To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, neck, back, arm, and leg, due to strain from exercise or normal household and work activities.
Device Story
Electronic Pulse Stimulator is a portable, microprocessor-controlled TENS device for OTC pain relief. Device delivers electrical pulses via connecting wires and adhesive electrode pads to user's skin; activates underlying nerves to provide temporary relief for sore/aching muscles. User operates device via front control panel to adjust pulse intensity. Device features two output channels and automatic shut-off. Intended for home use by patients to manage muscle strain from daily activities. Healthcare providers do not operate the device; output is used by the patient to manage pain symptoms.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC 60601-1-2, and software verification per FDA guidance.
Technological Characteristics
Battery-powered TENS device; 2 output channels; rectangular pulsed waveform; 100 Hz frequency; max charge 23 µC at 500Ω; max current density 1.4 mA/cm² at 500Ω. Microprocessor-controlled with user-adjustable intensity and automatic shut-off. Complies with IEC 60601-1 and IEC 60601-1-2 standards. Electrodes are adhesive pads.
Indications for Use
Indicated for temporary relief of pain in shoulder, waist, neck, back, arm, and leg muscles caused by exercise or normal household/work activities. For over-the-counter use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K|3|92|
## NOV 1 5 2013
### 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: Shenzhen Jingkehui Electronic Co., Ltd.
Address: 3/F, Building 1, B Area, Xifa Industrial Area, Yintian Village, Xixiang Town, Baoan District, Shenzhen. China
Contact Person: Pu Jiang Tel: +86-755-29970323 Fax: +86-755-23493443 Email: jkdz@vip.163.com Date of Preparation: 06/22/2013
## 2. Proposed Device
Trade Name: Electronic Pulse Stimulator Common Name: Transcutaneous electrical nerve stimulator Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter (OTC) Regulation Description: Transcutaneous electrical nerve stimulator for pain relief Regulation Medical Specialty: Neurology Review Panel: Neurology Product Code: NUH. NGX Regulation Number: 21 CFR 882.5890 Device Class: II Use: Over-The-Counter
## 3. Predicate device
Predicate Device: Prospera OTC TENS Electronic Pulse Massager 510(k) Number: K122744 Use: Over-The-Counter Submitter: Prospera Corporation
## 4. Description of Proposed Device
The Electronic Pulse Stimulator is a Transcutaneous Electrical Nerve Stimulator (TENS), intended for the over-the-counter use to relieve pain in different body areas (waist, back, neck, shoulders, legs, and arms). The proposed Electronic Pulse Stimulator, which is compact, portable, and microprocessorcontrolled, delivers a gentle electrical pulse through the connecting wires and electrode pads to the user's skin for pain relief. According to the need of users, the pulse intensity can be adjustable on the front control panel of the device.
#### 5. Intended Use Statement of Proposed Device
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, neck, back, arm, and leg, due to strain from exercise or normal household and work activities.
#### 6. Substantial Equivalence
The operational principle of the above predicate device is to generate small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying
Page 1 of 3
{1}------------------------------------------------
nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
Identically, the proposed device generates small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves are activated and the pain associated with sore and aching muscles is temporarily relieved.
Table 1 below summarizes the comparison between the proposed device and predicate device, indicating the technical characteristics, specifications, and intended use of the proposed device are substantially equivalent to those of the predicate device.
| | Proposed Device | Predicate Device |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K131921 | K122744 |
| Device Name | Electronic Pulse Stimulator | Prospera OTC TENS Electronic<br>Pulse Massager |
| Intended Use | To be used for temporary relief<br>of pain associated with sore and<br>aching muscles in the shoulder,<br>waist, neck, back, arm, and leg,<br>due to strain from exercise or<br>normal household and work<br>activities. | To be used for temporary relief<br>of pain associated with sore and<br>aching muscles in the shoulder,<br>waist, back, neck, upper<br>extremities (arm), and lower<br>extremities (leg) due to strain<br>from exercise or normal<br>household work activities. |
| Power Source | Battery | Battery |
| Number of Output Channels | 2 | 2 |
| Automatic Overload Trip | No | No |
| Automatic No-Load Trip | No | No |
| Automatic Shut Off | Yes | Yes |
| User Override Control | Yes | Yes |
| Indicator | Yes | Yes |
| Waveform | Pulsed | Pulsed |
| Shape | Rectangular | Rectangular |
| Frequency (Hz) | 100 | 100 |
| Maximum charge (µC) at<br>500Ω | 23 | 23 |
| Maximum current density<br>(mA/cm²) at 500Ω | 1.4 | 1.4 |
| Compliance with Voluntary<br>Standards | IEC60601-1, IEC60601-1-2 | IEC60601-1, IEC60601-1-2 |
| Compliance with 21 CFR<br>898 | Yes | Yes |
Table 1. Comparison between the proposed device and predicate device
## 7. Non-Clinical Tests Performed
The Electronic Pulse Stimulator does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the proposed device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.
{2}------------------------------------------------
- (a) IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance".
- (b) IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Safety -Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
In addition to the compliance of voluntary standards, the verification of software used in the proposed device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The electrodes, as the accessory of the proposed device, also meet the requirement of safety.
## 8. Conclusion
The tests performed and the comparison of technical characteristics, specifications, and intended use demonstrate the proposed device is substantially equivalent to the predicate device. Therefore, the proposed device is as safe, as effective, and performs as well as the foregoing identified OTC predicate devices that have been legally marketed in the United States.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 15, 2013
Shenzhen Jingkehui Electronic Co., Ltd. c/o Qaunqin (Bill) Dai Sunovo, LLC 513 Piazza Drive, Unit B Mountain View, CA 94043
Re: K131921
Trade Name: Electronic Pulse Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NGX Dated: October 28, 2013 Received: October 30, 2013
Dear Dr. Dai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 – Qaunqin (Bill) Dai
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K131921
Device Name: Electronic Pulse Stimulator
Indications For Use:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, neck, back, arm, and leg, due to strain from exercise or normal household and work activities.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.