Neck Care Therapy, Model: AST-905A, AST-905D, AST-905H
K210756 · Shenzhen Osto Technology Company Limited · NUH · Jun 11, 2021 · Neurology
Device Facts
Record ID
K210756
Device Name
Neck Care Therapy, Model: AST-905A, AST-905D, AST-905H
Applicant
Shenzhen Osto Technology Company Limited
Product Code
NUH · Neurology
Decision Date
Jun 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of neck, back, arm, and leg, due to strain from exercise or normal household and work activities. - Neck Pad is used on back of neck. - Meridian Pad is used on shoulder, waist, back, arm and leg.
Device Story
Household multifunctional device for pain relief; uses low-frequency electrical stimulation and circular traction vibration. Inputs: user-selected pulse modes, intensity levels (50 levels), and heating levels (3 levels) via remote control or main unit buttons. Outputs: electrical pulses delivered through electrode pads or neck pad; heating. Used in home setting by patient. Healthcare provider does not operate; output used by patient to manage muscle pain. Benefits: temporary pain relief via nerve stimulation and heat.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 62366), and software validation. Biocompatibility of electrode/neck pads confirmed via ISO 10993-5 and ISO 10993-10 testing, noting identity to previously cleared predicate components.
Indicated for temporary relief of pain in shoulder, waist, back of neck, back, arm, and leg muscles due to exercise or normal household/work strain. Over-the-counter use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Health Expert Electronic Stimulator (model: AST-300L) (K190783)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
June 11, 2021
Shenzhen OSTO Technology Company Limited % Cassie Lee Submission Correspondent Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District. Guangzhou, Guangdong, China 510663
Re: K210756
Trade/Device Name: Neck Care Therapy, Model: AST-905A, AST-905D, AST-905H Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: March 10, 2021 Received: March 15, 2021
Dear Ms. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210756
Device Name NECK CARE THERAPY (Model: AST-905A, AST-905D, AST-905H)
Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
- Neck Pad is used on back of neck.
- Meridian Pad is used on shoulder, waist, back, arm and leg.
Type of Use (Select one or both, as applicable)
|--|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K210756
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
510(k) Owner's Name: Shenzhen OSTO Technology Company Limited Establishment Registration Number: 3011564440 Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street, Longgang District, Shenzhen City, Guangdong Province, China Tel: +86-755-29769546 Fax: +86-755-29769540 Contact Person: Li Yang (General Manager) Email: annaosto@163.com
# Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
# 2. Subject Device Information
Common Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter, Stimulator Trade Name: NECK CARE THERAPY, Model: AST-905A, AST-905D, AST-905H Classification Name: Transcutaneous electrical nerve stimulator for pain relief, Review Panel: Neurology Product Code: NUH Regulation Number: 882.5890 Regulation Class: II
# 3. Predicate Device Information Predicate Device 1:
510(K) Number: K172897 Company Name: Shenzhen OSTO Technology Co., Ltd. Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street Longgang District, Shenzhen City Guangdong Province, CHINA Trade/Device Name: Neck Care Therapy, Models: SYK-509B Common Name: Electronic Stimulator Regulation Number: 882.5890 Regulatory Class: II Product Code: NUH
# Predicate Device 2:
510(K) Number: K190783 Company Name: Shenzhen OSTO Technology Company Limited Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street, Longgang District,
{4}------------------------------------------------
Shenzhen City, Guangdong Province, China Trade/Device Name: Health Expert Electronic Stimulator (model: AST-300L) Common Name: Transcutaneous electrical nerve stimulator for pain relief Regulation Number: 882.5890, 890.5850 Regulatory Class: II Product Code: NUH, NGX
### 4. Device Description
## For Model AST-905A and AST-905H:
This instrument is a new generation of a household multifunctional device based on physics, modern microelectronics and clinical practices, it uses low-frequency electronic therapy, and circular traction vibration to temporarily alleviate the pain associated with sore and aching muscles in the shoulder, waist, neck, back, arm, and leg. The main unit is used for the back of the neck and the streamlined ring design follows the physiological curvature of the human body. The Meridian Pad is used on the shoulder, waist, back, arm and leg.
The device has 2 operation modes: Stimulation and heating. The stimulation has 2 modes for stimulating, and 50 output Intensity Levels. The heating only has 1 mode for heating and 3 temperature levels. The device can give a certain electrical pulse through 2 pairs of electrode pads that are placed on the skin to help users to enjoy stimulation on the shoulder, waist, back, arm, and leg when using the individual Meridian pads or on the back of the neck when using the Neck Pad.
The remote control of this device is user-friendly controlled because it has the operating elements of the ON/OFF button. left or right button, and increase or decrease button.
The device is equipped with accessories of electrode pads and an electrode wire.
The LCD display screen can show the selected mode, the stimulation, and the time remaining of an application mode.
The device can be successfully opened only when both the switch button of the remote control and the main unit turned on.
When the device connected to the remote control, the remote control is the only controller to select the pulse modes, temperature modes, pulse intensity, temperature levels and adjust the treatment time.
### For Model AST-905D:
This instrument is a new generation of a household multifunctional device based on physics, modern microelectronics and clinical practices, it uses low-frequency electronic therapy, and circular traction vibration to temporarily alleviate the pain associated with sore and aching muscles in the shoulder, waist, neck, back, arm, and leg. The main unit is used for the back of the neck and the streamlined ring design follows the physiological curvature of the human body. The Meridian Pad is used on the shoulder, waist, back, arm and leg.
The device has 2 operation modes: Stimulation and heating. The stimulation has 2 modes for stimulating, and 50 output Intensity Levels. The heating only has 1 mode for heating and 3 temperature levels. The device can give a certain electrical pulse through 2 pairs of electrode pads that are placed on the skin to help users to enjoy stimulation on the shoulder, waist, back, arm, and leg when using the individual Meridian pads or on the back of the neck when using the Neck Pad..
The user can control the device effectively by the buttons on the main unit. There are 5 operation buttons: switch button, "M" button, ""C" button and "—" button. The switch button can help user to turn on/off the device, the "M" button intended to select the "C" button intended to select the temperature mode and the "-" button and "--" button intended to increase or decrease the Intensity Levels or temperature levels.
{5}------------------------------------------------
The device is equipped with accessories of Meridian pads and an electrode wire.
# 5. Intended Use / Indications for Use
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of neck, back, arm, and leg, due to strain from exercise or normal household and work activities. - Neck Pad is used on back of neck.
- Meridian Pad is used on shoulder, waist, back, arm and leg.
# 6. Test Summary
NECK CARE THERAPY, Model: AST-905A, AST-905D, AST-905H has been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1-11 and IEC 60601-2-10 standards �
- ◆ Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Usability test according to IEC 62366 standard
- ◆ Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- ◆ The waveform test has also been conducted to verify the output specifications of the device according to "Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning".
- � Biocompatibility:
The body-contacting components of this device are Electrode Pads and Neck Pad. And the Electrode Pads and Neck Pad comply with biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization) because the Electrode Pads and Neck Pad of the subject device are identical to those that had been cleared under K172897 (which clearance as K172897 on 12/07/2018) and been marketed to US market.
So. we have reason to believe that the Electrode Pads and Neck Pad are safe for the users. The Electrode Pads and Neck Pad comply with the following standards.
- ISO 10993-5, Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity;
- ISO 10993-10, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization.
### 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of NECK CARE THERAPY (Model: AST-905A, AST-905D, AST-905H) is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark | |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name and<br>Model | NECK CARE<br>THERAPY<br>Model: AST-905A,<br>AST-905D and AST-<br>905H | Neck Care Therapy<br>AST-905B or SYK-<br>509B | Health Expert<br>Electronic<br>Stimulator (model:<br>AST-300L) | -- | |
| 510(k) Number | K210756 | K172897 | K190783 | -- | |
| Intended Use &<br>Indications for Use | To be used for<br>temporary relief of pain<br>associated with sore<br>and aching muscles in<br>the shoulder, waist,<br>back of neck, back,<br>arm, and leg, due to<br>strain from exercise or<br>normal household and<br>work activities.<br>- Neck Pad is used on<br>back of neck.<br>- Meridian Pad is used<br>on shoulder, waist,<br>back, arm and leg. | To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist, back<br>of neck, back, arm,<br>and leg, due to strain<br>from exercise or<br>normal household<br>and work activities.<br>— Neck Pad is used<br>in back of neck.<br>— Meridian Pad is<br>used in shoulder,<br>waist, back, and arm. | PMS (Mode 1~8)<br>It is intended to<br>stimulate healthy<br>muscles in order to<br>improve and<br>facilitate muscle<br>performance.<br>TENS (Mode 9~25)<br>To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, back of the<br>neck, arm, leg, and<br>foot due to strain<br>from exercise or<br>normal household<br>work activities by<br>applying current to<br>stimulate nerve. | SE | |
| Power Source(s) | Main Unit: 3.7Vdc,<br>2200mAh lithium<br>battery<br>Power Adaptor:<br>Input: 100~240Vac,<br>50/60Hz, 0.2A;<br>Output: 5Vdc, 1A<br>Remote Control:<br>Battery: 3Vdc, 1.5V<br>AAA x 2 | Main Unit:<br>Power Adaptor:<br>Input:100~240Vac,<br>50/60Hz, 0.2A;<br>Output: 5Vdc, 1A<br>Battery: 3.7Vdc,<br>2200mAh<br>Remote Control:<br>Battery: 3Vdc, AAA x<br>2<br>Unit Input: 5Vdc, 1A | Adaptor Input:<br>100-240Vac, 50-<br>60Hz, 0.1A<br>Output: 5Vdc, 2A<br>Unit Input: 5Vdc, 2A | SE | |
| Method of Line<br>Current Isolation | Type BF Applied Part | Type BF Applied Part | Type BF Applied<br>Part | SE | |
| Patient<br>Leakage<br>Current | NC<br>SFC | AC: 54.5μA,<br>DC: 0.5μA<br>AC:120.0μA,<br>DC: 0.6μA | AC: 54.5μA,<br>DC: 0.5μA<br>AC:120.0μA,<br>DC: 0.6μA | AC: 54.5μA, DC:<br>0.5μA<br>AC: 120.0μA, DC:<br>0.6μA | SE |
| Average current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied | | < 0.01μA | < 0.01μA | < 0.01μA | SE |
| Number of Modes | | 3 | 2 | 25 | SE<br>Note 1 |
| Number of<br>Output Channel | | 1 | 1 | 2 | SE |
| Output Intensity Level | | For stimulating: 50<br>For heating: 3 | 50 steps | 99 steps | SE<br>Note 1 |
| Heating temperature | | 30-40°C | N/A | 30-40°C | SE |
| Synchronous or<br>Alternating? | | Synchronous | Synchronous | Synchronous | SE |
| Method of Channel<br>Isolation | | Voltage Transform<br>Isolation | Voltage Transform<br>Isolation | Voltage Transform<br>Isolation<br>"Body+" and "Body-"<br>buttons for body<br>channel,<br>" Sole+" and "Sole-"<br>buttons for feet<br>channel | SE |
| Regulated Current or<br>Regulated Voltage? | | Voltage Control | Voltage Control | Voltage Control | SE |
| Software/Firmware/Mi<br>croprocessor Control? | | Yes | Yes | Yes | SE |
| Automatic Overload<br>Trip | | No | No | No | SE |
| Automatic No-Load<br>Trip | | No | No | No | SE |
| Automatic Shut Off | | Yes | Yes | Yes | SE |
| User Override Control | | Yes | Yes | Yes | SE |
| Indicator<br>Display | On/Off<br>Status | Yes | Yes | Yes | SE |
| | Low<br>Battery | Yes | Yes | No | SE |
| | Voltage/<br>Current<br>Level | Yes | Yes | Yes | SE |
| Timer Range | | 5-30 min | 5-30 min | 25 to 60 min | SE |
| Weight | | | | | |
| | For Main Unit:<br>AST-905A: 193g;<br>AST-905D: 229g;<br>AST-905H: 178g<br>For Electrode:<br>Patch Electrode: 44g | Main Unit:<br>SYK-509B: 222g<br><br>Electrode:<br>Patch Electrode: 44g | 2.1Kg (Without<br>accessories) | SE<br>Note 2 | |
| Dimensions | | | | | |
| | For Main Unit:<br>AST-905D:<br>$166.0*159.5*55.6$ mm<br>AST-905H:<br>$173.4*156.3*43.4$ mm<br>For Electrode:<br>Meridian Pad: $8.9*5.8$ cm<br>Effective area: $50.04$ cm2<br>Neck Pad: $42.3$ mm x $29.5$ mm | Main Unit:<br>SYK-509B:<br>$187.2*169*67.3$ mm<br><br>Patch Electrode:<br>$8.9*5.8$ cm | $429.2$ mm x $401$ mm<br>x $152.8$ mm | SE<br>Note 2 | |
| Housing Materials and<br>Construction | For Electrode Pads:<br>White silica gel, Black<br>conductive silicone,<br>Transparent<br>conductive adhesive<br>silicone, Transparent<br>PET silicone | Electrode Pads:<br>White silica gel,<br>Black conductive<br>silicone, Transparent<br>conductive adhesive<br>silicone, Transparent<br>PET silicone | Main unit: ABS<br>plastic | SE | |
| | For Neck Pad:<br>Stainless steel | Sticky metal sheet:<br>Nickel plating | No publicly available | | |
| | For Unit Housing: ABS<br>plastic | Unit Housing: ABS<br>plastic | No publicly available | | |
| Waveform | Pulsed, symmetric,<br>biphasic | Pulsed, symmetric,<br>biphasic | Pulsed, symmetric,<br>biphasic | SE | |
| Shape | Rectangular, with<br>interphase interval | Rectangular, with<br>interphase interval | Rectangular, with<br>interphase interval | SE | |
| Maximum Output<br>Voltage | 44V±10% @ 500Ω | 44V±10% @ 500Ω | 44V±10% @ 500Ω | SE | |
| | 80V±10% @ 2KΩ | 80V±10% @ 2KΩ | 80V±10% @ 2KΩ | SE | |
| | 112V±10% @ 10KΩ | 112V±10% @ 10KΩ | 112V±10% @ 10KΩ | SE | |
| | | | | | |
| Maximum Output<br>Current | 88mA±10% @ 500Ω | 88mA±10% @ 500Ω | 88mA±10% @<br>500Ω | SE | |
| | 40mA±10% @ 2KΩ | 40mA±10% @ 2KΩ | 40mA±10% @ 2KΩ | SE | |
| | 11.2mA±10% @ 10KΩ | 11.2mA±10% @<br>10KΩ | 11.2mA±10% @<br>10KΩ | SE | |
| Pulse Duration | 120μs | 120μs | 120μs | SE | |
| Pulse frequency | 77.3Hz | 77.3Hz | 77.3Hz | SE | |
| Net Charge (per<br>pulse) | 0μC @ 500Ω<br>Method: Balanced<br>waveform | 0μC @ 500Ω<br>Method: Balanced<br>waveform | 0μC @ 500Ω<br>Method: Balanced<br>waveform | SE | |
| Maximum Phase<br>Charge @ 500Ω | 10.56μC | 12.78μC | 10.56μC | SE | |
| Maximum Average<br>Current@ 500Ω | 1.63 mA | 0.968 mA | 1.63 mA | SE | |
| Maximum Current<br>Density (r.m.s) @<br>500Ω | 0.0326 mA/cm² | 0.235 mA/cm²…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.