Ampa One System (AMPA-001)

K243319 · Neuromodulatory Devices & Applications · OBP · Feb 13, 2025 · Neurology

Device Facts

Record IDK243319
Device NameAmpa One System (AMPA-001)
ApplicantNeuromodulatory Devices & Applications
Product CodeOBP · Neurology
Decision DateFeb 13, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5805
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

Device Story

The Ampa One System is a computerized, electromechanical TMS device used in clinical settings to treat Major Depressive Disorder. It delivers non-invasive magnetic fields to induce electrical currents in the cerebral cortex, targeting the left dorsolateral prefrontal cortex (DLPFC) via the L Coil or the dorsomedial prefrontal cortex (DMPFC) via the M Coil. The system requires integration with a MagVenture MagPro R30 Pulse Generator. Operators use a tablet-based software interface to control treatment parameters, while a neuronavigation cap with integrated markings and an integrated camera in the coil assist in precise positioning. The system uses phase change material and software-controlled cutoffs for thermal management. By stimulating neural tissue, the device aims to provide therapeutic benefit to patients resistant to antidepressant medication. Healthcare providers use the system to deliver standardized stimulation protocols, with output affecting clinical management of depression.

Clinical Evidence

Bench testing only. No clinical data presented. Effectiveness was established by demonstrating that the Ampa One System's magnetic field strength and spatial characteristics are identical to the predicate devices, supported by high correlation coefficients (0.981–0.997) for induced electrical field distributions. Safety was validated through compliance with IEC 60601-1, IEC 60601-1-2, ISO 14971, and ISO 10993-1 standards.

Technological Characteristics

Electromechanical TMS system; figure-of-eight air-core coils (L and M). Energy source: external pulse generator (MagVenture MagPro R30). Thermal management: phase change material with software-controlled cutoff. Connectivity: tablet-based software interface. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 14971 (risk), ISO 10993-1 (biocompatibility), IEC 62304 (software).

Indications for Use

Indicated for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.

Regulatory Classification

Identification

A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 13, 2025 Neuromodulatory Devices & Applications % Tyler Ting Director of Regulatory Affairs Avio Medtech Consulting 2300 Myrtle Ave Suite 200 St. Paul, Minnesota 55114 Re: K243319 Trade/Device Name: Ampa One System (AMPA-001) Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive transcranial magnetic stimulation system Regulatory Class: Class II Product Code: OBP Dated: January 14, 2025 Received: January 14, 2025 Dear Tyler Ting: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # ROBERT KANG -S for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Submission Number (if known) K243319 Device Name Ampa One System (AMPA-001) Indications for Use (Describe) The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. Type of Use (Select one or both, as applicable) > | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Traditional 510(k) 510(k) Summary Image /page/4/Picture/2 description: The image shows the logo for Avio Medtech Consulting. The logo features a stylized paper airplane in shades of teal and gray on the left. To the right of the airplane is the company name, "AVIO," in a bold, sans-serif font, with the words "Medtech Consulting" in a smaller font underneath. #### 510(k) Summary – K243319 In accordance with 21 CFR 807.92 the following summary information is provided: ## Contact Details 21 CFR 807.92(a)(1) | Applicant Name: | Neuromodulatory Devices & Applications | |--------------------------|----------------------------------------------------------------------------------------------| | Applicant Address: | 501 W. Broadway<br>Suite 1025<br>San Diego, CA 82101<br>United States | | Applicant Contact | Dr. Don Vaughn | | Applicant Telephone | +1 619 846 5566 | | Applicant Contact Email: | don@ampahealth.com | | Correspondent Name: | Mr. Tyler Ting, MS, RAC-Devices<br>Director of Regulatory Affairs<br>Avio Medtech Consulting | | Correspondent Address: | 2300 Myrtle Ave<br>Suite 200<br>St. Paul, MN 55114<br>United States | | Correspondent Telephone: | +1 970 404 1555 | | Correspondent Email: | tyler@aviomedtech.com | #### Device Trade Name 21 CFR 807.92(a)(2) | Device Trade Name: | Ampa One System (AMPA-001) | |----------------------|-----------------------------------------------------| | Common Name: | Repetitive transcranial magnetic stimulation system | | Classification Name: | Transcranial Magnetic Stimulator | | Regulation Number | 882.5805 | | Product Code(s) | OBP | #### Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | 510(k) | Trade Name | Product Code | |---------|------------------------------------------------------|--------------| | K173620 | MagVita TMS Therapy System w/Theta Burst Stimulation | OBP | | K220819 | Brainsway Deep TMS System | OBP | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Avio Medtech Consulting. The logo consists of a blue paper airplane on the left and the word "AVIO" in large, dark gray letters on the right. Below the word "AVIO" is the phrase "Medtech Consulting" in smaller, blue letters. Traditional 510(k) 510(k) Summary #### Device Description Summary 21 CFR 807.92(a)(4) The Ampa One System is a computerized, electromechanical device that produces and delivers non-invasive, magnetic fields to induce electrical currents in brain tissues to target specific regions of the cerebral cortex. The system contains two different coils for targeting different regions of the brain. The L Coil is designed to target the left dorsolateral prefrontal cortex (DLPFC). The M Coil is designed to target the bilateral prefrontal cortex (DMPFC). Both coils are used within the Ampa One System to treat Major Depressive Disorder (MDD) in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons. The Ampa One System is an integrated system composed of the following hardware and software components with associated functions to enable its intended use: - . Ampa L and M Coils - MT determination and Depression treatment - . Bridge Box & cables - Allows connection to the pulse generator and control treatment protocols from the pulse generator and Ampa tablet with software - Ampa tablet with software - The software interface for operators to control TMS treatment - . Ampa Magic Arm - Support system to securely hold the coils in place during treatment - Neuronavigation Cap Fabric cap worn by patient with integrated markings for determination of . cerebral cortex treatment targets The system is intended to be used with the MagVenture MagPro R30 Pulse Generator (K230014) and cannot be used without a compatible pulse generator. The system comes with a tablet preloaded with a mobile app containing the software interface for operation of the system. #### Intended Use/Indications for Use 21 CFR 807.92(a)(5) The Ampa One System is intended for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode. ## Indications for Use Comparison 21 CFR 807.92(a)(5) The Ampa One System and the predicate devices have the same indications for use. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for AVIC Medtech Consulting. The logo features a stylized paper airplane in shades of blue and gray on the left. To the right of the airplane is the text "AVIC" in a bold, sans-serif font, with "Medtech Consulting" in a smaller, lighter font underneath. Traditional 510(k) 510(k) Summary Technological Comparison 21 CFR 807.92(a)(6) The AMPA One System has similar technological characteristics to the predicate devices and the minor ancillary differences between the two devices do not raise different questions regarding safety and effectiveness. The new device and predicate devices share the same technological characteristics to achieve their intended use of treating MDD using repetitive magnetic fields to induce electrical firing in the cerebral cortex. The Ampa One System coils and the predicate devices' coils are both wound coils of the same design where pulses of electric current are passed to deliver non-invasive, magnetic fields which induce electrical currents in targeted regions of the cerebral cortex. The design of the M and L coils were based on the selected predicate devices, the Mag Vita B70 (K173620) and the BrainsWay H7 (K220819). The process of designing the coil windings used state-ofthe-art simulations in collaboration with a renowned expert in TMS coil analysis. Because of the design process, the magnetic fields produced by the Ampa One System coils and the predicates are nearly indistinguishable from each other. The correlation coefficient between the L coil and the B70 coil is 0.981 and the correlation coefficient between the M coil and the H7 is 0.986. The concomitant induced electrical field distributions are also equally identical, with correlation coefficients of 0.985 and 0.997 for the L and M coil, respectively. The minor differences between the subject device and the predicate devices are: 1. the cooling systems of the coils; 2. the software used to control the treatment; and 3. the use of an integrated camera and premarked patient cap for patient positioning. The Ampa One System uses a phase change material alongside a software controlled cutoff to maintain both operational temperatures and prevent thermal harm whereas the predicate devices use active cooling svstems. The Ampa One System has its own software within a preloaded tablet to control treatment parameters whereas the predicate devices' software comes on their own capital equipment. The Ampa One System uses an integrated camera in the coil and software to control coil positioning whereas the predicate devices' use operator determined markings on the patient cap (K173620) and head positioning straps (K220819) to achieve accurate and reliable coil positioning. These represent the differences in technological characteristics. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Avio Medtech Consulting. The logo consists of a stylized paper airplane in blue and gray, followed by the word "AVIO" in a dark gray sans-serif font. Below "AVIO" is the text "Medtech Consulting" in a smaller, teal sans-serif font. The logo is clean and modern, suggesting innovation and expertise in the medical technology field. Traditional 510(k) Ampa One System 510(k) Summary #### Substantial Equivalence Table | | Subject Device:<br>Ampa One System | Predicate for Ampa One System, L<br>Coil<br>MagVita TMS Therapy System<br>w/Theta Burst Stimulation<br>K173620 | Predicate for Ampa One System,<br>M Coil<br>Brainsway Deep TMSTM System<br>K220819 | |-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code,<br>Device Risk Classification | OBP<br>Class II | OBP<br>Class II | OBP<br>Class II | | Indications for Use | The Ampa One System is indicated<br>for the treatment of Major<br>Depressive Disorder in adult patients<br>who have failed to achieve<br>satisfactory improvement from<br>previous anti-depressant medication<br>treatment in the current episode. | The MagVita TMS Therapy System<br>is indicated for the treatment of<br>Major Depressive Disorder in adult<br>patients who failed to achieve<br>satisfactory improvement from<br>previous anti-depressant medication<br>treatment in the current episode. | The BrainsWay Deep TMSTM is<br>indicated for the treatment of<br>depressive episodes and for<br>decreasing anxiety symptoms for<br>those who may exhibit comorbid<br>anxiety symptoms in adult patients<br>suffering from Major Depressive<br>Disorder (MDD) who failed to<br>achieve satisfactory improvement<br>from previous anti-depressant<br>medication treatment in the current<br>episode. | | Target Population | Adult subjects with Major<br>Depressive Disorder | Adult subjects with Major<br>Depressive Disorder | Adult subjects with Major<br>Depressive Disorder with<br>comorbid anxiety symptoms | | Energy Used / Delivered | Electromagnetic Energy is delivered | Electromagnetic Energy is delivered | Electromagnetic Energy is<br>delivered | | Area of brain to<br>stimulated | Ampa L-Coil: Head - theta burst<br>stimulation to the left dorsolateral<br>prefrontal cortex (DLPFC)<br>Ampa M-Coil: Head – stimulation to<br>the dorsomedial prefrontal cortex | Theta burst stimulation to the left<br>DLPFC | Stimulation to the DMPFC | | | Subject Device:<br>Ampa One System | Predicate for Ampa One System, L<br>Coil<br>MagVita TMS Therapy System<br>w/Theta Burst Stimulation<br>K173620 | Predicate for Ampa One System,<br>M Coil<br>Brainsway Deep TMS System<br>K220819 | | | (DMPFC) | | | | Applicator:<br>Configuration<br>Core Material | Ampa L-Coil:<br>Figure-of-eight coil<br>Air core | Figure-of-eight coil<br>Air core | Figure-of-eight coil<br>Air core | | | Ampa M-Coil:<br>Figure-of-eight coil<br>Air Core | | | | Performance of Treatment<br>Protocol | | Intermittent Theta Burst<br>Stimulation (iTBS) Treatment | Brainsway H7 Coil for treatment<br>of MDD and Depression with<br>Comorbid Anxiety | | Physical unit of amplitude<br>setting (e.g., coil current,<br>peak magnetic field) at<br>coil and its relation to the<br>SMT unit | L Coil<br>0 - 2.03 SMT (Standard Motor<br>Threshold)<br>M Coil<br>0 - 1.76 SMT | 0 - 1.7 SMT | 0-1.4 SMT | | Patient Contacting<br>Materials | Treatment Cap | Treatment Cap | Treatment Cap | | Coil Positioning System | Cap positioned in relation to<br>nasion | Cap positioned in relation to<br>nasion | Cap positioned in relation to<br>nasion | | Sterility | Not Applicable | Not Applicable | Not Applicable | | Electrical safety | Complies with IEC 60601-1 and IEC<br>60601-1-2 | Complies with IEC 60601-1 and IEC<br>60601-1-2 | Complies with IEC 60601-1 and<br>IEC 60601-1-2 | | Treatment stimulation<br>parameters | L Coil<br>Waveforms: 275 us biphasic | Waveforms: 290 us biphasic | | | Subject Device:<br>Ampa One System | Predicate for Ampa One System, L<br>Coil<br>MagVita TMS Therapy System<br>w/Theta Burst Stimulation<br>K173620 | Predicate for Ampa One System,<br>M Coil<br>Brainsway Deep TMS System<br>K220819 | | | sinusoid | sinusoid | | | | Magnetic field strength: 120% | Magnetic field strength: 120% | | | | MT (motor threshold) | MT | | | | Repetition rate: 50 Hz | Repetition rate: 50 Hz | | | | Train duration: 2 sec* | Train duration: 2 sec | | | | Inter-train interval: 8 secs | Inter-train interval: 8 secs | | | | Burst pulses: 3 | Burst pulses: 3 | | | | Bursts: 200 | Bursts: 200 | | | | Inter-pulse interval: 20 msec | Inter-pulse interval: 20 msec | | | | Number of trains: 20 | Number of trains: 20 | | | | Numbers of pulses/session: 600 | Numbers of pulses/session: 600 | | | | Total duration: 3 min 9 sec | Total duration: 3 min 9 sec | | | | M Coil | | | | | Waveform: 370 µs biphasic sinusoid | | Waveform: 370 µs biphasic<br>sinusoid | | | Magnetic Field Strength: 120% MT | | Magnetic Field Strength: 120%<br>MT | | | Repetition rate: 18 Hz | | Repetition rate: 18 Hz | | | Train duration: 2 sec* | | Train duration: 2 sec | | | Inter-train interval: 20 sec | | Inter-train interval: 20 sec | | | Number of trains: 55 | | Number of trains: 55 | | | Magnetic Pulses per Session: 1980 | | Magnetic Pulses per Session: 1980 | | | Treatment Session Duration: ~ 20<br>min | | | | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for AVIO Medtech Consulting. The logo features a stylized paper airplane in teal and gray on the left. To the right of the airplane is the word "AVIO" in a bold, sans-serif font, with the words "Medtech Consulting" in a smaller font underneath. Traditional 510(k) 510(k) Summary Ampa One System {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for AVIO Medtech Consulting. The logo features a stylized paper airplane in teal and gray on the left. To the right of the airplane is the company name, "AVIO", in a dark gray sans-serif font, with the words "Medtech Consulting" in a smaller teal font underneath. Treatment Session Duration: ~ 20 min #### Traditional 510(k) 510(k) Summary Ampa One System {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for AVIO Medtech Consulting. The logo features a blue paper airplane graphic on the left. To the right of the graphic is the company name, "AVIO", in a bold, sans-serif font, with the words "Medtech Consulting" in a smaller font underneath. Traditional 510(k) Ampa One System 510(k) Summary * Maximum Safe Train Duration (seconds) Limits for Avoiding Seizure limits from Table 2 of the FDA Special Controls Guidance' are equivalent to the predicate devices, and therefore safe as prior clinical testing has established a minimal risk of seizure for these durations. <sup>1</sup> https://www.fda.gov/medical-devices/guidance-documents-mediation-emitting-products/repetitive-transcranial-magnetic-stimulation-rtmssystems-class-ii-special-controls-guidance {11}------------------------------------------------ Traditional 510(k) 510(k) Summary Image /page/11/Picture/2 description: The image shows the logo for Avio Medtech Consulting. The logo consists of a blue paper airplane on the left and the word "AVIO" in dark gray on the right. Below the word "AVIO" is the phrase "Medtech Consulting" in a smaller, light blue font. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Nonclinical testing evaluated the Ampa One System's performance to be equivalent in terms of safety and effectiveness compared to the predicate devices. The suite of testing to demonstrate substantial equivalence can be divided into two aspects: safety (function and compatibility) and effectiveness (magnetic characteristics). Safety testing established that the Ampa One System is able to function in its configuration with the MagVenture R-30 pulse generator to perform its intended use. Safety is defined as functioning with risks assessed of TMS therapy mitigated through design. Effectiveness testing established that the Ampa One System produced equivalent magnetic characteristics compared to the predicate devices. Specifically, testing demonstrated that the amplitude of the magnetic field strength as a function of power input, and the spatial characteristics of the magnetic field produced were identical between the Ampa One coils vs. their respective predicate device coils. Development and testing of the Ampa One System was performance with the FDA Special Controls Guidance on Transcranial Magnetic Stimulator devices. Bench testing was performed in alignment with this Guidance document in order to support substantial equivalence. #### Standards Conformance The Ampa One System Complied with the following voluntary consensus standards: Risk - ISO 14971 Third Edition 2019-1 | Medical devices Application of risk management to medical ● devices Electrical Safety - IEC 60601-1 Edition 3.2 2020-08 Consolidated Version | Medical electrical equipment Part 1: . General requirements for basic safety and essential performance Electromagnetic Compatibility (EMC) - IEC 60601-1-2 Edition 4.1 2020-09 Consolidated Version | Medical electrical equipment Part ● 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests Software Development - IEC 62304 Edition 1.1 2015-06 Consolidated Version | Medical device software - Software life cycle processes Biocompatibility - ISO 10993-1 Fifth edition 2018-08 | Biological evaluation of medical devices Part 1: Evaluation ● and testing within a risk management process Shipping and Transportation - ASTM D4169-22 | Standard Practice for Performance Testing of Shipping Containers and . Systems {12}------------------------------------------------ Image /page/12/Picture/1 description: The image shows the logo for AVIO Medtech Consulting. The logo consists of a stylized paper airplane in blue and gray on the left, followed by the word "AVIO" in a dark gray sans-serif font. Below "AVIO" is the phrase "Medtech Consulting" in a smaller, teal sans-serif font. The overall design is clean and modern. Traditional 510(k) 510(k) Summary - ASTM D4332-22 | Standard Practice for Conditioning Containers Packages or Packaging ● Components for Testing #### Conclusion The evaluation of the subject device's safety and effectiveness has been performed as per the information that is listed above. Based upon this information, the subject device is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...