TENS

K133979 · Sunmas (H.K.) Trading Limited · NUH · Aug 26, 2014 · Neurology

Device Facts

Record IDK133979
Device NameTENS
ApplicantSunmas (H.K.) Trading Limited
Product CodeNUH · Neurology
Decision DateAug 26, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

It is intended to be used for temporary relief of pain associated with sore and aching muscle in the shoulder, wast, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.

Device Story

TENS device (models SM9066, SM9088, SM9090, SM9098, SM9128, SM9198) provides electrical nerve stimulation for pain relief. Device consists of LCD, control keys, switch, battery, output jacks, and USB port; uses external electrode pads (cleared in K092546) applied to patient skin. Device generates biphasic, rectangular pulses to deliver electrical current. Operated by patient for OTC use. Healthcare providers do not operate; output is used by patient to manage muscle pain. Benefits include temporary pain relief without medication.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-10 (nerve/muscle stimulators) verified. Usability study conducted in the US confirmed labeling is well-understood by patients for safe and effective use. No clinical trials performed.

Technological Characteristics

Powered by 3.7V DC Lithium battery. Delivers biphasic, rectangular pulses (100us duration). Output parameters: 42-140V and 14-84mA depending on load (500Ω-10kΩ). Connectivity includes USB port. Biocompatibility per ISO 10993-5 and ISO 10993-10. Form factor includes LCD and control interface.

Indications for Use

Indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, arms, and legs due to exercise or household work strain.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines above and below them, resembling a bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 26, 2014 Sunmas (H.K.) Trading Limited c/o Diana Hong Mid-Link Consulting Co., Ltd 13th Floor, Wah Kit Commercial Centre, 302 Des Voeux Road Central, G.P.O. Box 2878 Hong Kong, China Re: K133979 Trade/Device Name: TENS Models SM9066, SM9090, SM9090, SM9098, SM9128, and SM9198 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: July 26, 2014 Received: July 29, 2014 Dear Ms. Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Felipe Aguel -S Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K133979 Device Name TENS Model: SM9066, SM9088, SM9090, SM9098, SM9128 and SM9198 #### Indications for Use (Describe) It is intended to be used for temporary relief of pain associated with sore and aching muscle in the shoulder, wast, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities: Type of Use (Select one or both, as applicable) - Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CER 801 Subpart C) ### PLEASE DO NOT WRITE BECOW THIS LINE - CONTINUE ON A SEPARATE PAGE IE NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Felipe Date: 2014.08.26 22:12:08 -04'00' Aquel -S This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED EORM IO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect: of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staffl PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92. - 1. Date of Preparation: Nov 20, 2013 - 2. Sponsor Sunmas (H.K.) Trading Limited 13th Floor, Wah Kit Commercial Centre, 302 Des Voeux Road Central, G.P.O. Box 2878, Hong Kong Contact Person: Chunyi Liao Contact Title: Vice General Manager Tel: +852-25433505 Fax: +852-28152074 Email: sunmas@139.com - 3. Submission Correspondent Ms. Diana Hong & Mr. Tarzan Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023. Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net - 4. Proposed Device Identification Proposed Device Name: TENS Proposed Device Model: SM9066, SM9088, SM9090, SM9098, SM9128 and SM9198 Common Name: Stimulator, Nerve, Transcutaneous, Over-the-Counter Classification: II Product Code: NUH Regulation Number: 21 CFR 882.5890 Review Panel: Neurology Intended Use Statement: It is intended to be used for temporary relief of pain associated with sore and aching muscle in the {4}------------------------------------------------ shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. - న్. Predicate Device Identification 510(k) Number: K102598 Product Name: JQ-5C Manufacturer: Hi-Dow International, Inc - 6. Device Description The TENS is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. The TENS consists of LCD, Control Key, Switch, Battery, Output jack and USB port, its accessories includes electrode, electrode cable, USB wire and Battery charger. Only the electrode pads cleared in submission, K092546, are to be used with this device The proposed device has six models: SM9066, SM9088, SM9090, SM9098, SM9128 and SM9198. The six models have the same design principle and functional components, including battery, LCD, Control Key, Switch, channel number and USB port and accessories. The difference between the six models specifically expressed in the battery specification, LCD dimension, the position of Control key, Switch, and USB port on the proposed device, PCB(Printed Circuit Board) dimension and the number of Output jack numbers. {5}------------------------------------------------ | ITEM | Proposed device | Predicate device | |-----------------------------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------| | | | JQ-5C | | Power source | DC 3.7V Lithium Battery | DC 3.7V Lithium Battery | | Waveform | Pulse, biphasic | Pulse, biphasic | | Shape | Rectangular | Rectangular | | Pulse Duration(Pulse width) | 100us | 100us | | Net Charge | 0 µC@500Ω, balanced waveform | 0 µC(@500Ω, balanced waveform | | Same between proposed device and predicate device: | | | | Same power source, waveform, shape, pulse duration and net charge. | | | | Maximum Output Voltage | 42V ± 10%@500Ω | 62.4V@500Ω | | | 84V±10%@2kΩ | 79.2V@2.2kΩ | | | 140V ± 10%@10kΩ | 84V%@10kΩ | | Maximum Output Current | 84mA±10%@500ΩΩ | 124.8mA@500Ω | | | 42mA±10%@2kΩ | 39.65mA@2.2kΩ | | | 14mA±10%@10kΩ | 8.4 mA@10kΩ | | Differences between proposed device and predicate device: | | | | The maximum output voltage of the proposed device is covered by that of the predicate device; correspondingly the | | | | maximum output current of the proposed device is also covered by that of the predicate device. Meanwhile | | | | considering the ±10% deviation of each value, the maximum output voltage value and maximum output current | | | | value of the proposed device could be considered to be accepted and substantially equivalent to that of the predicate | | | | devices. | | | | Frequency | 68Hz,12.5~55.5 Hz, 1.17 Hz, 5.8 | Similar | | | Hz, 108 Hz, 29 Hz | | | Pulse period, msec | 14.7 ms, 18~80ms; 850ms; 172ms, | Similar | | | 9.2ms, 16.9ms | | | Similarities between proposed device and predicate device: | | | | Per IEC 60601-2-10, 51.104, it states that for equipment intended for therapeutic applications: with a load | | | | resistance of 5002 the output current r.m.s shall not exceed the limit of 50mA in case of Frequency ≤400Hz; and the | | | | RMS output current of the proposed device which is calculated based on the frequency parameters meets the IEC | | | | 60601-2-10 requirements, therefore although there is slight difference on pulse period and frequency between | | | | proposed device and predicate device, the two parameter is considered to be acceptable and substantially equivalent | | | | to the predicate devices | | | | Stimulus delivered for mode | Similar | Similar | | Similar stimulus delivery: | | | | Although there are no graphical drawings to assist in the description of the stimulus delivered in each mode for | | | | proposed device and predicate device, based on the SE discussion we could see that this stimulus delivered | | | from proposed device and predicate device are similar.; In other words, this stimulus delivery between proposed device and predicate device which adopt the same fundamental output technology will offer the similar treatment Table 3-1 the difference between proposed device and predicate device {6}------------------------------------------------ effect. Therefore, this item is considered to be substantially equivalent. Comparison Summary: The proposed device is compared to the predicate device in terms of power source, waveform, shape, pulse duration, net charge, frequency, pulse period and stimulus delivered of each mode. Based on the analysis of similar/different points between the proposed device and predicate device, the proposed device is finally considered to be substantially equivalent to the predicate device; and the differences will not affect the safety and effectiveness of the proposed device. - 7. Biocompatibility Summary The patient-contact component of the proposed device is the GMDASZ TENS Electrodes. The biocompatibility of the GMDASZ TENS Electrodes has been demonstrated in K092546. It was tested for cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation sensitivity (ISO 10993-10) with the test result in K092546. The biocompatibility of predicate device was demonstrated in term of cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation sensitivity (ISO 10993-10) in K102598. Therefore the biocompatibility of the proposed device is considered to be substantially equivalent to that of the predicate device (K102598). - Usability Discussion 8. In order to validate the usability performance of TENS device for US market, the sponsor organized a usability study in America. The study focused on the labeling information to support the safe and effective use of the proposed device. The usability study demonstrated that the final version of the labeling for the TENS, were well understood by a variety of "patients", and provided sufficient information for safe and proper use of the device. - Non-Clinical Test Conclusion 9. Bench tests were conducted to verify that the proposed device met all proposed design specifications. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1: 1988+ A1: 1991+ A2: 1995, Medical Electrical Equipment- Part 1: General requirements for safety. IEC 60601-1-2: 2007. Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility- Requirements and tests {7}------------------------------------------------ IEC 60601-2-10: 1987/A1: 2001, Medical electrical equipment, Part 2: Particular requirements for the safety of nerve and muscle stimulators. #### 10. Substantially Equivalent Conclusion The electrical stimulation provided by a series of model SM9066, SM9090, SM9098, SM9128 and SM9198 are similar to the commonly employed TENS devices that have been cleared for marketing without prescription labeling. These models have the same intended use and the similar technological characteristics as those OTC predicates. Moreover, verification and validation tests contained in this submission demonstrate that the difference in these models still maintain the same safety and effectiveness as that of the cleared devices. In other words, those engineering differences do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. The adequate guidance to the user on electrode application and electrode placement are presented in the User Manual; the information is used to insure the electrode can be used safely and effectively. We believe that there are no new safety or effectiveness issues concerning this device to be introduced. Technological characteristics, features, specifications, materials and intended uses of the series of model SM9066, SM9088, SM9090, SM9098, SM9128 and SM9198 are substantially equivalent to the quoted predicated devices. The differences that exist between the devices are insignificant in the terms of safety or effectiveness. The proposed device, TENS, Model of SM9066, SM9088, SM9090, SM9098, SM9128 and SM9198 is determined to be Substantially Equivalent (SE) to the predicate device, K102598, in respect of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...