K252951 · Neurofield, Inc. · QJQ · Dec 31, 2025 · Neurology
Device Facts
Record ID
K252951
Device Name
Genesis Sleep
Applicant
Neurofield, Inc.
Product Code
QJQ · Neurology
Decision Date
Dec 31, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Genesis Sleep is a non-invasive, clinical- and home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older.
Device Story
Genesis Sleep is a non-invasive, prescription-only cranial electrotherapy stimulator (CES) for chronic insomnia. It delivers electrical vestibular nerve stimulation (VeNS) via two self-adhesive electrode pads placed on the skin behind the ears (mastoid processes). The device is battery-powered and controlled via PC-based software, allowing users to adjust pulse intensity up to 1 mA. It operates by applying a symmetrical biphasic rectangular waveform to the vestibular nerves. The device is used by patients in clinical or home settings, typically while sitting before sleep. Healthcare providers use the output to manage insomnia symptoms; the device provides a non-pharmacological therapeutic option. Safety features include automatic overload monitoring and shut-off. The device is not wearable and cannot be recharged during stimulation.
Clinical Evidence
No new clinical data; relies on leveraged data from the predicate (Modius Sleep, K230826). The predicate's pivotal randomized, double-blind, sham-controlled trial (n=149) evaluated chronic insomnia treatment. Results showed a mean ISI score decrease of 5.80 in the treatment group vs. 3.52 in sham (p=0.005 between-group). Adverse events were minor, self-resolving, and comparable between groups (e.g., headache, peripheral vision flashes).
Indicated for the treatment of chronic insomnia in adults aged 22 and older. For clinical and home use.
Regulatory Classification
Identification
A cranial electrotherapy stimulator is a prescription device that applies electrical current that is not intended to induce a seizure to a patient's head to treat psychiatric conditions.
Special Controls
In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:
*Classification.* (1) Class II (special controls) when intended to treat insomnia and/or anxiety. The special controls for this device are:(i) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device to treat insomnia and/or anxiety.
(ii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iii) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC) in its intended use environment.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge, and energy, must be fully characterized and verified.
(vi) The labeling for the device must include the following:
(A) The intended use population and the intended use environment;
(B) A warning that patients should be monitored by their physician for signs of worsening;
(C) A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur;
(D) A warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur;
(E) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(F) Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions. This information must be based on the results of clinical studies for the device;
(G) A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and
(H) Information on validated methods for reprocessing any reusable components between uses.
(vii) Cranial electrotherapy stimulator devices marketed prior to the effective date of this reclassification must have an amendment submitted to the previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
(2) Class III (premarket approval) when intended to treat depression.
(c) Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 19, 2020, for any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 19, 2020, been found to be substantially equivalent to any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
K070412 — ELEXOMA MEDIC · Redplane AG · May 21, 2008
K060158 — NET-2000 MICROCURRENT STIMULATOR · Auri-Stem Medical, Inc. · Oct 13, 2006
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 31, 2025
Neurofield, Inc.
% Tom Renner
Quality, Efficiency & Regulatory Affairs Consultant
Vision 28
915 SW Rimrock Way
Suite 201 Pmb 402
Redmond, Oregon 97756
Re: K252951
Trade/Device Name: Genesis Sleep
Regulation Number: 21 CFR 882.5800
Regulation Name: Cranial electrotherapy stimulator
Regulatory Class: Class II
Product Code: QJQ
Dated: September 15, 2025
Received: September 16, 2025
Dear Tom Renner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252951 - Tom Renner
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252951 - Tom Renner
Page 3
Sincerely,
Heather L. Dean -S
for Pamela Scott
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252951 | |
| Device Name Genesis Sleep | |
| Indications for Use (Describe) Genesis Sleep is a non-invasive, clinical- and home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
# Contact Details
## Manufacturer
NeuroField, Inc.
1836 State St
Santa Barbara CA 93101 United States
Phone: 7608724200
Contact: Mr. Nicholas Dogris
Email: drdogris@neurofield.com
## Regulatory Correspondent
Vision28
915 SW Rimrock Way
STE 201 PMB 402
Redmond OR 97756 United States
Phone: (541) 359-5939
Contact: Mr. Tom Renner
Email: tom@vision28.com
## Device Name
Trade Name: Genesis Sleep
Common Name: Cranial electrotherapy stimulator
Classification Name: Cranial Electrotherapy Stimulator To Treat Insomnia And/Or Anxiety
Regulation Number: 882.5800
Product Code: QJQ
## Legally Marketed Predicate Devices
510(k) Number: K230826
Device Name: Modius Sleep
Product Code: QJQ
## Device Description Summary
Genesis Sleep is a non-invasive transdermal neurostimulation device to treat chronic insomnia. The device utilizes a technology known as electrical vestibular nerve stimulation (VeNS). It is indicated for clinical- and home-use to treat chronic insomnia in adults aged 22 and older.
It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1 mA) to the skin behind the ears, over the mastoid processes. The
{5}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
delivery of this neurostimulation is through two single-use self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area).
The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse which stimulates the vestibular nerve. When the device is not being used, the battery can be charged using the charging station provided. Therefore, it is not physically possible to recharge the battery while the device is in use in stimulation mode.
The Genesis Sleep device exerts its therapeutic effect through electrical stimulation of the vestibular nerves. Electrical vestibular nerve stimulation (VeNS) is a non-invasive therapeutic method that applies pulsed, alternating microcurrent transcutaneously to the vestibular nerves via stimulation pads.
Genesis Sleep delivers this neurostimulation through two self-adhesive stimulation pads which are placed on the skin behind each ear (mastoid processes). When turned on, the device delivers a small electrical impulse which can be adjusted up or down by the user.
## Substantial Equivalence Discussion
The comparison between the proposed Genesis Sleep device and the predicate Modius Sleep (K230826) below consists of a series of tables followed by explanations of the similarities and differences described in each table.
## I. Summary of Substantial Equivalence
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Indications for Use | Genesis Sleep is a non-invasive, clinical- and home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older. | Modius Sleep is a non- invasive, home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older. | The indications for use are materially the same. The proposed device indications include clinical use, whereas the predicate does not. Given that the labeling requirements for effective home use exceed those of clinical use, and considering that the |
{6}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| | | | device is prescription use only, this does not constitute a new intended use. |
| Environment | Clinical and Home | Home | *see Indications comment above. |
| Regulatory Class | Class II | Class II | Identical. |
| Classification Name | 882.5800 - Cranial electrotherapy stimulator | 882.5800 - Cranial electrotherapy stimulator | Identical. |
| Product Code | QJQ | QJQ | Identical. |
| Target Population | Adults 22 and older | Adults 22 and older | Identical. |
| Waveform | Symmetrical Biphasic Rectangular Wave | Symmetrical Biphasic Rectangular Wave | Identical. |
| Current Intensity Range | 0 μA - 1000 μA | 0 μA - 1000 μA | Identical. |
| Pulse Width Range | 1 s | 1 s | Identical. |
| Number of electrodes | 2 | 2 | Identical. |
| Electrode placement | Mastoid | Mastoid | Identical. |
| Power Source | 12 V Li-Ion Battery | 3.75 V Lithium Polymer Battery | |
| Frequency | 0.25 Hz | 0.25 Hz | Identical. |
| Treatment Range | 30 min | 30 min | Identical. |
| Unit Controls | PC-based Software | Built into the device | Different. |
| Dimensions | 16.8 cm x 17.5 cm x 6.1 cm | 16.5 cm x 15.1 cm x 6.6 cm | Similar. |
| Enclosure | Extruded Aluminum | Plastic | Different. |
## Discussion of Similarities and Differences
The two systems are identical in the following parameters:
- Regulatory Class
- Classification Name
- Product Code
- Target population
- Waveform
Page 3 of 15
{7}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
- Current intensity range
- Pulse width range
- Number of electrodes
- Electrode placement
The two systems differ in the following ways:
- The proposed Genesis Sleep indications include clinical use, whereas the predicate does not. Given that the labeling requirements for effective home use exceed those of clinical use, and considering that both devices are prescription use only, this does not constitute a new intended use.
- Both systems have been tested for safety and EMC. The proposed Genesis Sleep is also tested for use in the home healthcare environment.
- Although they have similar overall dimensions, the form factors of the two devices are different. The proposed Genesis Sleep is not intended to be worn.
- The enclosures are made of different materials.
- The controls of the proposed Genesis Sleep are not built into the device but are rather implemented via PC-based software.
These differences are minor, and do not materially affect their substantial equivalence with respect to technological basis or use.
## II. Safety Parameters
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Electrical Safety | Complies with IEC 60601-1. Additionally, complies with IEC 60601-1-11. | Complies with IEC 60601-1. | The proposed device and predicate devices are identical with regards to 60601-1. The Genesis Sleep has additionally been tested to be used in the home environment. |
| EMC | Complies with IEC 60601-1-2. | Complies with IEC 60601-1-2. | Identical. |
| Software LOC/Documentation | Basic. | Moderate. | The proposed device and predicate devices are different. |
{8}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Software V&V | Complies with FDA Guidance Requirement. | Complies with FDA Guidance Requirement. | Identical. |
| Biocompatibility | Complies with ISO 10993. | Complies with ISO 10993. | Identical. |
## Discussion of Similarities and Differences
The two systems are identical with respect to the following safety parameters:
- IEC 60601-1 compliance.
- IEC 60601-1-2 compliance.
- Software V&V compliance.
- Biocompatibility compliance.
The two systems differ in the following ways:
- The proposed Genesis Sleep has additionally been tested to the IEC 60601-1-11 standard.
- The Software LOC is different, reflecting evolving FDA guidance and terminology.
These differences are minor, and do not materially affect their substantial equivalence with respect to technological basis or use.
## III. Technical Parameters
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Power Source | 12 V Li-Ion Battery | 3.75 V Lithium Polymer Battery | Differs in supplied voltage, but both use lithium batteries. |
| Method of Line Current Isolation | DC:DC converter | DC:DC transformer | Same. Both use transformers for isolation. |
| Patient Leakage Current (as per ANSI/AAMI 60601-1) | 0.8 μA | 0 μA | Higher leakage current, but still within the allowable range. |
| Normal Condition (μA) | 0-1000 μA | 0-1000 μA | Identical. |
{9}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Single Fault Condition (μA) | 0.54 μA | 2.3 μA | Better than predicate. |
| Average DC current through the electrodes when the device is on but no pulses are being delivered (μA) | 0 μA | 0 μA | Identical. |
| Number of output channels | 1 | 1 | Identical. |
| If more than one channel, is the stimulus delivered to each channel synchronous or alternating between each channel? | N/A | N/A | Identical. |
| If more than one channel, describe method of channel isolation | N/A | N/A | Identical. |
| Software/Firmware/Microprocessor Control | Yes | Yes | Identical. |
| Automatic Overload Trip? | Yes | No | Better than predicate. Software continuously monitors current when delivering pulses |
| Automatic Shut Off? | Yes | Yes | Identical. |
| User Override Control? | Yes | Yes | Identical. |
| Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | Identical. |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular | Identical. |
Page 6 of 15
{10}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Maximum Output Voltage (volts) (+/-1%) | 500 mV @ 500 Ω
2 V @ 2 kΩ
10 V @ 10 kΩ | 500 mV @ 500 Ω
2 V @ 2 kΩ
10 V @ 10 kΩ | Identical. |
| Maximum Output Current (specify units) (+/-2%) | 1000 μA @ 500 Ω, 2 kΩ, and 10 kΩ | 1000 μA @ 500 Ω, 2 kΩ, and 10 kΩ | Identical. |
| Duration of primary (depolarizing) phase (msec) | 4000
(4 seconds) | 4000
(4 seconds) | Identical. |
| Pulse Duration (msec) | 2000
(2 second) | 2000
(2 second) | Identical. |
| Frequency (Hz) [or Rate (pps)] | 0.25 Hz | 0.25 Hz | Identical. |
| For interferential modes only: Beat Frequency^{†} (Hz) | N/A | N/A | Identical. |
| For multiphasic waveforms only: Symmetrical phases?
Phase Duration (include units), (state range, if applicable), (both phases, if asymmetrical) | N/A | N/A | Identical. |
| Net Charge (microcoulombs (μC) per pulse) (If zero, state method of achieving zero net charge.) | 1000 μC @ 2kΩ | 1000 μC @ 2kΩ | Identical. |
| Maximum Phase Charge, (μC) | 1000 μC @ 2kΩ | 1000 μC @ 2kΩ | Identical. |
| Maximum Current Density (mA/cm^{2}, r.m.s.) | 0.5 mA /cm^{2} @ 500 Ω | 0.5 mA /cm^{2} @ 500 Ω | Identical. |
| Maximum Average Current (average absolute value), mA | 1 mA @ 2 kΩ | 1 mA @ 2 kΩ | Identical. |
Page 7 of 15
{11}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Maximum Average Power Density, (W/cm²), (using smallest electrode conductive surface area) | 0.99 mW/cm² @ 2 kΩ | 0.99 mW/cm² @ 2 kΩ | Identical. |
| Burst Mode (i.e., pulse trains) pulses per burst: | 2 | 2 | Identical. |
| Burst Mode bursts per second: | 0.25 | 0.25 | Identical. |
| Burst Mode burst duration (seconds): | 4 | 4 | Identical. |
| Burst Mode duty cycle: | 0.5 (50%) | 0.5 (50%) | Identical. |
| ON Time (seconds) | 2 sec | 2 sec | Identical. |
| OFF Time (seconds) | 2 sec | 2 sec | Identical. |
## Discussion of Similarities and Differences
The two systems are identical with respect to the following technical parameters:
- Method of Line Current Isolation
- Normal Condition current
- Single Fault Condition current
- Average DC current through the electrodes when the device is on but no pulses are being delivered
- Number of output channels
- Software/Firmware/Microprocessor Control
- Automatic Shut Off
- User Override Control
- Waveform (e.g., pulsed monophasic, biphasic)
- Shape (e.g., rectangular, spike, rectified sinusoidal)
- Maximum Output Voltage
- Maximum Output Current
- Duration of primary (depolarizing) phase
- Pulse Duration
- Frequency
- Net Charge
- Maximum Phase Charge
Page 8 of 15
{12}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
- Maximum Current Density
- Maximum Average Current
- Maximum Average Power Density
- Burst Mode (i.e., pulse trains) pulses per burst
- Burst Mode bursts per second
- Burst Mode burst duration (seconds)
- Burst Mode duty cycle
- ON Time (seconds)
- OFF Time (seconds)
The two systems differ in the following technical parameters:
- The proposed Genesis Sleep uses a different lithium battery voltage.
- The proposed Genesis Sleep has a higher leakage current, but still within the allowable range.
- The proposed Genesis Sleep is better in the single fault condition.
- The proposed Genesis Sleep has an Automatic Overload Trip, increasing safety.
These differences are minor, and do not materially affect their substantial equivalence with respect to technological basis or use.
IV. Output Parameters
| Parameter | Proposed Device Genesis Sleep | Predicate Modius Sleep (K230826) | Comment |
| --- | --- | --- | --- |
| Output value at 500 Ω, 2 kΩ and 10 kΩ load conditions | 1 mA at all loads
0.25 Hz | 1 mA at all loads
0.25 Hz | Identical. |
| Electrode surface area in cm² | Conductive Area (16mm diameter)
2.0 cm² | Conductive Area (16mm diameter)
2.0 cm² | Identical. |
| Current density | 0.5 mA/cm² @ 500 Ω, 2kΩ, and 10 kΩ | 0.5 mA/cm² @ 500 Ω, 2kΩ, and 10 kΩ | Identical. |
| Charge density | 497.63 μC/cm² per pulse@ 500 Ω, 2 kΩ, and 10 kΩ | 497.63 μC/cm² per pulse@ 500 Ω, 2 kΩ, and 10 kΩ | Identical. |
| Power density | 0.25 mW/cm² @ 500 Ω
0.99 mW/cm² @ 2 kΩ
4.97 mW/cm² @ 10 kΩ | 0.25 mW/cm² @ 500 Ω
0.99 mW/cm² @ 2 kΩ
4.97 mW/cm² @ 10 kΩ | Identical. |
Page 9 of 15
{13}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Max phase charge (pulse width x peak current) | 1000 μC per pulse @ 500 Ω, 2 kΩ, and 10 kΩ | 1000 μC per pulse @ 500 Ω, 2 kΩ, and 10 kΩ | Identical. |
| --- | --- | --- | --- |
| Max phase charge density (pulse width x peak current) / electrode surface area | 497.63 μC/cm² per pulse @ 500 Ω, 2 kΩ, and 10kΩ | 497.63 μC/cm² per pulse @ 500 Ω, 2 kΩ, and 10kΩ | Identical. |
| Max average power density (Duty cycle x peak current)² x (load Ω) ÷ electrode surface area) | 0.12 mW/cm² @ 500 Ω
0.5 mW/cm² @ 2 kΩ
2.49 mW/cm² @ 10 kΩ | 0.12 mW/cm² @ 500 Ω
0.5 mW/cm² @ 2 kΩ
2.49 mW/cm² @ 10 kΩ | Identical. |
| Number of electrodes | 2 | 2 | Identical. |
## Discussion of Similarities and Differences
The two systems are identical with respect to all output parameters:
- Output value at 500 Ω, 2 kΩ and 10 kΩ load conditions
- Electrode surface area
- Current density
- Charge density
- Power density
- Max phase charge
- Max phase charge density
- Max average power density
- Number of electrodes
## V. Comparative Performance Evaluations
Comparative performance evaluations were performed. The results demonstrate substantial equivalence of the proposed Genesis Sleep to the predicate device.
## VI. Leveraged Clinical Data from Predicate Device (Modius Sleep, K230826)
To meet the special control requirement under 21 CFR 882.5800(b)(1)(vi)(E), the following provides a detailed summary of the clinical testing for the predicate device (Modius Sleep, K230826), including the clinical outcomes associated with its use and a summary of adverse events and complications that occurred with the device. This information is drawn from the predicate's 510(k) summary and the associated pivotal clinical study (published as Curry et al., Brain Stimulation 17 [2024] 782-793). A comparison of the predicate's stimulation parameters, treatment protocol, and device usage to those of the Genesis
Page 10 of 15
{14}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
Sleep is provided, followed by a justification for why the predicate's clinical data is pertinent and directly applicable to the Genesis Sleep.
## Summary of Predicate Device Clinical Testing
The predicate device's clinical performance was evaluated in a pivotal randomized, double-blinded, sham-controlled trial to assess the safety and efficacy of the Modius Sleep for treating chronic insomnia. The study was multi-site, conducted from May 2022 to January 2023 at Ulster University (UK, including remote recruitment from the Republic of Ireland) and Hong Kong Polytechnic University (HK). It enrolled 149 participants aged 19-67 years with moderate to severe chronic insomnia (Insomnia Severity Index [ISI] score ≥15), confirmed via screening to align with ICSD and DSM-V criteria (e.g., sleep disturbance duration/frequency, daytime symptoms, adequate sleep opportunity, no co-existing conditions). Participants were randomized 1:1 to the Modius Sleep (n=75) or sham device (n=74). Both devices appeared identical, ensuring blinding.
## Stimulation Parameters and Treatment Protocol Used in the Study:
- Waveform: Symmetrical biphasic rectangular.
- Frequency: 0.25 Hz.
- Current Intensity: 0.1 mA to 1.0 mA (user-adjustable in 0.1 mA increments until a gentle swaying sensation was felt, indicating vestibular nerve modulation).
- Pulse Duration: 2 seconds.
- Primary (Depolarizing) Phase Duration: 4 seconds.
- Burst Mode: 2 pulses per burst; 0.25 bursts per second; 4-second burst duration; 50% duty cycle.
- ON/OFF Time: 2 seconds each.
- Session Length: 30 minutes.
- Usage Frequency: Daily (28 consecutive days in UK; 5 days/week for 20 total days in HK).
- Placement: Bilateral on mastoid processes.
- Device Usage: At home, while sitting, immediately before sleep; self-adhesive electrode pads disposed after each use; intensity adjusted via device buttons; 16-hour post-session lockout to prevent overuse; device not rechargeable during stimulation for safety.
The sham device used a 0.8 Hz frequency (reducing vestibular stimulation likelihood), with stimulation ramping down to 0 mA after 30 seconds (total 50 seconds), exploiting user accommodation to current sensations for blinding.
## Clinical Outcomes:
- Primary Outcome (ISI Score Change): In the complete case analysis (n=126), the Modius Sleep group had a mean ISI decrease of 5.80 (95% CI: -6.79 to -4.81;
Page 11 of 15
{15}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
p<0.001 within-group), vs. 3.52 (95% CI: -4.74 to -2.30; p<0.001) in sham. Between-group difference: -2.28 (95% CI: -3.85 to -0.71; p=0.005). Intention-to-treat (ITT) analyses (multiple imputation [MI] and last observation carried forward [LOCF]) showed similar trends, with Modius Sleep reductions of 5.78 (MI) and 4.95 (LOCF), approaching or meeting the 6-point clinically meaningful threshold for over half of the treatment group.
- Secondary Outcomes: Superior improvement in SF-36 energy/fatigue (ITT p=0.006; complete case p=0.004). Pittsburgh Sleep Quality Index (PSQI) global score improvements were greater in Modius Sleep but not statistically significant after corrections.
- Over 4 weeks, ISI scores decreased most from Week 0 to Week 2, then stabilized; scores remained low at Week 8 and 16 (exploratory follow-up in HK).
# Adverse Events and Complications:
Twenty-two non-anticipated adverse events (AEs) were reported during the 4-week intervention (16 in Modius Sleep group, 6 in sham), all minor, self-resolving, and causing minimal discomfort. No serious device-related AEs or complications occurred. One non-device-related serious AE (minor cerebrovascular accident due to undiagnosed hypertension) was reported post-withdrawal. AE breakdown:
| Body System / Reported Term | Total n (%) | Modius Sleep n (%) | Sham Control n (%) |
| --- | --- | --- | --- |
| Nervous System Disorders - Headache/migraine | 7 (4.7) | 6 (4.0) | 1 (0.7) |
| Eye Disorders - Flashes in peripheral vision | 2 (1.3) | 2 (1.3) | 0 (0) |
| Eye Disorders - Shadow in peripheral vision | 1 (0.7) | 1 (0.7) | 0 (0) |
| Eye Disorders - Tingling in eye | 1 (0.7) | 1 (0.7) | 0 (0) |
| Ear Disorders - Ear pain | 2 (1.3) | 0 (0) | 2 (1.3) |
| Ear Disorders - Tinnitus | 2 (1.3) | 1 (0.7) | 1 (0.7) |
| Ear Disorders - Itching in ear | 1 (0.7) | 0 (0) | 1 (0.7) |
| Mood Disorders - Low Mood | 2 (1.3) | 2 (1.3) | 0 (0) |
| Mouth/Dental Disorders - Metal fillings pulsing | 1 (0.7) | 0 (0) | 1 (0.7) |
| Mouth/Dental Disorders - Grinding teeth | 1 (0.7) | 1 (0.7) | 0 (0) |
{16}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Body System / Reported Term | Total n (%) | Modius Sleep n (%) | Sham Control n (%) |
| --- | --- | --- | --- |
| Gastrointestinal Disorders - Nausea | 1 (0.7) | 1 (0.7) | 0 (0) |
| Other - Tingling in arm | 1 (0.7) | 1 (0.7) | 0 (0) |
Rates were similar to sham, indicating low device-specific risk. No withdrawals due to AEs except one (treatment group: nausea and headaches). Generalizability: Baseline ISI scores (HK 19.11, UK 18.82) comparable to US studies (e.g., 17-17.8); insomnia prevalence in HK (20.7%) aligns with US/global estimates (10-20%); no evidence of racial/ethnic differences in insomnia presentation.
## Comparison of Stimulation Parameters, Treatment Protocol, and Device Usage
The Genesis Sleep delivers identical stimulation output to that used in the predicate's study, as verified by comparative testing (e.g., oscilloscope tracings and digital multimeter measurements under 500 Ω, 2 kΩ, and 10 kΩ loads). The treatment protocol and usage are directly comparable, supporting applicability of the predicate's data.
| Parameter/ Aspect | Predicate (Modius Sleep) - Pivotal Study | Genesis Sleep | Comparison/ Comment |
| --- | --- | --- | --- |
| Waveform | Symmetrical biphasic rectangular | Symmetrical biphasic rectangular | Identical |
| Frequency | 0.25 Hz | 0.25 Hz | Identical |
| Current Intensity | 0.1 mA - 1.0 mA (user-adjustable) | 0 μA - 1000 μA (0-1 mA, user-adjustable) | Identical range |
| Pulse Duration | 2 seconds | 2 seconds | Identical |
| Primary Phase Duration | 4 seconds | 4 seconds | Identical |
| Burst Mode | 2 pulses/burst; 0.25 bursts/second; 4-second duration; 50% duty cycle | 2 pulses/burst; 0.25 bursts/second; 4-second duration; 50% duty cycle | Identical |
| ON/OFF Time | 2 seconds each | 2 seconds each | Identical |
| Session Length | 30 minutes | 30 minutes | Identical |
| Usage Frequency | Daily (28 days UK; 20 days HK over 4 weeks) | Indicated for daily use (up to 30 min/session) | Equivalent; Genesis supports similar |
Page 13 of 15
{17}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
| Parameter/Aspect | Predicate (Modius Sleep) - Pivotal Study | Genesis Sleep | Comparison/Comment |
| --- | --- | --- | --- |
| | | | home/clinical regimens |
| Electrode Placement | Bilateral mastoid processes | Mastoid | Identical |
| Device Usage | At home, while sitting, before sleep; user adjusts intensity via built-in buttons; disposable pads; 16-hour lockout | At home/clinical, while sitting, before sleep; user adjusts via PC-based software; disposable pads; software-monitored overload/shut-off | Equivalent stimulation delivery; minor control difference (PC vs. built-in) does not affect output or protocol; Genesis adds clinical use, but labeling for home use exceeds clinical requirements, and both are prescription-only (no new intended use) |
| Power Source/Isolation | 3.75 V Lithium Polymer Battery; DC:DC transformer | 12 V Li-Ion Battery; DC:DC converter | Different voltage but equivalent isolation and safety; output remains identical |
# Justification for Applicability of Clinical Outcomes, Adverse Events, and Complications
The Genesis Sleep is substantially equivalent to the predicate in all technological characteristics directly impacting stimulation delivery (e.g., waveform, frequency, current, pulse duration, burst mode, output voltage/current under loads, net charge, current/charge/power density - see sections III and IV above). Minor differences (e.g., PC-based controls vs. built-in, aluminum enclosure vs. plastic, non-wearable form factor vs. wearable) do not alter the stimulation output, electrode interface, or user protocol, raising no new questions of safety or effectiveness. Both devices are prescription-only, non-invasive cranial electrotherapy stimulators for chronic insomnia in adults aged 22+.
The predicate's clinical data, from a sham-controlled trial in a comparable population (moderate/severe chronic insomnia, similar demographics), directly applies because the
Page 14 of 15
{18}
NeuroField, Inc.
510(k) Summary - K252951
Genesis Sleep
delivered therapy is identical, ensuring similar efficacy (e.g., ISI reductions approaching clinical meaningfulness) and safety (low rate of minor, self-resolving AEs; no complications). Data generalizability to US populations is supported by comparable baseline ISI scores (18-19 vs. US 17-18), insomnia prevalence (HK 20.7% vs. US/global 10-20%), and lack of evidence for racial/ethnic differences in insomnia presentation/diagnosis.
CES devices collectively demonstrate a class effect of CES for treating anxiety and/or insomnia. However, it cannot be concluded, based on available information alone, that specific CES devices or stimulation parameters are effective for treating anxiety and/or insomnia. As such, individuals using this device should work with the prescribing medical provider to determine the best treatment settings to use.
## VII. Conclusion
The two devices have the same intended use and the same main classification. They have most of the same features and technical attributes, including an exact match of their output waveforms. They are used on the same populations. The two products are different in minor ways that do not materially affect their technological basis or use.
Based upon comparisons of regulatory parameters, technical and safety features, comparative performance evaluations, and the applicability of predicate clinical evaluations, the proposed Genesis Sleep device is substantially equivalent to the predicate Modius Sleep (K230826).
Page 15 of 15
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.