #### SUBMITTER I. | Company Name: | Bayer HealthCare LLC, Consumer Care<br>100 Bayer Boulevard<br>Whippany, NJ 07981-0915 | |------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | William R. Walsh, R.Ph.<br>Phone: (862)-404-6355<br>bill.walsh@bayer.com | | Date Prepared: | September 29, 2015 | | II.<br>DEVICE | | | Trade Name:<br>Common Name: | ALEVE Direct Therapy<br>Transcutaneous electrical nerve stimulator for pain relief intended<br>for over the counter use (TENS) | | Classification Name:<br>Device Classification: | Stimulator, Nerve, Transcutaneous, Over-The-Counter<br>21 CFR 882.5890, Class II, NUH | | III.<br>PREDICATE DEVICE | | | Predicate Devices: | The Pain Pilot™ (a.k.a. Pain Pilot™) / WiTouch®<br>K120500<br>21 CFR 882.5890, Class II, NUH | |--------------------|----------------------------------------------------------------------------------------------| |--------------------|----------------------------------------------------------------------------------------------| #### IV. DEVICE DESCRIPTION The ALEVE Direct Therapy device is a battery powered transcutaneous electrical nerve stimulator (TENS) for applying an electrical current to electrodes on a patient's skin to relieve pain. The device reduces the perception of pain by electrically stimulating peripheral nerves across the skin. The design of the device limits the application for use to the anatomical site of the back. The device is comprised of a TENS unit with integral electrodes, one pair of replaceable electroconductive hydrogel pads, batteries for the remote and TENS unit, and a remote control. The user can turn the device on/off by pressing a button on the TENS unit. The hydrogel pads are adhesive and gently adhere the TENS unit to the user's skin on the lower back. There is also a remote control for the device, which the user turns on/off, and by which the user adjusts the intensity of stimulation. {4}------------------------------------------------ #### V. INDICATIONS FOR USE The ALEVE Direct Therapy is indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE This product is nearly identical to the predicate device except for minor differences: trade name, color of the non-conductive area of the integral silicone electrodes (change to ultramarine blue), non-significant modifications to the labeling and label designs, and an updated version of the software due to non-significant revisions made during the marketing of the PainPilot (Neubac) / WiTouch compared to the version cleared in 2012. All device components and accessories, basic unit characteristics, and output specifications are the same as the predicate device. A comparison of the Basic Unit Characteristics is provided below. {5}------------------------------------------------ # Basic Unit Characteristics Comparison | Characteristic | Bayer HealthCare, LLC<br>ALEVE Direct Therapy<br>Subject of this Submission | Hollywog Pain Pilot<br>(Neubac) / WiTouch<br>K120500 | Comparison | |-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Power Source(s) | Battery Powered | Battery Powered | IDENTICAL | | Number, Size, and Type<br>of Batteries | 2 AAA 1.5v DC batteries (TENS device)<br>1 Lithium coin battery 3.0v DC (wireless remote) | 2 AAA 1.5v DC batteries (TENS device)<br>1 Lithium coin battery 3.0v DC (wireless remote) | IDENTICAL | | Average DC current<br>through electrodes when<br>device is on but no pulses<br>are being applied (µΑ) | 0 | 0 | IDENTICAL | | Number of Output<br>Modes | 1 | 1* | IDENTICAL | | Number of Output<br>Channels | 1 | 1 | IDENTICAL | | Regulated Current or<br>Regulated Voltage? | Voltage | Voltage | IDENTICAL | | Software/Firmware/<br>Microprocessor Control? | Yes | Yes | IDENTICAL | | Automatic Overload<br>Trip? | No | No | IDENTICAL | | Automatic No-Load<br>Trip? | No | No | IDENTICAL | | Automatic Shut Off? | Yes | Yes | IDENTICAL | | User Override Control? | Yes | Yes | IDENTICAL | | Indicator Display | On/Off Status? Yes<br>Low Battery? Yes<br>Voltage/Current Level? No | On/Off Status? Yes<br>Low Battery? Yes<br>Voltage/Current Level? No | IDENTICAL | | Timer Range | Nonadjustable<br>30 minutes 42 seconds | Nonadjustable<br>30 minutes 42 seconds | IDENTICAL | | Compliance with | ISO 14971:2007 | ISO 14971:2007 | Substantially Equivalent | | Characteristic | Bayer HealthCare, LLC<br>ALEVE Direct Therapy<br>Subject of this Submission | Hollywog Pain Pilot<br>(Neubac) / WiTouch<br>K120500 | Comparison | | Voluntary Standards | AAMI ANSI ES60601-<br>1:2005/(R)2012 And A1:2012 IEC 60601-1-2 Edition 2014-02 IEC 60601-1-11 Edition 1.0<br>2010-04 IEC 60601-2-10 Edition 1.0<br>2012-06 AAMI/ANSI/ISO 10993-<br>1:2009/(R) 2013 AAMI/ANSI/ISO 10993-5:2009<br>(R)2014 AAMI/ANSI/ISO 10993-10:2010 | IEC 60601-1: 1995 IEC 60601-1-2: 2001 IEC 60601-1-4:2000 IEC 60601-2-10: 2001 ISO 10993-1:2009 ISO 10993-5:2009 ISO 10993-10:2010 | Current device complies with<br>updated versions of standards. | | Compliance with 21 CFR<br>898 | N/A, device does not contain<br>electrode lead wires or patient cables. | N/A, device does not contain<br>electrode lead wires or patient<br>cables. | IDENTICAL | | Weight (lbs., oz.) | 4.8 oz.<br>w/ batteries included | 4.8 oz.<br>w/ batteries included | IDENTICAL | | Dimensions (in.) [W x H<br>x D] | 7.5 (w) x 3.5 (h) x 0.7 in (d) | 7.5 (w) x 3.5 (h) x 0.7 in (d) | IDENTICAL | | Patient Contacting<br>Materials including<br>Housing Materials and<br>Construction | Hydrogel (Gel pads)<br>ABS plastic enclosure (Housing)<br>Integral silicone electrodes with<br>conductive (carbon rubber) area and<br>non-conductive area with ultramarine<br>blue colorant (CAS # 57455-37-5) | Hydrogel (Gel pads)<br>ABS plastic enclosure (Housing)<br>Conductive (carbon) silicone<br>electrodes<br>Non-conductive area of the<br>integral silicone electrodes, of<br>green color | Substantially Equivalent. The<br>safety of this colorant has<br>been demonstrated (Section<br>12, Biocompatibility). A<br>difference in colorant does<br>not affect the performance<br>characteristics of the device,<br>and the difference does not<br>raise new questions of safety<br>or effectiveness. | {6}------------------------------------------------ {7}------------------------------------------------ ### VII. PERFORMANCE DATA Conformance with the following voluntary, FDA-recognized standards is provided in this 510(k) submission: - . Recognition Number 5-40: ISO 14971 Second Edition 2007-03-01, Medical devices application of risk management to medical devices. - . Recognition Number 19-4: AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (consolidated text) Medical electrical equipment part 1: General requirements for basic safety and essential performance. - . Recognition Number 19-8: IEC 60601-1-2 Edition 4.0 2014-02. Medical electrical equipment - part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic disturbances - requirements and tests. - Recognition Number 19-14: IEC 60601-1-11 Edition 1.0 2010-04, Medical electrical ● equipment - part 1-11: General requirements for basic safety and essential performance collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. - Recognition Number 17-11: IEC 60601-2-10 Edition 2.0 2012-06, Medical electrical equipment - part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. - Recognition Number 2-156: AAMI/ANSVISO 10993-1:2009/(R) 2013, Biological evaluation . of medical devices - part 1: Evaluation and testing within a risk management process. - Recognition Number 2-153, AAMI/ANSI/ISO 10993-5:2009/(R) 2014, Biological evaluation . of medical devices - part 5: Tests for in vitro cytotoxicity. - Recognition Number 2-173, AAMI/ANSI/ISO 10993-10:2010, Biological evaluation of ● medical devices - part 10: Tests for irritation and skin sensitization. # Performance Testing Data in support of the device waveform, and verification of output characteristics, was provided in this submission. # Software Verification and Validation Testing Software verification and validation testing was conducted, and documentation in accordance with a Moderate level concern device was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." # Usability A 15 subject usability study was conducted by Hollywog and reported to the FDA during the review of predicate K120500, based on feedback provided from the FDA during Pre-IDE submission correspondence. The submitter has a right of reference to this study, as the devices {8}------------------------------------------------ and labeling are sufficiently similar, in support of the usability of the ALEVE Direct Therapy device. ### VIII. CONCLUSIONS The electrical safety, EMC, biocompatibility, software verification and validation, basic unit characteristics, and output specifications information provided in the 510(k) submission is sufficient to demonstrate substantial equivalence to the predicate device. As the ALEVE Direct Therapy TENS device is nearly identical to the predicate device, with identical indications for use and essentially identical technological characteristics, the ALEVE Direct Therapy TENS device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...