K111609 · Boston Brace Intl., Inc. · MVA · Aug 22, 2011 · Neurology
Device Facts
Record ID
K111609
Device Name
BOSTON BAND
Applicant
Boston Brace Intl., Inc.
Product Code
MVA · Neurology
Decision Date
Aug 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Boston Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Device Story
The Boston Band is a custom-fabricated cranial orthosis for infants (3-18 months). It functions by applying static pressure to prominent cranial regions while providing space for growth in flattened areas to improve symmetry. The device is prescribed by a physician and fitted by a clinician. Input data includes head shape captured via negative plaster cast or laser scanning (Fastscan or Creaform/Omega). The 3D model is modified using either manual plaster techniques or Rodin4D CAD software. The orthosis consists of a thermoformable plastic outer shell with a hypoallergenic polyethylene foam (Aliplast/Plastazote) inner liner and securing straps. It is used in a clinical setting. The device is worn by the infant to guide skull growth, with fit monitored by the clinician. It benefits patients by correcting cranial deformities, potentially reducing long-term aesthetic or developmental concerns associated with abnormal head shapes.
Clinical Evidence
No clinical data. Bench testing only. Accuracy and repeatability testing compared manual modification vs. Rodin4D CAD modification and evaluated the performance of the Fastscan Class I laser scanner. Results confirmed that CAD-modified models and Class I scanned models met predetermined acceptance criteria and were equivalent to traditional methods.
Indicated for infants 3-18 months old with nonsynostotic positional plagiocephaly (plagiocephalic, brachycephalic, scaphocephalic shapes) to improve cranial symmetry. Also indicated as adjunctive treatment for infants 3-18 months old post-surgical correction of synostosis with residual moderate to severe cranial deformities. Contraindicated in infants with craniosynostosis or hydrocephalus.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
K072862 — BOSTON-BAND CRANIAL REMODLING ORTHOSIS · Boston Brace Intl., Inc. · Jan 22, 2008
K021594 — COPC BAND · Center For Orthotic and Prosthetic Care,Llc · Nov 13, 2003
K213587 — MyCRO Band · Otto Bock Healthcare LP · Jun 9, 2022
K063098 — STATIC CRANIOPLASTY ORTHOSIS · Boston Brace Intl., Inc. · Apr 11, 2007
K223238 — STARband 3D · Orthomerica Products, Inc. · Apr 20, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K11609
Section 5-1
Boston Brace AUG 2 2 2011
### 510(k) Summary
The following summary is provided in pursuant to Section 513 (I)(3)(A) of the Federal Food, Drug and Cosmetic Act.
### 1) Applicant Information
- Applicant's Name and Address: Boston Brace International, Inc., 20 Ledin Drive, Avon, MA 02322, Telephone: (508) 588 - 6060, Fax: (800) 634 - 5048
- FDA Establishment Registration Number: 3001967835 ●
- Contact: James H. Wynne, CPO, Vice President, Director of Education, ● Telephone: (508) 588 - 6060, x-244
- Submission Correspondent: Same as Contact ●
- . Summary Date - Oct 1, 2010
### 2) Submission Information
- . Type: Traditional 510(k) Submission
- Proprietary Name: The Boston Band .
- . Common Name: Cranial Orthosis
- Classification: Class II (Special Controls); OAN, CFR 882.5970 .
- . Classification Name: Cranial Orthosis
- Predicate Devices: STARband Cranial Remolding Orthosis K082950 . Boston Band Cranial Remolding Orthosis - K072862 Hanger Cranial Band - K072566
# 3) Manufacturing Site:
20 Ledin Drive, Avon, MA 02322, Telephone: (800) 262-2235, Facsimile: (800) 634-5048
{1}------------------------------------------------
FDA Establishment Registration Number: 3001967835
# 4) Device Description
The Boston Band is a cranial orthosis used to treat abnormally shaped craniums in infants three to eighteen months of age. This condition is clinically known as Positional or Deformational Plagiocephaly. The orthosis contains the protruding aspects of the cranium in a static equilibrium while guiding the growth of the flattened-areas of the skull into the created spaces to improve proportion and symmetry. The Boston Band is only available if prescribed by a physician. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
The Boston Band is essentially the same device as released by K072862. It is an orthosis designed for each patient from a cast or scan of the infant's head. The mold, either plaster from a cast, or foam, carved from the scan, is modified and prepared for fabrication according to the instructions provided by the treating practitioner using plaster/CAD modification techniques. Each orthosis is composed of an outer shell of thermoformable plastic, a single layer of ½ inch or 5-6 layers of 1/8 inch, or a combination of 1 - 2 1/4 inch layers with 1/8 inch layers of hypoallergenic polyethylene foam (Aliplast/Plastazote) and a strap for securing the orthosis. It has a top and side opening. Optimum fit and alignment is insured and monitored by the same clinical practitioner.
# 5) Statement of Indications and Intended Use
Statement of Indications:
The Boston Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
{2}------------------------------------------------
#### Section 5- 3
Intended Use:
The Boston Band is design to treat infants with abnormally shape craniums from age three to eighteen months. It is available by prescription only. The most common head deformities are positional plagiocephaly, brachycephaly, and scaphocephaly. However, due to new surgical techniques for infants with craniosynostosis. post-surgical plagiocephaly, brachycephaly,and and scaphocephaly are a growing patient group.
#### 6) Summary of Technological Characteristics
The proposed changes involve: A) expansion of the indications for use to include infants of the same age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads, B) how the infants head is captured (class of laser change), C) the method to modify the 3- dimensional (3D) mode (CAD vs. hand modification), D) the use of different thicknesses of the foam inner liner (both aliplast and plastazote) and E) the use of an updated carver machine that is an industry standard. There have been some minor changes to the original K072862 that were determined to not require a 510(k) submission. The overall design of the cranial orthosis will remain the same. The following table illustrates the minor differences between the cleared device (K072862) and the device as it is currently marketed.
{3}------------------------------------------------
# Table 1 Comparison of Boston Band Cranial Remolding Orthosis (K072862) to proposed device: The Boston Band
Note: No changes will be made to the current device as a result of the proposed indications and shape capture change in this submission
| Feature | Predicate Device<br>K072862 | The Boston Band |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | Intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. | Intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads. |
| Materials | - outer shell of thermoformable plastic (1/4 or 3/16 copolymer plastic) | - outer shell of thermoformable plastic (1/4 or 3/16 copolymer plastic) |
| | - 5-6 layers of 1/8 inch hypoallergenic polyethylene foam (Aliplast) | - a single layer of ½ inch or 5-6 layers of 1/8 inch, or a combination of 1 - 2 1/4 inch layers with 1/8 inch layers of pelite hypoallergenic polyethylene foam or Aliplast |
| | - Velcro strap (pile) sticky back Velcro hook | Velcro strap (pile) sticky back Velcro hook |
| Product<br>Design | Cranial Orthosis -made to individual's specifications | Cranial Orthosis - made to individual's specifications |
| Production | -Model of the infants head from a negative impression (cast) of the infants head<br>-Scan of the infants head from a Fastscan hands free class II laser scanner<br>- Scan of the infants head using the Orthomerica Starscaner class I laser scanner<br>- Foam model of the infants head carved using an industry standard 5 axis scanner | -Model of the infants head from a negative impression (cast) of the infants head<br>-Scan of the infants head from a Fastscan hands free class II laser scanner<br>-Scan of the infants head using a Fastscan handheld class I laser scanner<br>- Scan of the infants head using the Creaform class I laser scanner sold as Ohio Willow Wood Omega/handycam<br>- Foam model of the infants head carved using an industry standard 5 axis scanner |
| Labeling | Predicate Device (K072862) | The Boston Band |
| Indications | Infants from three to eighteen months of age,<br>with nonsynostotic positional plagiocephaly,<br>including infants with plagiocephic,<br>brachycephalic and scaphocephalic patterned<br>head shapes | Same and to include the<br>adjunctive use for infants from<br>three to eighteen months of age<br>whose synostosis has been<br>surgically corrected, but who<br>still have moderate to severe<br>cranial deformities including<br>plagiocephalic, brachycephalic,<br>and scaphocephalic shaped<br>heads. |
| Contraindications | Infants with Craniosynostosis or hydrocephalus | Same |
| Warnings | Evaluate head circumference measurements<br>and neurological status at intervals appropriate<br>to the infant's age and rate of growth. Describe<br>to caregiver steps that should be taken to<br>reduce the potential restrictions of cranial<br>growth and possible impairments to brain<br>growth and development. A clinician or<br>caregiver must evaluate the patient's skin at<br>frequent intervals. | Same |
| Precautions | If the positional plagiocephaly is associated<br>with torticollis, the torticollis must also be<br>treated. Evaluate the device's structural<br>integrity and fit it according to fitting<br>instructions. Wear and care guide to be<br>reviewed and provided to caregiver.(copy<br>attached) | Same |
| Adverse Effects | This device may cause skin irritations or<br>breakdown | Same |
| Instructions for<br>Use | Wear and care guide provided to caregiver | Same |
| Expiration date | When the infant outgrows the cranial helmet or<br>the orthosis is discontinued for any reason, the<br>referring physician is contacted. If another<br>helmet is requested by the family to try to<br>further correction, refer back to their physician<br>for reevaluation and a new prescription. | Same |
| Feature | Predicate Device<br>(K082950) | The Boston Band |
| Intended<br>Use | The STARband is intended for medical<br>purposes for use on infants from three to<br>eighteen months of age, with moderate to<br>severe non-synostotic positional<br>plagiocephaly, including infants with<br>plagiocephalic, brachycephalic, and<br>scaphocephalic shaped heads by applying<br>mild pressure to prominent regions of the<br>infant's cranium in order to improve<br>cranial symmetry and/or shape. The<br>device is also indicated for adjunctive use<br>for infants from three to eighteen months<br>of age whose synostosis has been<br>surgically corrected but who still have<br>moderate to severe cranial deformities<br>including plagiocephalic brachycephalic<br>and scaphocphalic shaper heads | Intended for medical purposes to passively<br>hold prominent cranial regions of an<br>infant's skull in order to improve cranial<br>proportion and symmetry in infants from<br>three to eighteen months of age, with<br>nonsynostotic positional plagiocephaly,<br>including infants with plagiocephic,<br>brachycephalic and scaphocephalic<br>patterned head shapes. The device is also<br>indicated for adjunctive use for infants<br>from three to eighteen months of age<br>whose synostosis has been surgically<br>corrected, but who still have moderate to<br>severe cranial deformities including<br>plagiocephalic-, brachycephalic-, and<br>scaphocephalic-shaped heads. |
| Materials | - outer shell of .156 copoly plsitc)<br>- An inner liner of $½$ " pelite<br>polyethylene foam or $½$ Aliplast<br>- Strap of 1 $½$ " Dacron<br>- A 1 $½$ " chafe buckle<br>- Large Flange, Blind Rivet<br>- A Gap Block made from $½$ ' firm<br>pelite polyethylene foam<br>- A nylon washer | outer shell of thermoformable plastic (1/4 or<br>3/16 copolymer plastic)<br>- a single layer of $½$ inch or 5-6 layers of 1/8<br>inch, or a combination of 1 - 2 $1/4$ inch layers<br>with 1/8 inch layers of pelite hypoallergenic<br>polyethylene foam or Aliplast<br>- Velcro strap (pile) sticky back Velcro hook |
| Product<br>Design | Cranial orthosis - made to individual's<br>specifications - approx 6 oz in weight | Cranial orthosis -made to individual's specifications |
| Production | - Form orthosis from a positive mold<br>of infant's head<br>- Positive mold is formed based upon<br>measurements of the infant's head<br>taken by the STARscanner the<br>OWW Omega Scanner from which<br>a 3-dimensional image is made or<br>from a traditional plaster case<br>- The 3-ddimensional image is used to<br>produce a positive mold using 5-axis<br>routing machine | Model of the infants head from a negative<br>impression (cast) of the infants head<br>-Scan of the infants head from a Fastscan hands free<br>class II laser scanner<br>-Scan of the infants head using a Fastscan handheld<br>class I laser scanner<br>- Scan of the infants head using the Creaform class 1<br>laser scanner sold as Ohio Willow Wood<br>Omega/handycam<br>- Foam model of the infants head carved using an<br>industry standard 5 axis scanner |
| Labeling | Predicate Device (K082950) | The Boston Band |
| Indications | The STARband is intended for medical purposes<br>for use on infants from three to eighteen months<br>of age, with moderate to severe non-synostotic<br>positional plagiocephaly, including infants with<br>plagiocephalic, brachycephalic, and<br>scaphocephalic shaped heads by applying mild<br>pressure to prominent regions of the infant's<br>cranium in order to improve cranial symmetry<br>and/or shape. The device is also indicated for<br>adjunctive use for infants from three to eighteen<br>months of age whose synostosis has been<br>surgically corrected but who still have moderate<br>to severe cranial deformities including<br>plagiocephalic brachycephalic and scaphocephalic<br>shaper heads | Same and to include the<br>adjunctive use for infants from<br>three to eighteen months of age<br>whose synostosis has been<br>surgically corrected, but who still<br>have moderate to severe cranial<br>deformities including<br>plagiocephalic, brachycephalic,<br>and scaphocephalic shaped heads. |
| Contraindications | Infants with Craniosynostosis or hydrocephalus | Same |
| Warnings | Evaluate head circumference measurements and<br>neurological status at intervals appropriate to the<br>infant's age and rate of growth. Describe to<br>caregiver steps that should be taken to reduce the<br>potential restrictions of cranial growth and<br>possible impairments to brain growth and<br>development. A clinician or caregiver must<br>evaluate the patient's skin at frequent intervals. | Same |
| Precautions | If the positional plagiocephaly is associated with<br>torticollis, the torticollis must also be treated.<br>Evaluate the device's structural integrity and fit it<br>according to fitting instructions. Wear and care<br>guide to be reviewed and provided to<br>caregiver.(copy attached) | Same |
| Adverse Effects | This device may cause skin irritations or<br>breakdown | Same |
| Instructions for<br>Use | Wear and care guide provided to caregiver | Same |
| Expiration date | When the infant outgrows the cranial helmet or<br>the orthosis is discontinued for any reason, the<br>referring physician is contacted. If another<br>helmet is requested by the family to try to further<br>correction, refer back to their physician for<br>reevaluation and a new prescription. | Same |
{4}------------------------------------------------
# Table 2 Comparison of Boston Band Cranial Remolding Orthosis (K072862) to proposed device: The Boston Band
{5}------------------------------------------------
# Table 3 Comparison of STARband Cranial Remolding Orthosis - K082950 to proposed device: The Boston Band
Note: No changes will be made to the current device as a result of the proposed indications and shape capture change in this submission
{6}------------------------------------------------
Table 4 Comparison of STARband Cranial Remolding Orthosis - K082950 to proposed device: The Boston Band
{7}------------------------------------------------
# Table 5 Comparison of Hanger Cranial Band (K072566) to proposed device: The Boston Band
| Note: No changes will be made to the current device as a result of the proposed |
|---------------------------------------------------------------------------------|
| indications and shape capture change in this submission |
| Feature | Predicate Device<br>K072566 | The Boston Band |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use | Intended for medical purposes to passively<br>hold prominent cranial regions of an infant's<br>skull in order to improve cranial proportion<br>and symmetry in infants from three to<br>eighteen months of age, with nonsynostotic<br>positional plagiocephaly, including infants<br>with plagiocephic, brachycephalic and<br>scaphocephalic patterned head shapes. | Intended for medical purposes to passively hold<br>prominent cranial regions of an infant's skull in<br>order to improve cranial proportion and symmetry<br>in infants from three to eighteen months of age,<br>with nonsynostotic positional plagiocephaly,<br>including infants with plagiocephic,<br>brachycephalic and scaphocephalic patterned head<br>shapes. The device is also indicated for adjunctive<br>use for infants from three to eighteen months of<br>age whose synostosis has been surgically<br>corrected, but who still have moderate to severe<br>cranial deformities including plagiocephalic-,<br>brachycephalic-, and scaphocephalic-shaped<br>heads. |
| Materials | - outer shell of thermoformable plastic<br>- hypoallergenic polyethylene foam<br>- Velcro strap (pile) sticky back Velcro<br>hook | - outer shell of thermoformable plastic (1/4 or<br>3/16 copolymer plastic)<br>- a single layer of ½ inch or 5-6 layers of 1/8<br>inch, or a combination of 1 - 2 1/4 inch<br>layers with 1/8 inch layers of<br>hypoallergenic polyethylene foam<br>- Velcro strap (pile) sticky back Velcro hook |
| Product<br>Design | Cranial Orthosis -made to individual's<br>specifications | Cranial Orthosis -made to individual's<br>specifications |
| Production | - -Scan of the infants head from a Fastscan<br>hands free class II laser scanner<br>- Foam model of the infants head carved<br>using an industry standard 5 axis scanner | - Model of the infants head from a negative<br>impression (cast) of the infants head<br>-Scan of the infants head from a Fastscan hands<br>free class II laser scanner<br>-Scan of the infants head using a Fastscan<br>handheld class I laser scanner<br>- Scan of the infants head using the Creaform<br>class I laser scanner sold as Ohio Willow Wood<br>Omega/handycam<br>- Foam model of the infants head carved using an<br>industry standard 5 axis scanner |
| Labeling | Predicate Device<br>K072566 | The Boston Band |
| | | |
| Indications | Intended for medical purposes to apply<br>static or gentle pressure to prominent<br>regions of an infant's cranium to<br>improve cranial symmetry or shape. To<br>treat infants from three to eighteen<br>months of age with moderate to severe<br>non-synostotic positional plagiocephaly,<br>including plagiocephalic-,<br>brachycephalic-, scaphocephalic-shaped<br>heads. | Intended for medical purposes to passively<br>hold prominent cranial regions of an infant's<br>skull in order to improve cranial proportion<br>and symmetry in infants from three to<br>eighteen months of age, with nonsynostotic<br>positional plagiocephaly, including infants<br>with plagiocephalic, brachycephalic and<br>scaphocephalic patterned head shapes. The<br>device is also indicated for adjunctive use<br>for infants from three to eighteen months of<br>age whose synostosis has been surgically<br>corrected, but who still have moderate to<br>severe cranial deformities including<br>plagiocephalic-, brachycephalic-, and<br>scaphocephalic-shaped heads. |
| Contraindications | Infants with Craniosynostosis or<br>hydrocephalus | Same |
| Warnings | Evaluate head circumference<br>measurements and neurological status at<br>intervals appropriate to the infant's age<br>and rate of growth. Describe to caregiver<br>steps that should be taken to reduce the<br>potential restrictions of cranial growth<br>and possible impairments to brain<br>growth and development. A clinician or<br>caregiver must evaluate the patient's<br>skin at frequent intervals. | Same |
| Precautions | If the positional plagiocephaly is<br>associated with torticollis, the torticollis<br>must also be treated. Evaluate the<br>device's structural integrity and fit it<br>according to fitting instructions. Wear<br>and care guide to be reviewed and<br>provided to caregiver.(copy attached) | Same |
| Adverse Effects | This device may cause skin irritations or<br>breakdown | Same |
| Instructions for<br>Use | Wear and care guide provided to<br>caregiver | Same |
| Expiration date | When the infant outgrows the cranial<br>helmet or the orthosis is discontinued for<br>any reason, the referring physician is<br>contacted. If an other helmet is requested<br>by the family to try to further correction,<br>refer back to their physician for<br>reevaluation and a new prescription. | Same |
{8}------------------------------------------------
Table 6 Comparison of Hanger Cranial Band (K072566) to proposed device: The Boston Band
{9}------------------------------------------------
### 7) Summary and Conclusions of Non-Clinical Performance Data
The Boston Band Cranial Remolding Orthosis has been marketed in the US since its original clearance in 2007. Since that time advances have been made in surgical techniques to treat craniosynostosis, scanning technologies, computer aided design (CAD) software, and multi-axis carves.
The Fastscan hands free class II laser scanner was cleared in K072862. Fastscan now offers hands free class I laser scanner. With the laser scanners the safety issue is the class of laser. Like the predicate device (K082950), this class I laser is inherently safe for use without eye protection under all normal operating conditions. The effectiveness of this scanning device was evaluated through accuracy and repeatability testing. Positive foam models were scanned using the Fastscan hands free class I laser scanner. The scanned models were then carved and measured to compare to the original model. Each device met the predetermined acceptance criteria and was found acceptable.
Rodin4D CADCAM software ( http://www.rodin4d.com/en) is a 3D rectification software for manufacturing orthopedic devices. Currently carved positive models are modified by hand according to the needs of the patient. The software has now developed to where these modifications can be done in CAD. The original scanned model is saved and each step of modification is recorded in the software. The modifications are still to the needs of the patient and follow the same steps as if done by hand. The difference is that each step is now saved in the software. The effectiveness of this modification software was evaluated through accuracy and repeatability testing. A comparison was made between hand modified and Rodin4D modifications of the same models. Four models were selected. Hand modifications were done on carved models and CAD modifications were done on the electronic models. The electronic models were then carved. All models where then measured and compared. The results showed that both modification techniques produced an acceptable model for fabrication. Practitioners having CAD software will have the ability to modify the model electronically according to the needs of the patient.
Since the time of the predicated device approval (K072862) the 3-axis carver has been upgraded to remain current with carving machinery. The only major change is the speed of carving.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circle around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Boston Brace International, Inc. c/o James H. Wynne Vice President. Direction of Education 20 Ledin Drive Avon, MA 02322
Re: K111609
AUG 2 2 2011
Trade/Device Name: Boston Band Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial orthosis Regulatory Class: Class II Product Code: MV A Dated: May 24, 2011 Received: June 9, 2011
Dear Mr. Wynne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements; including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{11}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{12}------------------------------------------------
[Type text]
Boston Brace
20 Ledin Drive
Avon, MA 02322
T: 800 262.2235
T: 508.588.6060
F: 800.634 5048
F: 508.587.8119
www.bostonbrace.com
# Indications for Use Statement
510(k) Number (if known):
KIII 609
Device Name: Boston Band
Indications for Use:
The Boston Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial proportion and symmetry in infants from three to eighteen months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephic, brachycephalic and scaphocephalic patterned head shapes. The device is also indicated for adjunctive use for infants from three to eighteen months of age whose synostosis has been surgically corrected, but who still have moderate to severe cranial deformities including plagiocephalic-, and scaphocephalic-shaped heads.
AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
JOE HUTTER
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
KIII 609 510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.