K102014 · Koalaty Products, Inc. · GZJ · Aug 31, 2010 · Neurology
Device Facts
Record ID
K102014
Device Name
TENS STIMULATOR, MODELS 2800 AND 3000
Applicant
Koalaty Products, Inc.
Product Code
GZJ · Neurology
Decision Date
Aug 31, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain
Device Story
TENS 2800 and TENS 3000 are portable, battery-operated transcutaneous electrical nerve stimulators for pain relief. Device delivers electrical current to nerves via skin-surface electrodes and lead wires. User adjusts intensity and treatment mode (normal, burst, or modulation) using manual rotate buttons and switches. Device housed in molded plastic case with battery compartment. Intended for use by patients to manage pain; output parameters (pulse width, frequency) controlled by internal circuitry. Healthcare providers use device to provide non-invasive pain management, potentially reducing reliance on pharmacological interventions. Device provides symptomatic relief; does not treat underlying pathology.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing and adherence to voluntary safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10).
Indicated for symptomatic relief of chronic intractable pain, acute post traumatic pain, or acute post surgical pain in patients requiring transcutaneous electrical nerve stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
\$\qquad 510(k):\$
KOALATY PRODUCTS, INC.
A Current Solutions Company
File No: WMI-04-TENS -FDA-05 Version: 1.1
AUG 3 1 2010
# 510(k) SUMMARY
TENS Electro-Stimulator, K(102014 )
4/15/2010 Date of Submission:
Koalaty Products.,INC 510(k) Submitter's Name:
> 3814 Woodbury Dr Austin TX 78704 Address:
510(k) Correspondent contact: Kang Jian Ping microcong@gmail.com E-mail: Telephone: +86(755) 27652471 +86(755) 27652674 Fax:
Section 05-Page 1 of 4
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for KoalaTY Products, INC. The logo features a stylized image of a koala's head on the left. To the right of the koala image, the text "KoalaTY Products, INC." is written in a simple, sans-serif font. Below the company name, the text "A Current Solutions Company" is written in a smaller font.
File No: WMI-04-TENS -FDA-05 Version: 1.0
#### 1. Proposed Device:
Trade Name: TENS 2800 Stimulator TENS 3000 Stimulator Classification Name: Stimulator, Nerve, Transcutaneous. For pain relief Regulation Number: 882.5890 Product Code: GZJ Device Class: ||
#### 2. Predicate Device:
Predicate Device: LT3001 TENS Stimulator 510(k) Number: K100117 Manufacturer: Shenzhen Dongdixin Technology Co., Ltd.
#### Description of Proposed Device: 3.
TENS Series Electro-Stimulator, which includes models TENS 2800 and TENS 3000, are Transcutaneous Electrical Nerve Stimulator for pain relief. The stimulator sends gentle electrical current to underlying nerves through the cable and electrode placed on the skin. The parameters of units are controlled by the rotate buttons. Its intensity level is adjustable according to the needs of patients.
These TENS electro-stimulator have the same housing in a molded portable plastic case, an accessible, switch, and accessible battery storage compartment. The case shape is rectangular, The process to set the parameter and attach lead wires to the two models is also the same except the Housing printing artwork, Mode No.
TENS 3000 stimulator has three treatment mode: normal mode, burst mode and modulation mode. The treatment mode can by selected by switch. TENS 2800 only one treatment mode: normal mode. The difference on the two device can be identified by panel, Mode No.
#### Proposed Device Intended Use Statement: 4.
Device Name:
TENS 2800 Stimulator, TENS 3000 Stimulator
### Indications for Use:
- Symptomatic relief of chronic intractable pain, 1)
- Post traumatic pain 2)
- Post surgical pain 3)
Section 05-Page 2 of 4
{2}------------------------------------------------
KOALATY PRODUCTS, INC.
A Current Solutions Company
File No: WMI-04-TENS -FDA-05 Version: 1.0
#### 5. Biocompatibility Certification:
Electrodes to be provided with this device are from the manufacturer Top-Rank Health Care Equipment Co., Ltd (K070612) who submitted in 2007.
The shell of device is used ABS material; this material has passed Biocompatibility testing in Jiangsu TUV Product Service Ltd. Shanghai Branch. Identification No: 080960.
#### 6. Technological Characteristics and Substantial Equivalence
Both the TENS Series Electro-Stimulator and the Predicate device Stimulator have the same intended use and fundamental technology. A side-by-side comparison of the TENS Series Electro-Stimulator and the cited predicate devices is included in the 510(k) submission. The TENS Series Electro-Stimulator is substantially equivalent to the technological features as the predicate devices.
| | | New device | Predicate device |
|----|---------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------|
| 1 | 510K# | K | K100117 |
| 2 | Device Name | TENS 2800 Stimulator<br>TENS 3000 Stimulator | LT3001 TENS Stimulator |
| 3 | Manufacturer | Koalaty Products., INC | Shenzhen Dongdixin<br>Technology Co., Ltd. |
| 4 | Power Source | 9V Battery | 9V Battery |
| | -Method of Line current isolation | Battery Supply N/A | Battery Supply N/A |
| | - Patient Leakage Current<br>-Normal condition<br>-Single fault condition | 1.2uA<br>1.3uA | 1.1uA<br>1.3uA |
| 5 | Number of Output Modes | 3 | 6 |
| 6 | Number of Output Channels | 2 | 2 |
| | Method of channel isolation | By enclosure | By enclosure |
| 7 | Regulated Current or regulated<br>Voltage? | Voltage control | Voltage control |
| 8 | Software/Firmware/<br>Microprocessor Control? | Yes | Yes |
| 9 | Automatic Overload Trip? | No | No |
| | Automatic Over Current Trip? | No | No |
| 10 | Automatic No Load Trip? | No | No |
| 11 | Automatic Shut off? | No | No |
| 12 | Patient Override Control? | No | No |
| 13 | Indicator Display | | |
| | -On/Off Status? | Yes | Yes |
| | -Voltage/Current Level? | Yes | Yes |
| | -Low Battery? | Yes | Yes |
Basic technological characteristics, new device vs. Predicate device
Section 05-Page 3 of 4
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for KoalaTY Products, INC. The logo features a stylized image of a koala's head on the left. To the right of the koala image, the text "KoalaTY Products, INC." is written in a bold, sans-serif font. Below the company name, the text "A Current Solutions Company" is written in a smaller, lighter font.
| 14 | Timer Range (minutes) | 15min, 30min, continue | 30min, 60min, continue |
|----|-------------------------------------|---------------------------------------------|--------------------------------------------|
| 15 | Waveform | Biphasic or Monophasic<br>Rectangular pulse | Biphasic Rectangular pulse |
| 16 | Pulse Width Range | 30-260us | 50-300us |
| 17 | Frequency | 2-150Hz | 2-120Hz |
| 18 | Compliance with Voluntary Standards | IEC60601-1, IEC60601-1-2,<br>IEC60601-2-10 | IEC60601-1, IEC60601-1-2,<br>IEC60601-2-10 |
| 19 | Compliance with 21 CFR 898? | Yes | Yes |
| 20 | Weight (grams.) | 115 grams(battery included) | 128 grams (battery included) |
| 21 | Dimensions (mm.) H x W x T | 95x65x23.5 | 102x64x26 |
| 22 | Housing Materials & Construction | Enclosure: ABS,94, V-1,80°C,UL Approved | Enclosure: ABS,94, V-1,80°C,UL<br>Approved |
### Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:
TENS Series Electro-Stimulator did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design according with the company's specified design requirements, and to assure conformance with the following voluntary design standards:
- IEC 60601-1 "Medical electrical equipment Part 1: General requirements for safety". >
- IEC 60601-1-2 "Medical electrical equipment Part 1-2: General requirements for safety -A Collateral Standard"
- IEC 60601-2-10 "Medical electrical equipment Part 2: Particular requirements for the safety of > nerve and muscle stimulators"
#### Conclusions: 8.
The TENS Series Stimulator, which includes models TENS 2800 and TENS 3000, has the same intended use and technological characteristics as the predicate device. Moreover, bench testing, safety report and Risk Analysis Report documentation supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, the TENS Series Electro-Stimulator is substantially equivalent to the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is smaller than the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Koalaty Products, Inc. c/o Mr. Jeffrey D. Rongero Senior Project Engineer, UL Health Sciences Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle, NC 27709
AUG 3 1 2010
Re: K102014
Trade/Device Name: TENS 2800, TENS 3000 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Codes: GZJ Dated: August 11, 2010 Received: August 16, 2010
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm].15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director
Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KID2014
File No: WMI-04-TENS-FDA-04 Version: 1.1
## Indications for Use
AUG 31 2010
510(k) Number ( < 102014 ):
Device Name:
TENS 2800 Stimulator TENS 3000 Stimulator
## Indications for Use:
Symptomatic relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Raffi
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K102014
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.