The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
Device Story
Relivion is a non-invasive, wearable neurostimulator for migraine treatment. Device headset features three pairs of electrodes contacting the forehead and occiput; frontal electrodes target trigeminal nerve branches (supraorbital/supratrochlear); posterior electrodes target greater occipital nerve branches. User wets single-use electrode pads for conductivity; activates device via on-board interface to deliver electrical stimulation pulses. Treatment runs automatically for 60 minutes; intensity adjustable by user. Optional mobile app connects via Bluetooth to display status/alerts. Device used at home by patient. Neuromodulation of trigeminal and occipital pathways reduces migraine pain and associated symptoms. Clinical benefit demonstrated by pain relief and pain freedom post-treatment.
Clinical Evidence
Three prospective studies: OS-TNS (N=30), SP-301 (N=55), and RIME (N=131). RIME study (multicenter, randomized, sham-controlled) primary endpoint: pain relief at 2 hours post-treatment (60% active vs 37% sham, p=0.0180). Secondary endpoints (pain freedom, MBS improvement) also favored active device. Safety profile consistent with other nerve stimulators; no serious adverse events reported.
Technological Characteristics
Transcutaneous electrical nerve stimulator; rechargeable 3.7V Li-Po battery. Housing: Plastic PA and silicone. Electrodes: 3 pairs (frontal/occipital) with replaceable pads. Connectivity: Bluetooth (low energy) to mobile app. Standards: IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-10, 62366, 62304. Software: 1 fixed program. Constant current output.
Indications for Use
Indicated for acute treatment of migraine with or without aura in patients 18 years of age or older. Prescription device for home self-use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.
In combination with the general controls of the Food Drug & Cosmetic Act, the Transcutaneous Electrical Nerve Stimulator to Treat Headache is subject to the following special controls:
{1}------------------------------------------------
February 16, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Neurolief Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2320 Philadelphia, Pennsylvania 19103
Re: K203419
Trade/Device Name: Relivion Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous Electrical Nerve Stimulator to Treat Headache Regulatory Class: Class II Product Code: PCC Dated: November 19, 2020 Received: November 19, 2020
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
## Indications for Use
Expiration Date: 06/30/2023 See PRA Statement below.
Form Approved: OMB No. 0910-0120
510(k) Number (if known) K203419
Device Name Relivion®
Indications for Use (Describe)
The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
Type of Use (Select one or both, as applicable)
△ Prescription Use (Part 21 CFR 801 Subpart D)
_Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unlessit displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{4}------------------------------------------------
#### 510(k) SUMMARY
Image /page/3/Picture/1 description: The image shows the logo for Neuroleaf Brain Innovation. The logo consists of a stylized brain icon on the left and the text "Neuroleaf" above "Brain Innovation" on the right. The brain icon and text are all in a light blue color.
# RELIVION®
| Applicant Name: | Neurolief Ltd.<br>12 Giborei Israel St.<br>Netanya, Israel 4250412<br>Tel: +972-9-3730288 |
|-----------------|-------------------------------------------------------------------------------------------|
|-----------------|-------------------------------------------------------------------------------------------|
Contact Person: Michal Kedar-Datel Clinical & Regulatory Affairs Director Neurolief Ltd. Tel: +972-9-3730288
Date Prepared: November 19, 2020
Relivion® Trade Name:
Classification Name: 21 CFR 882.5891 Transcutaneous electrical nerve stimulator to treat headache
Product Code: PCC
Classification: Class II
Classification Panel: Neurology
Predicate Device: CEFALY Technology's Cefaly® Acute (K171446)
Intended Use/Indication for Use:
{5}------------------------------------------------
The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
## Device Description:
The Relivion® is an external non-invasive neurostimulator designed for transcutaneous electrical nerve stimulation. The Relivion® headset integrates three pairs of output electrodes which come in contact with the subject's scalp at the forehead and occiput. The electrodes deliver the stimulation pulses produced by the headset's stimulation unit to the subject's scalp. The frontal electrodes stimulate the trigeminal (supraorbital and supratrochlear) nerve branches and the posterior electrodes stimulate the greater occipital nerve branches.
The Relivion® includes single-use replaceable electrode pads that are positioned on-top of the electrodes prior to treatment and are wetted by the user before use, in order to provide proper conductivity between the electrodes and the scalp.
The Relivion® is powered by a rechargeable battery and the headset incorporates an on-board user interface that enables the user to activate/deactivate the device and to adjust the stimulation intensity. Upon treatment activation, the treatment automatically runs and ends after 60 minutes or alternatively, the user can stop the treatment when desired.
The Relivion® can communicate via a low energy Bluetooth link with the Relivion® dedicated mobile application on the user's smartphone. The Relivion® mobile application is optional and it is used to display the device status and provide indications and alerts.
## Technological Characteristics:
The Relivion® treats migraines by stimulating the trigeminal and occipital nerve branches by a transcutaneous electrical nerve stimulation. Trigeminal and occipital electrical nerve stimulation induces neuromodulation of these nerve pathways and by that reduces pain associate with migraine and associated symptoms.
The Relivion® includes single-use, replaceable electrode pads that are positioned on-top of the headset electrodes and are wetted by the user before each use. Water releasing covers are located on the outer side of each back occipital electrode and are used to release moisture from the electrode pads to the scalp in order to provide proper electrical conductivity between the electrodes and the scalp.
The Relivion® headset adjusts to various head sizes and contours and can be worn comfortably. The headset includes two flexible arms that penetrate under the hair layers while the headset is worn.
The Relivion® headset incorporates an on-board user interface and can communicate via a low
{6}------------------------------------------------
energy Bluetooth link with the Relivion® dedicated mobile application on the user's smartphone, which displays the device status and provides indications and alerts.
## Performance Data:
Neurolief conducted several performance tests to demonstrate that the Relivion® device complies with performance standards and that it functions as intended.
Performance - Bench Testing: The Relivion® device underwent performance testing, including software validation and device verification tests. It was successfully verified that the Relivion® output parameters meet the product's specifications.
Electrical Safety and Electromagnetic Compatibility: in addition, the system was tested per the applicable electrical safety and electromagnetic compatibility standards listed below, and all results were passing.
- . IEC 60601-1 Ed. 3.1, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2 Ed. 4.0, Medical Electrical Equipment Part 1-2: General requirements for . safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
- IEC 60601-1-11 Ed. 2.0, Medical electrical equipment Part 1-11: General requirements . for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-10 Ed. 2.1, Medical electrical equipment Part 2-10: Particular requirements ● for the basic safety and essential performance of nerve and muscle stimulators
- IEC 60601-1-6 Ed. 3.1, Medical electrical equipment Part 1-6: General requirements for . basic safety and essential Performance - Collateral Standard: Usability
- IEC 62366 Ed. 1.0 Medical devices Part 1: Application of usability engineering to medical ● devices
- IEC 62304 Ed. 1.1, Medical device software Software life-cycle processes .
Software Testing: The software was also subject to verification and validation testing, and results demonstrated that the system performed as intended. Cybersecurity risks were also identified and addressed.
Biocompatibility: The Relivion® is a device that contacts intact skin surface for a limited time (<24 hours). The patient-contacting components of the device headset and electrode pads were tested for
{7}------------------------------------------------
cytotoxicity, sensitization, and irritation or intracutaneous reactivity as defined for surface devices in the FDA guidance (2020) and ISO 10993-1: 2009. Based on results of the conducted testing, the final device was determined to be biocompatible for its intended use.
Human Factors Usability Study: The device was also subject to formative and summative Human Factors testing, and results demonstrated that users could complete all essential tasks using the provided labeling with acceptable residual risk.
Performance - Clinical Testing: Two previous versions of Neurolief's device were used in preliminary prospective studies.
The OS-TNS Study was conducted at Meir Medical Center, Israel. The study assessed the safety and efficacy of combined occipital and supraorbital neurostimulation for treatment of migraine. It was approved by the institution's ethics committee on June 5, 2014 and the Israeli Ministry of Health Ethical committee and was concluded in November 2015. This clinical investigation was conducted in accordance with good clinical practice (GCP) as described in ISO 14155. Adults suffering from episodic migraine, aged 21-62 years, were enrolled. Subjects were randomly allocated in a 1:1 ratio to receive active or sham occipital and supraorbital stimulation. The primary endpoint was defined based on relative change (%) in VAS pain score from baseline to end of treatment without using pain medication. 30 patients treated one acute migraine episode with active device (N=15) or sham device (N=15). At the end of treatment there was a significant reduction of the average Pain VAS score in the treatment group vs. an increase in Pain VAS score in the control group (-79.2% vs. +14.9%, respectively; P=0.0002). Pain-free response rates significantly favored the active device at 2-hours (P=0.0031) and at 24-hours (P<0.05) post treatment. Superiority of the active device was also shown for functional disability (P=0.0004) and photophobia (P=0.002). No device-related serious adverse events were recorded.
The SP-301 Study was a prospective, randomized, double-blind, parallel-group, placebo-controlled investigation conducted to evaluate the clinical performance and safety of a self-administered abortive treatment for migraine headache using combined occipital and supraorbital transcutaneous nerve stimulation. The study was conducted at the Laniado Medical Center in Israel and approved by the Laniado Helsinki committee on February 8, 2018. The first patient enrollment was February 22, 2018, and the study concluded on November 2018. This clinical investigation was conducted in accordance with good clinical practice (GCP) as described in ISO 14155. Fifty-five (55) patients were randomized to either the active or placebo groups. To maintain blinding, measures included concealed allocation, use of an identical sham device and the same treatment protocol in all the study stages: device training, self-practice period and treatment period. Additionally, sham stimulation was set to a level well above the sensory threshold to further enhance subject's blinding. Subjects were instructed to use the device during 1 migraine episode within 14 weeks after a practice run in their home. Additionally, subjects were asked not to take any pain medication prior to treatment and until 2 hours post treatment. Mean pain intensity decreased significantly more in the treatment group than in the sham group at 1-hour (group difference: 41.4%; p=0.0002) and at 2-hours (group difference: 32.8%; p=0.0324) and 24-hours (group difference: 36.2%; p=0.0220). Responders rate
{8}------------------------------------------------
was also statistically significantly higher in the treatment group than in the sham group at 1-hour (66.7% vs. 20%, p=0.0014), 2-hours (66.7% vs. 32%, p= 0.0227) and 24-hours (78.3% vs. 48%, p= 0.0401). No Serious Adverse Events (SAEs) were reported throughout the study period. Safety data collected by the research team was recorded and reported to the sponsor. Safety monitoring, including device related assessment was performed by qualified sponsor personnel in consultation with the study PI. Post hoc review was conducted by an independent medical advisor. There were no complaints of unpleasant stimulation sensation during or within 24 hours after the treatment.
The results of these studies demonstrated that combined occipital and supraorbital transcutaneous nerve stimulation (OSTNS) is a safe and highly effective treatment for the treatment of migraine and may serve as a safe, fast acting treatment method.
The device was also subject to a prospective, randomized, parallel-group, sham-controlled clinical trial- the RIME study. The device used in the RIME study is the commercial version for which the company is seeking clearance. The study included 131 patients (109 females and 22 males) having migraine attacks with or without aura. Patients were enrolled at 12 sites in the US and Israel. The study was initiated in November 2018 and concluded in August 2020. The US study sites of the investigation complied with 21 CFR parts 50, 56, and 812 and the OUS clinical investigations were conducted in accordance with good clinical practice (GCP) as described in 21 CFR 812.28(a)(1). To maintain blinding, measures included concealed allocation, use of an identical sham device and the same treatment protocol in all the study stages: device training, self-practice period and treatment period. Additionally, sham stimulation was set to a level well above the sensory threshold to further enhance subject's blinding. These measures were proven to be effective as the blinding assessment revealed that the subjects' feeling regarding the treatment received did not influence the study outcome. Baseline demographics were consistent with the migraine population in terms of age, 83% (109/131) of the participants were females. The mean age was 40.3 years gender, etc. (SD=12.7). The mean age of migraine onset was 18.6 (SD=8.54) years and ranged from 4 to 50. 62.6% (82/131) of the subjects had migraine without aura, 37.4% (49/131) had migraine with aura, 26.7% (35/131) of the subjects were using migraine preventive medication at baseline. Statistical analysis showed no significant differences between the US and Israeli populations.
The primary outcome was the proportion of subjects reporting reduction of migraine headache pain (i.e., pain relief) 2 hours post treatment initiation from severe or moderate to mild or no pain, or from mild to no pain. Pain relief at 2 hours post-treatment was found statistically significantly higher in the active group than in the sham group: 60% of subjects in the active group compared to 37% (p value=0.0180) (mITT) or 36% (p value=0.0135)(PP) in the sham group met the primary endpoint, with similar between-group differences of 23% and 24% (mITT and PP, respectively). All statistical tests for the between-group difference were statistically significant in both populations. Therefore, the study was deemed successful.
{9}------------------------------------------------
Image /page/8/Figure/0 description: The image is a bar graph that compares the reduction in pain at 2-hours between an active treatment and a sham treatment. The y-axis represents the percentage of pain reduction, while the x-axis represents the study arm. The active treatment shows a 60% reduction in pain, while the sham treatment shows a 37.29% reduction in pain. The Chi-Square test p-value is 0.0180.
Figure 1: Pain relief 2 Hours Post-Treatment - mITT
Furthermore, the active arm was superior to the sham arm for all 3 secondary endpoints, as follows:
- Pain Freedom rate at 2 hours post-treatment initiation without use of rescue medication . (46.00% vs. 11.86% for active and sham arms, respectively, p-value: <0.0001).
- . Improvement in MBS 2-hours post-treatment initiation without use of rescue medication (80.56% vs. 60.00% for active and sham arms, respectively, p-value: 0.0466).
- . Reduction in pain level - 1-hour post-treatment initiation without use of rescue medication (-0.6 (-28.7%) vs. -0.3 (-14.4%) for active and sham arms, respectively, p-value: 0.0121).
In addition, the active arm was superior to the sham arm in the following additional parameters:
- . Complete MBS Freedom 1- and 2-hours post-treatment initiation without use of rescue medication.
- . Pain Freedom at 1-hour post-treatment initiation without use of rescue medication.
- . Pain and MBS Relief at 1- and 2-hours post-treatment initiation without use of rescue medication.
- Complete symptoms Free (Pain and MBS)- 1- and 2-hours post-treatment initiation without ● use of rescue medication.
- . Sustained 2-24 hours headache Pain Relief and Pain Freedom without use of rescue medication.
- . Pain Relief and Pain Freedom consistency of response across multiple treated episodes.
{10}------------------------------------------------
With respect to safety, safety monitoring, including device related assessment was performed by qualified sponsor personnel in consultation with the study PI and the medical advisor. In the ITT analysis set, 21 AE's were reported in 10 subjects, 8 in the active arm (incidence: 11.94%, 8/67), and 2 in the sham arm (incidence: 3.13% (2/64). None of the adverse events were serious, 7 AE's were considered moderate (4 in the active arm and 3 in the sham arm), and the other 14 were mild. 16 AE's were considered as at least possibly related to the study device (7 in 5 subjects randomized to active arm; 9 in 2 subjects in the sham arm). All the reported AEs were fully reversible and resolved without intervention. The nature of the observed events in the active group were anticipated and very similar to other nerve stimulators for migraine, including adverse events that are directly related to the neuromodulatory action of the device such as transient scalp numbness sensation, tingling and twitching. The rate and severity of events was as anticipated and similar to other previously cleared devices. An initial device malfunction resulted in exclusion of the first 50 cases from the above analysis. Because the malfunction resulted in failure to deliver stimulation, these patients were not analyzed for efficacy but were analyzed for safety, and the results were consistent with those reported for the ITT population.
Although COVID-19 resulted in interruption of study enrollment and early termination, FDA guidance was followed in the completion and analysis of study results in light of the pandemic and met the study hypothesis, as presented above, supporting substantial equivalence.
### Substantial Equivalence:
The following table compares the Relivion® device to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
{11}------------------------------------------------
### Table 1: Neurolief, Ltd.'s Relivion® Substantial Equivalence
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device) | Comparison |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Characteristics | | | |
| 510(k) number | Pending | K171446 | N/A |
| Classification | 21 CFR § 882.5891 | 21 CFR § 882.5891 | Same |
| Product Code | PCC | PCC | Same |
| Product Class | Class II | Class II | Same |
| Regulation Name | Transcutaneous Electrical Nerve<br>Stimulator to Treat Headache | Transcutaneous Electrical Nerve<br>Stimulator to Treat Headache | Same |
| Indications for Use | The Relivion® transcutaneous electrical nerve<br>stimulator is indicated for the acute treatment of<br>migraine with or without aura in patients 18<br>years of age or older. It is a prescription device<br>to be self-used at home. | Cefaly® Acute is indicated for<br>the acute treatment of migraine<br>with or without aura in patients<br>18 years of age or older. | Same<br>(Both devices are<br>intended for self-<br>administered, at<br>home use) |
| Technology | Transcutaneous Electrical<br>Nerve Stimulator | Transcutaneous<br>Electrical Nerve<br>Stimulator | Same |
| Invasiveness | Non-Surgical | Non-Surgical | Same |
| Electrode Locations | Forehead and Occiput | Forehead | Similar<br>(Forehead location is<br>the same for both<br>devices; addition of<br>occiput location does<br>not raise new<br>questions—see<br>below.) |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device) | Comparison |
| Nerves over which<br>electrodes are placed | Supratrochlear and supraorbital branches of<br>the trigeminal nerve bilaterally and the<br>occipital nerves | Supratrochlear and supraorbital<br>branches of the trigeminal nerve<br>bilaterally | Similar<br>(Relivion® treats the<br>occipital nerves in<br>addition to the<br>trigeminal nerve, but<br>the therapeutic effect<br>is the same) |
| Energy | Electric | Electric | Same |
| Power Source | Rechargeable Li-Po 3.7 V Battery | Rechargeable LiPo 3.7 V<br>battery | Same |
| Software-controlled | Yes, 1 fixed program | Yes, 1 fixed program | Same |
| Constant Current | Yes | Yes | Same |
| Constant Voltage | No | No | Same |
| Software Function | Controls the output of the device<br>and device indicators | Controls the output of the<br>device<br>and device indicators | Same |
| Timer Settings | Yes | Yes | Same |
| Patient override<br>control method | On/Off button | On/Off button | Same |
| Button Types | On/Off Button and buttons to adjust intensity<br>of electrical stimulus | On/Off Button and buttons to<br>adjust intensity of electrical<br>stimulus | Same |
| Functional features | Visual and auditory indicators inform the<br>user when the device is on vs. off and help<br>them troubleshoot if it is not working<br>properly (e.g., indicates if device is<br>active/non-active, low battery indication<br>and if electrical connection between device<br>and skin is unacceptable) | Visual and auditory<br>indicators inform the<br>user when the device is<br>on vs. off and help them<br>troubleshoot if it is not<br>working properly (e.g.,<br>indicates if device is<br>active/non-active , low | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device)<br>battery indication and if<br>electrical connection<br>between device and skin<br>is unacceptable) | Comparison |
| Bluetooth Capable | Yes | No | Different |
| Associated mobile<br>application to<br>display device<br>status, treatment<br>duration, and<br>battery status | Yes | No | (Relivion® includes a<br>mobile app and<br>Bluetooth capability,<br>but these features do<br>not alter the<br>therapeutic effect) |
| Max leakage current | None (battery operated) | None (battery operated) | Same |
| Electrodes | Relivion® electrode | Cefaly® electrode | Similar<br>(Each device uses its<br>own designated<br>electrode, but the<br>components perform<br>similarly, as<br>established by bench<br>testing) |
| Indicator display:<br>Unit functioning | Yes | Yes | Same |
| Low battery<br>indicator | Yes | Yes | Same |
| Standards:<br>IEC 60601-1 | Yes | Yes | Same |
| IEC 60601-1-2 | Yes | Yes | Same |
| IEC 60601-1-6 | Yes | Yes | Same |
| IEC 60601-1-11 | Yes | Yes | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device) | Comparison |
| IEC 60601-2-10 | Yes | Yes | Same |
| IEC 62366 | Yes | Yes | Same |
| Weight | 90 gr | 12 gr | Different<br>(The difference in<br>weight is not<br>clinically relevant to<br>the treatment) |
| Dimensions | 209mm x 128mm x 39mm | 55 mm x 40 mm x 15 mm | Different<br>(The difference in<br>size is not clinically<br>relevant to the<br>treatment) |
| Housing materials | Plastic PA + Silicone | Plastic PC | Similar<br>(Manufacturing the<br>device housing from<br>Plastic PA and<br>Silicon rather than<br>Plastic PC does not<br>raise different<br>questions of safety<br>and effectiveness) |
| Maximum Time<br>Device Used | 60 minutes | 60 minutes | Same |
| Net Charge (µC) per<br>pulse | 0 | 0 | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device) | Comparison |
| Pulse Duration<br>(µsec) | 850 | 505 | Similar<br>(Differences in the<br>stimulation<br>parameters were<br>evaluated in the<br>clinical testing, and<br>the results confirmed<br>that the devices have<br>equivalent safety and<br>efficacy<br>performance) |
| Frequency (Hz) | 80 | 100 | |
| Maximum output<br>voltage (V): @500<br>ohms @2000 ohms<br>@10000 ohms | 3 front electrodes / 6 back electrodes<br>12 front electrodes / 24 back electrodes<br>60 front electrodes / 100 back electrodes | 8<br>32<br>60 | |
| Maximum output<br>current (mA): @500<br>ohms @2000 ohms<br>@10000 ohms | 6 front electrodes /12 back electrodes<br>6 front electrodes /12 back electrodes<br>6 front electrodes /10 back electrodes | 16<br>16<br>6 | |
| Max Phase<br>Amplitude | 6 mA front electrodes/ 12 mA back<br>electrodes; with a load of a 4.7 uF<br>capacitor parallel<br>with 2.2K ohms resistance | 16 mA<br>with a load of a 4.7 uF<br>capacitor parallel with<br>2.2K ohms resistance | |
| Maximum phase<br>charge (µC) @500Ω | 4.8 | 4 | |
| Maximum Current<br>Density, (mA/cm²,<br>r.m.s.) @500Ω | 1.93 front electrodes/ 2.78 back electrodes | 2.37 | |
| Maximum Average<br>Power Density,<br>(W/cm²) @500Ω | 0.0000116 front electrodes /<br>0.000034 back electrodes | 0.000047 | |
| Maximum Average<br>Current (average<br>absolute value, mA)<br>@500Ω | 0.38 front electrodes / 0.76 back electrodes | 0.8 | |
| Phase rise time | 5 µS | 2 µS | |
| Modulation Options<br>Amplitude<br>Frequency Duration | 0- 12 mA<br>Fixed @ 80 Hz<br>330- 400 µS | 0-16 mA<br>Fixed @ 100 Hz<br>Fixed @ 250 µS | |
| Phase decay time | 2 µS | 2 µS | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | CEFALY Technology's<br>Cefaly® Acute<br>(Predicate Device) | Comparison |
| Ramp Modulations<br>Ramp Up<br>Ramp Down | Manually<br>Manually | 14 minutes<br>1 minute | Different<br>(Relivion®'s use of<br>manual ramp<br>modulation does not<br>raise different<br>questions of safety<br>and effectiveness) |
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
As described in the comparison table above, the subject Relivion® and the predicate Cefaly® Acute (K171446) share the same intended use and similar indications, technological characteristics, and same principles of operation. The minor differences in the technological characteristics do not alter the overall therapeutic effect of the device. Any differences between the Relivion® and its predicate (K171446) were carefully evaluated through performance testing. The Relivion® device underwent performance testing, including bench testing, clinical testing, software validation testing, electrical safety according to IEC 60601-1, electromagnetic compatibility testing according to IEC 60601-1-2 and other tests. These performance tests confirmed that the Relivion® complies with the same special controls and the same consensus and performance standards, on which FDA based its clearance of the Cefaly® Acute (K171446), and demonstrated that the differences in the technological characteristics between the subject and predicate device do not adversely impact performance and that the subject Relivion® is substantially equivalent to its predicate device (K171446).
#### Conclusions:
Therefore, based on the same intended use and similar indications, technological characteristics, and same principles of operation, the Relivion® device is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.