HIVOX OTC Electrical Stimulator, EM59-1 TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities. SH: This function is designed to be used for temporary relief of minor aches and pains. HIVOX OTC Electrical Stimulator, EM59-2 TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg)due to strain from exercise or normal household work activities. EMS: This function is designed to be used for stimulating healthy muscles in order to improve and facilitate muscle performance. SH: This function is designed to be used for temporary relief of minor aches and pains.
Device Story
OTC electrostimulation device; models EM59-1 (TENS, SH) and EM59-2 (TENS, EMS, SH). Inputs: user-selected pre-set or custom stimulation programs. Operation: two independent output channels deliver biphasic electrical pulses via four self-adhesive electrode gel pads; includes superficial heat (SH) function. Used by patients in home/non-clinical settings for pain relief, muscle revitalization, and fitness. Output: electrical stimulation and heat to target tissues. Healthcare provider role: none (OTC). Benefit: temporary pain relief and improved muscle performance.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility, software validation, electromagnetic compatibility, electrical safety, and usability testing.
Indicated for temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, upper/lower extremities due to exercise or household strain; minor aches and pains; and stimulation of healthy muscles to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 10, 2021
Hivox Biotek Inc. Ruby Lu Regulatory Affairs Specialist 5F, No. 123, Xingde Road, Sanchong District, New Taipei City, 241 Taiwan
Re: K203574
Trade/Device Name: HIVOX OTC Electrical Stimulator, EM59-1, HIVOX OTC Electrical Stimulator, EM59-2 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX, IRT Dated: November 30, 2020 Received: December 7, 2020
Dear Ruby Lu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K203574
Device Name
HIVOX OTC Electrical Stimulator (EM59-1, EM59-2)
Indications for Use (Describe)
HIVOX OTC Electrical Stimulator, EM59-1
TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
SH: This function is designed to be used for temporary relief of minor aches and pains.
HIVOX OTC Electrical Stimulator, EM59-2
TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (leg) due to strain from exercise or normal household work activities.
EMS: This function is designed to be used for stimulating healthy muscles in order to improve and facilitate muscle performance.
SH: This function is designed to be used for temporary relief of minor aches and pains.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
| 1. Type of Submission | Traditional |
|-----------------------|-----------------------------------------------|
| 510(k) Number | K203574 |
| 2. Date of Summary | Mar. 2, 2021 |
| 3. Submitter | HIVOX BIOTEK INC. |
| Address: | 5F., No. 123, Xingde Rd., Sanchong Dist., New |
| | Taipei City 24158, Taiwan, R.O.C. |
| Phone: | +886-2-8511-2668 |
| Fax: | +886-2-8511-2669 |
| Contact: | Ruby Lu |
| | (Ruby.Lu@hivox-biotek.com) |
# 4. Identification of the Subject Device
| 510(k) number | K203574 |
|------------------------------|------------------------------------------------------------------|
| Proprietary/Trade name: | HIVOX OTC Electrical Stimulator |
| Models: | EM59-1, EM59-2 |
| Classification product code: | NUH |
| Subsequent product code: | NGX, IRT |
| Regulation number: | 1) 882.5890 |
| | 2) 890.5850 |
| | 3) 890.5740 |
| Regulation description: | 1) Transcutaneous electrical nerve stimulator for<br>pain relief |
| | 2) Powered muscle stimulator |
| | 3) Power heating pad |
| Review panel: | 1) Neurology |
| | 2) Physical Medicine |
| | 3) Physical Medicine |
| Device class: | II |
# 5. Identification of the Predicate Device #1
| 510(k) number: | K162517 |
|------------------------------|-----------------------------|
| Proprietary/Trade name: | Electrical Pulse Stimulator |
| Models: | PL-029K13 |
| Classification product code: | NUH, NGX, NYN, IRT |
| Regulation number: | 882.5890 |
{4}------------------------------------------------
| Regulation description: | Transcutaneous electrical nerve stimulator for pain<br>relief |
|-------------------------|---------------------------------------------------------------|
| Review panel: | Neurology |
| Device class: | II |
#### 6. Identification of the Predicate Device #2
| 510(k) number: | K190347 |
|------------------------------|------------------------------------------------------------------|
| Proprietary/Trade name: | HIVOX OTC Electrical Stimulator |
| Models: | EM49-1, EM49-2 |
| Classification product code: | NUH |
| Subsequent product code: | NGX |
| Regulation number: | 1) 882.5890 |
| | 2) 890.5850 |
| Regulation description: | 1) Transcutaneous electrical nerve stimulator for<br>pain relief |
| | 2) Powered muscle stimulator |
| Review panel: | 1) Neurology |
| | 2) Physical Medicine |
| Device class: | II |
## 7. Intended Use / Indications for Use of the Device
HIVOX OTC Electrical Stimulator, EM59-1
TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
SH: This function is designed to be used for temporary relief of minor aches and pains.
HIVOX OTC Electrical Stimulator, EM59-2
TENS: This function is designated to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg)due to strain from exercise or normal household work activities.
EMS: This function is designed to be used for stimulating healthy muscles in order to improve and facilitate muscle performance.
SH: This function is designed to be used for temporary relief of minor aches and pains.
{5}------------------------------------------------
# 8. Device Description
HIVOX OTC Electrical Stimulators, EM59-1 and EM59-2, fall into the
electrostimulation device category.
EM59-1 provides two basic functions, TENS and SH; EM59-2 provides three basic functions, TENS, EMS and SH:
(1) Electrical stimulation of nerve tracts (TENS)
(2) Electrical stimulation of muscle tissue (EMS)
(3) Superficial heat (SH)
Both two models feature two independent output channels and four self-adhesive electrode gel pads. They offer a wide range of functions for increasing general wellbeing, pain relief, maintaining physical fitness, relaxation, muscle revitalisation and combating tiredness. For these purpose, the user can either choose from pre-set programs or specify their own to suit the user's individual needs.
The size of the self-adhesive electrode gel pad is 45x45 mm supplied by Shaoxing DL Healthcare Co., Ltd. It is an OTC medical device cleared by FDA under K182111. On the other hand, the electrode made by PET, sponge, PCB and protect cases is designed by HIVOX BIOTEK INC.
# 9. Non-clinical Testing
A series of safety and performance tests, as follows, were conducted on the subject device in accordance with FDA recognized consensus standards and/or guidance:
- Shelf life (ASTM F1980-16)
- Biocompatibility (ISO 10993-1 Edition 4.0, ISO 10993-5 Edition 3.0 and ISO 10993-10 Edition 3.0)
- Software validation (IEC 62304 Edition 1.1)
- Electromagnetic compatibility and electrical safety (ANSI/AAMI ES60601-1:2015/(R)2012, IEC 60601-1-2 Edition 4.0, IEC 60601-1-11 Edition 2.0 and IEC 60601-2-10 Edition 2.1)
- Function test (Guidance Document for Powered Muscle Stimulator 510(K)s. Document issued on: June 9, 1999)
- Usability test (IEC 60601-1-6 Edition 3.1 and IEC 62366-1 Edition 1.0)
All the test results demonstrate the subject device, HIVOX OTC Electrical Stimulator (EM59-1, EM59-2), meets the requirements of its pre-defined acceptance criteria and intended use, and its substantially equivalent to the predicate device.
## 10. Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
{6}------------------------------------------------
### 11. Substantial Equivalence Comparison
The subject device, HIVOX OTC Electrical Stimulator (EM59-1, EM59-2), was compared to a predicate device and reference device respectively in the tables below:
| Comparison item | Subject device | | Predicate device #1 |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K203574 | | K162517 |
| Device Name | HIVOX OTC Electrical<br>Stimulator | | Electronic Pulse<br>Stimulator |
| Model | EM59-1 | EM59-2 | PL-029K13 |
| Manufacturer | HIVOX BIOTEK INC. | | JKH Health Co., Ltd. |
| Intended Use | EM59-1:<br>TENS: This function is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>upper extremities (arm)<br>and lower extremities<br>(leg) due to strain from<br>exercise or normal<br>household work activities.<br>SH: This function is<br>designed to be used for<br>temporary relief of minor<br>aches and pains.<br>EM59-2:<br>TENS: This function is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>upper extremities (arm) | | TENS Mode<br>To be used for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the shoulder,<br>waist, back, arm and leg<br>due to strain from exercise<br>or normal household and<br>work activities.<br>It is also intended for<br>symptomatic relief and<br>management of chronic,<br>intractable pain and relief<br>of pain associated with<br>arthritis.<br>PMS Mode<br>To stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. To<br>be used for the<br>improvement of muscle<br>tone and firmness, and for<br>strengthening muscles in<br>the arm, abdomen, legs, |
| Prescription or OTC | | OTC | OTC |
| FDA Product Code | | NUH, NGX, IRT | NUH, NGX, NYN, IRT |
| Power Source(s) | | Rechargeable battery | Rechargeable battery |
| Function and Design | | Electrical stimulation and heat | Electrical stimulation and heat |
| | | Heating Setting | |
| Maximum Temperature Setting | | 43°C | 43°C |
| Maximum | @ 500 Ω | 50.0 | 46.0 |
| Output Voltage | @ 2 kΩ | 90.0 | 90.4 |
| (Vp, ±20%) | @ 10 kΩ | 125 | 124 |
| Maximum | @ 500 Ω | 100 | 92.0 |
| Output Current | @ 2 kΩ | 45.0 | 45.2 |
| (mAp, ±20%) | @ 10 kΩ | 12.5 | 12.4 |
| Pulse Period (µs) | | 50 to 450 | 5.6 to 806 |
| Frequency (Hz) | | 1 to 150 | 1.24 to 178.5 |
| Maximum Phase Charge<br>(µC @ 500Ω) | | 45 | 16.9 |
| Maximum Current Density<br>(mA/cm² @ 500Ω) | | 0.667 | 3.29 |
| Maximum Power Density | | 0.0046 | 0.00165 |
| (W/cm² @ 500Ω) | | | |
| Output Patterns | ● Electrical stimulation<br>only<br>● Heat only<br>● Electrical stimulation<br>+ Heat simultaneously | ● Electrical stimulation<br>only<br>● Heat only<br>● Electrical stimulation<br>+ Heat simultaneously | |
Table 1 - Comparison to Predicate Device #1
{7}------------------------------------------------
{8}------------------------------------------------
# Table 2 – Comparison to Predicate Device #2
| Comparison item | Subject device | Predicate device #2 | |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K203574 | K190347 | |
| Device Name | HIVOX OTC Electrical<br>Stimulator | HIVOX OTC Electrical<br>Stimulator | |
| Model | EM59-1 EM59-2 | EM49-1 EM49-2 | |
| Manufacturer | HIVOX BIOTEK INC. | HIVOX BIOTEK INC. | |
| Intended use | EM59-1:<br>TENS: This function is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>upper extremities (arm)<br>and lower extremities<br>(leg) due to strain from<br>exercise or normal<br>household work activities.<br>SH: This function is<br>designed to be used for<br>temporary relief of minor<br>aches and pains.<br>EM59-2:<br>TENS: This function is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and | EM49-1:<br>TENS: This device is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>neck, upper extremities,<br>lower extremities,<br>abdomen and bottom due<br>to strain from exercise or<br>normal household work<br>activities.<br>EM49-2:<br>TENS: This device is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>neck, upper extremities,<br>lower extremities, | |
| | | aching muscles in the<br>shoulder, waist, back,<br>upper extremities (arm)<br>and lower extremities<br>(leg) due to strain from<br>exercise or normal<br>household work activities.<br><br>EMS: This function is<br>designed to be used for<br>stimulating healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance.<br><br>SH: This function is<br>designed to be used for<br>temporary relief of minor<br>aches and pains. | abdomen and bottom due<br>to strain from exercise or<br>normal household work<br>activities.<br><br>EMS: This device is<br>designed to be used for<br>stimulating healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. |
| Prescription or OTC | OTC | OTC | |
| FDA product code | NUH, NGX, IRT | NUH, NGX | |
| Power Source(s) | Battery powered, d.c.<br>3.7 V, 1 × built-in<br>rechargeable lithium-ion<br>battery | Battery powered, d.c.<br>4.5 V, 3 × AAA batteries | |
| Method of Line Current<br>Isolation | N/A<br>(internal power source) | N/A<br>(internal power source) | |
| Patient<br>Leakage<br>Current | Normal<br>condition<br>(μΑ) | 6.0 | 6.0 |
| | Single fault<br>condition<br>(μΑ) | 5.6 | 5.6 |
| Number of Output Modes | | TENS: 15<br>SH: 1 | TENS: 15<br>EMS: 35<br>SH: 1 |
| | | | |
| Number of<br>output<br>Channels | Synchronous<br>or<br>Alternating? | 2 Synchronous | 2 Synchronous |
| | Method of<br>Channel<br>Isolation | By electrical circuit and<br>software | By electrical circuit and<br>software |
| Regulated Current or<br>Regulated Voltage? | | Regulated current | Regulated current |
| Software/Firmware/Micropr<br>ocessor Control? | | Yes | Yes |
| Automatic Overload Trip? | | Yes | Yes |
| Automatic No-Load Trip? |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.