TENS: [RS-18, RS-28, RS-38 models]: For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities. [RS-18, RS-38 models]: For temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medications. [RS-18, RS-28 models]: It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. EMS: [RS-18, RS-28 models]: To stimulate healthy muscles in order to improve and facilitate muscle performance. Environments of Use: Clinics, hospital and home environments (over-the-counter) Patient Population: Adult
Device Story
Battery-powered wireless electrical stimulator (TENS/EMS) for non-pharmacological pain relief and muscle performance improvement. System comprises remote control (1.77" TFT LCD, 4 buttons) and stimulator pods (2 output channels, 3 buttons). Inputs: user-selected therapy programs via remote or pod buttons. Transformation: microprocessor-controlled biphasic symmetrical rectangular pulses delivered to skin via self-adhesive electrodes or knee/elbow braces. Output: electrical stimulation pulses (1-150 Hz, 40-400 µsec pulse width). Used in clinics, hospitals, or home (OTC) by adult patients. Remote control manages multiple pods; pods can operate independently. Healthcare providers/patients use output to manage pain or facilitate muscle recovery. Benefits: non-invasive pain management and muscle conditioning.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601-1, 60601-1-2, 60601-2-10, 60601-1-11), software verification/validation, human factors/usability (IEC 62366-1), and mechanical/stability testing.
Technological Characteristics
Battery-powered (3.7V LiPo) wireless stimulator. PC/ABS plastic housing. Biphasic symmetrical rectangular waveforms. 2 output channels per pod. Microprocessor-controlled output. Connectivity: wireless remote to pod. Sterilization: N/A (non-sterile). Standards: IEC 60601-1, 60601-1-2, 60601-2-10, 60601-1-11, ISO 10993-1/5/10.
Indications for Use
Indicated for adults requiring temporary relief of pain from sore/aching muscles (shoulder, waist, back, neck, extremities) due to exercise/household strain, dysmenorrhea (with OTC meds), chronic intractable pain, or arthritis; and for stimulation of healthy muscles to improve performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Well Life Wireless TENS/EMS Stimulator, Models WR-2605A/2605 (K161453)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
September 1, 2022
ZMI Electronics Ltd. Yuta Lee President 6F-1, 286-4, Shin Ya Road Kaohsiung, R.O.C. 806 Taiwan
Re: K220997
Trade/Device Name: Wireless TENS/EMS, Bruno, Aela Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NGX, NYN Dated: July 29, 2022 Received: August 2, 2022
Dear Yuta Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K220997
Device Name
Wireless Electrical Stimulator (Models RS-18, RS-28, RS-38)
## Indications for Use (Describe)
TENS:
[RS-18, RS-28, RS-38 models]: For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
[RS-18, RS-38 models]: For temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medications.
[RS-18, RS-28 models]: It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
#### EMS:
[RS-18, RS-28 models]: To stimulate healthy muscles in order to improve and facilitate muscle performance.
Environments of Use: Clinics, hospital and home environments (over-the-counter) Patient Population: Adult
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="checked" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### 1. Submitter Information
Applicant Name: ZMI Electronics Ltd. Address: 6F-1, 286-4, Shin-Ya Road, Kaohsiung, Taiwan 806 Phone: +886-7-8150053 Fax: +886-7-8150057 E-mail: mail@zmi-electronics.com
### Contact Person
Name: Lawrence Liu Title: Regulatory Affairs Manager Phone: +886-7-8150053 Ext 351 Fax: +886-7-8150057 E-mail: lawrence@zmi-electronics.com
#### Date Summary Prepared: February 10, 2022
#### 2. Device Information
Device Name: Wireless Electrical Stimulator Model: RS-18, RS-28, RS-38 Regulation Number: 21 CFR 882.5890 Classification Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Subsequent Product Code: NGX, NYN
### 3. Predicate device
#### Primary Predicate
510K Number: K161453 Device Name: Well Life Wireless TENS/EMS Stimulator, Models WR-2605A/2605 Regulation Number: 21 CFR 882.5890 Classification Name: Transcutaneous electrical nerve stimulator for pain relief Requlatory Class: Class II Product Code: NUH Subsequent Product Code: NGX, NYN
### Secondary Predicate
510K Number: K183215 Device Name: Focus TENS Therapy, Model PM710-M/-L Requlation Number: 21 CFR 882.5890
Classification Name: Transcutaneous electrical nerve stimulator for pain relief
{4}------------------------------------------------
Requlatory Class: Class II Product Code: NUH Subsequent Product Code: NYN
#### Third Predicate
510K Number: K183110 Trade/Device Name: Livia Regulation Number: 21 CFR 882.5890 Classification Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Subsequent Product Code: NGX
#### 4. Device Description
"Wireless Electrical Stimulator" is a battery-powered wireless electrical stimulator designed and intended to provide Transcutaneous Electrical Nerve Stimulation (TENS) and Electrical Muscular Stimulation (EMS) for the use of non-pharmacological pain relief and improve muscle performance and recovery.
The "Wireless Electrical Stimulator" includes a remote control, stimulator pod(s) and several accessories. Both the remote control and the stimulator pod are powered by a 3.7V lithium-ion polymer battery. The remote control has a 1.77" TFT LCD screen and 4 buttons to control multiple stimulator pod(s) at the same time. Each stimulator pod has two output channels, and will perform electrical stimulation according to the parameter settings sent by the remote control. The stimulator pods can also be used independently via its own 3 buttons based on preset parameter.
The accessories include self-adhesive electrodes, kneelelbow brace, and the USB Type-C charging cable. The adhesive electrodes for "Wireless Electrical Stimulator" are connected with the Stimulator Pod(s) via magnetic snap and lead wire. There are replaceable hydrogel pads on the self-adhesive electrode pads. Users can replace the hydrogel when it exceeds its useful life, but does not need to discard the entire electrode. To apply electrical stimulation to body joints (e.g. knees and elbows), users can easily place electrodes on specific body parts using the Knee/elbow brace. The remote control and stimulator pods are the same for all models. The 3 models have different accessories to match their specific usage instructions. Below are the accessory variants for each model.
| Product Name | Wireless TENS/EMS | Brace TENS/EMS | Menstrual Pain Relief<br>TENS |
|---------------------|-------------------|--------------------------------------|----------------------------------------|
| Model | RS-18 | RS-28 | RS-38 |
| Indication of Use | General TENS/EMS | TENS/EMS for joint<br>(knee & elbow) | TENS for period pain<br>(dysmenorrhea) |
| Adhesive Electrodes | ● | ● | ● |
| Electrode Lead Wire | ● | ● | ● |
| Knee/Elbow Brace | n/a | ● | n/a |
{5}------------------------------------------------
| USB Type C Charge Cable | • | • | • |
|-------------------------|---|---|---|
|-------------------------|---|---|---|
Besides the accessory variants, the programs of all models also have some variants depend on the indication. RS-18 model has a "General TENS/EMS" indication, it includes all therapy programs. RS-28 model is indicated for "TENS/EMS for joint (knee & elbow)" and the RS-38 model is indicated for "TENS for period pain (dysmenorrhea)." Following are the program variants for each models.
| Therapy Type | Program | RS-18 | RS-28 | RS-38 |
|--------------|-----------------------|-------|-------|-------|
| Pain Relief | Acute Pain | ● | ● | ● |
| | Chronic Pain | ● | ● | ● |
| | Soreness | ● | ● | ● |
| Pain Relief | Back Pain | ● | n/a | ● |
| | Menstrual<br>Cramping | ● | n/a | ● |
| Pain Relief | Arthritis | ● | ● | n/a |
| Training | Warm-up | ● | ● | n/a |
| | Recovery | ● | ● | n/a |
| | Atrophy | ● | ● | n/a |
| | Strength | ● | ● | n/a |
| | Endurance | ● | ● | n/a |
| Massage | Resistance | ● | ● | n/a |
| | Comfort | ● | ● | ● |
| | Kneading | ● | ● | ● |
| | Tapping | ● | ● | ● |
| | Relaxation | ● | ● | ● |
| | Tingling | ● | ● | ● |
| | Alternating | ● | ● | ● |
{6}------------------------------------------------
#### 5. Indications for Use
TENS:
[RS-18, RS-28, RS-38 models]: For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
[RS-18, RS-38 models]: For temporary relief of pain associated with dysmenorrhea (menstrual cramps) when used with over-the-counter pain medications.
[RS-18, RS-28 models]: It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
EMS:
[RS-18. RS-28]: To stimulate healthy muscles in order to improve and facilitate muscle performance.
Environments of Use: Clinics, hospital and home environments (over-the-counter) Patient Population: Adult
{7}------------------------------------------------
## Comparison of Technological Characteristics with The Predicate Device
The "Wireless Electrical Stimulator" has been carefully compared to legally marketed devices with respect to intended use, indications for use (Table 1), technological characteristics (Table 2), and Output Specification (Table 3).
#### Table 1. Substantial Equivalence Table – Regulatory Information
| | Subject Device | Primary Predicate | Secondary Predicate | Third Predicate |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name | Wireless Electrical<br>Stimulator | Well Life Wireless<br>TENS/EMS Stimulator,<br>Models WR-<br>2605A/2605 | Focus TENS Therapy,<br>Model PM710-M/-L | Livia |
| 510(k) Number | K220997 | K161453 | K183215 | K183110 |
| Manufacturer | ZMI Electronics Ltd. | Well-life Healthcare<br>Limited | Omron Healthcare, Inc. | LifeCare Ltd. |
| Classification<br>Product Code | NUH | NUH | NUH | NUH |
| Subsequent<br>Product Codes | NGX, NYN | NGX, NYN | NYN | NGX |
| Regulation<br>Number | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 | 21 CFR 882.5890 |
| Regulation Name | Transcutaneous<br>electrical nerve<br>stimulator for pain relief | Transcutaneous<br>electrical nerve<br>stimulator for pain relief | Transcutaneous<br>electrical nerve<br>stimulator for pain relief | Transcutaneous<br>electrical nerve<br>stimulator for pain relief |
| Regulatory Class | Class II | Class II | Class II | Class II |
| Indications for | For temporary relief of | For temporary relief of | The device is intended | For temporary relief of |
| Use | | | | |
| | pain associated with<br>sore and aching<br>muscles in the shoulder,<br>waist, back, neck, upper<br>extremities (arm) and<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work<br>activities. | pain associated with<br>sore and aching<br>muscles in the lower<br>back due to strain from<br>exercise or normal<br>household and work<br>activities.<br><br>For temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the upper<br>and lower extremities<br>(arm and/or leg) due to<br>strain from exercise or<br>normal household and<br>work activities. | for the relief of pain<br>associated with sore or<br>aching muscles of the<br>lower extremities (leg)<br>due to strain from<br>exercise or normal<br>household work<br>activities.<br><br>It is also intended for<br>the use of symptomatic<br>relief and management<br>of chronic, intractable<br>pain associated with<br>arthritis. | pain associated with<br>sore and aching<br>muscles in the shoulder,<br>waist, back, neck, upper<br>extremities (arm) and<br>lower (extremities) leg<br>due to strain from<br>exercise or normal<br>household work<br>activities. |
| | For temporary relief of<br>pain associated with<br>dysmenorrhea<br>(menstrual cramps)<br>when used with over-<br>the-counter pain<br>medications. | For symptomatic relief<br>and management of<br>chronic, intractable pain<br>and relief of pain<br>associated with arthritis. | | For temporary relief of<br>pain associated with<br>dysmenorrhea<br>(menstrual cramps)<br>when used with over-<br>the-counter pain<br>medications. |
| | For symptomatic relief<br>and management of<br>chronic, intractable pain<br>and relief of pain<br>associated with arthritis. | For the stimulation of<br>healthy muscles in<br>order to improve or | | |
| | To stimulate healthy<br>muscles in order to | | | |
{8}------------------------------------------------
{9}------------------------------------------------
| | improve and facilitate<br>muscle performance. | facilitate muscle<br>performance. | | |
|---------------------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------|---------------------------------------------------|
| Environments of Use | Clinics, hospital and<br>home environments<br>(over-the-counter) | for home use, indoor,<br>not intended for<br>professional use (over-the-counter) | Clinics, hospital and<br>home environments<br>(over-the-counter) | home healthcare<br>environment (over-the-counter) |
| Patient<br>Population | Adult | At least 15 years old<br>and 8 years intensive<br>reading experience<br>(school), no maximum. | Adult | Be used only by women<br>aged 16 and above. |
| Over-the-Counter<br>(OTC) | Yes | Yes | Yes | Yes |
{10}------------------------------------------------
| | | Subject Device | Primary Predicate | Secondary<br>Predicate | Third Predicate |
|----------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Trade/Device Name | | Wireless Electrical<br>Stimulator | Well Life Wireless<br>TENS/EMS<br>Stimulator, Models<br>WR-2605A/2605 | Focus TENS<br>Therapy, Model<br>PM710-M/-L | Livia |
| 510(k) Number | | K220997 | K161453 | K183215 | K183110 |
| Power Source(s) | | 3.7V Lithium<br>polymer (LiPo)<br>battery | 3.7V Lithium<br>polymer (LiPo)<br>battery | Rechargeable<br>Lithium-ion battery | 3.7V Lithium ion<br>battery<br>(rechargeable) |
| Method of Line Current | | Output is electrically<br>disabled when<br>connect to charger,<br>by means of<br>microprocessor<br>charging circuit | Output is electrically<br>disabled when<br>connect to charger,<br>by means of<br>microprocessor<br>charging circuit | N/A (internal power<br>source) | Output is electrically<br>disabled when<br>connect to charger,<br>by means of<br>microprocessor<br>charging circuit |
| Patient Leakage Current | - Normal Condition (µA) | <10uA | <10uA | <10uA | <10uA |
| Patient Leakage Current | - Single Fault Condition (μΑ) | <50uA | <50uA | <50uA | <50uA |
| Average DC current through | electrodes when device is on<br>but no pulse are being<br>applied (uA) | 0 | 0 | 0 | 0 |
| Number of output Modes | | 1 | 1 | 1 | 1 |
| Number of | Synchronous or | 2 channel each | 1 channel each | 1 channel each | 1 channel each |
{11}------------------------------------------------
| Output<br>Channels: | Alternating | Stimulator Pod<br>Alternating in one<br>Stimulator Pod | Stimulator Pod | Stimulator device | Stimulator device |
|-----------------------------------------------|-----------------------------------|--------------------------------------------------------|-------------------|-------------------|------------------------------------------------------------------------------------------------|
| | Method of<br>Channel<br>Isolation | Microprocessor<br>stagger output time<br>of 2 channels | N/A | N/A | N/A |
| Regulated Current or<br>Regulated Voltage | | Regulated Voltage | Regulated Current | Regulated Current | Regulated Current |
| Software/Firmware/<br>Microprocessor Control? | | Yes | Yes | Yes | Yes |
| Automatic Overload Trip | | No | No | No | Yes |
| Automatic No-Load Trip | | Yes | No | Yes | Yes |
| Automatic Shut Off | | Yes | Yes | Yes | Yes |
| User Override Control | | Yes | Yes | Yes | Yes |
| Indicator<br>Display: | On/Off Status | Yes | Yes | Yes | Yes |
| | Low Battery | Yes | Yes | Yes | Yes |
| | Voltage/Current<br>Level | Yes | Yes | Yes | Yes |
| Timer Range (minutes) | | 5~60 | 5 ~ 60 | 30 | The Livia has no<br>internal timer as<br>there is no<br>treatment time<br>limitation for using |
{12}------------------------------------------------
| | | | | the Livia. |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Compliance with 21 CFR 898? | Yes | N/A (no patient cable) | N/A (no patient cable) | Yes |
| Weight (g) | Remote Control:<br>44.1g (1.6oz)<br>Stimulator Pod:<br>34.2g (12.1oz)<br>Brace: | Remote Control:<br>60g (2.1oz)<br>Stimulator Pod: 30g (1.1oz) | Device: 55g (1.9oz)<br>Knee pad: 20g (0.7oz)<br>Knee band M: 27g (1.0oz)<br>Knee band L: 29g (1.0oz)<br>Charger: 100g (3.5oz) | 36g (1.3oz) |
| Dimensions (mm) | Remote Control:<br>120 (4.73") x<br>40 (1.58") x<br>12.5 (0.49")<br><br>Stimulator Pod:<br>$φ$ 61 (2.36") x<br>14.77 (0.58")<br><br>Brace:<br>461 (18.15") x<br>207 (8.15") | Remote Control:<br>156.98 (6.18") x<br>47 (1.85") x<br>15.50 (0.61")<br><br>Stimulator Pod:<br>$φ$ 60 (2.36") x<br>15.65 (0.62") | Device:<br>60 (2.36") ×<br>72 (2.83") ×<br>16 (0.63")<br><br>Charger:<br>90 (3.54") x<br>80 (3.15") ×<br>23.5 (0.93")<br><br>Knee pad:<br>130 (5.12") ×<br>60 (2.36") ×<br>16 (0.63") | 55 (2.17") x<br>55 (2.17") x<br>18 (0.71") |
{13}------------------------------------------------
| | | | Knee Band M:<br>385 (15.16") ×<br>64 (2.52")<br><br>Knee Band L:<br>450 (17.72") ×<br>64 (2.52") | |
|---------------------------------------|---------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Housing Materials and<br>Construction | PC/ABS plastic | PC/ABS plastic | Not publicly<br>available | PC/ABS plastic |
| Operating conditions | 0 to 40°C (32 to<br>104°F)<br>10 to 90% RH<br>700 to 1060 hPa | 10 to 40°C (50 to<br>104°F)<br>40-90% RH<br>700 to 1013 hPa | 10 to 40°C (50 to<br>104°F)<br>30 to 80 %RH<br>700 to 1060 hPa<br>(noncondensing) | 5 to 40°C (41 to<br>104°F)<br>15 to 93% RH<br>700 to 1060 hPa |
| Storage conditions | -5 to 40°C (23 to<br>104°F)<br>0 to 90% RH | -10 to 60°C (14 to<br>140°F)<br>30-95% RH. | 0 to 40 °C (32 to<br>104°F)<br>30 to 80 % RH<br>(noncondensing) | -25 to 70°C (-13 to<br>158°F)<br>15 to 93% RH |
| Transporting conditions | -5 to 40 °C (23 to<br>104°F)<br>0 to 90% RH | -10 to 60°C (14 to<br>140°F),<br>30-95% RH. | -20 to 60 °C (-4 to<br>140°F)<br>10 to 90 % RH<br>(noncondensing) | -25 to +70°C (-13 to<br>158°F)<br>15 to 93% RH |
| Electrode style | Self-adhesive<br>Reusable | Self-adhesive<br>Reusable | HV-KNPAD-Z<br>Reusable | Self-adhesive<br>Reusable |
| Patient Contact Accessory | Yes | Yes | Yes | Yes |
{14}------------------------------------------------
#### Table 3. Substantial Equivalence Table – D Comparison
| | Subject Device | Primary Predicate | Secondary Predicate | Third Predicate |
|------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------|-------------------------------------------|
| Trade/Device Name | Wireless Electrical<br>Stimulator (RS-18,<br>RS-28, RS-38) | Well Life Wireless<br>TENS/EMS<br>Stimulator, Models<br>WR-2605A/2605 | Focus TENS<br>Therapy, Model<br>PM710-M/-L | Livia |
| 510(k) Number | K220997 | K161453 | K183215 | K183110 |
| Waveform | Biphasic,<br>Symmetrical | Biphasic,<br>Symmetrical | Biphasic,<br>Symmetrical | Biphasic,<br>Symmetrical |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular |
| Maximum<br>Output Voltage<br>(volts) | @500Ω<br>40<br>@2KΩ<br>40<br>@10kΩ<br>40 | @500Ω<br>40<br>@2KΩ<br>75<br>@10kΩ<br>133.5 | @500Ω<br>45<br>@2KΩ<br>68.6<br>@10kΩ<br>78.5 | @500Ω<br>50<br>@2KΩ<br>64<br>@10kΩ<br>64 |
| Maximum<br>Output Current<br>(mA) | @500Ω<br>80<br>@2kΩ<br>20<br>@10kΩ<br>4 | @500Ω<br>80<br>@2kΩ<br>37.5<br>@10kΩ<br>1.34 | @500Ω<br>90<br>@2kΩ<br>34.3<br>@10kΩ<br>7.9 | @500Ω<br>50<br>@2kΩ<br>31<br>@10kΩ<br>6.4 |
| Pulse Width (usec) | 40-400 | 100-520 | 60 | 100 |
| Frequency (Hz) | 1-150 | 2-60 | 1-250 | 100 |
| For<br>multiphasic<br>waveforms<br>only: | Symmetrical<br>phases<br>Phase<br>Duration<br>(µsec) | Yes | N/A | Yes |
| Net Charge(µC per pulse)<br>(@500Ω) (uC) | 0 | 0 | 0 | 0 |
| Maximum Phase Charge | 16 | 20.8 | 5.4 | 6.4 |
{15}------------------------------------------------
| (@500Ω) (μC) | Maximum Current Density<br>(@500Ω) (mA/cm²) r.m.s. | Maximum Average Current<br>(average absolute value),<br>(mA) | Maximum Average Power<br>Density (@500Ω) (W/cm²) | Burst Mode | Electrical Safety & EMC | Biocompatibility |
|-------------------------|----------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------|------------------|
| 0.96 | 4.8 | 9.4E-03 | (a) Pulses per<br>burst | IEC 60601-1, IEC<br>60601-1-2, IEC<br>60601-2-10, IEC<br>60601-1-11 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | |
| 0.25 | 2.496 | 1.76E-03 | (b) Bursts per<br>second | ES 60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10, IEC<br>60601-1-11 | Not publicly<br>available | |
| 0.97 | 2.7 | 7.59E-03 | (c) Burst<br>duration | ES 60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10, IEC<br>60601-1-11 | Not publicly<br>available | |
| 0.38 | 1.19 | 2.05E-03 | (d) Duty cycle:<br>Line(b) x Line<br>(c) | IEC 60601-1, IEC<br>60601-1-2, IEC<br>60601-2-10 | ISO 10993-5<br>ISO 10993-10 | |
| ON Time (seconds) N/A | | | | N/A | | |
| OFF Time (seconds) N/A | | | | N/A | | |
| Additional Features N/A | | | | N/A | | |
{16}------------------------------------------------
| | Subject Device | | Predicate /<br>Reference Device | | | | |
|-----------------------|------------------------|-------------------|---------------------------------|------------------|----------------------|--------------------|-------------------|
| Program | Pulse<br>Width(<br>us) | Frequency<br>(Hz) | Trade/Device<br>Name | 510(k)<br>Number | Program | Pulse<br>Width(us) | Frequency<br>(Hz) |
| Acute Pain | 175 | 2, 4 | DJO<br>Primera | K15322<br>4 | P03 | 175 | 2 |
| Chronic<br>Pain | 210 | 2/10/20/40/8<br>0 | Well-Life | K16145<br>3 | TENS-P8 | 210 | 2.45~245 |
| Soreness | 200 | 150 | DJO<br>Primera | K15322<br>4 | P04 | 200 | 150 |
| Back Pain | 250/150 | 2/70 | DJO<br>Primera | K15322<br>4 | HAN | 250/150 | 2/70 |
| Menstrual<br>Cramping | 100 | 100 | Livia | K18311<br>0 | N/A | 100 | 100 |
| Arthritis | 210 | 2.45~245 | Well-Life | K16145<br>3 | TENS-P8 | 210 | 2.45~245 |
| Warm-up | 200 | 12 | DJO<br>Primera | K15322<br>4 | P10 | 200 | 12 |
| Recovery | 220 | 10/8/6/4/2/1 | PowerDot | K17287<br>6 | Active<br>Recovery | 200 | 10/8/6/4/2/<br>1 |
| Atrophy | 250 | 50/1 | Cefar | K02080<br>3 | P10 | 250 | 50/1 |
| Strength | 400 | 85/100 | PowerDot | K17287<br>6 | Strength | 400 | 75~100 |
| Endurance | 300 | 5/10/15/20 | PowerDot | K17287<br>6 | Endurance | 300 | 10~25 |
| Resistance | 300 | 35/50 | PowerDot | K17287<br>6 | Resistance<br>I & II | 300 | 35/50 |
| Comfort | 220 | 7/5/3/1 | PowerDot | K17287<br>6 | General | 200 | 7/5/3/1 |
| Kneading | 300 | 40-99 | Well-Life | K16145<br>3 | EMS-P1 | 300 | 40-99 |
| Tingling | 200/100 | 65/100 | DJO<br>Primera | K15322<br>4 | P06 | 200/100 | 65/100 |
| Tapping | 300 | 5 | Well-Life | K16145<br>3 | EMS-P3 | 300 | 5 |
## Table 4. Comparison of Programs
{17}------------------------------------------------
| Relaxation | 200 | 12 | DJO Primera | K15322<br>4 | P08 | 200 | 12 |
|-------------|--------|----|-------------|-------------|-----|--------|----|
| Alternating | 70/180 | 80 | Cefar | K02080<br>3 | P7 | 70/180 | 80 |
Based on the comparison of indications for use (Table 1), all subject device and primary predicate devices (Well Life Wireless TENS/EMS K161453) have the same intended use on TENS and EMS. Subject device and third predicate (Livia K183110) both have the indications for use for temporary relief of pain associated with dysmenorrhea. And subject device also has the same environments of use and patient population as secondary predicate (Focus TENS K183215).
From the technical characteristics (Table 2) and technical characteristics comparison (Table 3), the following conclusions can be drawn. Although there are some differences in mechanical, environmental conditions, and output specifications among the subject device and the predicate devices, the subject device complies with a range of FDA- Recognized consensus standards and guidelines. This demonstrates that these differences do not raise any new questions about safety or efficacy.
#### Performance Data
The following performance data are provided to substantial equivalence determination.
#### Non-clinical Testing
A series of safety and performance tests, as follows, were conducted on the subject device in accordance with FDA recognized consensus standards and/or guidance:
#### Biocompatibility testing
The biocompatibility evaluation for the "Wireless Electrical Stimulator" was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,", and International Standard ISO 10993-1 as recognized by FDA. The self-adhesive reusable electrode for "Wireless Electrical Stimulator" is identical to the reference device (ZMI Self-Adhesive Electrodes K180865). The brace was tested for the following:
- ISO 10993-5 Cytotoxicity ●
- . ISO 10993-10 Sensitization & Irritation
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the "Wireless Electrical Stimulator". The system complies with the IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC. The stimulation function test has also been conducted to verify the output specifications of the device according to "Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning".
{18}------------------------------------------------
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documented which provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" The software for this device was considered as a "Moderate" level of concern, since a failure in the software could result in minor injury to the patient or operator.
#### Human Factor and Usability Testing
Usability test were conducted and documented according to the recognized consensus standards of IEC 62366-1. The result of the usability demonstrated all the users in testing did not show any critical errors and represented the performance of using can be smoother by extra practices without specific usability problems. The "Wireless Electrical Stimulator" has been found to be reasonably safe and effective for the intended users, uses and use environments.
#### Performance Bench Testing (Devices)
To make sure the device has met the intended functions and specifications, the function performance tests had proven the Remote Control and Stimulator (or Receiver) Pod of "Wireless Electrical Stimulator" is safe and effective for the intended users, uses and use environments.
#### Performance Bench Testing (Electrode)
AC Impedance, Current Dispersion, Retention Force Test, and Reusability Test proof the Self-Adhesive Electrode for "Wireless Electrical Stimulator" is identical to and even better than the Referencing Device. The results also indicating that the electrodes are safe and effective for the intended users, uses and use environments.
### Mechanical and Stability Testing
- Button reliability testing ●
- . Lead wire connector and bending life testing
- Drop/Vibration Testing .
- . Temperature and humidity cycle test
- Device life test .
By conducting those mechanical and stability test, the result shown that the device can work well during the whole service life and being safe and effective for the intended users, uses and use environments.
#### Clinical Studies
Not applicable. Clinical testing was not performed to support this 510(k) submission.
#### Conclusions
The information submitted to the "Wireless Electrical Stimulator" does not raise new questions about safety or effectiveness and demonstrates with reasonable assurance based on established controls that the device is as safe and effective as a legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.