← Product Code [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP) · K223164

# Aquilo Sports CCT1500 System (K223164)

_Aquilo Sports, LLC · IRP · Nov 4, 2022 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K223164

## Device Facts

- **Applicant:** Aquilo Sports, LLC
- **Product Code:** [IRP](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP.md)
- **Decision Date:** Nov 4, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5650
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Aquilo Sports CCT1500 System is indicated for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The CCT1500 simulates kneading and stroking of tissues by using an inflatable garment.

## Device Story

The Aquilo Sports CCT1500 System is a powered inflatable tube massager for OTC use in clinics, hospitals, athlete training, and home environments. It consists of an air compressor unit with a microprocessor-based control system, plastic tubing, and an inflatable leg garment. The device operates by sequentially inflating and deflating six segments of the garment to simulate kneading and stroking of tissues. Users operate the device via a front panel display to adjust treatment time (20-150 minutes) and pressure (20-150 mmHg). The system provides temporary relief of minor muscle aches and increases local circulation. It is powered by an external IEC 60601-1 compliant power supply or an integral rechargeable battery. Safety features include a user-accessible button to stop or pause therapy. The device does not transfer energy to the user or maintain electrical contact.

## Clinical Evidence

No clinical testing was performed. Evidence is based on bench testing, including control, indicator, and performance testing, and compliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-1-11 standards.

## Technological Characteristics

Powered inflatable tube massager; Class II, Type BF applied part. Materials: Nylon with Polyurethane laminate (garment), molded ABS (enclosure). Sensing/Actuation: Air compressor and valve system. Pressure range: 20-150 mmHg. Connectivity: Standalone. Power: 100-240V AC or internal rechargeable battery. Software: Microprocessor-controlled. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11.

## Regulatory Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

## Predicate Devices

- Rapid Reboot ([K182668](/device/K182668.md))

## Reference Devices

- NormaTec Pulse and NormaTec Pulse Pro ([K160608](/device/K160608.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 4, 2022

Aquilo Sports LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K223164

Trade/Device Name: Aquilo Sports CCT1500 System Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: October 06, 2022 Received: October 07, 2022

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223164

Device Name Aquilo Sports CCT1500 System

#### Indications for Use (Describe)

The Aquilo Sports CCT1500 System is indicated for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The CCT1500 simulates kneading and stroking of tissues by using an inflatable garment.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

|                                        | 510(k) Summary<br>Page 1 of 10                                                           |
|----------------------------------------|------------------------------------------------------------------------------------------|
|                                        | K223164                                                                                  |
| Date Prepared:                         | November 2, 2022                                                                         |
| Applicant                              | Aquilo Sports<br>1902 Campus Pl Ste 12<br>Louisville, KY 40299<br>Tel - 502 290 8994     |
| Official Contact:                      | John Paul Spence VP Sales and Marketing                                                  |
| Proprietary or Trade Name:             | Aquilo Sports CCT1500 System                                                             |
| Common/Usual Name:                     | Powered Inflatable Tube Massager                                                         |
| Classification Name:                   | IRP - Massager Powered Inflatable Tube (CFR 890.5650)                                    |
| Predicate Device:<br>Reference Device: | IRP Rapid Reboot 510(k) K182668<br>NormaTec Pulse and NormaTec Pulse Pro 510(k) K160608: |

#### Device Description:

This submission is for the Aquilo Sports CCT1500 System is a powered inflatable tube massager. It is being submitted as an over the counter device.

The device is indicated for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in people who are in good health. The CCT1500 simulates kneading and stroking of tissues by using an inflatable garment.

The device is a Class II, type BF applied part that receives power a separately approved external IEC 60601-1 compliant power supply or integral battery pack.

The CCT1500 consists of an air compressor unit with a control system, an inflatable "boot" for the leg, plastic air tubing with a proprietary connector for connecting the device to the boot.

The user interface is a front panel display and buttons.

The CCT1500 contains an air compressor with a system control that that allows the user to adjust the amount of air coming from the air compressor and going to the individual segments of the inflatable boot.

There is no electrical contact with the user and the device does not transfer or detect energy to or from the user.

{4}------------------------------------------------

## 510(k) Summary Page 2 of 10

The user interface of the CCT1500 provides for starting and stopping the massage treatment, allows for adjusting time and intensity (pressure) of the treatment. The device also provides a proprietary keyed connector to the tubing which connects to the boot.

Pressure level is selectable between 20 and 150 mmHg

#### Indications for Use:

The Aquilo Sports CCT1500 System is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The CCT1500 simulates kneading and stroking of tissues by using an inflatable garment.

# Patient Population:

Adults

## Environments of Use:

Clinics, hospital, athlete training, and home environments

{5}------------------------------------------------

## 510(k) Summary Page 3 of 10

# Table of the Similarities and Differences of Predicate vs. Proposed Device Over The Counter Device

| Model Name<br>510(k) Number                | New Device<br>Aquilo Sports CCT1500 System                                                                                                                                                                                                                                                                                | Predicate Device<br>Rapid Reboot<br>510(k) K182668                                                                                                                                                                                                                                                                                                                      | Comment                                                                                                                                         |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification                             | Class II Device, IRP (21<br>CFR890.5650)                                                                                                                                                                                                                                                                                  | Class II Device, IRP (21<br>CFR890.5650)                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                                       |
| Indications for use                        | The Aquilo Sports CCT1500<br>System is indicated for the<br>temporary relief of minor muscle<br>aches and pains<br>and for temporary increase in<br>circulation to the treated areas in<br>people who are in good health.<br>The CCT1500 simulates<br>kneading and stroking of tissues<br>by using an inflatable garment. | The Rapid Reboot Compression<br>Therapy System is indicated for the<br>temporary relief of minor muscle<br>aches and pains and for temporary<br>increase in circulation to the treated<br>areas in people who are in good<br>health. The Rapid Reboot<br>Compression Therapy System<br>simulates kneading and stroking of<br>tissues by using an inflatable<br>garment. | Identical                                                                                                                                       |
| OTC or Prescription                        | OTC                                                                                                                                                                                                                                                                                                                       | OTC                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                       |
| Environment of Use                         | Clinics, hospital, athlete<br>training, and home<br>environments                                                                                                                                                                                                                                                          | Clinics, hospital, athlete<br>training, and home<br>environments                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                       |
| Compliance with Voluntary<br>standards     | ES 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11                                                                                                                                                                                                                                                                              | IEC 60601-1, IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                              | The CCT1500 meets<br>home use standard                                                                                                          |
| Mode of Operation                          | Sequential                                                                                                                                                                                                                                                                                                                | Sequential/Peristaltic                                                                                                                                                                                                                                                                                                                                                  | No difference.<br>Peristaltic and<br>Sequential are<br>synonymous                                                                               |
|                                            |                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                 |
| Cycle Time                                 | Between 8 sec. and 95 sec. (full cycle - including hold time) depending on device settings*.                                                                                                                                                                                                                              | 1 min 20 sec                                                                                                                                                                                                                                                                                                                                                            | Similar                                                                                                                                         |
| Hold Time                                  | Between 4 sec. and 95 sec. depending on device settings*.                                                                                                                                                                                                                                                                 | Not available                                                                                                                                                                                                                                                                                                                                                           | Similar                                                                                                                                         |
| Power                                      | 100-240 V, 50-60 Hz, 2.5A Max<br>Internal rechargeable battery                                                                                                                                                                                                                                                            | 110 V, 60Hz                                                                                                                                                                                                                                                                                                                                                             | The addition of the battery does not impact the efficacy of the device. The device and battery have been tested for safety to current standards |
| Device<br>Pressure range                   | 20-150mmHg                                                                                                                                                                                                                                                                                                                | 0 - 200 mmHg                                                                                                                                                                                                                                                                                                                                                            | The CCT1500 has a more restricted range                                                                                                         |
| Treatment Time                             | Treatment time programmable from 20-150 minutes in 5-minute intervals. Time default is 30 minutes                                                                                                                                                                                                                         | User determines therapy time. Choose from 10, 20, or 30 minutes session time, with option to add additional 10 minutes to any therapy time.                                                                                                                                                                                                                             | Maximum selectable time is 200 minutes which is longer that the predicate but shorter than the reference device as below                        |
| Garments contact surface<br>Material       | Nylon with a Polyurethane laminate                                                                                                                                                                                                                                                                                        | Nylon with a Polyurethane laminate                                                                                                                                                                                                                                                                                                                                      | Same materials                                                                                                                                  |
| Leg Attachment                             | Yes                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                     | Identical in size and construction                                                                                                              |
| Number of Inflatable appliance<br>Segments | 6                                                                                                                                                                                                                                                                                                                         | 4                                                                                                                                                                                                                                                                                                                                                                       | Similar.                                                                                                                                        |
| Weight                                     | 18.1 pounds                                                                                                                                                                                                                                                                                                               | 5.8 pounds                                                                                                                                                                                                                                                                                                                                                              | Similar. CCT1500 weighs more in part because of battery                                                                                         |
| Dimensions (W x H x D)                     | 16" x 8.25" x 10.25"                                                                                                                                                                                                                                                                                                      | 10" x 6.5" x 5"                                                                                                                                                                                                                                                                                                                                                         | Similar                                                                                                                                         |
| Housing Materials and<br>Constructions     | Molded ABS enclosure                                                                                                                                                                                                                                                                                                      | Molded ABS enclosure                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                       |
|                                            | Patient contact                                                                                                                                                                                                                                                                                                           | Non-conductive appliances                                                                                                                                                                                                                                                                                                                                               | Non-conductive appliances                                                                                                                       |
| Safety Features                            | Button on display allows<br>user to stop or pause<br>therapy session at any time.                                                                                                                                                                                                                                         | Button on display allows<br>user to stop or pause<br>therapy session at any time.                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                       |
| Modes                                      | 5 modes, sequential and uniform<br>(all segments inflated deflated at<br>once).                                                                                                                                                                                                                                           | 2 Modes: “A” mode inflates and<br>deflates chambers from bottom to<br>up (distal to proximal chambers),<br>one at a time "B" mode also inflates<br>from bottom up, but maintains<br>pressure in lower chambers as it<br>works it way to the top. Then all<br>chambers release pressure at the<br>same time once all chambers have<br>sequentially inflated              | Similar                                                                                                                                         |
| Accessories                                | One size Length: 36"<br>Max Width: 27"                                                                                                                                                                                                                                                                                    | X-Short: 14" x 41"<br>Short: 14" x 43"<br>Medium: 14" x 45"<br>Long: 14" x 48"<br>X-Long: 14" x 52"                                                                                                                                                                                                                                                                     | CCT1500 only offers<br>leg attachment                                                                                                           |
| SW/Firmware/<br>Microprocessor<br>Control  | Microprocessor                                                                                                                                                                                                                                                                                                            | Microprocessor                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                       |
| Technology                                 | Compressor and valve<br>system which sequentially<br>inflates cells of appliance                                                                                                                                                                                                                                          | Compressor and valve<br>system which sequentially<br>inflates cells of appliance                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                       |

{6}------------------------------------------------

# 510(k) Summary Page 4 of 10

{7}------------------------------------------------

# 510(k) Summary

Page 5 of 10

*The cycle time and hold time of the device are contings programmed by the user. These times will fluctuate within the described range depending on the user's input for the pressure, and compression mode settings. For example, using only one compression chamber at a low pressure will result in cycle and hold times on the range, while using all six chambers at a high pressure will result

{8}------------------------------------------------

## 510(k) Summary Page 6 of 10

in cycle and hold times at the upper end of the range. Additionally, the hold time will be different chambers in the compression boot during sequential compression mode 1, compression chamber 1 at the foot will inflate and maintain the target pressure while the rest of the boot inflates, meaning the hold time in chamized (up to 95 seconds), while the hold time in chamber 6 will be minimized (4 seconds).

# Reference Device: NormaTec Pulse and NormaTec Pulse Pro 510(k) K160608:

The NormaTec Pulse and NormaTec Pulse Pro are used as a reference device for "Treatment Time". The subject device has a treatment time of 20-150 minute intervals. Time default is 30 minutes while the reference device is user controlled 10 minutes to 175 minutes or continuous. As the reference device includes a continuous treatment that the subject device range is within the reference.

{9}------------------------------------------------

#### 510(k) Summary Page 7 of 10

#### Substantial Equivalence Discussion

In the above tables we have compared the CCT1500 to the predicate for equivalence of: Indications -

The Aquilo Sports CCT1500 System is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The CCT1500 simulates kneading and stroking of tissues by using an inflatable garment.

#### These indications are identical to the predicate device.

Design, Technology and Principle of Operation - The CCT15000 is equivalent in design and features when compared to the predicate and has identical technology

Performance and Specifications – The CCT1500 has equivalent specifications of performance when compared to the predicate device. This includes cycle time and hold time.

Compliance with standards - The CCT1500 is compliance with IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2 which is identical to the predicate devices.

#### Patient Population -

The CCT1500 and predicate are indicated for adults

#### Environment of Use -

The CCT1500 environments are identical to the predicate

#### Differences –

There are no differences between the proposed device and the predicate device that raise any new safety and efficacy concerns.

#### Performance Testing

#### Bench:

The device has been tested to ensure that it all requirements have been met, this includes:

- . Testing of all controls
- Testing of all indicators .
- Testing of performance .

The device has also been tested to the requirements of the following standards:

- AAMI / ANSI ES60601-1:2005 + A1: 2012 Medical electrical equipment part 1: general requirements for basic safety and essential performance
- IEC 60601-1-2: 2014, Collateral standard: Electromagnetic Disturbances -● Requirements and Tests

{10}------------------------------------------------

#### 510(k) Summary Page 8 of 10

- IEC 60601-1-11: 2015, Collateral standard: requirements for medical . electrical equipment and medical electrical systems used in the home healthcare environment
#### Animal:

No animal testing was performed

## Clinical:

No clinical testing was performed

#### Differences -

There are no differences between the proposed device and the predicate device that raise any new safety and efficacy concerns.

## Substantial Equivalence Rationale

The CCT1500 is viewed as similar to the predicate device because:

Indications - are identical to the predicate

Prescriptive - The CCT1500 is over the counter and same as the identified predicate.

Design, Technology and Principle of Operation - The CCT1500 has equivalent design and features when compared to the predicate and has the identical technology.

Performance and Specifications – The CCT1500 has equivalent specifications of performance when compared to the predicate.

Compliance with standards - The CCT1500 declares compliance with IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2 which is identical to the predicate.

Environment of Use - Identical to the predicate.

Features - The CCT1500 has equivalent features when compared to the predicate device

#### Conclusion

{11}------------------------------------------------

#### 510(k) Summary Page 9 of 10

The CCT1500 is similar to the predicate in: indications for use, patient population, environment of use, technology characteristics, materials, specifications / performance and compliance with international standards. Minor differences as detailed in the substantial equivalence table above do not raise questions of safety and effectiveness. Therefore, subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K223164](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94physical-medicine-therapeutic-devices/IRP/K223164)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
