The Livia is designed for symptomatic relief and management of chronic intractable pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower (extremities) leg due to strain from exercise or normal household work activities. It should be applied to normal, healthy, dry and clean skin of adult patients, and is to be used for stimulate healthy muscles in order to improve and facilitate muscle performance
Device Story
Livia is a TENS device for pain management and muscle stimulation. It delivers high-frequency electrical pulses via skin-surface electrodes to block pain signals to the brain and low-frequency bursts to stimulate natural pain control (beta-endorphins). Device is intended for OTC use by adults. Input: user-controlled intensity settings. Output: biphasic, symmetrical, rectangular electrical pulses. Operation: battery-powered (3.7V Li-ion) unit with microprocessor control; electrodes placed on peripheral nerves. Healthcare providers do not operate; patient self-administers. Output affects clinical decision-making by providing non-pharmacological pain relief. Benefits include symptomatic pain management and improved muscle performance without systemic drug side effects.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-5, -10), and a usability study (IEC 62366-1) confirming no user errors.
Technological Characteristics
TENS/EMS device; PC/ABS plastic housing; 3.7V rechargeable Li-ion battery; microprocessor-controlled; biphasic, symmetrical, rectangular waveform; 100µs pulse duration; 60-100Hz random frequency; output electrically disabled during charging. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for adult patients for symptomatic relief/management of chronic intractable pain and temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, neck, upper/lower extremities due to exercise or household strain; also for stimulation of healthy muscles to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2017
Lifecare Ltd. % Irving Wiesen Official Correspondent Cohen, Tauber, Spievack & Wagner 420 Lexington Avenue-suite 2400 New York, New York 10170
Re: K163153
Trade/Device Name: Livia Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: May 25, 2017 Received: May 30, 2017
Dear Irving Wiesen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
William J. Heetderks -S 2017.06.28 15:57:47 -04'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163153
Device Name Livia
Indications for Use (Describe)
The Livia is designed for symptomatic relief and management of chronic intractable pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower (extremities) leg due to strain from exercise or normal household work activities.
It should be applied to normal, healthy, dry and clean skin of adult patients, and is to be used for stimulate healthy muscles in order to improve and facilitate muscle performance
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 10px;">☐</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10px;">☑</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(K) SUMMARY
The assigned 510(K) Number: K163153
#### 1. Submitter's Identification:
### LifeCare Ltd.
Zipori St.2 1424602, Tiberias, Israel
Date Summary Prepared: May 25, 2017
Contact Person: Irving L. Wiesen, Esq. Law Offices of Irving L. Wiesen P.C. 420 Lexington Avenue - Suite 2400 New York, New York 10170 Tel 212-381-8774 Fax 646-536-3185 Email: iwiesen@wiesenlaw.com
#### 2. Name of the Device:
Livia
#### 3. Common Name and Classification:
Stimulator, Nerve, Transcutaneous, Over-The-Counter Requlation: 21 CFR Part 882.5890 Product Code: NUH
#### 4. Predicate Device Information:
- Primary Predicate: Philips Consumer Lifestyle PulseRelief (K151035) ●
- Reference Predicate: NeuroMetrix ASCEND (K10433)
#### 5. Device Description:
The LIVIA is a TENS device. TENS is an acronym for Transcutaneous Electrical Nerve Stimulator, and it works as a pain treatment system through electrotherapy. The unit sends light electrical pulses into the body through the skin by the use of electrodes. which are placed over peripheral nerves. The TENS unit works by sending high frequency electrical signals that are both continuous and mild to block out those pain signals that are being delivered to the brain. The Livia device was designed with specific pulse frequency and pulse length that are suitable for its intended use. When these pain signals are halted, pain isn't felt by the reactive area and the patiernt get relief. Low frequency bursts of mild electrotherapy also help the natural pain control response to activate, and these beta endorphins ease the pain that are being felt by the patient.
{4}------------------------------------------------
#### 6. Indication For Use:
The Livia is designed for symptomatic relief and management of chronic pain, and for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower (extremities) leg due to strain from exercise or normal household work activities. It should be applied to normal, healthy, dry and clean skin of adult patients, and is to be
used for stimulate healthy muscles in order to improve and facilitate muscle performance.
#### 7. Comparison to Predicate Devices:
The Livia device indications for use and technological characteristics are substantially equivalent to the Philips Consumer Lifestyle PulseRelief device application (K151035).
The primary predicate device is the Philips Consumer Lifestyle PulseRelief device and the reference predicate is NeuroMetrix ASCEND device application (K104333).
# TABLE 1: Indication For Use Comparison Table
| Parameter | Primary Predicate | Subject Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Philips Consumer<br>Lifestyle PulseRelief<br>(K151035) | LIVIA |
| Indication<br>For Use | The OTC TENS/EMS<br>stimulator PulseRelief is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>neck, upper extremities<br>(arm) and lower<br>extremities (leg) due to<br>strain from exercise or<br>normal household work<br>activities. It should be<br>applied to normal,<br>healthy, dry and clean<br>skin of adult patients, and<br>is to be used for stimulate<br>healthy muscles in order | The Livia is designed for symptomatic relief and<br>management of chronic pain, and for temporary<br>relief of pain associated with sore and aching<br>muscles in the shoulder, waist, back, neck, upper<br>extremities (arm) and lower (extremities) leg due<br>to strain from exercise or normal household<br>work activities.<br>It should be applied to normal, healthy, dry and<br>clean skin of adult patients, and is to be used for<br>stimulate healthy muscles in order to improve<br>and facilitate muscle performance. |
Comparison Tables - Technological Characteristics
{5}------------------------------------------------
| | to improve and facilitate<br>muscle performance. | |
|-------------------|--------------------------------------------------|----------------------------------------|
| Class | II | II |
| Product Code | NUH; NGX | NUH; NGX |
| Regulation Number | 21 CFR 882.5890(b);<br>21 CFR 890.5850 | 21 CFR 882.5890(b);<br>21 CFR 890.5850 |
# Table 2: Basic Unit Characteristics Comparison Table
| Parameter | Primary Predicate<br>Philips Consumer<br>Lifestyle<br>PulseRelief<br>(K151035) | Livia | Significant<br>Differences | |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | K151035 | K163153 | N/A | |
| Device Name and Model Number<br>Manufacturer | PulseRelief<br>Philips Consumer<br>Lifestyle | LIVIA<br>Life Care Ltd | N/A | |
| Power Source(s) | 3.7V Lithium-ion | 3.7V Lithium-ion<br>battery<br>(rechargeable) | Identical | |
| Method of Line Current Isolation | N/A | Output is<br>electrically<br>disabled when<br>connect to<br>charger, by<br>means of<br>microprocessor<br>charging circuit | Similar to the<br>reference predicate<br>Neurometrix<br>device cannot be used<br>while battery is<br>charging | |
| Patient Leakage Current<br>- Normal Condition (μΑ) | Battery powered<br>(< 10μΑ) | Battery powered<br>(< 10μΑ) | Identical | |
| Patient Leakage Current<br>- Single Fault Condition (μΑ) | Battery powered<br>(< 50μΑ) | Battery powered<br>(< 50μΑ) | Identical to Philips | |
| Average DC current through<br>Electrodes when device is on but no<br>pulses<br>are being applied (μA) | 0 μΑ | 0 μΑ | Identical to Philips | |
| Number of Output Modes | 15 TENS, 5 EMS | 1 | Similar to the<br>reference predicate<br>Neurometrix | |
| Number of Output<br>Channels: | Synchronous or<br>Alternating<br>Method of Channel | 1<br>N/A | 1<br>N/A | Identical |
{6}------------------------------------------------
| | Isolation | | | |
|----------------------------------------------|--------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulated Current or Regulated | | Current | Current | Identical |
| Voltage | | | | |
| Software/Firmware/Microprocessor<br>Control? | | Yes | Yes | Identical |
| Automatic Overload Trip | | Yes | Yes | Identical |
| Automatic No-Load Trip | | Yes | Yes | Identical |
| Automatic Shut Off | | Yes | No | |
| User Override Control | | Yes | Yes | Identical |
| Indicator<br>Display: | On/Off Status | Yes | Yes | Identical |
| | Low Battery | Yes | Yes | Identical |
| | Voltage/Current<br>Level | Yes | Yes | Identical to Philip |
| Timer Range (minutes) | | 1-59 minutes | | The Livia has no<br>interal timer, a minor<br>difference, as there is<br>no treatment time<br>limitation for using the<br>Livia-see Instructions<br>for Use. |
| Compliance with Voluntary<br>Standards? | | IEC 60601-1, IEC<br>60601-1-2, IEC<br>60601-2-10, ISO<br>10993-5 and -10 | IEC 60601-1,<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-10,<br>ISO 10993-5 and -<br>10 | Identical |
| Compliance with 21 CFR 898? | | Yes | Yes | Identical |
| Weight (g) | | 62 g | 36 g | None, all devices are<br>of hand-held weight |
| Dimensions (mm) (W x H x D) | | 2 units, each<br>54 x 54 x14 | 55 x 55 x 18 | None, all devices are<br>of hand-held size |
| Housing Materials and<br>Construction | | PC/ABS plastic | PC/ABS plastic | Identical |
Table 3: Output Specification for TENS Mode Comparison Table
| Parameter | Primary Predicate<br>Philips Consumer<br>Lifestyle<br>PulseRelief<br>(K151035) | Livia | Significant<br>Differences | |
|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Waveform (e.g., pulsed | Biphasic,<br>Symmetrical | Biphasic,<br>Symmetrical | Identical | |
| monophasic, | | | | |
| biphasic) | | | | |
| Shape (e.g., rectangular, spike) | Rectangular | Rectangular | Identical | |
| Maximum Output Voltage | 50V @ 500Ω | 50V @ 500Ω | Identical to the reference | |
| (volts) | | | predicate NeuroMetrix | |
| (+/-10%) | 100V @ 2kΩ | 64V @ 2kΩ | Similar to both devices and within | |
| | | | | their range |
| | | 100V @ 10kΩ | 64V @ 2kΩ | Similar; the maximum output<br>voltage of the Livia, although<br>slightly lower than the predicates,<br>provides the required intensity for<br>an effective treatment with lower<br>risk of discomfort to the user due<br>to high voltage |
| Maximum Output Current<br>(mA)<br>(+/-10%) | | 100mA @ 500Ω | 50 mA @ 500Ω | Similar to both devices and within<br>their range<br>The Livia was designed to supply<br>the minimum required output for<br>providing an effective treatment.<br>This in fact lowers the risk to the<br>patient |
| | | 50mA @ 2kΩ | 31mA @ 2kΩ | The Livia was designed to supply<br>the minimum required output for<br>providing an effective treatment.<br>This in fact lowers the risk to the<br>patient |
| | | 10mA @ 10kΩ | 6.4mA @ 10kΩ | The Livia was designed to supply<br>the minimum required output for<br>providing an effective treatment.<br>This in fact lowers the risk to the<br>patient |
| Duration of primary<br>(depolarizing)<br>phase (usec) | | 100 µs | 100 µs | Same as the reference predicate<br>NeuroMetrix |
| Pulse Duration (both phases)<br>(usec) | | 200µs, additional<br>30µs<br>inter-phase delay | 100 µs | Similar, within the range of the<br>predicate devices |
| Frequency (Hz) | | Random, mean<br>80Hz,<br>uniform<br>distribution<br>60-100Hz | 100 Hz | Identical to PulseRelief |
| For<br>multiphasic<br>waveforms<br>only: | Symmetrical<br>phases<br>Phase<br>Duration<br>(include units),<br>(state range, if<br>applicable),<br>(both phases,<br>if<br>asymmetrical) | Yes | Yes | Identical |
| | | 100µs (each<br>phase) | 100 µs | Identical to the reference<br>predicate NeuroMetrix |
| Net Charge (microcoulombs<br>(µC) | | Nominally 0uC<br>@ 500Ω | 0uC<br>@ 500Ω | Identical to PulseRelief |
| per pulse) (If zero, state how<br>this<br>was achieved) | zero net current | | | |
| Maximum Phase Charge, (µC) | 10 µC @<br>500Ω | 6.4 µC @<br>500Ω | Similar and within the range of<br>PulseRelief | |
| Maximum Current Density,<br>(mA/cm2, r.m.s.) | 0.51 mA/cm2<br>@ 500Ω | 0.38<br>mA/cm2<br>@500Ω | The minor differences are<br>resulting from the minor spec<br>difference as well as the<br>difference in electrode area. | |
| Maximum Average Current<br>(average absolute value), mA | 1.6 mA @ 500Ω | 1.19 mA @<br>500Ω | Similar PulseRelief | |
| Maximum Average Power<br>Density,<br>(W/cm2), (using smallest<br>conductive<br>surface area) | 3.6 mW/cm2 @<br>500Ω | 2.05<br>mW/cm2<br>@500Ω | Similar PulseRelief | |
| NBurst<br>Mode;<br>(i.e., pulse<br>trains) | (a) Pulses per<br>N/A | N/A | N/A | N/A |
| | (b) Bursts per<br>second | N/A | N/A | N/A |
| | (c) Burst<br>duration<br>(seconds) | N/A | N/A | N/A |
| | (d) Duty Cycle:<br>Line (b) x Line<br>(c) | N/A | N/A | N/A |
{7}------------------------------------------------
{8}------------------------------------------------
The candidate device is similar or very close to the predicate devices in its all major claims. The minor differences between predicate and candidate device do not affect the safety and effectiveness.
#### 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Non-clinical verification testing of the LIVIA device included electrical, mechanical, and software tests to show the device meets its design specifications. Validation and performance testing validates that the device meets its user needs. Verification and validation test results established that the device meets its intended use, that it is as safe, as effective, and performs as well as the predicate devices, and that no new issues of safety and effectiveness were raised. The LIVIA device was designed, verified, and validated according to the company's Design Control process and has been
{9}------------------------------------------------
subjected to extensive safety and performance testing as shown in the test results provided in this submission. Verification and Validation testing data demonstrate that the device meets all of its specifications including compliance with the following standards:
### Safety :
- IEC 60601-1:2005/A1:202, EN 60601-1:2006/A1:2013
- EN 60601-2-10:2012/A1:2012
- IEC 60601-1-11:2015
## EMC:
- IEC 60601-1-2:2007
## Software:
- IEC 62304:2006
Biocompatibility:
- ロ ISO 10993-5:2009
- ISO 10993-10:2010
## Usability Study:
- ロ IEC 62366-1: 2015
The Livia device and its accompanying documentation, i.e., the IFU have met all usability specifications that were defined. No user errors or near errors were observed.
#### 9. Discussion of Clinical Tests Performed:
Livia Inc. determined that bench and non clinical testing were sufficient to demonstrate that the LIVIA device is as safe and effective as the predicate devices.
#### 10. Conclusions:
Livia is substantially equivalent in its intended use, performance, safety, effectiveness and underlying scientific and operating principal used, to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.