IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit
K200807 · Irras USA, Inc. · JXG · Apr 29, 2020 · Neurology
Device Facts
Record ID
K200807
Device Name
IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit
Applicant
Irras USA, Inc.
Product Code
JXG · Neurology
Decision Date
Apr 29, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
Device Story
System monitors intracranial pressure (ICP) and drains cerebrospinal fluid (CSF). Input: CSF via catheter; gravity-driven flow. Transformation: Control unit uses peristaltic pump and pinch valve to regulate drainage speed based on user settings; includes bolus injection capability for catheter flushing. Output: Real-time ICP monitoring, drainage of fluid into aspiration bag. Used in clinical settings by healthcare providers. Laser leveler accessory assists in aligning control unit with patient's external auditory meatus for accurate drainage height. Benefits: Automated pressure management and monitoring; reduces risk of catheter clogging via flushing.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing for the laser leveler accessory included dimensional, functional, laser safety (IEC 60825-1), fluid ingress (IEC 60529 IP22), and environmental/distribution testing (ASTM D4332, D4169, D5276, D642, D999, D6653, D4729, D6344, F1980).
Technological Characteristics
System includes control unit, tube set, catheter, and laser leveler accessory. Sensing: Pressure transducers for ICP measurement. Actuation: Stepper-motor controlled peristaltic pump and pinch valve. Connectivity: Standalone. Power: Battery back-up. Laser Leveler: IEC 60825-1 compliant, IP22 rated. Software: Embedded firmware for UI, alarms, and data display. Sterilization: Sterile disposable components.
Indications for Use
Indicated for patients requiring intracranial pressure monitoring and external drainage of intracranial fluid to reduce intracranial pressure.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 29, 2020
IRRAS USA Inc. Niloufa Insanally, Ph.D., RAC Head of Regulatory Affairs 11975 El Camino Real, Suite 304 San Diego, California 92130
Re: K200807
Trade/Device Name: IRRAflow® CNS System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG. GWM Dated: March 28, 2020 Received: March 30, 2020
Dear Dr. Niloufa Insanally:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200807
Device Name IRRAflow® CNS System
Indications for Use (Describe)
The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k)Summary
Applicant:
IRRAS USA, Inc. 11975 El Camino Real, Suite 304 San Diego, CA 92130 U.S.A Phone Number: 858-220-3761 Fax number: 866-575-1002 FDA Registration #: 3013508628
Primary Contact Personnel:
Niloufa Insanally, Ph.D. RAC Head of Regulatory Affairs Phone Number: 858-220-3761 Fax Number: 866-575-1002 Email: niloufa.insanally@irras.com
Secondary Contact Personnel:
| Kleanthis G. Xanthopoulos, Ph.D., | |
|------------------------------------|---------------------------------------------------|
| President and CEO | |
| Phone Number: | 858-336- 8093 |
| Fax Number: | 866-575- 1002 |
| Email: | Kleanthis.xanthopoulos@irras.com |
| Date prepared: | February 07, 2020 |
| Trade name: | IRRAflow® CNS System |
| Common Name: | CSF Drainage System with ventricular catheter |
| Primary Classification: | |
| Name: | Central Nervous System Fluid Shunt and Components |
| Product Code: | JXG |
| Regulation: | 21 CFR 882.5550 |
| Secondary Classification: | |
| Name: | Intracranial Pressure Monitoring Device |
| Product Code: | GWM |
| Regulation: | 21 CFR 882.1620 |
| Predicate and Reference Device(s): | |
IRRAflow® CNS System (K171880, K192289)
510(k) Summary
{4}------------------------------------------------
#### DEVICE DESCRIPTION – IRRAflow CNS System
The IRRAflow CNS System is an intracranial pressure (ICP) monitoring and drainage system, and remains identical, as cleared under predicate 510(k) K192289. The IRRAflow® CNS System consists of an IRRAflow Control Unit and two sterile disposable parts, the IRRAflow Tube Set and the IRRAflow Catheter.
The drainage flow of cerebrospinal fluid (CSF) into the IRRAflow Catheter is uni-directional and gravity-driven; there is no recirculation of the CSF. A parallel line from the saline infusion bag is used in case clearance at the tip of the catheter is required. The IRRAflow Tube Set has a cassette that clicks on to the IRRAflow Control Unit and aligns the tubing against a peristaltic pump and pinch valve. An aspiration bag is attached to the Control Unit tape measure, defining the height of the bag relative to the catheter's tip position in the patient's head and thus controlling the speed of drainage. The tubing and catheter can be disconnected and connected by standard Luer-Lock connectors. Settings can be changed via the user interface on the Control Unit.
The default mode provides drainage and measuring ICP, allowing single bolus injections when indicated. The bolus injections allow the catheter to be flushed when it becomes clogged. CSF or intracranial fluid samples can be taken from the aspiration port.
#### INDICATIONS FOR USE
The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed.
{5}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE
The IRRAflow® CNS system's intended use, technological characteristics and principles of operation are the same as the predicate device described in K192289. For ease of reading, both earlier predicate devices described in K171880 and K192289 are shown for comparison in Table 1 below.
Comparison of these Monitoring Systems will show that the only difference in the device presented in this Special 510(k) is the inclusion of the Laser Leveler Accessory tool, which supports and augments the IRRAflow CNS system by providing a more accurate and easier method for positioning the control unit (or drainage system) with the patients external auditory meatus (or external ear canal). Using the laser leveler tool, the drainage system can be raised or lowered until the laser light points horizontally to the correct landmark on the patient's head. The common landmark used is the external auditory canal.
The earlier methods used for this alignment involved the placement of a taut string from the patient's head to the drainage system. This method for alignment may still be used. However, the laser leveler alignment tool presents a better method for this alignment step which is performed when the instrument is set up. The IRRAflow Control Unit was originally designed with a screw connection port which would hold the laser leveler tool, and IRRAS's earlier intention was to submit the laser leveler tool along with the most recent 510(k) submission (K192289) in August 2019, but the testing for the laser leveler tool was not completed at the time of the K192289 submission.
Therefore, this Special 510(k) captures the information related to the laser leveler accessory which is part of the overall IRRAflow CNS System that was originally classified by FDA as a Class II medical device. Traditionally, FDA has been able to classify accessories within the same classification as the parent device and for this reason, IRRAS is submitting this Special 510(k) as a follow up adjunct to the IRRAflow CNS System device described in K192289.
{6}------------------------------------------------
| Items | PREDICATE<br>IRRAflow® CNS System<br>(K171880)<br>(JXG), (GWM) | PREDICATE<br>IRRAflow® CNS System<br>(K192289)<br>(JXG), (GWM) | IRRAflow® CNS System<br>Special 510(k)<br>(JXG), (GWM)<br>Equivalence |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Primary Product Code | JXG | JXG | Same |
| Primary Regulation<br>Number: | 21 CFR 882.5550 | 21 CFR 882.5550 | Same |
| Secondary Product Code | GWM | GWM | Same |
| Secondary Regulation<br>Number: | 21 CFR 882.1620 | 21 CFR 882.1620 | Same |
| Indications for Use | The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed for ≤ 24 hours. | The use of IRRAflow® CNS System is indicated when intracranial pressure monitoring is required and for externally draining intracranial fluid as a means of reducing intracranial pressure in patients where an external drainage and monitoring system is needed. | Same as for K192289 |
| Injection/ CSF Sampling<br>Ports | Yes | Yes | Same |
| Unidirectional<br>Flow of Drained Fluid | Yes | Yes | Same |
| Fluid Injection Capability | Yes | Yes | Same |
| Attaches to separate,<br>commercially available<br>EVD Catheter | Yes<br>The IRRAflow system attaches to IRRAflow Catheter which is an EVD Catheter part of the complete system. | Yes<br>The IRRAflow system attaches to IRRAflow Catheter which is an EVD Catheter part of the complete system. | Same |
| Sterile Disposable tubing<br>set | Yes | Yes | Same |
| CSF Drainage Bag | Yes | Yes | Same |
| Gravity drainage of CSF | Yes | Yes | Same |
| Method to control gravity<br>drainage of CSF | Automated adjustment based on user settings via a stepper-motor controlled, tube-pinching mechanism to either compress or release the compliant drainage tubing contained within the sterile, disposable Cartridge. | Automated adjustment based on user settings via a stepper-motor controlled, tube-pinching mechanism to either compress or release the compliant drainage tubing contained within the sterile, disposable Cartridge. | Same |
| | | | Table 1 - Substantial Equivalence Comparison of the Monitoring Systems | | | | |
|--|--|--|------------------------------------------------------------------------|--|--|--|--|
|--|--|--|------------------------------------------------------------------------|--|--|--|--|
{7}------------------------------------------------
| Items | PREDICATE<br>IRRAflow CNS System<br>(K171880)<br>(JXG), (GWM) | PREDICATE<br>IRRAflow CNS System<br>(K192289)<br>(JXG), (GWM) | IRRAflow CNS System<br>Special 510(k)<br>(JXG), (GWM)<br>Equivalence |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Pressure Transducer for<br>ICP Measurement | Yes<br>(The IRRAflow system<br>integrates transducers into its<br>design for measurement and<br>visual display of ICP) | Yes<br>(The IRRAflow system<br>integrates transducers into its<br>design for measurement and<br>visual display of ICP) | Same |
| Software-based, Powered<br>Console for User Interface,<br>User Settings and Alarm<br>Adjustments, Data Storage<br>and Display, and Alarms for<br>ICP Monitoring | Yes | Yes | Same |
| Method to account for<br>location of ventricles via<br>patient head position | Reference marks on the device<br>to allow for the system to be<br>aligned with patient's head<br>positioning. | Reference marks on the device<br>to allow for the system to be<br>aligned with patient's head<br>positioning. | "Same" + Alternative method<br>to use the Laser Leveler<br>accessory to position the<br>system (control unit) with the<br>patient's head. |
| Measured Pressure Range | -80 mmHg to +100 mmHg | -80 mmHg to +100 mmHg | Same |
| Displayed ICP | Yes | Yes | Same |
| Battery Back-up | Yes | Yes | Same |
An inspection of Table 1 will show that there is no change in the actual IRRAflow CNS System so that no additional testing (or verification and validation testing) on the CNS system was required. There are no changes to the technological characteristics or principles of operation of the drainage system itself; no new safety and effectiveness issues are raised with this new device. The only change was to add a separate tool, the IRRAflow laser leveler, to the overall system to facilitate the alignment of the drainage system with the patient's head at the beginning of the system (refer to redlines in Table 1).
{8}------------------------------------------------
| Method to account for location of ventricles via patient head position | | |
|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IRRAflow CNS system<br>(K171880) | IRRAflow CNS system<br>(K192289) | IRRAflow CNS system<br>Special 510(k) |
| Reference marks on the device to<br>allow for the system to be aligned<br>with patient's head positioning. | Reference marks on the device to allow<br>for the system to be aligned with<br>patient's head positioning. | <b>Discussion:</b> Same as for<br>predicate(s) but also added the use<br>of the Laser Leveler tool to adjust<br>positioning of IRRAflow CNS<br>system so that it is aligned with<br>patient's head (external auditory<br>meatus). |
## Discussion of differences in Table 1
It should be noted that the earlier methods involving a line level measurement which employs a taut string can still be used for alignment, but the laser leveler tool presents an alternative and better method for alignment of the drainage system with the patient's head.
Therefore, there is no change in the overall IRRAflow CNS system, except for the addition of an accessory to aid in positioning of the drainage system in alignment with the location of the ventricles in the patient's head. The use of the laser leveler is an additional option for location of the ventricles via the patient head position.
## Use of the The IRRAflow® Laser Leveler
The IRRAflow® laser leveler is a reusable accessory to be utilized in conjunction with the IRRAflow® CNS system. The laser leveler generates a laser mark to assist in setting the height of the control unit. The IRRAflow® Laser Leveler device is provided with a mounting screw for use with the The IRRAflow® CNS control unit. The laser leveler has a power button and auto shuts off after approximately 20 seconds.
The control unit has a screw connection on the side of the unit and the laser leveler has a shoulder screw which can be screwed into the control unit using an Allen wrench. Once attached to the control unit, the control unit assembly can be moved up and down on an IV pole, and the laser leveler should swing freely (it is weighted down). The laser leveler may then be used to auto align the control unit assembly to the patient's head.
The laser light is turned on and remains on for a maximum of 20 seconds, during which time the emitted light is pointed directly to the patient's external auditory meatus (in order to align with the patient's ventricles). Once the control unit drainage system is aligned horizontally with the patient's auditory meatus, the assembly is locked into position. The laser automatically shuts off after 20 seconds, the laser light will disappear, and the drainage system will be ready to operate.
{9}------------------------------------------------
No changes were made to the EVD catheter or the IRRAflow CNS system.
## 1. EVD Catheter:
The IRRAflow® CNS System's Catheter technological characteristics and principles of operation are identical to the IRRAflow® CNS System Catheter (K192289). There have been no changes to the system catheter. The materials, design and principle of operations remain the same.
### 2. Verification and Validation Documentation for the IRRAflow CNS System:
Similarly, the addition of the laser leveler accessory does not change or impact any of the verification and validation testing performed earlier on the IRRAflow® CNS System. The verification and validation performance testing which was performed earlier (refer to K192289) has not changed. In conclusion, the IRRAflow® CNS System is substantially equivalent to the predicate device.
### 3. Verification and Validation testing for the Laser Leveler:
While no changes were made to the IRRAflow CNS System as a whole, and therefore no additional verification and validation testing was required for the system, there was verification and validation testing performed during the design of the laser leveler accessory. Development and testing for the laser leveler was performed at a contract manufacturer, Meraqi Medical, Inc. (Freemont, CA). The laser testing was performed at a sub-contractor, Intertek (Menlo Park, CA). The design of the laser leveler was performed at IRRAS and transferred to Meraqi Medical for development (including design controls), testing and subsequently, manufacturing.
{10}------------------------------------------------
#### SUMMARY OF VERIFICATION & VALIDATION FOR THE LASER LEVELER
No additional verification and validation performance testing was done on the IRRAflow® CNS System itself since no new safety and effectiveness issues were raised with the 'new' device. The IRRAflow CNS System remains the same as described in 510(k) K192289. The only change is the addition of the laser leveler accessory. The only new risks to the overall system with the inclusion of the laser leveler accessory, is the introduction of laser energy in the system and the risk posed by exposing the patient to laser light. The risk was evaluated separately, as well as, in conjunction with the laser leveler tool being used in the initial setup of the IRRAflow CNS system. Since no new concerns or risks were identified that could not be mitigated by the actual proper use and function of the laser leveler, and there were built-in fail safes in the laser leveler (i.e. the laser light automatically turns off after 20 seconds), one may conclude that the IRRAflow® CNS System is substantially equivalent to the predicate device (K192289).
However, a full analysis of the laser leveler was performed, and a summary of the testing done at Meraqi Medical and Intertek is presented in Table 2 below.
It is important to note that since no new performance data for the actual IRRAflow CNS system are needed to evaluate the change (i.e. use of the laser leveler accessory with the IRRAflow CNS system), the submission of a Special 510(k) is appropriate in this case. Furthermore, in Table 2 below, IRRAS has identified the various well-established and recognized standards which allow one to evaluate the performance of the laser leveler accessory separately. The data or results are provided in Table 2 in a summary format.
{11}------------------------------------------------
| Test | Test Method Summary | Results |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimensional | Verify units meet dimensional specification. | Pass |
| Visual (laser) | Verify units have legible labeling and are free of damage and/or<br>foreign particulate. | Pass |
| Visual (Packaging) | Verify packaging is free of damage and labels are legible | Pass |
| Functional Testing | Functional testing includes: button press, vertical position test,<br>laser spot alignment, auto shutoff test, trunnion articulation. | Pass |
| IEC 60825-1 | Laser safety testing: Wavelength and conformance to IEC 60825-<br>1 standard | Pass |
| IEC 60529 (IP22) | Fluid ingress testing: Verify conformance to the IP2X<br>requirements of IEC 60529 Edition 2.2. | Pass |
| ASTM D4332-14-<br>Package<br>environmental<br>conditioning | 2014 Version - Standard Practice for Conditioning<br>Containers, Packages, or Packaging Components for<br>Testing<br>Conduct the conditioning test per ASTM D4332-14 on the<br>packages:<br>Extreme Cold: -30°C, uncontrolled RH for 24 hours<br>Tropical: +40°C, 90% RH for 24 hours<br>Desert: +60°C, 15% RH for 24 hours<br>Chamber Class: 3<br>Notes: Westpak calls out tolerances of ±2°C and ±5% RH. | Pass |
| ASTM D4169-16,<br>Distribution cycle<br>13 | 2016 Version- Air (intercity) and motor freight (local),<br>single package up to 150 lb. (61.8 kg) | Pass |
| Schedule A Initial<br>manual handling -<br>ASTM D5276-98<br>Box | Impact Quantity: Six (6)<br>Top, Two adjacent bottom edges, two diagonally opposite<br>bottom corners, bottom<br>Drop height: dependent on weight per procedure | Pass |
| Schedule C Vehicle<br>stacking- ASTM<br>D642-15 | The package systems will be compressed per the<br>computed load value in section 11.4. | Pass |
| Schedule F Loose<br>load vibration -<br>ASTM D999-08 | Method A2 - Repetitive Shock Test (Rotary Motion)<br>Duration: 60 min (AL I)<br>Dwell time distributed 50% along predetermined bottom<br>orientation facing down and remaining 50% evenly along<br>side and end orientations. | Pass |
| Test | Test Method Summary | Results |
| Schedule I Low<br>Pressure Test-<br>ASTM D6653-13 | 14,000 Feet (446 torr)<br>60 minute dwell, 1,000 to 2,000 feet per minute ramps | Pass |
| Schedule E Vehicle<br>Vibration – ASTM<br>D4729-17 Truck<br>and Air | Truck Spectrum Low<br>Frequency Range: 1-200 Hz<br>Overall Intensity: 0.40 Grms<br>Truck Spectrum Medium<br>Frequency Range: 1-200 Hz<br>Overall Intensity: 0.54 Grms<br>Truck Spectrum High<br>Frequency Range: 1-200 Hz<br>Overall Intensity: 0.70 Grms<br>Air Spectrum<br>Frequency Range: 2-300 Hz<br>Overall Intensity: 1.49 Grms (Assurance Level I)<br>Durations as outlined per standard. | Pass |
| Schedule J<br>Concentrated<br>Impact - ASTM<br>D6344-04 | Cylindrical mass<br>Drop Height: 32 inches (5.4 J)<br>Box: All 6 faces<br>Plastic Wrapped Unitized Loads: All wrapped faces<br>(excluding base)<br>Note: This test will be omitted if the shipper is a doublewall<br>shipper OR a singlewall shipper and has a Mullen Burst<br>Strength equal to or greater than 275 psi/44 ECT. | Pass |
| Schedule A Final<br>Manual handling –<br>ASTM D5276-99<br>Box | Box<br>Impact Quantity: Six (6)<br>Vertical edge, Two adjacent side faces, one top corner,<br>one adjacent top edge, *bottom<br>Drop height: dependent on weight per procedure<br>*The last impact should be made at twice the specified<br>height. | Pass |
| ASTM F19880-16<br>Accelerated Aging<br>Testing | 2016 Version - Standard Guide for Accelerated Aging of<br>Sterile Barrier Systems for Medical Devices<br>Conduct the accelerated aging test per ASTM F1980-16<br>on the packages:<br>6-Month Accelerated Aging (Desired Real Time)<br>Accelerated Aging Duration: 10 (days)<br>Accelerated Aging Temperature (TAA): +65°C<br>Accelerated Aging Relative Humidity (RH): Uncontrolled<br>Real Time Room Temperature (TRT): +23°C<br>Q10: 2<br>Volume: 3 cu ft | Pass |
{12}------------------------------------------------
{13}------------------------------------------------
## Conclusion
The IRRAflow® CNS System is substantially equivalent to the predicate device (K192289). The IRRAflow® CNS System has the same indications for use, technological characteristics, and principles of operation as the predicate device. The addition of the laser leveler accessory does not change or impact the operational parameters of the IRRAflow CNS System, but it does add an additional option to assess the position of the drainage system with the landmark target on the patient's head during initial setup. The positioning of the drainage system with the external auditory meatus of the patient's head is facilitated by use of the laser leveler.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.