TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Wireless Pain Relieve Device (LT5018C) provides TENS and EMS therapy via self-adhesive electrodes placed on skin. Device generates biphasic, rectangular electrical pulses to stimulate sensory nerves (TENS) for pain perception modification or motor neurons (EMS) for passive muscle contraction. Controlled via mobile app (iOS/Android) or remote control using Bluetooth Low Energy. Used in home setting by adult patients. Output intensity, pulse duration, and frequency are adjustable via pre-set programs. Device includes automatic overload/no-load contact trips and auto-shutoff. Output affects muscle performance and pain management; benefits include drug-free pain relief and muscle conditioning.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with IEC 60601-1 (safety), IEC 60601-1-11, IEC 60601-2-10 (performance), and IEC 60601-1-2 (EMC) demonstrated. Biocompatibility supported by use of materials identical to previously cleared devices (K162479, K130802).
Technological Characteristics
Materials: ABS enclosure, standard electrodes. Energy: 3.7V battery. Waveform: Biphasic rectangular. Connectivity: Bluetooth Low Energy (mobile app/remote). Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-2, FCC Part 15. Software: Moderate level of concern, IEC 62304 compliant.
Indications for Use
Indicated for adults of all genders for temporary relief of pain in shoulder, waist, back, neck, arms, and legs due to exercise or household strain; symptomatic relief/management of chronic, intractable pain; relief of pain associated with arthritis; and stimulation of healthy muscles to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2017
Shenzhen Dongdixin Technology Co.,Ltd. Siping Yuan R.A. Specialist No. 3 Building Xilibaimang Xusheng Industrial Estate, Nanshan Shenzhen, Guangdong 51808 China
Re: K173462
Trade/Device Name: Wireless Pain Relieve Device, Model LT5018C Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX, NYN Dated: October 31, 2017 Received: November 8, 2017
Dear Siping Yuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173462
Device Name Wireless Pain Relieve Device Model: LT5018C
#### Indications for Use (Describe)
TENS: The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
EMS: The device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
| Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Residential Use (Part 21 CFR 601 Subpart E) <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 601 Subpart C) |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "DONGDIXIN TECHNOLOGY CO., LTD." is written in a sans-serif font. There are also three Chinese characters to the right of the English text.
001_510 (k) Summary Version: 1.0
## 510(k) SUMMARY
as required by section 21 CFR 807.92
Wireless Pain Relieve Device
| Date of Submission: | 10/31/2017 |
|---------------------|------------------------------------------------------------------------------------------|
| Submitter's Name: | Shenzhen Dongdixin Technology Co., Ltd |
| Address: | No. 3 Building, Xilibaimang Xusheng Industrial<br>Estate Nanshan, Shenzhen, CHINA 518108 |
| E-mail: | yuansp@dundex.com |
| Tel: | +86(755) 27652471 |
| FAX: | +86(755) 27652674 |
| Contact: | Siping Yuan |
{4}------------------------------------------------
## DONGDIXIN翡
001 510 (k) Summary ı: 1.0
| | TECHNOLOGY CO., LTD.<br>厅灾 | Version: | | |
|----|-------------------------------------------|-------------------------------------------------------------|--|--|
| 1. | Proposed Device: | | | |
| | Device Name: Wireless Pain Relieve Device | | | |
| | Model: LT5018C | | | |
| | Device classification Name: | Stimulator, Nerve, Transcutaneous, Over-The-Counter | | |
| | Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief | | |
| | Regulation Medical Specialty: | Neurology | | |
| | Review Panel: | Neurology | | |
| | Regulation Number: | 882.5890 | | |
| | Product Code: | NUH | | |
| | Device Class: | ll | | |
| | Device classification Name: | Stimulator, Muscle, Powered, For Muscle Conditioning | | |
| | Regulation Description: | Powered muscle stimulator. | | |
| | Regulation Medical Specialty: | Physical Medicine | | |
| | Review Panel: | Physical Medicine | | |
| | Regulation Number: | 890.5850 | | |
| | Product Code: | NGX | | |
| | Device Class: | = | | |
| | Device classification Name: | Stimulator, Electrical, Transcutaneous, For Arthritis | | |
| | Regulation Description: | Transcutaneous electrical nerve stimulator for pain relief. | | |
| | Regulation Medical Specialty: | Neurology | | |
| | Review Panel: | Neurology | | |
| | Regulation Number: | 882.5890 | | |
| | Product Code: | NYN | | |
| | Device Class: | = | | |
| 2. | Predicate Device: | | | |
| | | Lorally Markatad Davira. Smart Dain Rollovar IT501Q | | |
Legally Marketed Device: Smart Pain Reliever LT5019 510(k) Number: K162479 Manufacturer: Shenzhen Dondixin Technology Co., Ltd.
#### Description of Proposed Device: 3.
LT5018C is designed to be used at home, by adults of all genders.
#### Explanation of how the device functions:
The Wireless Pain Relieve Device includes TENS, EMS (including MASSAGE Program) mode. Transcutaneous Electrical Nerve Stimulation (TENS) is a noninvasive, drug free method of controlling pain. EMS works by sending electronic pulses to the muscle needing treatment; this
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and orange, with the company name in large letters. Below the company name is the phrase "TECHNOLOGY CO., LTD." in smaller letters. There is also some text in another language to the right of the company name, and the phrase "causes the muscle to exercise passively." is at the bottom.
## Scientific concepts that from the basis for the device:
TENS uses tiny electrical impulses sent through the skin to nerves to modify your pain perception. TENS does not cure any physiological problem; it only helps control the pain, this activates the underlying sensory nerves. EMS is a product derived from the square waveform (ladder-shaped), through the square wave pattern it is able to work directly on muscle motor neurons, contraction helps to condition the muscle in order to facilitate performance.
For Transcutaneous Electrical Nerve Stimulation (TENS) Self-adhesive electrodes are placed on the skin close to the area of pain. The user can choose 6 pre-set TENS programs with different pulse settings. In each program, the intensity of the pulse can be adjusted.
For Electrical Muscle Stimulation (EMS), the electrodes are placed near the muscle to be stimulated. The user can choose 5 pre-set EMS programs with different pulse settings. In each program, the intensity of the pulse can be adjusted.
There have 3 massage programs, and the massage stimulation program is a stimulation program that is used with the EMS mode of stimulation to facilitate recovery from muscle fatigue and to help recover muscle strength after training sessions and competitions.
The Wireless Pain Relieve Device is controlled by means of an APP on a mobile device (phone or table) and remote control. The communication is done via Bluetooth Low Energy. The APP operates on IOS and Android platforms (IOS 8.0 or greater, Android 5.0 or greater).
#### Description of All Device Modification(s) 4.
Compared with the existing device (LT5019), the following main parts of Wireless Pain Reliever LT5018C have changed:
- 1) Added the remote control to control the operation of device.
- 2) Change the dimensional specification.
- 3) Change the hardware based on the new dimensional of enclosure. But the fundamental scientific technology is not change.
Based on the 21CFR820.30 requirement, this is belong to design change and shall have been evaluated according to the standard IEC60601-2-10, and IEC 60601-1-2. We have conducted these standard test and the results is Passed. Meanwhile, for add the remote control; we added the wireless coexistence test and the results display the remote control also safely and effectively.
#### 5. Proposed Device Intended for Use Statement:
Device Name: Wireless Pain Relieve Device, Model: LT5018C
## Indications for Use:
- � TENS: The device is designed to be used for temporary relief of pain associated with
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue, except for a diagonal line in the middle of the word "DONGDIXIN" which is in yellow. To the right of the English text is the same company name in Chinese.
sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm) and lower extremities (leg) due to strain from exercise or normal household work activities. And to be used for the symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
- EMS: The device is designed to be used for stimulate healthy muscles in order to ◆ improve and facilitate muscle performance.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold, blue letters, with a yellow design element in the middle. Below that, the text "TECHNOLOGY CO., LTD." is in smaller, blue letters.
#### 6. Technological Characteristics and Substantial Equivalence
Both the LT5018C and Predicate Device the application of electrical current through electrodes placed on the skin for pain control, or electrical muscle stimulation (EMS) the elicitation of muscle contraction using electric impulses are generated by the device and delivered through electrodes on the skin in direct proximity to the (painful) muscles to be stimulated.
Basic technological characteristics, new device vs. Predicate device
| | | New device | Predicate device | S.E. Discussion |
|----|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| 1 | 510K# | To be assigned | K162479 | N/A |
| 2 | Device Name and<br>Model | Wireless Pain Relieve device<br>Mode: LT5018C | Smart Pain Reliever<br>Mode: LT5019 | N/A |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same |
| 4 | Intended for use | TENS: The device is designed to be used for<br>temporary relief of pain associated with<br>sore and aching muscles in the shoulder,<br>waist, back, neck, upper extremities (arm)<br>and lower extremities (leg) due to strain<br>from exercise or normal household work<br>activities. And to be used for the<br>symptomatic relief and management of<br>chronic, intractable pain and relief of pain<br>associated with arthritis.<br>EMS: The device is designed to be used for<br>stimulate healthy muscles in order to<br>improve and facilitate muscle performance. | TENS: The device is designed to be used<br>for temporary relief of pain associated<br>with sore and aching muscles in the<br>shoulder, waist, back, neck, upper<br>extremities (arm) and lower extremities<br>(leg) due to strain from exercise or normal<br>household work activities. And to be used<br>for the symptomatic relief and<br>management of chronic, intractable pain<br>and relief of pain associated with arthritis.<br>EMS: The device is designed to be used<br>for stimulate healthy muscles in order to<br>improve and facilitate muscle<br>performance. | Same |
| 5 | Power Source | 3.7V battery supply | 3.7V battery supply | Same |
| | -Method<br>of<br>current isolation | Line N/A | N/A | Same |
| | - Patient Leakage<br>Current (μΑ)<br>-Normal condition | 0uA | 0uA | Same |
| | -Single fault<br>condition | 2.0uA | 2.0uA | |
| 6 | DC current<br>Average<br>electrodes<br>through<br>when device is on but<br>no pulses are being<br>applied (μΑ) | 0 | 0 | Same |
| 7 | Number of Output<br>Modes | 2 (TENS/EMS(including MASSAGE)) | 2 (TENS/EMS(including MASSAGE)) | Same |
| 8 | Number of Output Channels | | | |
| | Synchronous or<br>Alternating? | N/A | N/A | Same |
| | Method of Channel<br>Isolation | N/A | N/A | Same |
| ਰੇ | Regulated Current or<br>Regulated Voltage? | Current control | Current control | Same |
| 10 | Software/Firmware/Mi<br>cro processor Control? | Yes | Yes | Same |
| 11 | Automatic Overload<br>Trip | Yes | Yes | Same |
| 12 | Automatic No Load<br>contact Trip | Yes | Yes | Same |
| 13 | Automatic Shut off | Yes | Yes | Same |
| 14 | User Override Control? | Yes<br>Power on/off button on the device<br>Power on/off on the remote control<br>Power on/off in the APP software | Yes<br>Power on/off button on the device, and<br>Power on/off in the APP software | Similar, the new device added<br>the remote control, doesn't<br>affect the safety and<br>effectiveness. |
| 15 | Indicator Display: | | | |
| | On/Off Status? | Yes | Yes | Same |
| | Low Battery? | Yes | Yes | Same |
| | Voltage/ Current | Yes | Yes | Same |
#### Table 1: Substantial Equivalence Comparison Table
{8}------------------------------------------------
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DONGDIXIN Technology Co., LTD. The logo is in dark blue, except for a stylized "X" in the middle of the name, which is in orange. To the right of the English name is the company name in Chinese characters.
| Level? | | |
|--------|--|--|
|--------|--|--|
## Table 2: Output Specification TENS mode
| | | New device | Predicate device | S.E. Discussion | |
|----|---------------------------------------------------------------------|---------------------------------------------|----------------------------------------|--------------------------------|--|
| 1 | 510K# | To be assigned | K162479 | N/A | |
| 2 | Device Name or<br>Program Name | Wireless Pain Relief Device<br>Mode:LT5018C | Smart Pain Reliever<br>Mode: LT5019 | N/A | |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same | |
| 4 | Waveform | Biphasic | Biphasic | Same | |
| 5 | Shape | Rectangular | Rectangular | Same | |
| 6 | Max Output Voltage (V) ±20% | | | | |
| 7 | 500Ω | 31 | 31.2 | Similar, the new device have | |
| 8 | 2kΩ | 66 | 69.6 | been evaluated and passed the | |
| 9 | 10kΩ | 66 | 69.6 | test according to the | |
| | | | | requirement of IEC60601-2-10 | |
| 10 | Pulse Duration<br>(µsec) | 150~250us | 150~250us | Same | |
| 11 | Frequency (Hz) | 2~100Hz | 2~100Hz | Same | |
| 12 | Maximum Phase<br>Charge (uC) 500Ω | 31 | 30 | Similar, the new device have | |
| | | | | been evaluated and passed | |
| 13 | Maximum Current<br>Density 500Ω | 0.48mA/cm² | 0.32mA/cm² | the test according to the | |
| | | | | requirement of | |
| | | | | IEC60601-2-10. | |
| 14 | Maximum Average<br>Current (average<br>absolute value),<br>mA, 500Ω | 1.2 | 1.2 | The maximum average power | |
| | | | | density of new device is far | |
| | | | | lower than 250mW/cm2 as | |
| | | | | required by the FDA guidance | |
| | | | | for Powered Muscle | |
| | | | | Stimulator. The safety and | |
| | | | | effectiveness of the device is | |
| | | | | not affected. | |
| 15 | Maximum Average<br>Power Density,<br>(mW/cm²),500Ω | 2.88 | 1.9 | | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold, blue letters, with a yellow "X" in the middle. Below the company name is the text "TECHNOLOGY CO., LTD." in smaller, blue letters. To the right of the company name are three Chinese characters.
## Table 3: Output Specification EMS mode
| | | New device | Predicate device | S.E. Discussion | |
|----|---------------------------------------------------------------------|---------------------------------------------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 1 | 510K# | To be assigned | K162479 | N/A | |
| 2 | Device Name or<br>Program Name | Wireless Pain Relief Device<br>Mode:LT5018C | Smart Pain Reliever<br>Mode: LT5019 | N/A | |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same | |
| 4 | Waveform | Biphasic | Biphasic | Same | |
| 5 | Shape | Rectangular | Rectangular | Same | |
| 6 | | Max Output Voltage (V) ±20% | | | |
| 7 | 500Ω | 31 | 31.2 | Similar, the new device have<br>been evaluated and passed the<br>test according to the<br>requirement of IEC60601-2-10 | |
| 8 | 2kΩ | 66 | 69.6 | Same | |
| 9 | 10kΩ | 66 | 69.6 | Similar, the new device have<br>been evaluated and passed the<br>test according to the<br>requirement of IEC60601-2-10 | |
| 10 | Pulse Duration (usec) | 200~370us | 200~370us | Same | |
| 11 | Frequency (Hz) | 3-55 Hz | 3~75Hz | Similar, the new device have<br>been evaluated and passed the<br>test according to the<br>requirement of IEC60601-2-10 | |
| 12 | Maximum Phase<br>Charge (uC) 500Ω | 44.4 | 24 | Similar, the new device have<br>been evaluated and passed the<br>test according to the<br>requirement of IEC60601-2-10 | |
| 13 | Maximum Current<br>Density 500Ω | 0.484mA/cm2 | 0.248mA/cm2 | Similar, the new device have<br>been evaluated and passed the<br>test according to the<br>requirement of IEC60601-2-10. | |
| 14 | Maximum Average<br>Current (average<br>absolute value), mA,<br>500Ω | 1.22 | 0.72 | The maximum average power<br>density of new device is far<br>lower than 250mW/cm2 as<br>required by the FDA guidance<br>for Powered Muscle<br>Stimulator. The safety and<br>effectiveness of the device is<br>not affected. | |
| 15 | Maximum Average<br>Power Density,<br>(mW/cm²),500Ω | 2.93 | 1.2 | | |
| | | New device | Predicate device | S.E. Discussion | |
| 1 | 510K# | To be assigned | K162479 | N/A | |
| 2 | Device<br>Name<br>or | Wireless Pain Relief Device | Smart Pain Reliever | N/A | |
| | Program Name | Mode:LT5018C | Mode: LT5019 | | |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd | Shenzhen Dongdixin Technology Co., Ltd | Same | |
| 4 | Waveform | Biphasic | Biphasic | Same | |
| 5 | Shape | Rectangular | Rectangular | Same | |
| 6 | Max Output Voltage (V) ±20% | | | Similar, the new device have | |
| 7 | 500Ω | 31 | 31.2 | been evaluated and passed the | |
| 8 | 2kΩ | 66 | 69.6 | test according to the | |
| 9 | 10kΩ | 66 | 69.6 | requirement of IEC60601-2-10 | |
| 10 | Pulse Duration (usec) | 50-250us | 250us | Similar, the new device have | |
| 11 | Frequency (Hz) | 2~90Hz | 10~80Hz | been evaluated and passed the | |
| | | | | test according to the | |
| | | | | requirement of IEC60601-2-10 | |
| 12 | Maximum Phase | 31 | 30 | Similar, the new device have | |
| | Charge (uC) 500Ω | | | been evaluated and passed the | |
| 13 | Maximum<br>Current | 0.51mA/cm2 | 0.32mA/cm2 | test according to the | |
| | Density 500Ω | | | requirement of IEC60601-2-10. | |
| 14 | Maximum<br>Average | 3.2 | 1.9 | The maximum average power | |
| | Power<br>Density, | | | density of new device is far | |
| | (mW/cm²),500Ω | | | lower than 250mW/cm2 as | |
| | | | | required by the FDA guidance | |
| | | | | for Powered Muscle | |
| | | | | Stimulator. The safety and | |
| | | | | effectiveness of the device is | |
| | | | | not affected. | |
{11}------------------------------------------------
#### Table 4: Output Specification MASSAGE program
{12}------------------------------------------------
TECHNOLOGY CO., LTD.
#### Discussion
The timer range, weight and dimension of the new device are different from the predicate devices, but the new devices are evaluated and passed the testing according to IEC60601-1 and IEC60601-2-10, this Different does not pose any new questions of safety and effectiveness. The Housing Material, Max Output Voltage/Current, Maximum Phase Charge, Maximum Current Density and Maximum Average Current/Power Density are similar with the predicate devices, but the new devices are evaluated and passed the testing according to IEC60601-2-10 and compliance with the guidance Document for Powered Muscle Stimulator, this difference doesn't pose any new questions of safety and effectiveness.
#### 7. Performance Data:
The following performance data are provided in support of the substantial equivalence determination:
#### 7.1 Biocompatibility testing
Compared with unmodified device, the material of electrodes for the new product LT5018C is the same material of electrodes (K162479) that got the 510(k) clearance on 2017. The material of enclosure is ABS, and it is the same as the material of enclosure (k130802) that got the 510(k) clearance on 2013. These materials have the same product process and injection process.
#### 7.2 Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the LT5018C. The system complies with the IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
For FCC part 15 RADIO FREQUENCY DEVICES, Subpart C-Intentional Radiators.
#### 7.3 Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this
device was considered as a "Moderate" level of concern. IEC 62304 was followed.
#### 8. Conclusions
The LT5018C has the same intended use and technological characteristics as the predicate device Smart Pain Reliever LT5019. Moreover, bench testing and safety report supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, LT5018C is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.