K192773 · Nu Eyne Co., Ltd. · PCC · Dec 6, 2019 · Neurology
Device Facts
Record ID
K192773
Device Name
ALLIVE
Applicant
Nu Eyne Co., Ltd.
Product Code
PCC · Neurology
Decision Date
Dec 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5891
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Allive is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.
Device Story
Allive is a transcutaneous electrical nerve stimulator (TENS) for episodic migraine prophylaxis. Device consists of an electrical stimulation unit, leads, and self-adhesive electrodes applied bilaterally to the forehead over the upper branches of the trigeminal nerve. Powered by a 3.7V rechargeable Li-po battery, it delivers a fixed-program, constant-current biphasic waveform (60Hz, 250µS pulse duration, 16mA max output). User operates via an On/Off button. Device provides non-invasive stimulation to afferent cranial nerves to reduce migraine frequency. Used in home environments by patients. Output is controlled by the user; clinical benefit is migraine prophylaxis.
Clinical Evidence
No clinical data provided for the subject device. Clinical evaluation report based on literature and clinical data of the predicate device (CEFALY) was submitted to support safety and effectiveness.
Indicated for prophylactic treatment of episodic migraine in patients 18 years of age or older.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.
In combination with the general controls of the Food Drug & Cosmetic Act, the Transcutaneous Electrical Nerve Stimulator to Treat Headache is subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2019
Nu Eyne Co., Ltd % Dongha Lee Regulatory Affairs Consultant KMC, Inc. Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu Seoul, 08375 Kr
Re: K192773
Trade/Device Name: Allive Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous Electrical Nerve Stimulator To Treat Headache Regulatory Class: Class II Product Code: PCC, Dated: September 26, 2019 Received: September 30, 2019
Dear Dongha Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Timothy Marjenin Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192773
Device Name
ALLIVE (MODEL: ALLIVE01, AD-01)
Indications for Use (Describe)
The Allive is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.
Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nu Eyne. The logo features a stylized eye with the letters "NE" inside the pupil. Below the eye, the words "Nu Eyne" are written in a sans-serif font. Underneath the company name is the tagline "Advanced Vision Care, New Life."
# 510(k) SUMMARY
This summary of 510(k) -safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: SEPTEMBER 26, 2019
#### 1. INFORMATION
1.1 Submitter Information
- 트 Submitter Name: Nu Eyne Co., Ltd.
- . Address
- : #403, Seoul Biohub, 117-3 Hoegi-ro, Dongdaemun-gu, Seoul, 02455, Korea
- . . Telephone Number: +82-2-6953-8120 Fax: +82-303-347-0017
- I Email: dongseong.lee@nueyne.com
#### 1.2 Contact Person
- . Name: DongHa Lee (Consultant / KMC, Inc.)
- . Address: Room no. 904, 27, Digital-ro 27ga-gil, Guro-gu, Seoul, 08375, Korea
- Telephone Number: +82-70-8965-5554 ■ Fax: +82-2-2672-0579
- . E-mail: dhlee@kmcerti.com
#### 2. DEVICE INFORMATION
- 2.1 Trade Name / Proprietary Name: ALLIVE (Model: ALLIVE01, AL-2HP)
- 2.2 Common Name: Transcutaneous electrical nerve stimulator to treat headache
- 2.3 Classification Name: Stimulator, Nerve, Electrical, Transcutaneous, For Migraine
- 2.4 Product Code: PCC
- 2.5 Classification Regulation: 21CFR 882.5891
- 2.6 Device Class: Class II (Special Control)
- 2.7 Classification Panel: Neurology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nu Eyne. The logo features a stylized eye with a blue iris and a white pupil. Inside the pupil is a stylized "N" and "E". Below the eye is the text "Nu Eyne" in a sans-serif font. Below that is the text "RESTORING SIGHT WITH LIGHT THERAPY" in a smaller font.
## 3. PREDICATE DEVICE
| Predicate Device | |
|--------------------------|-------------------|
| Manufacturer | CEFALY Technology |
| Device Name (Trade Name) | CEFALY® |
| 510(k) Number | K160237 |
# 4. SUBJECT DEVICE DESCRIPTION
The Allive device is a transcutaneous electrical nerve stimulator (TENS) that is applied to the forehead using a self-adhesive electrode positioned over the upper branches of the trigeminal verve bilaterally. It is intended to stimulate the upper branches of the trigeminal verve in order to reduce the frequency of migraine attack.
# 5. INTENDED USE
The Allive is indicated for the prophylactic treatment of episodic migraine in patients 18 years of age or older.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for NuEyene. The logo features a stylized eye with a blue outline and a stylized "N" inside the eye. Below the eye is the text "NuEyene" in a sans-serif font, with the tagline "Innovative ocular drug delivery" written in a smaller font below the company name.
# 6. SUBSTANTIAL EQUIVALENCE
| Items | Subject Device | Predicate Device | Comparison Result | |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Nu Eyne Co., Ltd. | CEFALY Technology | Different | |
| Device | Stimulator, Nerve, Electrical,<br>Transcutaneous, For Migraine | Stimulator, Nerve, Electrical,<br>Transcutaneous, For Migraine | Same | |
| Trade/Device Name | Allive | CEFALY® | Different | |
| K Number | None | K160237 | Different | |
| Regulation Number | 21CFR 882.5891 | 21CFR 882.5891 | Same | |
| Regulation Description | Transcutaneous electrical nerve<br>stimulator to treat headache | Transcutaneous electrical nerve<br>stimulator to treat headache | Same | |
| Regulatory Class | Class II | Class II | Same | |
| Product Code | PCC | PCC | Same | |
| Definition | Used to apply an electrical current<br>to a patient's cranium through<br>electrodes placed on the skin. | Used to apply an electrical current<br>to a patient's cranium through<br>electrodes placed on the skin. | Same | |
| Review Panel | Neurology | Neurology | Same | |
| Physical State | Electrical stimulation unit with<br>leads and cutaneous electrodes. | Electrical stimulation unit with<br>leads and cutaneous electrodes. | Same | |
| Technical Method | Applies an electrical current<br>through electrodes on patient's<br>skin. | Applies an electrical current<br>through electrodes on patient's skin. | Same | |
| | | | | |
| Target Area | Afferent cranial nerves. | Afferent cranial nerves. | Same | |
| Intended use | The Allive is indicated for the<br>prophylactic treatment of episodic<br>migraine in patients 18 years of<br>age or older. | The Cefaly® device is indicated for<br>the prophylactic treatment of<br>episodic migraine in patients 18<br>years of age or older. | Same | |
| Picture | Image: Allive device | Image: Cefaly device | Different | |
| Power Source | 1rechargeable Lipo 3.7 V battery | 1rechargeable Lipo 3.7 V battery | Same | |
| Channels | 1 | 1 | Same | |
| Computerized | Yes | Yes | Same | |
| S/W provided | 1 fixed program | 1 fixed program | Same | |
| Constant current | Yes | Yes | Same | |
| Constant voltage | No | No | Same | |
| Max output current | 16mA | 16mA | Same | |
| Max output voltage (2kOhm) | 32 Volts | 32 Volts | Same | |
| Patient Override Control Method | On/Off button | On/Off button | Same | |
| | Max Leakage Current | None (battery operated) | None (battery operated) | Same |
| | Electrode | Self-adhesive | Self-adhesive | Same |
| | Indicator display: Unit functioning | Yes | Yes | Same |
| | Low battery indicator | Yes | Yes | Same |
| | Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-10<br>IEC 62366-1<br>IEC 62304<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| | Timer Setting | Yes | Yes | Same |
| Device | Weight | 110.2g | 12g | Different |
| | Dimensions | 160mm x 66.5mm x 55mm | 55mm x 35mm x 15mm | Different |
| | | Expected Service Life | 5 years | 7 years |
| | | IP Classification | IP22 | IP22 |
| | | Electrical Protection | Type BF | Type BF |
| | Battery Type | Lithium ion Battery | Lithium ion Battery | Same |
| Battery | Expected Service Life | 7 years (300Cycles of complete<br>discharge) | 7 years (300Cylces of complete<br>discharge) | Same |
| | | Maximum input voltage<br>(USB connector) | 5.25 Vdc | 5.25 Vdc |
| | Dimensions | 94mm (W) x 30mm (H) | 94mm (W) x 30mm (H) | Same |
| Electrode | Expected Service Life | 20 times | 20 times | Same |
| Special Requirement b.3) in accordance with 21CFR 882.5891 | | | | |
| Waveform | | Biphasic | Biphasic | Same |
| Maximum output voltage<br>(V) | 500 ohms | 8 | 8 | Same |
| | 2,000 ohms | 32 | 32 | Same |
| | 10,000 ohms | 65 | 60 | Same |
| Maximum output current<br>(mA) | 500 ohms | 16 | 16 | Same |
| | 2,000 ohms | 16 | 16 | Same |
| | 10,000 ohms | 6.5 | 6 | Same |
| Pulse duration (µS) | | 505 | 505 | Same |
| Frequency Range | | 60Hz fixed | 60Hz fixed | Same |
| Net charge (µC) per pulse | | 0 | 0 | Same |
| Duration of the primary (depolarizing) phase<br>(µS) | | 250 | 250 | Same |
| Standby duration between the two phase (µS) | | 5 | 5 | Same |
| Maximum current density<br>(mA/cm2, r.m.s.) | 500 ohms | 2.37 | 2.37 | Same |
| Maximum average power density<br>(W/cm2) | 500 ohms | 0.000017 | 0.000017 | Same |
| Maximum average current<br>(Average absolute value, mA) | 500 ohms | 0.48 | 0.48 | Same |
| Other Technical Item | | | | |
| Wave Shape | | Rectangular<br>Full compensated<br>Symmetrical | Rectangular<br>Full compensated<br>Symmetrical | Same |
| | | | | |
| Phase rise time | | 2 µS | 2 µS | Same |
| Phase decay time | | 2 µS | 2 µS | Same |
| Maximum Phase Charge (µS) at 500 ohms | | 4 | 4 | Same |
| Modulation option | Amplitude | 0 to 16 mA Fixed | 0 to 16 mA Fixed | Same |
| | Frequency | 60Hz fixed | 60Hz Fixed | Same |
| | Duration | 250 µS | 250 µS | Same |
| Ramp Modulation | Ramp up | 14min | 14min. | Same |
| | Ramp down | 45s | 45s | Same |
| Steady time | | 6min | 6min | Same |
| Session Duration | | 20min 45s | 20min 45s. | Same |
| Amplitude modulation | | Amplitude is adjusted by the user | Amplitude is adjusted by the user | Same |
| Material | | | | |
| Device housing materials | | Plastic ABS | Plastic ABS | Same |
| Electrode top layer | | Polyethylene terephthalate | Polyurethane | Different |
| Electrode intermediate layer | | Conductive silver carbon ink | Carbon with silver coating | Different |
| Electrode bottom layer | | Uv curable dielectric ink | x | Different |
| Central pin | | N/A | N/A | Same |
| Metallic parts for magnetic attraction | | Tinplate | Tinplate | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Nu Eyne. The logo features a stylized eye with a blue outline and a white fill. Inside the eye is the letter "N" in a stylized font. Below the eye is the text "NU Eyne" in a sans-serif font, with the word "NU" in a larger font size than "Eyne".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for NuEyene. The logo features a stylized eye with a blue outline and a stylized "N" inside. Below the eye is the text "NU Eyene" in a sans-serif font, with the word "NU" in a larger font size than "Eyene".
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Nu Eyne. The logo features a stylized eye with a blue outline and a stylized "N" inside. Below the eye is the text "Nu Eyne" in a sans-serif font. The text is in a dark color, possibly black or dark blue.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo for NuEyene. The logo features a stylized eye with a blue outline and a white fill. Inside the eye is the letter "N" in a stylized font. Below the eye is the word "NuEyene" in a sans-serif font. Below the word "NuEyene" is the tagline "Technology for better vision, better life."
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for NuEyene. The logo features a stylized eye with the letters 'NE' inside the pupil. Below the eye is the text 'NuEyene' in a sans-serif font, with the tagline 'Advanced retinal imaging & telehealth' underneath in a smaller font.
# 7. NON-CLINICAL DATA
- 7.1 Safety Test
- 1) Biocompatibility
The biocompatibility tests were performed to protect patients from undue risks arise from biological hazards associated with materials of manufacture and final device. The tests were performed in accordance with the following standards and FDA Guidance - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
| No. | Test Items | Standards |
|-----|--------------------------------|-------------------|
| 1 | Cytotoxicity | ISO 10993-5:2009 |
| 2 | Sensitization | ISO 10993-10:2010 |
| 3 | Intracutaneous Reactivity Test | ISO 10993-10:2010 |
#### 2) Electrical Safety and EMC
The electrical safety tests were performed to protect patients from undue risks arise from any hazards associated with final device. The tests were performed in accordance with the following standards.
| No. | Test Items | Standards |
|-----|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| 1 | General requirement for basic safety and<br>essential performance | IEC 60601-1:2005+A1: 2012<br>(AAMI/ANSI ES 60601-1: 2005+A1: 2012) |
| 2 | General requirement for safety -<br>Electromagnetic disturbances | IEC 60601-1-2:2014 |
| 3 | General requirement for safety - Medical<br>electrical equipment used in the home<br>healthcare environment | IEC 60601-1-11:2015 and<br>FDA Guidance ("Design Considerations for<br>Devices Intended for Home Use") |
| 4 | Particular requirement for safety - Nerve<br>and muscle stimulators | IEC 60601-2-10:2012/Amd1:2016 |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Nu Eyne Advanced Spine and Joint Labs. The logo features a stylized eye with the letters "NE" inside the pupil. The text "Nu Eyne" is written below the eye, and the words "Advanced Spine and Joint Labs" are written in a smaller font below "Nu Eyne."
# 7.2 Performance Test
The following tests were performed to assess effectiveness of performance of the device. The tests were performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|---------------------------------------------------------------------|-------------------------------|
| 1 | Particular requirement for safety – Nerve<br>and muscle stimulators | IEC 60601-2-10:2012/Amd1:2016 |
| 2 | Technical Test | Manufacturer's SOP |
## 7.3 Usability V&V
The following tests were performed to assess effectiveness of usability of the device. The test was performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for safety - Usability | IEC 60601-1-6:2013<br>IEC 62366-1:2015 and<br>FDA Guidance ("Applying Human Factors<br>and Usability Engineering to Medical<br>Devices") |
## 7.4 Software
The following tests were performed to assess effectiveness of software of the test was performed in accordance with following standards.
| No. | Test Items | Standards |
|-----|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General requirement for safety -<br>Programmable electrical medical systems<br>(PEMS) | • IEC 62304:2006/Amd1:2015<br>• FDA Guidance (“Guidance for the Content<br>of Premarket Submissions for Software<br>Contained in Medical Devices”) |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for NuEyene. The logo features a stylized eye with the letters "NE" inside the pupil. Below the eye is the text "NuEyene" in a modern sans-serif font. Underneath "NuEyene" is the tagline "Innovative Intraocular Lens LLC" in a smaller font.
## 8. CLINICAL DATA
Although clinical performance data are required to demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population on the special control according to clause b.6) of 21CFR882.5981, we consider that the subject device (Allive) is not applied with clause b.6) of 21CFR882.5981.
Although there are not the clinical performance data of the subject device, we prepare the clinical evaluation report by using the collected clinical data of the predicate device.
In this evaluation report, the subject device is safe and effective as a treatment for headache in the indicated patient population.
The clinical evaluation was performed in accordance with following standards.
| No. | Items | Standards |
|-----|---------------------|-----------------------|
| 1 | Clinical Evaluation | • MEDDEV 2.7.1. rev.4 |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for NuEyene. The logo features a stylized eye with a blue outline and a teal fill. Inside the eye is the letter 'N' in white, with a curved line extending from the top of the 'N' to form the upper part of the eye. Below the eye is the word 'NuEyene' in a modern, sans-serif font, with the tagline 'Advanced Vision Patient Care LLC' in a smaller font underneath.
# 9. CONCLUSION
Under the comparing substantial equivalence between the subject device and the predicate device, there are the same points such as general information, some technical and material information. Although there are some differences, the safety and performance test reports are supported to the safety and effectiveness of the subject device.
In this regard, we conclude that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.