Browse hierarchy: [Neurology (NE)](/submissions/NE) → [Subpart F — Neurological Therapeutic Devices](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices) → [21 CFR 882.5705](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/882.5705) → QMZ — Digital Therapy Device To Reduce Sleep Disturbance For Psychiatric Conditions

# QMZ · Digital Therapy Device To Reduce Sleep Disturbance For Psychiatric Conditions

_Neurology · 21 CFR 882.5705 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QMZ

## Overview

- **Product Code:** QMZ
- **Device Name:** Digital Therapy Device To Reduce Sleep Disturbance For Psychiatric Conditions
- **Regulation:** [21 CFR 882.5705](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/882.5705)
- **Device Class:** 2
- **Review Panel:** [Neurology](/submissions/NE)

## Identification

A digital therapy device to reduce sleep disturbance for psychiatric conditions is a prescription device that is intended to provide stimulation using a general purpose computing platform to reduce sleep disturbance in patients who experience this symptom due to psychiatric conditions such as nightmare disorder or post-traumatic stress disorder.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with the general controls of the FD&C Act, the digital therapy device to reduce sleep disturbance for psychiatric conditions is subject to the following special controls:

*(b) Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under the labeled conditions for use must evaluate the following:
(i) The ability of the device to provide therapy for patients with sleep disturbance due to psychiatric conditions, using a validated measure;
(ii) Worsening of any condition-specific symptoms using a validated measure for assessment of the particular condition; and
(iii) Increase in symptoms of disturbed sleep or sleepiness using a validated measure.
(2) Software must clearly describe all features and functions of the software implementing the digital therapy. Software verification, validation, and hazard analysis must also be provided.
(3) The labeling must include the following:
(i) Patient and physician labeling must include instructions for use, including images that demonstrate how to interact with the device;
(ii) Patient and physician labeling must list the minimum operating system and general purpose computing requirements that support the software of the device;
(iii) Patient and physician labeling must include a warning that the digital therapy device is not intended for use as a stand-alone therapeutic device;
(iv) Patient and physician labeling must include a warning that the digital therapy device does not represent a substitution for the patient's medication; and
(v) Physician labeling must include a summary of the clinical performance testing conducted with the device.

## Recent Cleared Devices (1 of 1)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [DEN200033](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QMZ/DEN200033.md) | NightWare Kit, Apple iPhone, Apple Watch, Apple iPhone Charging Cable, Apple Watch Charging Cable | Nightware, Inc. | Nov 6, 2020 | DENG |

## Top Applicants

- Nightware, Inc. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QMZ](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QMZ)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
