Longeviti PMMA Static Cranial Implant

K170410 · Longeviti Neuro Solutions, LLC · GXN · Mar 23, 2018 · Neurology

Device Facts

Record IDK170410
Device NameLongeviti PMMA Static Cranial Implant
ApplicantLongeviti Neuro Solutions, LLC
Product CodeGXN · Neurology
Decision DateMar 23, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5330
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Longeviti Static Cranial Implants are designed and manufactured individually for each patient to correct bony voids and/or defects of the cranium.

Device Story

Longeviti PMMA Static Cranial Implant is a patient-specific prosthetic plate for cranioplasty; used to correct cranial bony voids/defects. Device design derived from patient CT scan data; manufactured from polymethyl methacrylate (PMMA). Provided sterile; fixated to cranial bone using commercially available fasteners. Maximum surface area 411 cm2; nominal thickness 4mm (max 5mm). Intended for surgical implantation by neurosurgeons to restore cranial integrity.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical strength testing performed per ASTM D638-03 (tensile), D790-03 (flexural), and D4812-06 (impact). Sterilization validation performed using ISO-14937 Overkill approach 3.

Technological Characteristics

Material: Polymethyl methacrylate (PMMA). Dimensions: Max surface area 411 cm2, nominal thickness 4mm (max 5mm). Sterilization: Peracetic acid (ISO-14937). Fixation: Commercially available fasteners. Design: Patient-specific, derived from CT scan data.

Indications for Use

Indicated for patients requiring correction of bony voids and/or defects of the cranium.

Regulatory Classification

Identification

A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized caduceus symbol. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 23, 2018 Longeviti Neuro Solutions, LLC % Elaine Duncan President Paladin Medical, Inc. 303 International Circle Suite T128 Hunt Valley, Maryland 21030 Re: K170410 Trade/Device Name: Longeviti PMMA Static Cranial Implant Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Nonalterable Cranioplasty Plate Regulatory Class: Class II Product Code: GXN, PJN Dated: February 15, 2018 Received: February 16, 2018 Dear Ms. Duncan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael J. Hoffmann -S Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K170410 #### Device Name Longeviti PMMA Static Cranial Implant Indications for Use (Describe) The Longeviti Static Cranial Implants are designed and manufactured individually for each patient to correct bony voids and/or defects of the cranium. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | Manufacturer | Longeviti Neuro Solutions, LLC<br>303 International Circle #T128<br>Hunt Valley, MD 21030 | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Name: Elaine Duncan, Paladin Medical, Inc.<br>regulatory consultant for Longeviti Neuro Solutions<br>Phone: 715-549-6035<br>Email: duncan@paladinmedical.com | | Date Prepared | February 14, 2018 | | Device Name | Trade Name: Longeviti PMMA Static Cranial Implant<br>Common Name: Cranial Implant<br>Classification Name: Plate, Cranioplasty, Preformed, Non-alterable | | ProCode and Classification | Product Code: GXN, PJN<br>Class II | | Substantially Equivalent To | The Longeviti PMMA Static Cranial Implant is substantially<br>equivalent to the following legally marketed devices:<br>Primary Predicate Device:<br>OsteoSymbionics<br>Patient Specific Cranial Plate (a.k.a. ClearSheild) – K133082<br>Secondary Predicate Device:<br>OsteoSymbionics<br>Patient Specific Cranial Implant - K072601 | | Device Description | The Longeviti PMMA Static Cranial Implants are patient specific,<br>implantable prosthetic cranioplasty plates intended to correct<br>and/or restore bony voids and/or defects of the cranium. The<br>implants are manufactured from polymethyl methacrylate<br>materials and are designed using the patient's CT scan data. The<br>devices are provided sterile and can be fixated to cranial bone<br>using commercially available fasteners. Typical maximum single<br>plate sizes do not exceed a total surface area of 411 cm2 and<br>should have a nominal thickness of 4mm, not to exceed 5mm, to<br>ensure physical integrity. | | Indication for Use | The Longeviti Static Cranial Implants are designed and<br>manufactured individually for each patient to correct bony voids<br>and/or defects of the cranium. | {4}------------------------------------------------ ## 510(k) Summary | Summary of Testing and Basis for | | The Longeviti PMMA Static Cranial Implant are provided sterile. | | |----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|--| | Substantial Equivalence | Sterilization validation was conducted according to the Overkill<br>ISO-14937 approach 3 (conservative process definition based on<br>inactivation of reference microorganisms) product challenge<br>device method. The peracetic acid sterilization method<br>conducted by contract sterilizer Revox Sterilization Solutions is<br>the same method, cycle, and sterilizer as used by the primary<br>predicate device manufacturer Osteosymbionics. (K133082) | | | | | The Longeviti PMMA Static Cranial Implant are fabricated from<br>the same PMMA materials using the same methods as the<br>Osteosymbionics Patient Specific Cranial Plate.<br>Longeviti contracted a certified and approved materials testing<br>laboratory to perform a series of mechanical strength tests to<br>demonstrate that the material is physically and mechanically<br>substantially equivalent to the Osteosymbionics predicate<br>material. The mechanical tests conducted were the following: | | | | | | | | | | ASTM | Test Description | | | | D638-03 | Tensile properties of Plastics | | | | D790-03 | Flexural properties of unreinforced and<br>reinforced plastics and electrical insulating<br>materials | | | | D4812-06 | Unnotched cantilever beam impact | | | | Conclusion: The Longeviti PMMA Static Cranial Implants (design,<br>materials, properties and indication for use) are substantially<br>equivalent to predicate devices by Osteosymbionics; (K072601)<br>and (K133082) | | |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...