K181382 · St Jude Medical · GXR · Jul 20, 2018 · Neurology
Device Facts
Record ID
K181382
Device Name
GuardianTM Burr Hole Cover System
Applicant
St Jude Medical
Product Code
GXR · Neurology
Decision Date
Jul 20, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.5250
Device Class
Class 2
Indications for Use
The GuardianTM burr hole cover system is intended for use following cranial surgery as an implantable 14-mm (0.55-in) burr hole cover for the skull. It can also be used to secure a lead with a 1.29-mm (0.051-in) or 1.39-mm (0.055-in) diameter.
Device Story
Implantable burr hole cover system; used post-cranial surgery to close 14-mm burr holes; secures compatible leads (1.29-mm or 1.39-mm diameter). System consists of three components: base, clip, and cover. Base features grooved slots for lead placement; clip secures lead within base; cover snaps onto base to lock lead and close hole. Used by neurosurgeons in clinical settings. Device is MR Conditional; labeling updated to reflect safety in MR environment. Benefits include secure closure of cranial access site and stabilization of implanted leads.
Clinical Evidence
No clinical data. Bench testing performed per FDA guidance 'Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment' to support MR Conditional status.
Technological Characteristics
Passive implantable burr hole cover; 14-mm diameter; three-piece assembly (base, clip, cover). Mechanical locking mechanism for lead securement. MR Conditional.
Indications for Use
Indicated for patients undergoing cranial surgery requiring a 14-mm burr hole cover; optionally used to secure implanted leads of 1.29-mm or 1.39-mm diameter.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
July 20, 2018
St Jude Medical Rebecca Brunson Regulatory Affairs Specialist 6901 Preston Road Plano, Texas 75024
Re: K181382
Trade/Device Name: Guardian™ Burr Hole Cover System Regulation Number: 21 CFR 882.5250 Regulation Name: Burr Hole Cover Regulatory Class: Class II Product Code: GXR Dated: May 23, 2018 Received: May 25, 2018
Dear Ms. Brunson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181382
Device Name Guardian Burr Hole Cover System
Indications for Use (Describe)
The Guardian burr hole cover system is intended for use following cranial surgery as an implantable 14-mm (0.5-in) burr hole cover for the skull. It can also be used to secure a lead with a 1.29-mm (0.051-in) or 1.39-mm (0.055-in) diameter.
Type of Use (Select one or both, as applicable)
| <span style="font-family:Wingdings;"> </span> <span style="font-family:Arial;">Prescription Use (Part 21 CER 801 Subpart D)</span> |
|------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family:Wingdings;"></span> <span style="font-family:Arial;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| 510(k) Summary | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K181382 |
| Submitter Information: | |
| Date Prepared: | May 23, 2018 |
| Submitter Name &<br>Address: | St. Jude Medical<br>6901 Preston Road<br>Plano, TX 75024<br>USA |
| Contact Person: | Rebecca Brunson<br>Regulatory Affairs Specialist<br>Phone (972) 526-4658<br>Fax (855) 902-0767<br>Becky.Brunson@abbott.com |
| Device Information: | |
| Name of Device | GuardianTM Burr Hole Cover System |
| Common Name: | Cover, Burr Hole |
| Regulatory Class | II |
| Classification Name: | 882.5250 Burr Hole Cover |
| Predicate Device: | K152342: GuardianTM Burr Hole Cover System |
| Purpose of Submission | This Abbreviated 510(k) premarket notification is submitted to<br>obtain clearance for the modification to the device labeling for<br>the Guardian Burr Hole Cover System to include the “MR<br>Conditional” statement in accordance with the FDA Guidance,<br>“Establishing Safety and Compatibility of Passive Implants in<br>the Magnetic Resonance (MR) Environment”. |
| Device Description: | The GuardianTM burr hole cover system is used to close a<br>cranial burr hole and secure an implanted, compatible lead,<br>when applicable. The burr hole cover system is nonpyrogenic<br>and has three main features: base, clip, and cover. The base is<br>intended for burr holes with a 14-mm (0.55-in) diameter. It<br>contains two grooved slots to hold a lead in place. The clip fits<br>into the base to hold the lead. The locking mechanism<br>temporarily holds a lead in place before the burr hole cover is<br>secured. The cover snaps onto the base, closing the burr hole<br>and locking a lead in place. |
| Intended Use:<br>(Indications for Use) | The GuardianTM burr hole cover system is intended for use<br>following cranial surgery as an implantable 14-mm (0.55-in)<br>burr hole cover for the skull. It can also be used to secure a<br>lead with a 1.29-mm (0.051-in) or 1.39-mm (0.055-in)<br>diameter |
| Summary on Non-<br>Clinical Testing | Non-clinical testing following the FDA Guidance,<br>"Establishing Safety and Compatibility of Passive Implants in<br>the Magnetic Resonance (MR) Environment", demonstrated<br>that the GuardianTM Burr Hole Cover System is “MR<br>Conditional”. |
| Statement of Equivalence | The GuardianTM Burr Hole Cover System is substantially<br>equivalent to the predicate device in which the basic design<br>features and intended uses are the same. The difference, the<br>MR Conditional labeling, is supported by performance data.<br>Based on the indications for use, technological characteristics,<br>and the summary of data submitted, St. Jude Medical has<br>determined that the proposed device is substantially equivalent<br>to the currently marketed predicate device. |
{4}------------------------------------------------
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.