BioniCare Hand System

K142236 · Vision Quest Industries, Inc. · NYN · Mar 23, 2015 · Neurology

Device Facts

Record IDK142236
Device NameBioniCare Hand System
ApplicantVision Quest Industries, Inc.
Product CodeNYN · Neurology
Decision DateMar 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K142236 · Mar 23, 2015BioniCare Hand SystemVision Quest Industries, Inc.Prospective multi-center clinical study (investigating physicians' practices)The study evaluated the efficacy and safety of the BioniCare Hand System in treating osteoarthritis of the hand by measuring pain reduction (VAS), functional improvement (DASH), and physical strength metrics in a clinical setting.Osteoarthritis; Pain reduction; Functional assessment; Open-label study

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Prospective, multi-center, open-label study; Prospective, multi-center, open-label; Follow-up/Duration: 8 weeksPatients 18 years or older with osteoarthritis of the hand symptomatic for at least 3 months; Sample Size: 82; Number of Sites: Multi-centerNot applicable for this studyOsteoarthritis pain (VAS), patient/physician global assessment, DASH functional assessment, pinch force, grip strength

Indications for Use

The BioniCare Hand System, Model BIO-2000, is intended for use by patients with rheumatoid and/or osteoarthritis of the hand as: ● An adjunctive therapy in reducing the level of pain and stiffness associated with pain from rheumatoid arthritis of the hand. ● As an adjunctive therapy in osteoarthritis of the hand to reduce the level of pain and stiffness and to improve the function of the hand.

Device Story

Portable, battery-operated, single-channel transcutaneous electrical nerve stimulator (TENS). Inputs: user-controlled settings via interface. Transformation: voltage-regulated circuit generates spike-shaped monophasic pulses (100 Hz fixed frequency; 0-15V amplitude). Output: electrical stimulation delivered via electrodes to the hand. Used by patients in home or clinical settings. Healthcare providers monitor patient progress; output provides adjunctive pain/stiffness relief and functional improvement. Benefits: non-pharmacological pain management, improved grip/pinch strength, and reduced stiffness.

Clinical Evidence

Prospective, multi-center, open-label study (n=82; mean age 64). Primary endpoint: OA pain reduction in index hand over 48 hours. Results: 8-week treatment showed effect sizes of 1.3 for OA pain, 1.2 for patient global assessment, 1.1 for physician global assessment, and 0.5 for DASH functional score. Secondary measures (pinch/grip strength) showed smaller significant effect sizes (0.4/0.3). Study limitations include open-label design and potential placebo effect, partially mitigated by stable baseline NSAID/analgesic use.

Technological Characteristics

Molded ABS/PC housing; 3.7V battery-powered; single-channel; monophasic spike-shaped pulse; 100 Hz fixed frequency; 0-15V output. Connectivity: LCD interface for parameter display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, AAMI NS4. Embedded firmware controls user interface and stimulation parameters.

Indications for Use

Indicated for patients 18+ years with symptomatic rheumatoid arthritis or osteoarthritis of the hand to reduce pain and stiffness and improve hand function as an adjunctive therapy.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing right, with flowing lines extending from the bottom profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 23, 2015 Vision Quest Industries, Inc. Mr. Mohammed Ouerghi Director of OA/RA 1390 Decision Street, Suite A Vista. CA 92081 Re: K142236 Trade/Device Name: BioniCare Hand System, Model BIO-2000 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator Regulatory Class: Class II Product Code: NYN Dated: February 17, 2015 Received: February 19, 2015 Dear Mr. Ouerghi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {2}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Felipe Aquel -S Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K142236 Device Name BioniCare Hand System, Model BIO-2000 Indications for Use (Describe) The BioniCare Hand System, Model BIO-2000, is indicated for use as an adjunctive therapy in reducing the level of pain and stiffness associated with pain from rheumatoid arthritis of the hand. The BioniCare Hand System, Model BIO-2000, is indicated for use as an adjunctive therapy in osteoarthritis of the hand to reduce the level of pain and stiffness and to improve the function of the hand. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) SUMMARY | 510(k) Owner: | Vision Quest Industries, Inc.<br>18011 Mitchell South,<br>Irvine, CA, 92614 | |---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Mohamed Ouerghi<br>Director of QA/RA<br>Vision Quest Industries, Inc.<br>Phone 760-477-8201<br>Mobile 760-691-0168<br>Fax 760-727-5950<br>mouerghi@vqorthocare.com | | Date Summary Prepared: | 3/23/2015 | | Proprietary Name: | BioniCare Hand System, Model BIO-2000 | | Device Name and Classification: | Transcutaneous Electrical Nerve Stimulator<br>For pain relief, Class II, 21 CFR 882.5890, Product<br>Code NYN | | Predicate Devices: | BioniCare® Stimulator, Model BIO-1000, k052625<br>and k030332 | | Device Description: | The BioniCare Hand System, Model BIO-2000 is<br>portable, rechargeable, battery-operated, single<br>Channel device that utilizes a voltage regulated<br>output circuit to generate a spike-shaped<br>Monophasic pulse with adjustable amplitude of 0 -<br>15 volts peak and repeating at a single fixed<br>frequency of 100 ± 5 Hertz. The device consists of<br>electrodes, lead wires and a signal generator<br>(BioniCare® Stimulator). | | Statement of Intended Use: | The BioniCare Hand System, Model BIO-2000, is<br>intended for use by patients with rheumatoid and/or<br>osteoarthritis of the hand as: | | | ● An adjunctive therapy in reducing the level of<br>pain and stiffness associated with pain from<br>rheumatoid arthritis of the hand. | | | ● As an adjunctive therapy in osteoarthritis of the<br>hand to reduce the level of pain and stiffness<br>and to improve the function of the hand. | {5}------------------------------------------------ ### Clinical Study Results and Summary: The study was designed as a prospective, multi center, open-label study to evaluate the efficacy and safety of the BioniCare Hand System in the treatment of Osteoarthritis of the hand. Patients were entered from the investigating physicians practice, without advertising, if they were 18 years or older, had osteoarthritis of the hand that was symptomatic for at least 3 months with a pain score of at least 3 out of 10 on a VAS scale despite stable NSAIDs and/or analgesics for at least a month prior to study entry. Eighty two patients were enrolled, 66 females and 16 males, with a mean age of 64 and a range of 45 to 89 years. When both hands were effected the more symptomatic hand was designated the index hand to be treated. The primary outcome measure was osteoarthritis pain in the last 48 hours. Additional efficacy outcomes were pain in the study thumb in the last 48 hours, patient global assessment and the physicians global assessment. Efficacy assessment of function included the validated DASH functional assessment questionnaire and the traditional measures of strength, pinch force and grip strength as measured by a JAMAR Hand Assessment Kit which includes a squeeze (grip) dynamometer and a pinch dynamometer. Efficacy was expressed for each variable as the effect size. Effect sizes are generally recognized to be small if they are from 0.2-0.49, moderate from 0.5 to 0.99 and large if they are 1.0 or greater. In the intent-to-treat statistical analysis, after 8 weeks of treatment, the effect size for OA pain in the study hand in the past 48 hours was 1.3, for OA pain in the study thumb for the past 48 hours was 0.8, for patient global assessment was 1.2, and for physician global assessment was 1.1. Functional outcome was moderately improved by the DASH Score with an effect size of 0.5. Smaller but still significant effect sizes were seen for the additional functional outcomes of pinch force, effect size 0.4 and grip strength with an effect size of 0.3. A limitation of this trial is that it was an open label study which includes the likelihood of bias caused by placebo effect. This was partially compensated by maintaining patients on stable NSAIDs and/or analgesics for the month prior to and for the entire study. This makes it more difficult to show efficacy as the benefit must be in addition to that of NSAIDs and/or analgesics. For example meta-analytic studies have shown an effect size for acetaminophen of 0.21 (95% confidence interval 0.02-0.41) and an effect size for NSAIDs of 0.32 (0.24-0.39). It should be noted that these effect sizes were in comparison to placebo. However, it should also be noted that all of the trials required discontinuation of previous pharmacotherapy for 3 to 14 days prior to study entry. The majority of trials additionally required a pre-defined flare of symptoms when NSAID treatment was discontinued in the pretreatment washout. This pretreatment washout and flair requirement makes it easier to demonstrate effectiveness by any treatment. This was evidenced by the mete-analysis as when they removed the studies which excluded nonresponders to NSAID treatment the effect size for pain decreased to 0.23 (0.15-0.31). Thus, even the effect size for pinch force and grip strength (0.4 and 0.3 respectively) are respectable, particularly since multiple studies have demonstrated that surgery of the hand can improve pain but does not increase strength as measured by pinch force and grip strength. Open label clinical data was used to support the Substantial Equivalence of the BIO-2000 to the predicate. VQ OrthoCare conducted randomized clinical trials to support {6}------------------------------------------------ clearance of the BioniCare BIO-1000 for rheumatoid arthritis (RA) of the hand and Osteoarthritis (OA) of the knee. Given that the pathophysiological process of osteoarthritis of the knee is similar to the degenerative process that would occur in the hand, data was leveraged from the randomized controlled clinical trial of the knee. The device was previously cleared for Rheumatoid Arthritis of the hand which is a more destructive process (worst case scenario) when compared to OA of the hand. Hence, the clinical data presented in support of the BioniCare BIO-2000 was found to be adequate to support Substantial Equivalence of this device to the predicate. ### Substantial Equivalence The device is substantially equivalent to the predicate devices because all changes made are categorized as changes to Performance Specifications, Ergonomics of User Interface and/or Firmware as the following tables demonstrate. | 510(k) Number | K052625 | K142236 | |----------------------------------------------|--------------------------------------------|--------------------------------------------| | Device Name | BIO-1000 | BIO-2000 | | Manufacturer | BioniCare | VQ OrthoCare | | Power Source | One 9V battery | One 3.7V battery | | No. of Output Modes | 1 | 1 | | Channels | 2 | 1 | | Synchronous | Yes | N/A | | Reciprocal | No | N/A | | Computerized | No | Yes | | Software Provided | N/A | Yes - Embedded Firmware | | Constant Current | No | No | | Constant Voltage | Yes | Yes | | Channel Isolation as per AAMI NS4<br>3.2.3.2 | No | N/A | | Line Current Isolation | N/A | N/A | | Automatic Overload Trip | No | No | | Automatic No-Load Trip | Yes | Yes | | Patient Override Control | No | No | | Max Leakage Current | N/A | N/A | | Indicator Display | | | | Unit Functioning | Yes | Yes | | Low Battery | Yes | Yes | | Other | LCD panel displays all parameter settings. | LCD panel displays all parameter settings. | | Standards - AAMI NS4 | Yes - applicable parts | Yes - applicable parts | BIO-2000/BIO-1000 Equivalency Table {7}------------------------------------------------ | Timer Settings | No | No | |--------------------------------------------------------|-------------------------------|------------------------------------------| | Automatic Shut Off | N/A | N/A | | Frequency | 100 Hz. | 100 Hz. | | Waveform | Monophasic spike shaped pulse | Monophasic spike shaped pulse | | Voltage Output | 0-12 Volts | 0-12 Volts (0 - 15V mfr override option) | | Voltage Pulse Width (% of peak) | 1.8mS @ 10%; 0.64mS @ 50% | 1.8mS @ 10%; 0.64mS @ 50% | | Current Output Range (500 Ohms) | 0 - 24mA peak | 0 - 24mA peak (30mA option) | | Peak Charge (500 Ohm) | 20uC | 20uC (25uC optional) | | Average Power (500 Ohm) | .092W | .092W (0.115W optional) | | Minimum Electrode Size | >30cm2 | >30cm2 | | Peak Current Density | < 0.667uC/cm2 | < 0.667uC/cm2 (< 0.833uC/cm2 optional) | | Below AAMI NS4 Safe current limit (20uC + (0.8)(35t)uC | Yes (limit is 50.8uC) | Yes (limit is 50.8uC) | | Less than AAMI safe average power (0.25W/ cm2) | Yes: 0.003W/ cm2 | Yes: 0.003W/ cm2 (0.0036W/ cm2 optional) | | Area of stimulation | hand | hand | | Weight (with batteries) in grams | 136 | 72 | | Dimensions (mm) | 63.9 X 96.4 X 36.9 | 54.0 X 90.5 X 14.7 | | Construction | Molded ABS/PC housing | Molded ABS/PC housing | ## Summary of Non-Clinical Testing The device and accessories were tested thoroughly using standard test equipment and methods and in accordance with company SOP's. All testing verified conformance to specifications as provided in 510(k). The device is substantially equivalent to the predicate devices. The following table contains a list of some of the non-clinical testing performed. | Test or Standard | Name | Purpose | |---------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------| | IEC 60601-1:1998<br>+A1:1991 +A2:1995 | Medical electrical equipment — Part 1:<br>General requirements for safety and<br>Essential Performance | Safety and Essential<br>Performance | | IEC 60601-1-2:2001 | Medical electrical equipment. General<br>requirements for safety. Collateral | Electromagnetic<br>Compatibility | {8}------------------------------------------------ | +A1:2004 | standard. Electromagnetic<br>compatibility. Requirements and tests. | | |-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | IEC 60601-2-10:1987<br>1.0b<br>+A1:2001 | Medical electrical equipment. Part 2:<br>Particular requirements for the safety<br>of nerve and muscle stimulators | Safety and Essential<br>Performance | | IEC 60601-1-11:2010 | Medical electrical equipment. Part 1-<br>11: General requirements for basic<br>safety and essential performance –<br>Collateral Standard: Requirements for<br>medical electrical equipment and<br>medical electrical systems used in the<br>home healthcare environment | Safety and Essential<br>Performance | | P10-04-22R | Design Hardware Verification Report | Bench Verification and<br>Functional Testing for<br>substantial equivalence and<br>or verification of<br>performance enhancement | | P10-04-20R | Design Validation Report | Voice of the Customer | ### Summary The BioniCare Hand System stimulator generates a waveform with proven benefit that is substantially equivalent to the waveform provided by the predicate devices. This waveform used has been demonstrated in predicated devices to be effective as an adjunctive therapy in Osteoarthritis to reduce the level of pain and stiffness. The BioniCare Hand System delivers this same waveform in an effective manner to the hand.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...