EVERYWAY INTERFERNTIAL STIMULATOR, MODEL IF-908

K110509 · Everyway Medical Instruments Co.,Ltd · LIH · Jun 20, 2011 · Neurology

Device Facts

Record IDK110509
Device NameEVERYWAY INTERFERNTIAL STIMULATOR, MODEL IF-908
ApplicantEveryway Medical Instruments Co.,Ltd
Product CodeLIH · Neurology
Decision DateJun 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is an interferential stimulator with TENS indications used for symptomatic relief and management of chronic intractable pain. It is intended for use on the order of a physician only.

Device Story

Everyway IF-908 is an interferential stimulator generating true-sine electrical current pulses for pain relief. Device consists of a stimulus generator and electrodes; electrodes attach to patient skin to deliver current to underlying nerves. Operator (physician or patient under physician instruction) adjusts output amplitude and channel frequencies to tailor stimulation. Powered by 9V internal battery or rechargeable pack. Used in clinical or home settings for management of chronic intractable pain. Output provides symptomatic relief by stimulating nerves; clinical decision-making relies on physician prescription and patient response to stimulation parameters.

Clinical Evidence

No clinical data. Bench testing only, including performance compliance per ANSI/AAMI NS4, electrical safety per IEC 60601-1, and EMC compliance per IEC 60601-1-2.

Technological Characteristics

Interferential stimulator; true-sine wave output; dual-channel; adjustable frequency and amplitude; 9V battery or rechargeable pack power source; electrode-based delivery. Complies with ANSI/AAMI NS4, IEC 60601-1, and IEC 60601-1-2.

Indications for Use

Indicated for symptomatic relief and management of chronic intractable pain in patients requiring physician-prescribed electrical stimulation therapy.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### EVERYWAY EVERYWAY MEDICAL INSTRUMENTS CO.,LTD. JUN 2 0 2011 3F1., No. 5, Lane 155, Sec. 3,Peishen Rd, Shen Keng Hsiang, Taipei Hsien, Taiwan, ## 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 on February 10, 2011. The assigned 510(k) number is: K110509 . #### 1. Submitter's Identifications: Establishment: EVERYWAY MEDICAL INSTRUMENT CO., LTD. Address: 3Fl., No. 5, Lane 155, Sec. 3, Peishen Rd., Shenkeng Hsiang, Taipei Hsien 222, Taiwan Registration Number: 9616877 Operations: Manufacturer Owner/Operator: EVERYWAY MEDICAL INSTRUMENT CO., LTD. Address : 3FI .; No. 5, Lane 155, Sec. 3, Peishen Rd., Shenkeng Hsiang, Taipei Hsien 222, Taiwan ' Contact Person: Robert Tu Phone : 886-2-2662-0038 Fax No: 886-2-2664-5566 e-mail : tu922@ms35.hinet.net - 2. Name of the Device: Everyway Interferential Stimulator, model IF-908. - 3. Information of the 510(k) Cleared Device (Predicate Device): Theratech IF series True sine interferential stimulator, model WL-2206A(K050847). 4. Classification Information: Trade/Device Name: Everyway Interferential Stimulator, model IF-908. Requlation Number: Unclassified Regulation Name: Interferential Current Therapy Regulatory Class: II Product Code: LIH 5. Device Description: The Everyway IF-908 Interferential stimulator is the device which generates the small ture-sine pulses of electrical current. The generated current may be delivered to the patient skin and/or underlying nerves through the cable and electrode placed on skin. Through the current stimulation the chronic and/or acute pain is to be relieved The Everyway IF-809 interferential stimulator consists mainly of two parts: the stimulus generator, and electrode. The stimulus generator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the patient's skin so as to transmit this stimulus current to the patient for pain relief {1}------------------------------------------------ ## EVERYWAY EVERYWAY MEDICAL INSTRUMENTS CO.,LTD. 3F1., No. 5, Lane 155, Sec. 3,Peishen Rd, Shen Keng Hsiang, Taipei Hsien, Taiwan, The stimulation current is generated in such a way that the frequency in channel 1 is fixed, but frequency in channel 2 is adjustable. Also the output amplitude is also adjustable. Those adjustable parameters could provide user some stimulation output change depending upon the condition of patients according to the prescription of physician. In general IF-908 is the prescribtion device, which needs the physician instruction for stimulation treatment. IF-908 provides two different alternative power supply system. <1> 9V internal battery, <2> 9V rechargeable battery pack charged by adaptor. The operator may recharge 9V battery pack for long term use via the power rechargeable battery system of IF-908. The battery pack can not be charged during operation via using 9V battery pack power. ### 6. Intended Use: The device is an interferential stimulator with TENS indications used for symptomatic relief and management of chronic intractable pain. It is intended for use on the order of a physician only. #### 7. Comparison to the 510(k) Cleared Device (Predicate Device): The Everyway Interferential Stimulator, model IF-908 is substantially equivalent to the Theratech IF series True sine interferential stimulator, model WL-2206A(K050847) without any significant difference in main technological and operational feature. - 8. Discussion of Non-Clinical Tests Verification Activities Performed to Determine the Safety and Performance of IF-908 are as the followings: - 1> Performance Compliance Test according to ANSI/AAMI NS4 conducted by manufacturer - 2> Electrical Compliance Test according to IEC 60601-1 by accredited laboratory. - 3> EMC Compliance Test according to IEC 60601-1-2 by accredited laboratory. - 9. Discussion of Clinical Test Validation Activities Performed to Determine the Effectiveness of Device are as the followings: No particular Clinical Test was conducted for Everyway Interferential Stimulator, model IF-908. ### 10. Conclusions The Everyway Interferential Stimulator, model IF-908, has the same intended use and 1 technological characteristics as the cleared device of Theratech IF series True sine interferential stimulator, model WL-2206A(K050847). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted model could maintain the same safety and effectiveness as that of cleared device. In the other words, Everyway Interferential Stimulator, model IF-908 is substantial equivalent with the Theratech IF series True sine interferential stimulator, model WL-2206A(K050847). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings and body. The eagle is facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 JUN 2 0 2011 Everyway Medical Instrument Co., Ltd. Mr. Robert Tu 3 Fl., No. 5, Lane 155, Sec. 3 Peishen Road Shenkeng Hsiang Taipei Hsien 222 Taiwan Re: K110509 Trade/Device Name: Everyway Interferential Stimulator, model IF-908 Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: LIH Dated: May 17, 2011 Received: May 19, 2011 Dear Mr. Tu: ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Kesia Alexander Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ EVERYWAY EVERYWAY MEDICAL INSTRUMENTS CO.,LTD. 3F1., No. 5, Lane 155, Sec. 3,Peishen Rd, Shen Keng Hsiang, Taipei Hsien, Taiwan, # Indications For Use 510(k) Number (if known): K110509 Device Name: Everyway Interferential Stimulator, model IF-908. Indications For Use: The device is an interferential stimulator with TENS indications used for symptomatic relief and management of chronic intractable pain. It is intended for use on the order of a physician only Prescription Use V (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) OR Concurrence of CDRH, Office of Device Evaluation (ODE) f CDRH, Office of Device Evaluation (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K110509 Page 1 of 1. Page 3 of 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...