Shenzhen Siken 3D Technology Development Co., Ltd.
Product Code
NFO · Neurology
Decision Date
Dec 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Multi-functional Facial Beauty Device is a handheld portable device for over the counter aesthetic use including facial stimulation through MC(micro-current) mode.
Device Story
Handheld, portable, battery-operated device for home aesthetic use; provides micro-current facial stimulation, LED light therapy (red/blue), hot compress, and vibration. Device consists of main unit, massage heads, and Type-C charging cable. User selects modes via buttons; intensity and mode indicators provide feedback. Micro-current stimulation is the primary medical function; other functions (LED, heat, vibration) are either non-medical or cleared under separate submissions. Device operates via internal rechargeable lithium battery. Healthcare provider involvement is absent; patient self-operates. Output is intended for cosmetic facial stimulation. Benefits include aesthetic facial improvement. Device includes automatic shut-off and is non-sterile/reusable.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by biocompatibility testing (ISO 10993-5, -10, -23), electrical safety/EMC testing (IEC 60601-1, -1-2, -1-11, -2-10), and battery safety (IEC 62133-2). Software verification and validation confirmed requirements met and hazards mitigated.
Technological Characteristics
Handheld, battery-powered (lithium) stimulator. Output: pulse-symmetric, biphasic, rectangular wave (1.6kHz). Max output: 32mA @ 500Ω. Patient-contacting materials compliant with ISO 10993-5, -10, -23. Electrical safety per IEC 60601-1, IEC 60601-2-10. Type BF protection. Includes LED (red/blue), vibration, and hot compress (35.5-39°C). Software-controlled with automatic shut-off.
Indications for Use
Indicated for over-the-counter aesthetic facial stimulation using micro-current (MC) mode in a home environment.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
K201782 — NuFace Trinity Plus Device · Carol Cole Company Dba Nuface · Jan 22, 2021
K201906 — Trinity ELE Plus and Trinity ELE Plus Pro · Carol Cole Company Dba Nuface · Feb 24, 2021
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
Shenzhen Siken 3D Technology Development Co., Ltd.
% Bing Huang
Registration specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518052
China
Re: K252146
Trade/Device Name: Multi-functional Facial Beauty Device (SKB-1703,SKB-1803,SKB-1909,SKB-2003 SKB-2109,SKB-2129,SKB-2208PRO, SKB-2209)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: November 12, 2025
Received: November 12, 2025
Dear Bing Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252146 - Bing Huang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252146 - Bing Huang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
Multi-functional Facial Beauty Device
Page 9 of 41
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252146 | ? |
| Please provide the device trade name(s). | | ? |
| Multi-functional Facial Beauty Device (SKB-1703,SKB-1803,SKB-1909, SKB-2003
SKB-2109,SKB-2129,SKB-2208PRO, SKB-2209) | | |
| Please provide your Indications for Use below. | | ? |
| Multi-functional Facial Beauty Device is a handheld portable device for over the counter aesthetic use including facial stimulation through MC(micro-current) mode. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s -510(k) Summary
# 510(k) Summary of K252146
"510(k) Summary" as required by 21 CFR Part 807.92.
Date: 2025-12-19
## I. Submitter
Shenzhen Siken 3D Technology Development Co., Ltd.
Room 1104, Building 2, Xinweirun High-tech Park, No. 162, Shajiang Road, Xitou Community, Shajiang Road, Songgang Street, Bao'an District, Shenzhen, Guangdong, China.
Post code: 518105
Tel.: +86-0755-27697523
Jing Quan Liu
Title: Management representative
Tel.: +86-0755-27697523
Email: 1550420556@qq.com
## II. Device
Name of Device: Multi-functional Facial Beauty Device
Model(s): SKB-1703,SKB-1803,SKB-1909,SKB-2003,SKB-2109,SKB-2129,SKB-2208PRO, SKB-2209
Common or Usual Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes
Classification Name: Transcutaneous electrical nerve stimulator for pain relief
Regulatory Class: II
Product Code: NFO
Regulation Number: 21 CFR 882.5890
## III. Predicate Device & Reference Device
Primary predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| Xtreem Pulse | PureLift Pro Edition | K230506 | June 21, 2023 |
Predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Cleared Date |
| --- | --- | --- | --- |
| Shenzhen Aozemei Technology Co. Ltd | Micro-current Facial Beauty Device | K243430 | December 19, 2024 |
| BELEGA Co., Ltd. | Beagank 4T Plus | K233010 | November 21, 2023 |
{5}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s -510(k) Summary
# IV. Device Description
The Multi-functional Facial Beauty Device is a portable, non-sterile, reusable device designed to achieve the aesthetic effect. It consists of main unit, charging cable and user manual. The device is supplied by internal rechargeable lithium battery, which can be recharged by external charger through the Type-C charging cable. The device is unusable when charging.
The device is only home environment use, which has some massage heads, and LED light mode (Red and /or Blue lights are output independently) to provide following functions.
a. Micro-current stimulation function.
b. Red LED irradiation function. (Cleared under K252142).
c. Blue LED irradiation.(Cleared under K252142).
d. Hot compress and vibration function.(The vibration is classified as class I and not need for 510K, and the hot compress $(39^{\circ}\mathrm{C} \pm 0.5^{\circ}\mathrm{C})$ is not for medical purpose).
e. Hot compress function.(The hot compress $(35.5^{\circ}\mathrm{C} \sim 38.5 \pm 0.5^{\circ}\mathrm{C})$ is not for medical purpose).
f. Vibration function(motor vibration).(The vibration is classified as class I and not need for 510K).
Each mode of the device is independent from all other modes of the Multi-functional Facial Beauty Device.
The Multi-functional Facial Beauty Device consists of a main unit, USB charging cable, components shown as following illustration: (Here takes model SKB-2109 and model SKB-2129 as examples)
SKB-2109:

1 Intensity indicator
2 Mode indicator
3 Intensity button
4 OFF/ON& Mode switch button
5 Massage head
6 LED light
7 Massage head
8 USB charging cable
{6}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s -510(k) Summary

SKB-2129:
## V. Indications for Use
Multi-functional Facial Beauty Device is a handheld portable device for over the counter aesthetic use including facial stimulation through MC(micro-current) mode.
## VI. Comparison of Technological Characteristics With the Predicate Device
Multi-functional Facial Beauty Device is compared with the following Predicate Devices in terms of intended use, design, specifications, and performance:
Page 3 of 13
{7}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | K252146 | K230506 | K243430 | K233010 | / |
| Trade name | Multi-functional Facial Beauty Device | PureLift Pro Edition | Micro-current Facial Beauty Device
Model: AM-810B, AM-810W, AM-812B, AM-812W | Beagank 4T Plus | / |
| Manufacturer | Shenzhen Siken 3D Technology Development Co., Ltd | Xtreem Pulse | Shenzhen Aozemei Technology Co., LTD | BELEGA Co., Ltd. | / |
| Regulation number | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | Same |
| Product code | NFO | NFO | NFO | NFO | Same |
| Device classification | II | II | II | II | Same |
| Prescription or OTC | OTC | OTC | OTC | OTC | Same |
| Indication for use/ Intended use | Multi-functional Facial Beauty Device is a handheld portable device for over the counter aesthetic use including facial stimulation through MC(micro-current) mode. | Intended for facial stimulation and indicated for over-the-counter cosmetic use. | Micro-current Facial Beauty Device is a handheld portable device for over the counter aesthetic use including facial use including facial and neck stimulation. | The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation. | Same |
| Handheld or stationary | Hand-held Type | Hand-held Type | Hand-held Type | Hand-held Type | Same |
Page 4 of 13
{8}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| Treatment area | Face | Face | Face and neck | Face and neck or body | Same |
| Number of output modes | 1 | 1 | 1 | 4 | Same |
| Number of out channels | 1 | 1 output channel | 1 | 1 | Same |
| Regulated current or regulated voltage | Regulated current | Regulated current | unknown | Voltage | Same |
| Software/Firmware/Microprocessor Control | Yes | Yes | Yes | Yes | Same |
| Automatic Shut off | Yes | Yes | unknown | Yes | Same |
| Indicator Display | Yes | Yes | unknown | unknown | Same |
| Timer range | 10 minutes | 10 minutes only | MS-Micro current stimulation mode: It is recommended to use the product 10-15 minutes at a time / 2-3 times a week. | Maximum 5 minutes | Same |
| Type of protection | Type BF | Type BF | unknown | Type BF | Same |
| On/off status | Yes | Yes | unknown | unknown | Same |
Page 5 of 13
{9}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s –510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| Waveform | Pulse-symmetric, biphasic, rectangular wave | Pulses Monophasic, alternating polarity | Biphasic pulsed square wave | Rectangle, biphasic asymmetric | Similar Note 1 |
| Shape | Rectangular Pulses | Rectangular Pulses | Square Pulses | Rectangle, biphasic asymmetric | Same |
| Maximum output voltage | 16Vpp @500Ω
28Vpp @ 2kΩ
37Vpp @ 10kΩ | 16Vpp(@500Ω)
27.0Vpp(@2kΩ)
38.4Vpp(@10kΩ) | 225mV @500Ω
900mV @ 2kΩ
3.6mV @ 10kΩ | Mode 1:
127mV (0.12V)@500Ω
515mV (0.51V)@2KΩ
1.83V@10KΩ
Mode 2:
110mV(0.11V)@500Ω
507mV(0.50V)@2KΩ
1.87V@10KΩ
Mode 3:
16.6V@500Ω
21.0V@2KΩ
22.5V@10KΩ
Mode 4:
15.3V@500Ω
21.2V@2KΩ
22.9V@10KΩ | Same |
Page 6 of 13
{10}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s –510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| Maximum output current | 32mA @500Ω
14mA @ 2kΩ
3.7mA @10kΩ | 7.7mA(@500Ω)
3.6mA(@2kΩ)
1.0mA(@10kΩ) | 450μA @500Ω
450μA @ 2kΩ
360μA @10kΩ | Mode 1:
0.19mA@500Ω
0.18mA@2KΩ
0.16mA@10KΩ
Mode 2:
0.45mA@500Ω
0.39mA@2KΩ
0.24mA@10KΩ
Mode 3:
30.8mA@500Ω
8.52mA@2KΩ
2.26mA@10KΩ
Mode 4:
31.4mA@500Ω
8.60mA@2KΩ
2.15mA@10KΩ | Similar Note 2 |
| Output tolerance | +/- 10% | +/- 1mA | +/-10% | +/-15% | Same |
| Pulse Width | 4μs | 4μs | 4ms | Mode 1: 265μs | Same |
Page 7 of 13
{11}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s –510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| | | | | Mode 2: 265μs
Mode 3: 274μs
Mode 4: 610μs | |
| Frequency (Hz) | 1.6kHz | 1.37kHz~1.73kHz | 57Hz | Mode 1: 3.80kHz
Mode 2: 3.80kHz
Mode 3: 3.65kHz
Mode 4: 1.64kHz | Similar Note 3 |
| Symmetrical phases | Yes | Not multiphasic | Yes | No | Same |
| Net Charge (μC per pulse train) | 0μC@500Ω | 0μC per pulse train | 0μC@500Ω | Mode 1: 0.025μC@500Ω
Mode 2: 0.025μC@500Ω
Mode 3: -1.3μC@500Ω
Mode 4: 0μC@500Ω | Same |
| Maximum Phase Charge (μC) | 0.0774μC@500Ω | 4.97μC@500Ω | 2.3619μC@500Ω | Mode 1: 0.059μC@500Ω
Mode 2: 0.025μC@500Ω
Mode 3: 2.2μC@500Ω
Mode 4: 0.29μC@500Ω | Difference Note 4 |
| Maximum current Density (mA/cm²) | 0.4mA/cm²@500Ω | 7.5mA/cm²@500Ω | AM-810B, AM-810W: 0.10mA/cm²@500Ω
AM-812B, AM-812W: | With a wet cotton pad (Purified water) | Difference Note 5 |
Page 8 of 13
{12}
Shenzhen Siken 3D Technology Development Co., Ltd.
510(k)s –510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| | | | 0.06mA/cm²@500Ω | Mode 1:
0.15~1.36mA/cm²@500Ω
Mode 2:
0.10~0.92mA/cm²@500Ω
Mode 3:
4.26~38.43mA/cm²@500Ω
Mode 4:
3.13~28.21mA/cm²@500Ω | |
| Maximum Power Density | 0.26mW/cm²@500Ω | 29250μW/cm² | AM-810B, AM-810W:
0.0117mW/cm²@500Ω
AM-812B, AM-812W:
0.0072mW/cm²@500Ω | With a wet cotton pad:
Mode 1:
11.39μW/cm²@500Ω
Mode 2:
33.87μW/cm²@500Ω
Mode 3:
1124.0μW/cm²@500Ω | Difference Note 5 |
Page 9 of 13
{13}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s -510(k) Summary
| Comparison Elements | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Remark |
| --- | --- | --- | --- | --- | --- |
| | | | | Mode 4: 556.96μW/cm²@500Ω | |
| Electrical safety | IEC 60601-1
IEC 60601-1-11
IEC 60601-1-2
IEC 60601-2-10
IEC 62133-2 | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-10
IEC 60601-1-11 | IEC 60601-1
IEC 60601-1-11
IEC 60601-1-2
IEC 60601-2-10
IEC 62133-2 | IEC 60601-1
IEC 60601-1-2 | Same |
| Biocompatibility feature | All patient contacting materials are complied with ISO 10993-5, ISO 10993-10 and ISO 10993-23 | ISO 10993-5
ISO 10993-10 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | Same |
**Note 1:**
Though the waveform of the subject device is little different from the primary predicate device, but it is the same as predicate device 1(K243430). Therefore, this difference will not any raise safety or effectiveness issues.
**Note 2:**
The maximum output current of the subject device is close to the mode 4 of predicate device 2 (K233010). The subject device has been tested for electrical safety as per IEC60601-1, IEC 60601-2-10, so this difference does not impact its safety and effectiveness.
**Note 3:**
Though the Frequency of the subject device is little different from the primary predicate device, however it is within the rang of the primary predicate device. So this difference will not any raise safety or effectiveness issues.
**Note 4:**
There is a difference in the maximum phase charge between the subject devices and the primary predicate device, however the maximum
Page 10 of 13
{14}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s –510(k) Summary
phase value is within the range of the predicate device 2. Therefor this difference will not any raise safety or effectiveness issues.
**Note 5:**
Regarding the Maximum current density and the Maximum power density, the maximum current density of the subject device ( 0.4mA/cm² ) is within the rang of predicate devices. And the 0.4mA/cm² is far less than the limit of 2mA/cm² specified in IEC 60601-2-10. The value of Maximum power density of subject device is within the rang of the predicate devices. The subject device has provide adequate electrical safety and performance testing, which were conducted according to applicable standards for nerve and muscle stimulators (including IEC 60601-1 and IEC 60601-2-10) as well as lab bench performance evaluation. Therefore, these differences will not any raise safety or effectiveness issues.
Page 11 of 13
{15}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s –510(k) Summary
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### 1) Biocompatibility Safety
The materials of the patient-directly contacting components of the subject device is performed the biocompatibility evaluation in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recommended by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation
### 2) Electrical Safety and EMC Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
- IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems.
- IEC 60601-2-10 Medical electrical equipment – Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
### 3) Software Verification and Validation
Software documentation consistent with **Basic Documentation** of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
## Summary
Based on the above performance as documented in this application, the subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.
## VIII. Conclusions
{16}
Shenzhen Siken 3D Technology Development Co., Ltd. 510(k)s –510(k) Summary
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, it can be concluded that the Multi-functional Facial Beauty Device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Page 13 of 13
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.