← Product Code [NUH](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH) · K150386

# Tyece OTC TENS Device (K150386)

_Tyece , Ltd. · NUH · Nov 5, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K150386

## Device Facts

- **Applicant:** Tyece , Ltd.
- **Product Code:** [NUH](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH.md)
- **Decision Date:** Nov 5, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Tyece OTC TENS Device, Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household activities.

## Device Story

Tyece OTC TENS Device (Model SEM44) is a dual-channel, battery-operated (3 AAA) transcutaneous electrical nerve stimulator. Device delivers electrical pulses to muscles via two pairs of electrode pads to provide temporary pain relief. User interface includes a full-screen display for selecting 15 pre-programmed modes, adjusting intensity (0-50), and setting timers (5-100 minutes). Device is intended for use in clinics, hospitals, and home environments. Healthcare providers or patients use the device to manage muscle pain; output affects clinical decision-making by providing non-pharmacological pain management. Benefits include portable, adjustable pain relief for muscle strain.

## Clinical Evidence

Bench testing only. No clinical data was conducted or relied upon. Device performance was validated through testing against recognized consensus standards including IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, and IEC 60601-1-11. Biocompatibility of electrode pads was established via reference to K132588.

## Technological Characteristics

Dual-channel TENS device; ABS plastic housing; 4.5V DC power (3 AAA batteries). Delivers biphasic square-wave pulses; frequency 1-150Hz; pulse duration 50-360μS. Connectivity: Standalone. Software: Microprocessor-controlled with moderate level of concern. Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 60601-1-11. Sterilization: Not applicable (electrodes are patient-contacting components).

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Counter TENS Device, Model PTS-IV ([K132993](/device/K132993.md))

## Reference Devices

- Electrode Pads ([K132588](/device/K132588.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, with three curved lines above them, possibly representing the flow of information or services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 5, 2015

Tyece Ltd % Guenter Ginsberg Consultant Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, Florida 34135

Re: K150386

Trade/Device Name: Tyece OTC TENS Device, Model SEM44 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: April 4, 2015 Received: September 30, 2015

Dear Mr. Ginsberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J.Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150386

Device Name Tyece OTC TENS Device, Model SEM44

Indications for Use (Describe)

The Tyece OTC TENS Device, Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household activities.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY (revised 6.30.2015)

#### Date of Summary Prepared: 02/05/2015

#### 1. Submitter's Name: Tyece Ltd.

Unit 801, Block A, Po Lung Centre, 11 Wang Chiu Rd., Address: Kowloon Bay, Kowloon, Hong Kong, SAR

Contact person: Parshid Falahati

Tel: 852- 2 - 3497456

Fax: 852 -2 - 3499166

E-mail sales@tyece.com

#### 2. Proposed New Device:

Trade Name: Tyece OTC TENS Device, Model SEM44 Classification Name: TENS Regulation Number: 882.5890 Product Code: NUH Device Class: II

#### 3. Predicate (cleared) Device:

Counter TENS Device, Model PTS-IV

510(k) Number: K132993

Manufacturer: Counter Scientific Development Ltd.

{4}------------------------------------------------

### 4. Description of Proposed Device:

The Tyece OTC TENS Device, Model SEM44 is a dual channel TENS device operated by DC 4.5V (3 AAA batteries). It is made up of one main unit, two electrode cables, and two pairs of electrode pads. There are 12 selectable, pre-programmed output waveforms (modes) to choose from, blus 3 programmable modes. The two channels can be individually adjusted for intensity from 0 to 50. Running time can be selected from 5 to 100 minutes.

All parameters (Menu, Program Number, Impulse intensity for channels, Jow battery, key-lock, frequency, pulse width and timer) are displayed on a full screen display.

The only patient contacting components are the Electrode Pads. The patient contacting materials are FDA cleared under K132588.

The device has been tested to and meets the requirements of the following recognized consensus standards:

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1(2012). EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology)

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment - part 1-11: general requirement for basic safety and essential performance -- collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

#### 5. Indication for Use:

This Tyece OTC TENS Device. Model SEM44 is to be used for temporary relief of pain associated with sore and aching muscles in the lower back, arms, or legs due to strain from exercise or normal household work activities.

#### 6. Environment of Use:

Clinics. Hospitals and home environments

#### 7. Contraindications:

Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electrical shock, burns, electrical interference, or death.

{5}------------------------------------------------

Do not use this device together with a life-supporting medical electronic device such as an artificial heart or lung.

Do not use this device together with a body-worn medical electronic device such as an ECG.

# 8. Technological Characteristics Compared to the Predicate Device

The Tyece OTC TENS Device, Model SEM44 is substantially equivalent to the Counter healthcare product OTC TENS for muscle pain relief, Model PTS-IV, The function, performance, and intended use of the two devices are very similar. Both systems are designed to relief pain in the muscle areas. Outputs are almost the identical, each can produce square electronic pulses and positive-going, reverse and bidirectional waveform output. Each can produce an asymmetrical biphasic pulse with peak voltages up to 60/70 volts on 500 Ohms Joad. Both devices have selection modes to allow users to select programs and adjust intensity level by pressing the keypad on the main unit. Both devices are transiting pulses by the use of Electrode Pads. Both are portable devices. Technological characteristics, features, specifications, materials and intended uses of the new device are substantially equivalent to the quoted predicate device. The differences that exist between the devices are insignificant in terms of safety or effectiveness.

#### Differences:

1. Device SEM44 has 15 preset programs, while the predicate device has 7.

2. Intensity levels of SEM44 are selectable from 0 to 50, while the predicate device is from 0 to 16, however, the maximum intensity settings of both units produce very similar outputs to the electrode pads. The SEM44 model can be adjusted in finer increments.

3. All other parameters such as waveform, frequency, pulse width, output current and voltage, although not exactly similar to this particular predicate, they do not raise any new questions of safety or effectiveness.

Note: It is very difficult, if not impossible, to find predicate devices that completely match the specifications of the subject device.

# 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

Savia did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the SEM44 vs. the predicate. Non-clinical testing was performed in order to validate the design according to the company's specified design requirements, and to assure conformance with the following voluntary design standards:

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1 (2012), EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology) .. 1991 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 199

{6}------------------------------------------------

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment - part 1-11: general requirement for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

# a. EMC and electrical safety

IEC 60601-1:2005+Corr.1(2006)+Corr.2(2007) + AM1 (2012), EN 60601-1:2006/A1:2013 general requirements for basic safety and essential performance

IEC 60601-2-10 Edition 2.0 2012-08 basic safety and essential performance of nerve and muscle stimulators. (Neurology)

IEC 60601-1-2 Edition 2007-03 medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests. (General II (ES/EMC)

IEC 60601-1-11 First Edition: 2010 medical electrical equipment -- part 1-11: general requirement for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

# b. Biocompatibility

The electrode pads are the only body contacting parts. They have been FDA cleared for OTC under K132588. The patient contacting electrodes comply with ISO 10993 and are considered Class A ≤ 24 hrs contact.

# c. Software

Based upon the test results it was concluded that the software performs within specifications and is safe to the stated intended use. The software of this device carries a MODERATE level of concern. Since a permanent hazard analysis is implemented in the software development process, and due to the clear software architecture, it is believed that the test protocol sufficiently verifies the software's main functional operation.

{7}------------------------------------------------

# d. Cleaning

The cleaning instructions as described the Instruction Manual have been tested to be sufficient. Testing involved validation of the manual cleaning method as per the instructions. All testing concluded that that the Tyece OTC TENS Device, Model SEM44 can be cleaned by the use of the methods described in the Instruction Manual.

# 10. Conclusions:

The Tyece OTC TENS Device, Model SEM44 has the same intended use and similar technological characteristics as the predicate device. Moreover, bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and effectiveness as that of predicate device. In the other words, the differences do not affect the intended use and do not raise any new questions of safety or effectiveness or alter the fundamental scientific technology of the device. Thus, the Tyece OTC TENS Device, Model SEM44 is substantially equivalent to the predicate device.

a managar and consistence and consistence and consistence and consistences and consistences and consistences and consistences and consistences and consistences and consistenc

... .. . . . . . . . . . . . . . . . . . . . . . . .

{8}------------------------------------------------

#### Output Specifications:

| Parameter                                                                                             | Your device                                                                                                                 | Predicate Device                                  |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Mode or Program Name                                                                                  | PAIN THERAPY DEVICE,<br>SEM44                                                                                               | PAIN THERAPY DEVICE,<br>PTS-IV                    |
| Waveform (e.g. positive-going,<br>reverse, biphasic)                                                  | Biphasic                                                                                                                    | Positive-going, Reverse and<br>Biphasic           |
| Shape(e.g., spike, rectangular,<br>square ware)                                                       | Square ware                                                                                                                 | Square ware                                       |
| Maximum Output Voltage (Volts)<br>(+/-15%)                                                            | 70vp p@500 Ω<br>110vp p@2K Ω<br>190vp p@10K Ω                                                                               | 69.2vp p@500 Ω<br>86.4vp p@2K Ω<br>88.8vp p@10K Ω |
| Maximum Output Current (+/-15%)                                                                       | 86mA@500 Ω<br>23.3mA@2K Ω<br>3.75mA@10K Ω                                                                                   | 69.2mA@500 Ω<br>21.6mA@2K Ω<br>4.4mA@10K Ω        |
| Duration of primary Phase                                                                             | 0                                                                                                                           | 0                                                 |
| Pulse Duration                                                                                        | 50-360μS                                                                                                                    | 200μS                                             |
| Frequency                                                                                             | 1-150Hz                                                                                                                     | 1-136Hz                                           |
| For multi Symmetrical<br>Program Programs<br>waveforms Pulse duration<br>only:                        | N/A                                                                                                                         | N/A                                               |
| Net Charge (micro coulombs μC)<br>(per pulse) (if zero, State method<br>of achieving zero net charge) | 0.001μC@500 Ω                                                                                                               | 0.0013μC@500 Ω                                    |
| Maximum Phase Charge (mC)                                                                             | 0.0454mC@500 Ω                                                                                                              | 0.0135mC@500 Ω                                    |
| Maximum Current Density<br>(mA/cm2, r.m.s)                                                            | 0.790mA/cm2@500 Ω                                                                                                           | 3.270mA/cm2@500 Ω                                 |
| Maximum Average Current<br>(average absolute value), mA                                               | 16.0mA                                                                                                                      | 2.218mA                                           |
| Maximum Average Power Density,<br>(W/cm2)(using smallest electrode<br>conductive surface area)        | 0.00632W/cm2@500 Ω 1.0S                                                                                                     | 0.00116W/cm2@500 Ω 1.0S                           |
| Burst Mode(i.e.<br>pulse trains)                                                                      | (a) Pulses per burst 4<br>(b) Bursts per second 4/83Hz<br>(c) Burst duration<br>(seconds) 180μS<br>(d) Duty Cycle 35ms/60ms | 4/2<br>30/136Hz<br>800μS/400μS<br>24ms/54.4ms     |
| ON Time (seconds)                                                                                     | 2s                                                                                                                          | 0.5s                                              |
| OFF Time (seconds)                                                                                    | 2s                                                                                                                          | 0.5s                                              |
| Additional Features (specify, if<br>applicable)                                                       | N/A                                                                                                                         | N/A                                               |
| Electrode area:                                                                                       | 20.25sqcm x 4 (81sqcm)                                                                                                      | 21.16sqcm x 4(84.64sqcm)                          |
| Basic Unit Characteristics:                                                                           |                                                                                                                             |                                                   |
| Parameter                                                                                             | Your Device                                                                                                                 | Predicate Device                                  |
| 510(k) Number                                                                                         | K150386                                                                                                                     | K132993                                           |
| Device Name and See attached photo Program                                                            | PAIN THERAPY DEVICE<br>SEM44                                                                                                | PAIN THERAPY DEVICE<br>PTS-IV                     |
| Manufacturer                                                                                          | SAVIA LTD.                                                                                                                  | COUNTER Scientific Development (GZ) Limited       |
| Power Source                                                                                          | 4.5V(batteries, 3x1.5V AAA)                                                                                                 | 3.7v Lithium ( Power supply)                      |
| Method of Line Current Isolation                                                                      | N/A                                                                                                                         | N/A                                               |
| Patient Leakage Current                                                                               | N/A                                                                                                                         | N/A                                               |
| Normal Condition (µA)                                                                                 | N/A                                                                                                                         | N/A                                               |
| Single Fault Condition (µA)                                                                           | N/A                                                                                                                         | N/A                                               |
| Average DC current through electrodes when device is on but no pulse are being applied (µΑ)           | 0µA                                                                                                                         | 0µA                                               |
| Numbers of Output Modes                                                                               | 15 Modes(01-15)                                                                                                             | 7 modes(A,B,C,D,E,F,ALL)                          |
| Number Synchronous or Alternating?                                                                    | 2 Synchronous                                                                                                               | 2 Synchronous                                     |
| Channels Method of Channel Isolation                                                                  | PCB Insulation<br>Boost Isolation                                                                                           | Boost Isolation                                   |
| Regulated current or regulated voltage?                                                               | Regulated voltage                                                                                                           | Regulated voltage                                 |
| Software/firmware/microprocessor control?                                                             | Yes                                                                                                                         | Yes                                               |
| Automatic overload trip?                                                                              | No                                                                                                                          | No                                                |
| Automatic no-load trip?                                                                               | Yes                                                                                                                         | Yes                                               |
| Automatic shut off?                                                                                   | Yes                                                                                                                         | Yes                                               |
| Users override control?                                                                               | Yes                                                                                                                         | Yes                                               |
| Indicator On/off status?                                                                              | Yes                                                                                                                         | Yes                                               |
| display Low battery?                                                                                  | Yes                                                                                                                         | Yes                                               |
| Voltage/current level? Only voltage level display                                                     | Yes for voltage                                                                                                             | Yes for voltage                                   |
| Timer range (minutes)                                                                                 | 5-100min                                                                                                                    | 20-40 min                                         |
| Compliance with voluntary standards?                                                                  | Yes                                                                                                                         | Yes                                               |
| Compliance with 21CFR898-7?                                                                           | Yes                                                                                                                         | Yes                                               |
| Dimensions                                                                                            | L135xW65xT20mm                                                                                                              | L120xW70xT18mm                                    |
| Housing materials and construction                                                                    | ABS                                                                                                                         | ABS                                               |

:

6

: 多

.

:

{9}------------------------------------------------

#### Basic Unit Characteristics:

.

.

7

1 ---------------

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K150386](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K150386)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
