← Product Code [PBJ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/PBJ) · K150771

# RxG Distraction System (K150771)

_KLS Martin L.P. · PBJ · Aug 27, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/PBJ/K150771

## Device Facts

- **Applicant:** KLS Martin L.P.
- **Product Code:** [PBJ](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/PBJ.md)
- **Decision Date:** Aug 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5330
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The RxG Distraction System includes devices intended for use in bone stabilization and elongation (lengthening) in the correction of congenital or developmental defects in cranial reconstruction for the pediatric and adult populations.

## Device Story

The RxG Distraction System is an implantable device for cranial bone distraction osteogenesis. It consists of resorbable footplates (85:15 PLLA-PGA) secured to bone via SonicPins RxG, connected to a threaded drive screw and activation arm (TI-6AL-4V or stainless steel). The device is positioned internally with the activation arm extending through soft tissue. A physician performs the procedure in a hospital setting. Distraction is achieved by the patient or caregiver turning the external activation arm using a driver, causing the plates to separate and lengthen the bone. Drive screws allow up to 40 mm of distraction. After consolidation, the drive screw and activation arm are removed; footplates and pins remain to resorb over 12-14 months. The system facilitates gradual bone growth, correcting cranial defects and potentially reducing the need for more invasive reconstructive surgeries.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing comparing the subject device to the primary predicate. Bench testing simulated osteotomy on sawbone blocks in Ringer's solution, measuring deformation and distraction distance over 52 days. Results confirmed the subject device achieved greater distraction distance and exhibited less deformation than the predicate.

## Technological Characteristics

Materials: 85:15 PLLA-PGA (plates/pins), TI-6AL-4V or stainless steel (drive screw/arm). Dimensions: 2-5 mm plate thickness, 1.6-2.1 mm pin diameter, 3-9 mm pin length. Max distraction: 40 mm. Sterilization: Gamma radiation (SAL 10^-6). Resorption time: 12-14 months. Breakdown products: H2O and CO2. Standards: ISO 10993, ISO 11137, ISO 11607.

## Regulatory Identification

A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

## Predicate Devices

- LactoSorb Distraction ([K002083](/device/K002083.md))

## Reference Devices

- RxG Distraction System ([K133304](/device/K133304.md))
- Resorb-X G ([K121606](/device/K121606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol composed of three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 27, 2015

KLS-Martin L.P. Mr. Gary Moore Ouality MGT and Regulatory Affairs Manager 11201 Saint Johns Industrial Parkway South Jacksonville, Florida 32246

Re: K150771

Trade/Device Name: RxG Distraction System Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Nonalterable Cranioplasty Plate Regulatory Class: Class II Product Code: PBJ Dated: July 21, 2015 Received: July 24, 2015

Dear Mr. Moore:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -SFD/A

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150771

Device Name RxG Distraction System

Indications for Use (Describe)

The RxG Distraction System includes devices intended for use in bone stabilization and elongation (lengthening) in the correction of congenital or developmental defects in cranial reconstruction for the pediations.

| Type of Use (Select one or both, as applicable) | Registration Use (Part 21 CFR 201 Subpart D)    OTC System Use (Part 21 CFR 201 Subpart G) |
|-------------------------------------------------|--------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black and are smaller than the other letters.

P.O. Box 16369 · Jacksonville, FL 32245-6369 904-641-7746 or 800-625-1557 • Fax 904-641-7378 www.klsmartinnorthamerica.com

|                    | 510(k) Summary<br>21 CFR 807.92                                                                             |
|--------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter:         | KLS-MARTIN L.P.<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246                            |
| Contact Person:    | Jennifer Damato<br>Director of Quality MGT & Regulatory Affairs<br>Phone: 800-625-1557<br>Fax: 904-641-7378 |
| Date Prepared:     | August 19, 2015                                                                                             |
| Trade Name:        | RxG Distraction System                                                                                      |
| Common Name:       | Cranial Distraction System                                                                                  |
| Classification:    | Preformed nonalterable cranioplasty plate<br>Class II, 21 CFR 882.5330, Product Code PBJ                    |
| Primary Predicate: | LactoSorb Distraction (K002083)                                                                             |
| Reference Devices: | RxG Distraction System (K133304)<br>Resorb-X G (K121606)                                                    |

## Device Description:

The RxG Distraction System consists of implantable devices used to lengthen or increase the dimension of cranial bones through distraction osteogenesis. It is composed of multiple sizes and shapes of RxG footplates and a threaded drive screw connected to an activation arm. The device is positioned internally with the connected activation arm extending through the soft tissue for external activation. The RxG footplates are secured to the bone on either side of the osteotomy with SonicPins RxG. Distraction is achieved by turning the activation arm with the patient activation driver, causing the plates to separate. Various lengths of drive screws are available to achieve up to 40 mm of distraction. Upon completion of distraction and consolidation of the bone, the drive screw is detached from the RxG footplates and removed, while the RxG footplates and SonicPins RxG remain implanted and are resorbed in 12-14 months.

The purpose of this Traditional 510(k) is to expand the indications for use of the RxG Distraction System to include cranial use in the pediatric and adult populations.

#### Indications for Use:

The RxG Distraction System includes devices intended for use in bone stabilization and elongation (lengthening) in the correction of congenital or developmental defects in cranial reconstruction for the pediatric and adult populations.

{4}------------------------------------------------

# KLS martın LP.

P.O. Box 16369 • Jacksonville, FL 32245-6369 904-641-7746 or 800-625-1557 • Fax 904-641-7378 www.klsmartinnorthamerica.com

## Technological Characteristics / Substantial Equivalence:

|                                                | RxG Distraction System<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LactoSorb Distraction<br>(Primary Predicate – K002083)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | RxG Distraction System<br>(Reference Device – K133304)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Resorb-X G<br>(Reference Device – K121606)                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                            | The RxG Distraction System includes<br>devices intended for use in bone<br>stabilization and elongation<br>(lengthening) in the correction of<br>congenital or developmental defects in<br>cranial reconstruction for the pediatric<br>and adult populations.                                                                                                                                                                                                                                                      | The Lorenz Resorbable Distraction System<br>includes devices intended for use in bone<br>stabilization and elongation (lengthening)<br>when correction of oral (alveolar ridge),<br>cranial, maxillofacial deficiencies or<br>post-traumatic defects require gradual bone<br>distraction. The Lorenz Resorbable Distraction<br>System also includes devices intended for use<br>in bone stabilization and elongation<br>(lengthening) when correction of mandibular<br>deficiencies or post-traumatic defects require<br>gradual bone distraction patients two (2) years<br>old or younger. The mid-face distractor is<br>intended primarily for LeFort III osteotomies. | The RxG Distraction System includes<br>devices intended for use in bone<br>stabilization and elongation<br>(lengthening) in the correction of<br>congenital or developmental defects of<br>the midface and alveolar ridge.<br>The RxG Distraction System also<br>includes devices intended for use in<br>bone stabilization and elongation<br>(lengthening) in the correction of<br>congenital or developmental defects of<br>the mandible in patients 2 years old or<br>younger.<br>The RxG Distraction System is not<br>intended for load-bearing applications<br>in adult or adolescent populations. | Resorb-X G is intended for use in<br>non-load bearing fracture repair and<br>reconstructive procedures in<br>adolescent and adult populations. In<br>addition, resorbable meshes, plates<br>and pins may be used in non-loading<br>bearing applications for maintaining<br>the relative position of, and/or<br>containing, bony fragments, bone<br>grafts (autograft or allograft), or bone<br>graft substitutes in cranial<br>reconstruction. |
| Contraindications                              | 1. Active infection<br>2. Foreign body sensitivity. Where<br>material sensitivity is suspected,<br>testing is to be completed prior to<br>implantation.<br>3. Patients with limited blood supply,<br>insufficient quantity or quality of<br>bone, or latent infection.<br>4. Patients or parents/guardians of<br>patients are unwilling or incapable<br>of following postoperative care<br>instructions.<br>5. Patients with bleeding disorders or<br>poor wound healing.<br>6. Patients with metabolic disorders. | 1. Active infection<br>2. Foreign body sensitivity. Where material<br>sensitivity is suspected, testing is to be<br>completed prior to implantation.<br>3. Patients with limited blood supply,<br>insufficient quantity or quality of bone,<br>or latent infection.<br>4. Patients, or parents/guardians of patients<br>are unwilling or incapable of following<br>postoperative care instructions.<br>5. Patients with bleeding disorders or poor<br>wound healing.<br>6. Patients with metabolic disorders.                                                                                                                                                            | 1. Active infection<br>2. Foreign body sensitivity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1. High-load regions in the absence<br>of traditional rigid fixation<br>2. Active or latent infections<br>3. Patients in a bad general state of<br>health or suffering from<br>metabolic disorders                                                                                                                                                                                                                                             |
| Target Population                              | Pediatrics and Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Pediatrics and Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Pediatrics and Adults                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Pediatrics and Adults                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Anatomical Sites                               | Cranial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Cranial, Oral, Mandibular, and Maxillofacial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Oral, Mandibular, and Maxillofacial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Cranial                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                | RxG Distraction System<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LactoSorb Distraction<br>(Primary Predicate – K002083)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | RxG Distraction System<br>(Reference Device- K133304)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Resorb-X G<br>(Reference Device – K121606)                                                                                                                                                                                                                                                                                                                                                                                                     |
| Materials                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Bone Plates                                    | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 82:18 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Bone<br>Pins/Screws                            | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 82:18 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 85:15 PLLA-PGA                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Drive Screw                                    | TI-6AL-4V or Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | TI-6AL-4V or Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Activation Arm                                 | TI-6AL-4V or Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | TI-6AL-4V , Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | TI-6AL-4V or Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Sterility                                      | Provided Sterile (Gamma<br>Radiation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Provided Sterile (ETO)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Provided Sterile (Gamma<br>Radiation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Provided Sterile (Gamma<br>Radiation)                                                                                                                                                                                                                                                                                                                                                                                                          |
| Standards Met                                  | See Section 9.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | ASTM F136, ASTM F138                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same as current submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | ISO 10993-1, -5, -12, -18<br>ISO 11137-1, -2, 11737-1, -2                                                                                                                                                                                                                                                                                                                                                                                      |
| Where used (hospital, home,<br>ambulance, etc) | Health Care Facilities/ Hospitals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Health Care Facilities/ Hospitals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Health Care Facilities/ Hospitals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Health Care Facilities/ Hospitals                                                                                                                                                                                                                                                                                                                                                                                                              |
| Breakdown Products<br>(Plates, Pins/Screws)    | H2O and CO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | H2O and CO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | H2O and CO2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | H2O and CO2                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Resorption Time                                | 12 - 14 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 12 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 12 - 14 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 12 - 14 months                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Max. Distraction Distance                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Cranial<br>Distractor                          | 40 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 40 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Alveolar Ridge<br>Distractor                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 25 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 25 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Midface<br>Distractor                          | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 40 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 30 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Mandibular<br>Distractor                       | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 25 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 30 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Plate Thickness                                | 2 - 5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 2 - 5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2 - 5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 0.3 - 1.0 mm                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Cumulative Max. Plate<br>Volume                | 7840 mm3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 7840 mm3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 15876 mm3                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Pin/Screw Diameter                             | 1.6 mm or 2.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 1.5 – 2.5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 2.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 1.6 mm or 2.1 mm                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Pin/Screw Length                               | 3 - 9 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 6 - 9 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 3 - 7 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 3 - 7 mm                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{5}------------------------------------------------

# kıs martın L.P.

 P.O. Box 16369 • Jacksonville, FL 32245-6369 904-641-7746 or 800-625-1557 ● Fax 904-641-7378 www.klsmartinnorthamerica.com

## Technological Characteristics / Substantial Equivalence, Continued:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for KLS Martin L.P. The text "KLS" is in red, while the text "martin" is in gray. The text "L.P." is in black and is smaller than the other text. The logo is simple and modern.

#### Nonclinical Testing:

Nonclinical testing to support substantial equivalence of the RxG Distraction System is summarized in the following table:

| Test                   | Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Results                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Bench      | Comparative mechanical testing was performed between the RxG distractor and LactoSorb distractor to measure deformation and distraction distance. Acceptance Criteria: The RxG distractor should exhibit less deformation than the LactoSorb distractor. Devices were affixed to sawbone blocks to simulate an osteotomy, and then placed in a waterbath of Ringer's solution for the duration of testing. After the 48 hour latency period, distraction and consolidation were simulated over a period of 52 days. Deformation measurements were taken at regular intervals. | Final deformation was determined by measuring the distance between the two footplates of each distractor. The RxG Distraction System exhibited less deformation and achieved greater distraction distance than the LactoSorb distractor.<br><br>The results of the testing showed substantial equivalence to the primary predicate.                                                                                                                                                                          |
| Biocompatibility       | Biocompatibility requirements were assessed in accordance with FDA Blue Book Memorandum #G95-1 “Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. The battery of testing included the following tests:<br>Toxicology Cytotoxicity Chemical Analysis                                                                                                                                                                                                | The RxG Distraction System patient-contacting materials Resorb xG, Ti-6Al-4V, and Stainless Steel used for the components in the subject device are previously cleared in K133304, have the same chemical composition, undergo the same manufacturing processes, have the same body contact, and are subjected to the same sterilization methods as the reference device.<br><br>The results of the testing showed that the biocompatibility requirements have been met. No claim of "pyrogen free" is made. |
| Sterility              | Bioburden, dose verification, and tests of sterility for the gamma sterilization process were validated in accordance with ISO 11137-1, -2, as well as ISO 11737-1, -2 to achieve an SAL of 10-6.                                                                                                                                                                                                                                                                                                                                                                             | The subject device undergoes the same sterilization methods as the reference devices in K121606 and K133304.<br><br>The results of the testing showed that the sterilization doses achieve an SAL of 10-6 for the devices.                                                                                                                                                                                                                                                                                   |
| Packaging & Shelf-Life | Validation of package integrity was performed in accordance with ISO 11607-1, -2. Real-time stability testing was performed for the Resorb xG implants in K121606.                                                                                                                                                                                                                                                                                                                                                                                                            | The packaging & shelf-life for the subject device is identical to the reference device in K133304.<br><br>The results of the testing showed that the packaging is suitable for the intended sterilization process and can maintain sterility during the intended shelf life period of 5 years.                                                                                                                                                                                                               |

## Clinical Testing:

Clinical testing was not necessary for the determination of substantial equivalence.

#### Conclusions:

The RxG Distraction System has the same intended use as the primary predicate, the LactoSorb Distractor (K002083). In addition, the subject device has the same technological characteristics as the RxG Distraction System (K133304) and Resorb-X G (K121606) reference devices. The nonclinical testing results and similarities in technological characteristics do not raise new issues of safety or effectiveness and demonstrate substantial equivalence to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/PBJ/K150771](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/PBJ/K150771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
