To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis (Choose Mode B or C). To stimulate healthy muscles in order to improve and facilitate muscle performance (Choose Mode A). To temporarily increase local blood circulation in healthy muscles (Choose Mode A).
Device Story
Pain Therapy Device is a household multifunctional electronic stimulator; uses two channels to deliver electrical pulses via electrode pads to skin. Device features three operation modes (A, B, C) and user-controlled settings (ON/OFF, Mode, Time). Input: user-selected mode/time; Output: pulsed, symmetric, biphasic electrical current. Used by patients in home settings for pain relief, muscle performance improvement, and circulation enhancement. Healthcare providers do not operate; device is OTC. Output affects nerve/muscle stimulation; benefits include pain management and muscle conditioning.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-5, ISO 10993-10), usability (IEC 62366), and software verification/validation per FDA guidance.
Indicated for temporary relief of pain associated with sore/aching muscles in back, neck, upper/lower extremities due to exercise or household strain; symptomatic relief/management of chronic, intractable pain; relief of pain associated with arthritis; stimulation of healthy muscles to improve/facilitate muscle performance; and temporary increase of local blood circulation in healthy muscles.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Health Expert Electronic Stimulator, Models: AST-300C and AST-300D (K133929)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines that resemble a ribbon or banner beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2017
Guangzhou Xinbo Electronic Co., Ltd. % Ms. Cecilia Ceng Manager Guangzhou Glomed Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, Guangdong 510000 CN
Re: K163611
Trade/Device Name: Pain Therapy Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief, Over The Counter Regulatory Class: Class II Product Code: NUH, NYN, NGX Dated: August 3. 2017 Received: August 9, 2017
Dear Ms. Ceng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163611
Device Name Pain Therapy Device
#### Indications for Use (Describe)
To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis (Choose Mode B or C).
To stimulate healthy muscles in order to improve and facilitate muscle performance (Choose Mode A).
To temporarily increase local blood circulation in healthy muscles (Choose Mode A).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 201 Subpart D) | <span style="font-size: 20px;">☐</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 201 Subpart C) | <span style="font-size: 20px;">☑</span> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Chapter 5. 510(k) Summary
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
510(k) Owner's Name: Guangzhou Xinbo Electronic Co., Ltd. Establishment Registration Number: Applying Address: 2nd Building, Juntuo Industry Park, Xingye Dadao, Nancun Town, Panyu District, Guangzhou City, Guangdong Province, China Tel: +86-020-66393598 Fax: +86-020-34822409 Contact Person: Li Huifang (Manger) Email: drtvsammy@hotmail.com
## Application Correspondent:
Contact Person: Ms. Cecilia Ceng / Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com
### 2. Subject Device Information
Trade Name: Pain Therapy Device Common Name: Electronic Stimulator Classification name: Transcutaneous electrical nerve stimulator for pain relief Review Panel: Neurology, Physical Medicine Product Code: NUH, NYN, NGX Regulation Class: II Regulation Number: 882.5890, 890.5850
### 3. Predicate Device Information
| Sponsor | Shenzhen OSTO<br>Technology Company<br>Limited | Actegy, Ltd. | Counter Scientific<br>Development (GZ) Ltd. |
|--------------------------|---------------------------------------------------------------------------|-------------------------------------|---------------------------------------------|
| Device Name<br>and Model | Health Expert Electronic<br>Stimulator, Models: AST-<br>300C and AST-300D | Revitive IX (OTC), Model:<br>MT1101 | Pain Therapy System,<br>Model: PTS-II |
| 510(k) | K133929 | K143207 | K150277 |
{4}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|--------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-II |
| Document Name: | FDA 510(k) Submission Report |
| Number | | | (Primary Predicate) |
|----------------------|--------------------|--------------------|---------------------|
| Product Code | NUH, NGX | NUH, NGX | NUH, NGX, NYN, GZJ |
| Regulation<br>Number | 882.5890, 890.5850 | 882.5890, 890.5850 | 882.5890, 890.5850 |
| Regulation<br>Class | II | II | II |
### 4. Device Description
Pain Therapy Device is a household multifunctional device, it can stimulate healthy muscles in order to improve and facilitate muscle performance, and temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet) due to strain from exercise or normal household work activities by applying current to stimulate nerve.
Pain Therapy Device has three operation modes and two channels, which can give certain electrical pulse through 2 pairs of electrode pads on the skin.
The electronic stimulatory main units have the operating elements of ON/OFF knob, Mode Selection key and Time Selection key which can be user-friendly controlled.
The device is equipped with accessories of electrode pads and an electrode wire. The electrode wire is used to connect the pads to the main unit. All the accessories can only be changed by local distributor. "Large Patch Electrode" and "Small Patch Electrode" are cleared in K090198.
"Insole Electrode", "Sole Plant Electrode A" and "Sole Plant Electrode B" are cleared in K151693.
### 5. Intended Use / Indications for Use
To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back, back of the neck, upper extremities (shoulder and arm), lower extremities (leg and feet), due to strain from exercise or normal household work activities by applying current to stimulate nerve. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis (Choose Mode B or C).
To stimulate healthy muscles in order to improve and facilitate muscle performance (Choose Mode A). To temporarily increase local blood circulation in healthy muscles (Choose Mode A).
# 6. Test Summary
Pain Therapy Device has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
- � Usability test according to IEC 62366 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- � The electrode wire is compliance with 21 CFR 898 by IEC 60601-1 (version 3.1, clause 8.5.2.3) evaluation.
{5}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|-------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I |
| Document Name: | FDA 510(k) Submission Report |
#### 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, mode of operation, and intended use of Electronic Muscle Stimulator is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Predicate Device | Remark | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name<br>and Model | Pain Therapy<br>Device, Models:<br>P.T.S-II, P.T.S-IIA,<br>P.T.S-IIB, CP-I | Health Expert<br>Electronic<br>Stimulator<br>Models: AST-300C<br>and AST-300D | Revitive IX (OTC),<br>Model: MT1101 | Pain Therapy<br>System,<br>Model: PTS-II<br>(Primary Predicate) | -- |
| 510(k) Number | K163611 | K133929 | K143207 | K150277 | -- |
| Intended Use &<br>Indications for<br>Use | To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>upper and lower<br>back, back of the<br>neck, upper<br>extremities<br>(shoulder and arm),<br>lower extremities<br>(leg and feet) due<br>to strain from<br>exercise or normal<br>household work<br>activities by applying<br>current to stimulate<br>nerve.<br>To be used for<br>symptomatic relief<br>and management of<br>chronic, intractable<br>pain and relief of<br>pain associated with<br>arthritis (Choose<br>Mode B or C).<br>To stimulate healthy<br>muscles in order to<br>improve and<br>facilitate muscle<br>performance<br>(Choose Mode A).<br>To temporarily<br>increase local blood | PMS (Mode 1~8):<br>It is intended to<br>stimulate healthy<br>muscles in order to<br>improve or facilitate<br>muscle<br>performance.<br>TENS (Mode 9~25):<br>To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, back of the<br>neck, arm, leg, and<br>foot due to strain<br>from exercise or<br>normal household<br>work activities by<br>applying current to<br>stimulate nerve. | To temporarily<br>increase local blood<br>circulation in healthy<br>leg muscles.<br>To stimulate healthy<br>muscles in order to<br>improve and<br>facilitate muscle<br>performance.<br>For temporary relief<br>of pain associated<br>with sore and aching<br>muscles in the<br>shoulder, waist,<br>back, upper<br>extremities (arms)<br>and lower<br>extremities (legs)<br>due to strain from<br>exercise or normal<br>household duties. | To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the upper<br>and lower back due<br>to strain from<br>exercise or normal<br>household work<br>activities (Choose<br>Mode A, B, or C).<br>To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the upper<br>extremities (Arms)<br>due to strain from<br>exercise or normal<br>household work<br>activities (Choose<br>Mode A, B or C).<br>To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the lower<br>extremities (Legs)<br>due to strain from<br>exercise or normal<br>household work<br>activities (Choose<br>Mode A, B or C).<br>To be used for | SE<br>Note 4 |
{6}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|-------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark | | |
|--------------------------------------------------------------------------------------------------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------------|--------------|
| circulation in healthy<br>muscles (Choose<br>Mode A). | | symptomatic relief<br>and management of<br>chronic, intractable<br>pain and relief of<br>pain associated with<br>Arthritis (Choose<br>Mode A).<br>To stimulate healthy<br>muscles in order to<br>improve and<br>facilitate muscle<br>performance<br>(Choose Mode B). | | | |
| OTC or Rx | OTC | OTC | SE | | |
| <b>Basic Unit Characteristics</b> | | | | | |
| Power<br>Source(s) | DC 3.0V, 2 x AAA | Adaptor Input: 100-<br>240Vac, 50-60Hz,<br>0.1A<br>Main Unit Input:<br>5Vdc, 1A | Adaptor Input: 100-<br>240Vac, 50/60Hz,<br>0.18A<br>Main Unit Input:<br>5.0Vdc, 1.0A | DC 3.0V, 2 x AAA | SE |
| Method of Line<br>Current Isolation | Type BF Applied<br>Part | Type BF Applied<br>Part | -- | SE | |
| Patient<br>Leakage<br>Current | NC | DC: 0.5 μΑ | AC: 54.5 μΑ,<br>DC: 0.5 μΑ | -- | SE<br>Note 1 |
| | SFC | DC: 0.6 μΑ | AC: 120.0 μΑ,<br>DC: 0.6 μΑ | -- | |
| Average DC<br>current through<br>electrodes when<br>device is on but<br>no pulses are<br>being applied | | < 0.01 μA | < 0.01 μA | -- | SE |
| Number of<br>Output<br>Channels | | 2 Channels: for<br>models P.T.S-II,<br>P.T.S-IIA, P.T.S-IIB;<br>1 Channel: for<br>model CP-I | 2 | 2 | SE<br>Note 1 |
| Number of<br>Output Modes | | 3 | 25 | 1 | SE |
| Output Intensity<br>Level | | 5 steps | 99 steps | -- | SE |
| Synchronous or<br>Alternating? | | Synchronous | Synchronous | -- | SE |
{7}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP- |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | Predicate Device | Remark |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------------------------|--------------|
| Method of<br>Channel<br>Isolation | Parallel connection | -- | Parallel connection | SE |
| Regulated<br>Current or<br>Regulated<br>Voltage? | Regulated Voltage | Regulated Voltage | Regulated Voltage | SE |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | SE |
| Automatic<br>Overload Trip | No | -- | No | SE |
| Automatic No-<br>Load Trip | No | -- | No | SE |
| Automatic Shut<br>Off | Yes | -- | Yes | SE |
| User Override<br>Control | Yes | -- | Yes | SE |
| Indicator<br>Display | On/Off<br>Status | Yes | Yes | SE |
| Low<br>Battery | No | -- | No | SE |
| Voltage/<br>Current Level | Yes | -- | Yes | SE |
| Timer Range | 10, 20, 40 min | 25 min | 1 to 60 min | SE |
| Weight | Main Unit:<br>P.T.S-II: 75g<br>P.T.S-IIA: 100g<br>P.T.S-IIB: 100g<br>CP-I: 66g<br><br>Electrode:<br>Big Patch Electrode:<br>40g<br>Small Patch<br>Electrode: 10g<br>Insole Electrode:<br>200g<br>Sole Plant Electrode<br>A (only for CP-I):<br>900g<br>Sole Plant Electrode | 2kg (Without<br>accessories) | 1.725kg (Without<br>PSU) | SE<br>Note 2 |
{8}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|-------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | | | Remark |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------------------------|----------------------------|--------------|
| | B: 920g | | | | |
| Dimensions | Main Unit:<br>P.T.S-II: 110 x 78 x 20 mm<br>P.T.S-IIA: 135 x 82 x 20 mm<br>P.T.S-IIB: 135 x 82 x 20 mm<br>CP-I: 92 x 78 x 20 mm<br><br>Electrode:<br>Large Patch<br>Electrode: 120 x 80 mm<br>Small Patch<br>Electrode: 46 x 46 mm<br>Insole Electrode:<br>260 x 110 mm<br>Sole Plant Electrode A (only for CP-I):<br>450 x 450 x 90 mm<br>Sole Plant Electrode B:<br>450 x 450 x 90 mm | 428mm x 428.8mm<br>x 185mm | Ø360mm x 75mm | 110 x 78 x 20 mm | SE<br>Note 2 |
| Housing<br>Materials and<br>Construction | Main unit: ABS<br>plastic | Main unit: ABS<br>plastic | Casing/body: ABS,<br>Footpads: NBR | Main unit: ABS<br>plastic | SE |
| Output Specifications | | | | | |
| Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Positive-going,<br>Reverse | SE |
| Shape | Rectangular, with interphase interval | Rectangular, with interphase interval | Rectangular and bipolar | Square wave | SE |
| Maximum<br>Output Voltage | 40Vp @ 500Ω | 44V±10% @ 500Ω | -- | 88Vp @ 500Ω | SE |
| | 80Vp @ 2kΩ | 80V±10% @ 2kΩ | -- | 102Vp @ 2kΩ | SE<br>Note 3 |
| | 95Vp @ 10kΩ | 112V±10% @ 10kΩ | -- | 106Vp @ 10kΩ | |
| Maximum<br>Output Current | 80mA @ 500Ω | 88mA±10% @ 500Ω | -- | 176mA @ 500Ω | SE<br>Note 3 |
| | 40mA @ 2kΩ | 40mA±10% @ 2kΩ | -- | 51.0mA @ 2kΩ | |
| | 9.5mA @ 10kΩ | 11.2mA±10% @10kΩ | -- | 10.6mA @ 10kΩ | |
| Pulse Duration | 200µs | 120µs | -- | 170µs | SE<br>Note 3 |
{9}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP- |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | | Remark | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------|
| Pulse frequency | 13.7~48.5 Hz | 77.3Hz | -- | SE<br>Note 3 | |
| Net Charge (per<br>pulse) | 0µC @ 500Ω,<br>Method: Balanced<br>waveform | 0µC @ 500Ω,<br>Method: Balanced<br>waveform | -- | 1.63µC @ 500Ω | SE<br>Note 3 |
| Maximum<br>Phase Charge | 19.2µC @ 500Ω | 12.78µC @ 500Ω | -- | 29.9µC @ 500Ω | SE<br>Note 3 |
| Maximum<br>Average Current | 1.53mA @ 500Ω | 0.968mA @ 500Ω | -- | 2.22mA | SE<br>Note 3 |
| Maximum<br>Current Density<br>(r.m.s) | 0.073mA/cm² @<br>500Ω | 0.235mA/cm² @<br>500Ω | -- | 8.31mA/cm² @ 500Ω | SE<br>Note 3 |
| Maximum<br>Average Power<br>Density | 0.056mW/cm² @<br>500Ω | 1.38mW/cm² @<br>500Ω | -- | 0.115mW/cm² @<br>500Ω | SE<br>Note 3 |
| ON Time | 2s | 0.6s | -- | -- | SE<br>Note 3 |
| OFF Time | 2s | 0.6s | -- | -- | SE<br>Note 3 |
| Additional Features | | | | | |
| Operating<br>Environment | Temperature:<br>5~40°C,<br>Humidity: ≤80%RH,<br>Atmospheric<br>Pressure:<br>86~106kPa | Temperature:<br>5~45°C,<br>Humidity: 20~65%<br>RH | -- | -- | SE<br>Note 1 |
| Storage<br>Environment | Temperature:<br>Main Unit: -20 ~<br>55°C,<br>Electrode Pad: 10 ~<br>20°C<br>Humidity: 10~95%<br>RH,<br>Atmospheric<br>Pressure: 50~106<br>kPa | Temperature:<br>0~45°C,<br>Humidity: 10~90%<br>RH,<br>Electrode Pad:<br>10~20°C | -- | -- | SE<br>Note 1 |
| Standards | | | | | |
| Biocompatibility | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | SE |
{10}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|-------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | | Remark |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|--------------------------------------------------|--------|
| | Electrode" and<br>"Small Patch<br>Electrode" are<br>cleared in K090198.<br>"Insole Electrode",<br>"Sole Plant<br>Electrode A" and<br>"Sole Plant<br>Electrode B" are<br>cleared in K151693. | | | |
| Electrical Safety | Comply with IEC<br>60601-1 and IEC<br>60601-2-10 | Comply with IEC<br>60601-1 and IEC<br>60601-2-10 | Comply with IEC<br>60601-1 and IEC<br>60601-2-10 | SE |
| EMC | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | Comply with IEC<br>60601-1-2 | SE |
### Comparison in Detail(s):
### Note 1:
Although the "Patient Leakage Current". "Number of Output Channels". "Operating Environment", and "Storage Environment" are a little different from the predicate devices, they all comply with IEC 60601-1 requirements. So the differences will not raise any safety or effectiveness issue.
### Note 2:
Although the "Weight" and "Dimensions" of subject device are different from the predicate devices, they are all comply with IEC 60601-1 and IEC 60601-2-10 requirements. So the differences of the function specifications will not raise any safety or effectiveness issue.
### Note 3:
Although some output specifications "Maximum Output Voltage", "Maximum Output Current", "Pulse Duration", "Maximum Pulse Frequency", "Net Charge (per pulse)", "Maximum Phase Charge", "Maximum Average Current", "Maximum Current Density", "Maximum Average Power Density of subject device", "ON Time" and "OFF Time" are a little different from the predicate devices, they all comply with IEC 60601-1, IEC 60601-2-10 requirement, FDA guidance requirement for Transcutaneous Electrical Nerve Stimulator for Pain Relief and FDA guidance requirement for Powered Muscle Stimulator for Muscle Conditioning. So the differences of function specification will not raise any safety or effectiveness issue.
#### Note 4:
Although the Indication for Use (IFU) of subject device is a little different from that of predicate, its combines from 3 predicate devices' IFU, so the differences between them will not raise any safety or effectiveness issue.
#### Additional Modes Comparison Table:
| Elements of<br>Comparison | Subject Device (K163611) | Predicate Device (K133929) | Remark |
|---------------------------|--------------------------|----------------------------|--------|
|---------------------------|--------------------------|----------------------------|--------|
{11}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP- |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device (K163611) | Predicate Device (K133929) | Remark |
|----------------------------------------------|----------------------------------------------------------------------|----------------------------------------------|----------------|
| Device Name | Pain Therapy Device, Models:<br>P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I | Health Expert Electronic Stimulator | -- |
| Mode Comparison<br>Group 1 | Mode A | Mode 2 | -- |
| - Indications for Use<br>Claim | PMS claims | PMS claims | SE |
| - Waveform | Pulsed, Rectangular, Symmetric,<br>Biphasic | Pulsed, Rectangular, Symmetric,<br>Biphasic | SE |
| - Pulse Width | 200μs | 120μs | SE<br>Note (a) |
| - Pulse Frequency | 48.5Hz | 77.3Hz | SE<br>Note (a) |
| - Pulse Cycle<br>Number (T) for One<br>Burst | 200T ~ 206T | 200T | SE<br>Note (a) |
| - Maximum Output<br>Voltage | 40Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>95Vpk @ 10kΩ | 44Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>112Vpk @ 10kΩ | SE<br>Note (b) |
| Mode Comparison<br>Group 2 | Mode B | Mode 20 | -- |
| - Intended Use Claim | TENS claims | TENS claims | SE |
| - Waveform | Pulsed, Rectangular, Symmetric,<br>Biphasic | Pulsed, Rectangular, Symmetric,<br>Biphasic | SE |
| - Pulse Width | 200μs | 120μs | SE<br>Note (a) |
| - Pulse Frequency | 13.7Hz / 19.0Hz | 77.3Hz | SE<br>Note (a) |
| - Pulse Cycle<br>Number (T) for One<br>Burst | 82T ~ 113T | 100T | SE<br>Note (a) |
| - Maximum Output<br>Voltage | 40Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>95Vpk @ 10kΩ | 44Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>112Vpk @ 10kΩ | SE<br>Note (b) |
| Mode Comparison<br>Group 3 | Mode C | Mode 14 | -- |
| - Intended Use Claim | TENS claims | TENS claims | SE |
| - Waveform | Pulsed, Rectangular, Symmetric,<br>Biphasic | Pulsed, Rectangular, Symmetric,<br>Biphasic | SE |
| - Pulse Width | 200μs | 120μs | SE<br>Note (a) |
| - Pulse Frequency | 46.1Hz / 38.6Hz / 29.4Hz | 77.3Hz | SE<br>Note (a) |
| - Pulse Cycle | 150T ~ 184T | 200T | SE |
{12}------------------------------------------------
| Sponsor: | Guangzhou Xinbo Electronic Co., Ltd. |
|-----------------|-------------------------------------------------------------------|
| Subject Device: | Pain Therapy Device, Models: P.T.S-II, P.T.S-IIA, P.T.S-IIB, CP-I |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device (K163611) | Predicate Device (K133929) | Remark |
|-----------------------------|---------------------------------------------|----------------------------------------------|----------------|
| Number (T) for One<br>Burst | | | Note (a) |
| - Maximum Output<br>Voltage | 40Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>95Vpk @ 10kΩ | 44Vpk @ 500Ω<br>80Vpk @ 2kΩ<br>112Vpk @ 10kΩ | SE<br>Note (b) |
#### Comparison in Detail(s):
#### Note (a):
Although the "Pulse Duration", "Pulse Frequency" and "Pulse Cycle Number for One Burst" of subject device are different from the predicate device, they are very similar in waveform group, and are all comply with IEC 60601-1 and IEC 60601-2-10 requirements. So the differences of the function specifications will not raise any safety or effectiveness issue.
#### Note (b):
Although the "Maximum Output Voltage" of subject device is little different from the predicate device, their maximum peak voltage are very similar, and are all comply with IEC 60601-1 and IEC 60601-2-10 requirements. So the differences of the function specifications will not raise any safety or effectiveness issue.
### Finial Conclusion:
The subject device "Pain Therapy Device" is Substantial Equivalent to the predicate devices.
### 8. Date of the summary prepared: September 19, 2017
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.