K210223 · Top-Rank Health Care Co., Ltd. · NUH · Dec 15, 2021 · Neurology
Device Facts
Record ID
K210223
Device Name
Transcutaneous Electrical Nerve Stimulator
Applicant
Top-Rank Health Care Co., Ltd.
Product Code
NUH · Neurology
Decision Date
Dec 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TOPTENS-01 can be used for the Symptomatic relief of chronic intractable pain, Post traumatic pain adjunctive treatment, Post-surgical pain adjunctive treatment.
Device Story
TOPTENS-01 is a transcutaneous electrical nerve stimulator (TENS) for pain relief. Device applies low-frequency electrical current to patient skin via electrodes to treat pain through electrochemical and electrophysiological reactions. System comprises host unit, built-in DC 4.5V power supply (3 AAA batteries), lead wires, and pads. Features 2 independent output channels (A&B) with adjustable intensity; LCD screen displays remaining stimulation time (default 30 minutes). Offers 9 stimulation modes with varying frequency (60-150 Hz) and pulse width (100-260 μs) settings tailored to specific anatomical locations (neck, shoulder, lumbar, elbow, knee, hip). Used in home healthcare environments by patients. Output affects nerve stimulation to provide pain relief. Device includes automatic no-load trip and automatic shut-off safety features.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
ABS plastic enclosure; DC 4.5V battery-powered; Type BF applied part; 2 independent output channels; biphasic symmetrical square waveform; LCD interface. Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 60601-1-11, and IEC 62133.
Indications for Use
Indicated for symptomatic relief of chronic intractable pain, and as adjunctive treatment for post-traumatic and post-surgical pain in patients.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
December 15, 2021
Top-Rank Health Care Co., Ltd. % Yuling Chen Official Correspondent Microkn Business Consulting (Shanghai) Co., Ltd. Room 1219, Block A, No3699, Gonghexin Road, Jingan District Shanghai, Shanghai 200040 China
Re: K210223
Trade/Device Name: Transcutaneous Electrical Nerve Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: November 24, 2020 Received: January 28, 2021
Dear Yuling Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K210223
Device Name
Transcutaneous Electrical Nerve Stimulator
Indications for Use (Describe)
The TOPTENS-01 can be used for the Symptomatic relief of chronic intractable pain, Post traumatic pain adjunctive treatment, Post-surgical pain adjunctive treatment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Product name | Transcutaneous Electrical Nerve Stimulator |
|---------------|--------------------------------------------|
| Model | TOPTENS-01 |
| Applicants | Top-Rank Health Care Co.,Ltd. |
| Document Name | 510(k) Summary |
| Objective | 510 (k) Application |
{4}------------------------------------------------
## 510(k) Summary
## 1. Contact information
## 1.1. Applicant
Applicant Name: Top-Rank Health Care Co.,Ltd.
Address: CHECUN VILLAGE, DONGGUAN ST.
Shang-Yu District Shaoxing city, Zhejiang province, China CN 312352
Contact Person: Franck Lou
Title: Manager
Telephone: +0575-82563666
E-mail: info@etop-rank.com
### 1.2. Consultant
Company: Microkn Business Consulting (Shanghai) Co., Ltd.
Address: Room 1219, Block A, No 3699, Gonghexin Road, Jingan District,
Shanghai, China
Contact Person: Yuling.Chen
Telephone: +86 15021397762
Email: yuling.chen@microkn.com
{5}------------------------------------------------
## 2. Device information
- . Trade Name: Transcutaneous Electrical Nerve Stimulator
- Model(s): TOPTENS-01 ●
- Classification: II ●
- Product Code: NUH ●
- Regulation Number: 21 CFR 882.5890 ●
- 510(K)Number: K210223 .
## 3. Indications for Use
- Symptomatic relief of chronic intractable pain ●
- Post traumatic pain adjunctive treatment ●
- . Post surgical pain adjunctive treatment
## 4. Legally Marketed Predicate Device
Product name:Transcutaneous Electrical Nerve Stimulator (T.E.N.S)
510(k) Number: K152374
Product Code: NUH
Manufacture: Fuji Dynamics Ltd
## 5.Description of the device
The proposed device is transcutaneous electrical nerve stimulator for pain
{6}------------------------------------------------
relief. this is a device used to apply an electrical current to electrodes on a patient's skin to treat pain. It only has one model: TOPTENS-01.
TOPTENS-01 is mainly composed: host, built-in power supply, output lead and electrode.
TOPTENS-01 power supply is DC4.5V, using built-in removable 3 AAA batteries.
TOPTENS-01 comes with 2 independent output channels (A&B). Users can be adjust independently the two output channels Intensity by pressing the Left and right sides Intensity adjustment key.
LCD screen display the remaining stimulation time in the top left corner. The default stimulation time is 30 minutes
TOPTENS-01 has 9 types stimulation modes: Mode1, Mode2, Mode 3, Mode 4, Mode 5, Mode6, Mode 7, Mode 8, Mode 9. These stimulation modes can be selected by pressing the <P> key.The detailed difference shown in Table 1.
| Table 1 the Difference of Models | | | | | |
|----------------------------------|-------------------|---------------------|-------------------------------------------|--------------------------------------------------------|------------------|
| Mode | Frequency<br>(Hz) | Pulse Width<br>(μs) | Max<br>Amplitude<br>(V)(Vpeak,<br>@500 Ω) | Output Mode | Position for Use |
| Mode 1 | 60~100 | 100~150 | 33 | Frequency<br>Modulation &<br>Pulse Width | Back of Neck |
| Table 1 the Difference of Models | | | | | |
| Mode | Frequency<br>(Hz) | Pulse Width<br>(μs) | Max amplitude<br>(V)(Vpeak,<br>@500 Ω) | Output Mode | Position for Use |
| | | | | Modulation | |
| Mode 2 | 80~100 | 260 | 23 | Frequency<br>Modulation | Shoulder |
| Mode 3 | 100~150 | 100 | 33 | Frequency<br>Modulation | Lumbar pain |
| Mode 4 | 100 | 100 | 23 | Continuous wave | Elbow |
| Mode 5 | 120 | 100~150 | 33 | Pulse Width<br>Modulation | Knee |
| Mode 6 | 100~150 | 200 | 23 | Frequency<br>Modulation | Hip |
| Mode 7 | 100 | 260 | 23 | Continuous wave | Elbow 2 |
| Mode 8 | 100 | 100 | 33 | Continuous wave | Knee 2 |
| Table 1 the Difference of Models | | | | | |
| Mode | Frequency<br>(Hz) | Pulse<br>(μs) | Max amplitude<br>(V) (Vpeak,<br>@500 Ω) | Output Mode | Position for Use |
| Mode 9 | 60~100 | 100~160 | 33 | Frequency<br>Modulation &<br>Pulse Width<br>Modulation | Back of Neck 2 |
{7}------------------------------------------------
{8}------------------------------------------------
Note 1: Do not place the electrodes on the Spine.
Note 2: Do not place the electrodes directly over joints.
The main components of proposed device shown in Table 2.
| Table 2 Main Components of Proposed Device | |
|--------------------------------------------|-------------------------------------------------------------------|
| Components | Function Description |
| Host | Control circuit signal, user interface, key<br>interaction |
| Built-in power<br>supply | Energy Supply |
| Pads | Is responsible for sending electrical<br>signals to the skin |
| Set of Lead Wires | Is responsible for transmitting electrical<br>signals to the Pads |
{9}------------------------------------------------
## 6.Non-Clinical Test conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-2-10:2016, Medical electrical equipment Part 2-10: Particular requirements for the safety of nerve and muscle stimulators
- ANSI/AAMI/ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(r)2012 and a2:2010/(r)2012. Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- IEC 60601-1-11-2015 Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2: General requirements for basic safety and essential . performance-Collateral standard: Electromagnetic disturbances-Requirements and tests. The EMC property meets the requirements.
## 7.Clinical Test Conclusion
6
{10}------------------------------------------------
No clinical study is included in this submission.
## 8.Predicate Devices
Product name:Transcutaneous Electrical Nerve Stimulator (T.E.N.S)
510(k) Number: K152374
Product Code: NUH
Manufacture: Fuji Dynamics Ltd
## 9.Substantially Equivalent (SE) Comparison
The Transcutaneous Electrical Nerve Stimulator has been carefully compared to legally marketed devices with respect to intended use, configuration, principle of operation (Table 3), and performance specifications (Table 4).
| Table 3 General Comparison | | | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Item | Proposed Device | Predicate Device | Remark |
| | TOPTENS-01 | Transcutaneous Electrical<br>Nerve Stimulator (T.E.N.S) | |
| Product<br>Code | NUH | NUH | Same |
| 510(k)<br>Number | K210223 | K152374 | -- |
| Regulation | 21 CFR 882.5890 | 21 CFR 882.5890 | Same |
| Table 3 General Comparison | | | |
| Item | Proposed Device | Predicate Device | Remark |
| | TOPTENS-01 | Transcutaneous Electrical<br>Nerve Stimulator (T.E.N.S) | |
| Number | | | |
| Indications<br>for Use | Symptomatic relief of chronic intractable pain Post traumatic pain adjunctive treatment Post surgical pain adjunctive treatment | Used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. | Same |
| Principle of<br>Operation | The device applies low-frequency current to human tissue through the body surface skin, and treats specific parts of the human body through the generated electrochemical and electrophysiological reactions. | The device applies low-frequency current to human tissue through the body surface skin, and treats specific parts of the human body through the generated electrochemical and electrophysiological reactions. | Same |
| Table 3 General Comparison | | | |
| Item | Proposed Device | Predicate Device | Remark |
| | TOPTENS-01 | Transcutaneous Electrical<br>Nerve Stimulator (T.E.N.S) | |
| Type of use | OTC | OTC | Same |
{11}------------------------------------------------
{12}------------------------------------------------
# Table 4 Performance Comparison
| Table 4 Performance Comparison | | | | |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------|--------------|
| Item | Proposed Device | Predicate Device | Remark | |
| Power source | DC 4.5V, 3 AAA<br>batteries | DC 3.0V, 2 AAA<br>batteries | Difference 1 | |
| - Method of Line<br>Current Isolation | Type BF | Type BF | Same | |
| - Normal condition | 2μΑ | - | Difference 1 | |
| - Single fault<br>condition | N/A | N/A | Same | |
| User interface | By LCD display | By LCD display | Same | |
| Table 4 Performance Comparison | | | | |
| Item | Proposed Device<br>TOPTENS-01 | Predicate Device<br>Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | Remark | |
| Number of output models | 4 | 4 | Same | |
| Output channel | Two channels | Two channels | Same | |
| Number of output models | 1 | 1 | Same | |
| Number of output channels | 2 | 2 | Same | |
| Number of anatomical locations | 6 | 6 | Same | |
| Anatomical locations | Back of Neck, Shoulder, Lumbar pain, Elbow, Knee, Hip | Back of Neck, Shoulder, Lumbar pain, Elbow, Knee, Hip | Same | |
| -Synchronous or Alternating? | Synchronous | Alternating | Difference 2 | |
| -Method of channel isolation | By enclosure | By enclosure | Same | |
| Table 4 Performance Comparison | | | | |
| | Proposed Device | Predicate Device | | |
| Item | TOPTENS-01 | Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | Remark | |
| Regulated Current or<br>Regulated Voltage? | Regulated Voltage | Regulated Voltage | Same | |
| Software/Firmware/<br>Microprocessor<br>Control? | Software | Software | Same | |
| Automatic overload<br>trip | NO | NO | Same | |
| Automatic No Load<br>Trip? | Yes | Yes | Same | |
| Automatic shut off? | Yes | Yes | Same | |
| User Override<br>Control? | NO | NO | Same | |
| Indication Display | - | - | - | |
| -On/off status? | Yes | Yes | Same | |
| -Low battery? | Yes | Yes | Same | |
| Table 4 Performance Comparison | | | | |
| Item | Proposed Device<br>TOPTENS-01 | Predicate Device<br>Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | Remark | |
| -Voltage/<br>current<br>level? | Yes | Yes | Same | |
| Time range (min) | 5~60 minutes<br>(5,10,15,20,25,30,35,40,<br>45,50,55and 60 min<br>selectable) | Continuous, 15 min,<br>30 min, 45 min and<br>60 min selectable. | Difference 3 | |
| Compliance with<br>Voluntary<br>Standards? | Yes.<br>AAMI/ANSI ES<br>60601-1,<br>IEC 60601-1-2,<br>IEC 60601-2-10,<br>IEC 62133,<br>IEC 60601-1-11 | Yes.<br>AAMI/ANSI ES<br>60601-1, IEC<br>60601-1-2,<br>IEC 60601-2-10,<br>IEC 62133,<br>IEC 60601-1-11 | Same | |
| Compliance with 21<br>CFR 8988? | Yes | Yes | Same | |
| Weight (lbs.) | 0.18 (Battery Excluded) | 70g | | |
| Dimensions(in.) [W<br>× H × D]<br>For unit | 2.32x 4.96x1.18 | 38x127x20 | Difference 4 | |
| Housing materials | Plastic (ABS) enclosure | ABS | | |
| Table 4 Performance Comparison | | | | |
| Item | Proposed Device | Predicate Device | Remark | |
| | TOPTENS-01 | Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | | |
| construction | | | | |
| Waveform | biphasic symmetrical<br>Square waveform | biphasic symmetrical<br>Rectangular<br>waveform | Same | |
| Shape | Square | Square | Same | |
| Maximum<br>output<br>voltage<br>(Vp) | 33 @500Ω<br>60 @2kΩ<br>100 @10kΩ | 70 @500Ω<br>98 @2kΩ<br>108 @10kΩ | | |
| Max Output<br>Current (mA) | 66 @500Ω<br>30 @2kΩ<br>10 @10kΩ | 140 @500Ω<br>49 @2kΩ<br>10.8 @10kΩ | Difference 5 | |
| Pulse Width<br>Range(uS) | 100~260 | 104 | Difference 6 | |
| Frequency (Hz) | 60~150 | 90 | Difference 6 | |
| Table 4 Performance Comparison | | | | |
| Item | Proposed Device<br>TOPTENS-01 | Predicate Device<br>Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | Remark | |
| For<br>multi-prog<br>ram<br>waveforms<br>only | Symmetrical<br>phases? | Yes | Yes | Same |
| | Phase<br>Duration | 100~260 | 208 | Difference 6 |
| | Net Charge (mC per<br>pulse) | 0@500Ω,<br>Method: Balanced<br>waveform | 0@500Ω,<br>Method: Balanced<br>waveform | Same |
| | Maximum Average<br>Current (mA) | 2.22@500Ω | 0.9@500Ω | Difference 7 |
| | Maximum Phase<br>Charge, (µC) | 17.16@500Ω | 18.1@500Ω | Difference 7 |
| | Maximum Current<br>Density<br>(mA/cm², 500Ω,<br>r.m.s) | 0.14@500Ω | 2.28mA/cm²@500Ω | Difference 7 |
| | Maximum Power<br>Density | 0.00015@500Ω<br>(using smallest | 0.00225@500Ω<br>(using smallest | Difference 7 |
| Table 4 Performance Comparison | | | | |
| Item | Proposed Device | Predicate Device | Remark | |
| | TOPTENS-01 | Transcutaneous<br>Electrical Nerve<br>Stimulator (T.E.N.S) | | |
| , (W/cm²)<br>(using smallest<br>electrode conductive<br>surface area) | electrode 16cm²) | electrode 25cm²) | | |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
## Difference 1:
The new device TOPTENS-01 is substantial equivalence as the predicate device K152374. The differences in Power source specification(value) and Patient Leakage Current are the differences Because of the difference in structure and circuit layout, which only affect the safety performance of the device, the device is an internal power device that only uses battery power, which has lowest safety risks and has passed the safety test, so these differences will not cause any safety and effectiveness issues. Considering the same Insulation method and power source, they are substantial equivalence.
## Difference 2:
The new device TOPTENS-01 is substantial equivalence as the predicate device K152374. The difference between synchronous or alternating is that only the output method of the two channal is different, The final output
{19}------------------------------------------------
waveform of the two channal is the same 。 They has passed the safety test, so these differences will not cause any safety and effectiveness issues. Considering the same output waveform and Output Shape, they are substantial equivalence.
#### Difference 3:
The new device TOPTENS-01 is substantial equivalence as the predicate device K152374.
The time range of the new device is same, and the options can be increased, in order to facilitate different use requirements. The new device has passed the safety test and the maximum time value is the same, so these differences will not cause any safety and effectiveness issues.
The new device is just an expansion of the time range and the maximum time value is the same, it can be considered that they are essentially equivalent.
#### Difference 4:
The new device TOPTENS-01 is substantial equivalence as the predicate device K152374. The differences in Dimensions, Weight and Housing materials construction are the differences in design requirements between different devices, which only affect the safety performance of the device, but the new device passed Safety testing, so these differences will not cause any safety and effectiveness issues. Considering the same intended use, working
{20}------------------------------------------------
principle, output mode and output waveform, they are substantial equivalence.
### Difference 5:
The new device TOPTENS-01 is basically equivalent to the predicate device K152374. The difference between the maximum output voltage and maximum output current of the new device is only in the output amplitude range. This can be achieved by adjusting the appropriate output intensity to achieve the same output effect and meet the requirements of use.
Based on the same intended use, output waveform, and the new equipment has passed safety tests, these differences will not cause any new safety and effectiveness issues, they are essentially equivalent.
#### Difference 6:
The new device TOPTENS-01 is basically equivalent to the predicate device K152374. The difference between of Pulse Width, Frequency, Phase Duration of the new device is only in the pulse parameter . As the new device adapts to the needs of different treatment sites, a wider parameter range is required, and the parameter range of the new device can cover the predicate device, Based on the same output waveform,it can be considered to have the same validity and the predicate device.
There are no safety difference in the pulse parameters themselves. The safety risks related to these parameters are mainly as follows: Frequency is higher,
{21}------------------------------------------------
Pulse Width is wider, and Phase Duration is longer,The energy of the pulse per unit time is higher, but the pulse parameters meet the maximum current density <2mA/cm² And the maximum average power density <0.25W/cm² safety standard requirements, and passed the safety test.
Based on the same intended use, output waveform, and the new equipment has passed safety tests, these differences will not cause any new safety and effectiveness issues, they are essentially equivalent.
#### Difference 7:
The new device TOPTENS-01 is basically equivalent to the predicate device K152374. Maximum phase charge, maximum current density, and maximum power density are the main safety requirements. Although the some parameters of the new device is higher than that of the predicate device K152374, the value is close and at a low level. The maximum current density and the maximum power density are different because they are calculated by different electrodes, and the value is close and at a low level. But they all meet the safety standard requirements of maximum current density < 2mA/cm 2 and maximum average power density <0.25W/cm² ,and the new device has passed the safety test. These differences will not lead to any new safety and effectiveness issues. They are essentially equivalent.
{22}------------------------------------------------
Safety comparison has been done to validate the electrical safety of the
device (Table 5).
| Table 5 Safety Comparison | | | |
|----------------------------------|--------------------------------------------|-----------------------------------------------|--------|
| Item | Proposed Device | Predicate Device | Remark |
| EMC, Electrical and Laser Safety | | | |
| Electrical safety | Comply with IEC 60601-1,<br>IEC 60601-2-10 | Comply with IEC<br>60601-1, IEC<br>60601-2-10 | SE |
| EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
| Home<br>healthcare<br>enviroment | IEC 60601-1-11:2015 | IEC 60601-1-11:2015 | SE |
## Table 5 Safety Comparison
# 11.Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is
determined to be Substantially Equivalent (SE) to the predicate device.
{23}------------------------------------------------
{24}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.