Browse hierarchy: [Neurology (NE)](/submissions/NE) → [Subpart F — Neurological Therapeutic Devices](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices) → [21 CFR 882.5893](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/882.5893) → QAR — Thermal Vestibular Stimulator For Headache

# QAR · Thermal Vestibular Stimulator For Headache

_Neurology · 21 CFR 882.5893 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QAR

## Overview

- **Product Code:** QAR
- **Device Name:** Thermal Vestibular Stimulator For Headache
- **Regulation:** [21 CFR 882.5893](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/882.5893)
- **Device Class:** 2
- **Review Panel:** [Neurology](/submissions/NE)

## Identification

The thermal vestibular stimulator for headache is a prescription device used to stimulate the vestibular system by applying thermal waveforms through earpieces placed in a patient's ear canal for the treatment of headache.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with the general controls of the FD&C Act, the thermal vestibular stimulator for headache is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Performance testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform outputs and temperature limits, must be identified.
(4) Cleaning validation of earpieces must be conducted.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include the following:
(i) Information on how the device operates and the typical sensations experienced during treatment;
(ii) A detailed summary of the device's technical parameters; and
(iii) Instructions for maintenance and cleaning of the device.

## Recent Cleared Devices (1 of 1)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [DEN170023](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QAR/DEN170023.md) | ThermoNeuroModulation Device, TNM Device | Scion Neurostim, LLC | Mar 26, 2018 | DENG |

## Top Applicants

- Scion Neurostim, LLC — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QAR](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/QAR)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
