actiTENS is intended to be used as: · Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications: - Symptomatic relief and management of chronic, intractable pain - Adjunctive treatment for post-surgical and post-trauma acute pain - Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities - Relief of pain associated with arthritis Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode. · Electrical Muscle Stimulation (EMS), used for the following indications: - Temporary relaxation of muscle spasms - Prevent or retard disuse atrophy - Increase of local blood flow in the treatment area - Re-educate muscles - Maintain or increase the range of motion - Prevention of venous thrombosis of the calf muscles immediately after surgery Program P9 corresponds to EMS mode.
Device Story
actiTENS is a compact, wearable transcutaneous electrical nerve stimulator (TENS) and electrical muscle stimulator (EMS) for chronic pain management and muscle therapy. Device consists of an electrical impulse generator, rechargeable lithium-ion battery, and disposable electrodes. Operated via a downloadable iPhone mobile app for program selection and manual intensity adjustment (0.5mA increments). Used by patients at home or by medical professionals in clinical settings. Device delivers biphasic, asymmetric electrical stimulation waveforms to targeted nerves/muscles. Integrated impedance monitoring automatically stops stimulation if electrodes peel off or impedance is out of range to prevent discomfort. Benefits include non-invasive pain relief, muscle rehabilitation, and prevention of venous thrombosis. Prescription-only device.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via benchtop performance testing, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 62366-1), and software verification/validation (IEC 62304).
Technological Characteristics
Rechargeable lithium-ion battery-powered electrical stimulator. Biphasic, asymmetric waveforms. Connectivity via mobile app (iPhone). Impedance monitoring for safety. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10, IEC 62304, IEC 62366-1, ASTM D4169-16, UL 1642, IEC 62133-2. Non-sterile.
Indications for Use
Indicated for patients 22 years and older for symptomatic relief of chronic intractable pain, post-surgical/post-trauma acute pain, pain from muscle strain, arthritis, muscle spasms, disuse atrophy, local blood flow increase, muscle re-education, range of motion maintenance, and prevention of post-surgical venous thrombosis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 22, 2020
Sublimed % Robert Packard President Medical Device Academy, Inc. 345 Lincoln Hill Road Shrewsbury, Vermont 05738
Re: K202159
Trade/Device Name: actiTENS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ, NGX, NYN, IPF Dated: July 30, 2020 Received: August 3, 2020
Dear Robert Packard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202159
Device Name actiTENS
#### Indications for Use (Describe)
actiTENS is intended to be used as:
- · Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications:
- Symptomatic relief and management of chronic, intractable pain
- Adjunctive treatment for post-surgical and post-trauma acute pain
- Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities
- Relief of pain associated with arthritis
Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode.
· Electrical Muscle Stimulation (EMS), used for the following indications:
- Temporary relaxation of muscle spasms
- Prevent or retard disuse atrophy
- Increase of local blood flow in the treatment area
- Re-educate muscles
- Maintain or increase the range of motion
- Prevention of venous thrombosis of the calf muscles immediately after surgery
Program P9 corresponds to EMS mode.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> </span> </div> |
|----------------------------------------------|-----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
- -SUBMITTER SUBLIMED 137 rue de Mayoussard Moirans, F-38430 France Tel: +33(0)4-76-93-37-15
Contact Person: Corinne Bulteau Date Prepared: December 22, 2020
| II. DEVICE | |
|------------------------------|------------------------------------------------------------|
| Trade/Device Name: | actiTENS |
| Classification Name: | Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Regulation: | 21 CFR § 882.5890 |
| Regulatory Class: | Class II |
| Product Classification Code: | GZJ, NGX, NYN, IPF |
| III. PREDICATE DEVICE | |
|-------------------------|---------------------|
| Predicate Manufacturer: | JKH USA, LLC |
| Predicate Trade Name: | JKH Stimulator Plus |
| Predicate 510(k): | K182203 |
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
actiTENS is a transcutaneous electrical nerve stimulation (TENS) medical device designed to manage chronic pain in people ages 22 and older.
The compact design and flexible shape of the actiTENS electrical impulse generator allows to fix it for daily use directly to the body of the patient.
The impulse generator is delivered with a separate cradle which allows safely recharging its nonremovable battery.
The actiTENS is compatible with a selection of disposable electrodes and compatible connector cables transmit the electrical stimulation impulses to the targeted nerves according to the selected therapy. The electrical stimulation waveforms are biphasic and asymmetric.
The actiTENS is controlled via a downloadable mobile app for iPhone which allows the comfortable and simple selection of the adequate stimulation program.
The actiTENS is intended to be used by the patient at home and also for therapeutic application by medical professionals.
{4}------------------------------------------------
It is available for prescription only.
#### V. INDICATIONS FOR USE
#### actiTENS is intended to be used as:
· Transcutaneous Electrical Nerve Stimulator (TENS), used for the following indications:
- Symptomatic relief and management of chronic, intractable pain
- Adjunctive treatment for post-surgical and post-trauma acute pain
- Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities
- Relief of pain associated with arthritis
Programs P1, P2, P3, P4, P5, P6, P7, P8, P10, P11, P12 and P13 correspond to TENS mode.
· Electrical Muscle Stimulation (EMS), used for the following indications:
- Temporary relaxation of muscle spasms
- Prevent or retard disuse atrophy
- Increase of local blood flow in the treatment area
- Re-educate muscles
- Maintain or increase the range of motion
- Prevention of venous thrombosis of the calf muscles immediately after surgery
Program P9 corresponds to EMS mode.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE
The table below provides a comparison of indications for use and technological characteristics between the subject device and the predicate device in order to demonstrate substantial equivalence:
{5}------------------------------------------------
# Comparison with Predicate Device:
Table 1: Comparison of actiTENS with JKH Stimulator Plus (K182203).
| Feature | actiTENS | JKH Stimulator Plus -<br>K182203 | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | actiTENS is intended to be<br>used as:<br>• Transcutaneous Electrical<br>Nerve Stimulator (TENS),<br>used for the following<br>indications:<br>- Symptomatic relief and<br>management of chronic,<br>intractable pain<br>- Adjunctive treatment for<br>post-surgical and post-trauma<br>acute pain<br>- Temporary relief of pain<br>associated with sore and<br>aching muscles due to strain<br>from exercise or normal<br>household and work activities<br>- Relief of pain associated<br>with arthritis<br>Programs P1, P2, P3, P4, P5,<br>P6, P7, P8, P10, P11, P12 and<br>P13 correspond to TENS<br>mode.<br>• Electrical Muscle<br>Stimulation (EMS), used for<br>the following indications: | Over-the-Counter Use:<br>TENS (Modes 1, 2, 4, 6,<br>8):<br>PL-029K12 and PL-<br>029K13 are used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back, arm,<br>and leg, due to strain from<br>exercise or normal household<br>work activities. PL-0219K12<br>and PL-<br>029K13 are also intended<br>for symptomatic relief an<br>management of chronic,<br>intractable pain and relief of<br>pain associated with<br>arthritis.<br>The device of PL-029K12<br>may be used during sleep.<br>The device of PL-029K12 is<br>labeled for use only with its<br>own compatible electrodes.<br>PMS (also called EMS,<br>Modes 1, 3, 7):<br>PL-029K12 and PL- | The subject device is not<br>intended for OTC use, and<br>the prescription use of the<br>subject device is identical to<br>the indications of the<br>predicate model PL-<br>029K12. |
{6}------------------------------------------------
| - Temporary relaxation of<br>muscle spasms | 029K13 are also intended to<br>temporarily increase local<br>blood circulation in the<br>healthy muscles of lower<br>extremities. |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| - Prevent or retard disuse<br>atrophy | Heating:<br>The device of PL-029K13<br>is intended for temporary<br>relief of minor aches and<br>pains. |
| - Increase of local blood flow<br>in the treatment area | Prescription Use:<br>PL-029K12 and PL-<br>029K13 are intended for<br>the following use: |
| - Re-educate muscles | -Symptomatic relief and<br>management of chronic,<br>interactable pain |
| - Maintain or increase the<br>range of motion | -Adjunctive treatment for<br>post-surgical and post-<br>trauma acute pain |
| - Prevention of venous<br>thrombosis of the calf<br>muscles immediately after<br>surgery | -Relief of pain associated<br>with arthritis |
| Program P9 corresponds to<br>EMS mode. | -Temporary relaxation of<br>muscle spasm |
| | -Prevention or retardation of<br>disuse atrophy |
| | -Muscle re-education |
| | -Maintaining or increasing<br>range of motion |
| | -Increase of local blood<br>flow in the treatment area |
| | -Prevention of post-<br>surgical venous<br>thrombosis through |
{7}------------------------------------------------
| | | immediate stimulation of<br>calf muscles | |
|-----------------------------------------|------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription or OTC | Prescription | OTC and Prescription | The subject device does not<br>include OTC indications,<br>but this eliminates the risks<br>associated with OTC use. |
| Power Source(s) | Rechargeable | Rechargeable or non-<br>rechargeable battery | The subject device is<br>limited to use with the<br>rechargeable lithium-ion<br>battery, but this does not<br>introduce any new risks. |
| Compliance with Voluntary<br>Standards? | Yes | Yes | Same |
| Compliance With 21 CFR<br>898? | Yes | Yes | Same |
| Functions and design | Electrical stimulation | Electrical stimulation and<br>heat | The subject device does not<br>include the IRT product<br>classification, and this<br>eliminates the risks<br>associated with that<br>function. |
| Maximum skin temperature | N/A | 43°C | The subject device does not<br>include the IRT product<br>classification and does not<br>provide heat |
{8}------------------------------------------------
| | | actiTENS | JKH Stimulator Plus – K182203 | Comparison |
|-------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Maximum output<br>voltage (Volts+/- 20%) at<br>500Ω | P1 | | 30.2 PL- 029K12: 57.6 | Maximum output voltage of the subject<br>device is in the range or below this of the<br>predicate device. These technological<br>differences between the subject and<br>predicate device do not raise new safety<br>and effectiveness questions |
| | P2 | | 30.2 PL- 029K13: 46.0 | |
| | P3 | | 30 | |
| | P4 | | 30.2 | |
| | P5 | | 30.2 | |
| | P6 | | 30 | |
| | P7 | | 30.2 | |
| | P8 | | 30.2 | |
| | P9 | | 29.5 | |
| | P10 | | 29.1 | |
| | P11 | | 29.3 | |
| | P12 | | 30.2 | |
| | P13 | | 30.2 | |
| Maximum output<br>voltage (Volts+/- 20%) at<br>2KΩ | P1 | | 57.9 PL- 029K12: 96.0 | Maximum output voltage of the subject<br>device is in the range or below this of the<br>predicate device. These technological<br>differences between the subject and<br>predicate device do not raise new safety<br>and effectiveness questions |
| | P2 | | 57.9 PL- 029K13: 90.4 | |
| | P3 | | 57.9 | |
| | P4 | | 57.9 | |
| | P5 | | 57.9 | |
| | P6 | | 57.9 | |
| | P7 | | 57.9 | |
| | P8 | | 57.9 | |
| | P9 | | 57.9 | |
| | P10 | | 57.9 | |
| | P11 | | 57.9 | |
| | P12 | | 57.9 | |
| | P13 | | 57.9 | |
| Maximum output<br>voltage (Volts+/- 20%) at<br>$10KΩ$ | P1<br>P2<br>P3<br>P4<br>P5<br>P6<br>P7<br>P8<br>P9<br>P10<br>P11<br>P12<br>P13 | N/A | PL-029K12: 134<br>PL-029K13: 124 | The EIG monitors the impedance of each<br>channel and stop any running program (or<br>refuse to launch one) if the measured<br>impedance is out of a specified range. This<br>feature ensures a proper installation of the<br>EIG, the cables and electrodes. If an<br>electrode peels off, stopping the stimulation<br>program quickly enough prevents the<br>patient from feeling an electric discharge as<br>the attached surface diminishes. The 10<br>$kOhm$ is at the upper range and the EIG<br>refused to launch a stimulation program,<br>thus the measurement could not be<br>performed |
| Maximum output<br>current(mA+/- 20%) at<br>$500Ω$ | P1<br>P2<br>P3<br>P4<br>P5<br>P6<br>P7<br>P8<br>P9<br>P10<br>P11<br>P12<br>P13 | 60.4<br>60.4<br>60<br>60.4<br>60.4<br>60<br>60.4<br>60.4<br>59<br>58.2<br>58.6<br>60.4<br>60.4 | 60.4 PL-029K12: 115.2<br>60.4 PL-029K13: 92.0 | Maximum output current of the subject<br>device is in the range or below this of the<br>predicate device. These technological<br>differences between the subject and<br>predicate device do not raise new safety<br>and effectiveness questions |
| Maximum output<br>current(mA+/- 20%) at<br>2KΩ | P1 | 28.95 | PL-029K12: 48.0 | Maximum output current of the subject<br>device is in the range or below this of the<br>predicate device. These technological<br>differences between the subject and<br>predicate device do not raise new safety<br>and effectiveness questions |
| | P2 | 28.95 | PL-029K13: 45.2 | |
| | P3 | 28.95 | | |
| | P4 | 28.95 | | |
| | P5 | 28.95 | | |
| | P6 | 28.95 | | |
| | P7 | 28.95 | | |
| | P8 | 28.95 | | |
| | P9 | 28.95 | | |
| | P10 | 28.95 | | |
| | P11 | 28.95 | | |
| | P12 | 28.95 | | |
| | P13 | 28.95 | | |
| Maximum output<br>current(mA+/- 20%) at<br>10kΩ | P1 | N/A | PL-029K12: 13.4 | The EIG monitors the impedanceof each<br>channel and stop any running program (or<br>refuse to launch one) if the measured<br>impedance is out of a specified range. This<br>feature ensures a proper installation of the<br>EIG, the cables and electrodes. If an<br>electrode peels off, stopping the stimulation<br>program quickly enough prevents the<br>patient from feeling an electric discharge as<br>the attached surface diminishes. The 10<br>kOhm is at the upper range and the EIG<br>refused to launch a stimulation program,<br>thus the measurement could not be<br>performed |
| | P2 | | PL- 029K13: 12.4 | |
| | P3 | | | |
| | P4 | | | |
| | P5 | | | |
| | P6 | | | |
| | P7 | | | |
| | P8 | | | |
| | P9 | | | |
| | P10 | | | |
| | P11 | | | |
| | P12 | | | |
| | P13 | | | |
| | P1 | 197.5 | | |
| Pulse Width (µSec) | P2 | 147 PL-029K12: 100 | | These technological differences between<br>the subject and predicate device do not<br>raise new safety and effectiveness<br>questions |
| | P3 | 247.5 PL-029K13: 92 | | |
| | P4 | 197.5 | | |
| | P5 | 147.5 | | |
| | P6 | 200 or 147 | | |
| | P7 | 147.5 | | |
| | P8 | modulated 100-200 | | |
| | P9 | 247.5 | | |
| | P10 | 147.5 | | |
| | P11 | 147.5 | | |
| | P12 | 57.5 | | |
| | P13 | 177.5…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.