RADIONICS SIPHON LIMITING DEVICE (SLD)
Device Facts
| Record ID | K962990 |
|---|---|
| Device Name | RADIONICS SIPHON LIMITING DEVICE (SLD) |
| Applicant | Radionics, Inc. |
| Product Code | JXG · Neurology |
| Decision Date | Oct 23, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.5550 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
As a component of a cerebrospinal fluid shunt system, the SLD prevents over-drainage of CSF due to siphoning.
Device Story
Radionics Siphon Limiting Device (SLD) functions as a component within a cerebrospinal fluid (CSF) shunt system; prevents over-drainage caused by siphoning when ventricular catheter is elevated relative to distal catheter. Device utilizes a flexible diaphragm that opens or closes in response to changes in hydrostatic pressure; inlet and outlet ports facilitate fluid flow. Used in clinical settings by neurosurgeons during shunt implantation or revision. Output is mechanical regulation of CSF flow; assists in maintaining appropriate intracranial pressure; benefits patient by reducing risks associated with overdrainage.
Clinical Evidence
No clinical data provided; device relies on design and functional equivalence to the predicate device.
Technological Characteristics
Mechanical flow control device; features inlet and outlet ports with a flexible diaphragm; operates via hydrostatic pressure-sensitive mechanical actuation; no electronic components or software.
Indications for Use
Indicated for patients requiring cerebrospinal fluid shunt systems to prevent over-drainage caused by siphoning effects when the ventricular catheter is elevated relative to the distal catheter.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
- PS Medical Siphon Control Device
Related Devices
- K964386 — RADIONICS CONTOUR FLEX PLUS VALVE · Radionics, Inc. · Jan 31, 1997
- K042558 — EQUI-FLOW VALVE AND SHUNT SYSTEM · Integra Neurosciences Implants S.A. · Oct 6, 2004
- K012052 — MEDTRONIC PS MEDICAL STRATA VALVE, MODELS 42855, 42865, 92865, 92855 · Medtronic PS Medical · Feb 11, 2002
- K053350 — MODIFICATION TO CODMAN HAKIM SHUNT SYSTEMS · Codman & Shurtleff, Inc. · Jan 19, 2006
- K981246 — BEVERLY REFERENTIAL VALVE (NL850-9710) WITH INTEGRAL PERITONEAL CATHETER (NL850-9720) VALVE KIT (NL850-9730) · Integra Neurocare, LLC · May 26, 1999