K141978 · Omron Healthcare, Inc. · NUH · Dec 3, 2014 · Neurology
Device Facts
Record ID
K141978
Device Name
MAXPOWER RELIEF
Applicant
Omron Healthcare, Inc.
Product Code
NUH · Neurology
Decision Date
Dec 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder or foot due to strain from exercise or normal household and work activities.
Device Story
Maxpower relief (PM3032) is a battery-operated, single-channel TENS device for OTC pain relief. It delivers biphasic, rectangular electrical pulses to muscles via electrode pads. The device features 9 output modes with adjustable voltage/current levels and timer settings. It is intended for use in clinics, hospitals, and home environments by adult patients. The device is controlled by an internal microprocessor that manages waveform characteristics and safety features, including an automatic no-load trip that stops stimulation if pads are not properly adhered. Users operate the device via an On/Off control and monitor status via an indicator display. The device provides symptomatic pain relief by stimulating muscles, potentially reducing discomfort from strain.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including electrical safety (AAMI ANSI ES60601-1, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), and usability testing. Biocompatibility of patient-contacting materials (Long Life Pads) was established via reference to K120516.
Technological Characteristics
Battery-operated (2x AAA) TENS device; single-channel; biphasic rectangular waveform; microprocessor-controlled. Complies with AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Patient-contacting materials (Long Life Pads) tested per ISO 10993-5 and ISO 10993-10. Dimensions: ~52x112x25mm; Weight: ~100g.
Indications for Use
Indicated for adults for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder, or foot due to strain from exercise or normal household and work activities. Contraindicated for use with pacemakers, implanted metallic/electronic devices, life-supporting medical electronic devices (e.g., artificial heart/lung, respirator), in the presence of monitoring equipment, with high-frequency surgical units, in close proximity to shortwave/microwave therapy equipment, or with another TENS device.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Inside the circle is a stylized image of a human figure, with the head, torso, and legs represented by curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
MAR 16 2015
Omron Healthcare, Inc. % Paul Dryden Consultant ProMedic. Inc. 24301 Woodsage Dr. Bonita Springs, Florida 34134
Re: K141978
> Trade/Device Name: Maxpower relief (PM3032) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: July 19, 2014 Received: October 21, 2014
Dear Mr. Dryden,
This letter corrects our substantially equivalent letter of December 3, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Paul Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Felipe Aquel -S
Carlos L. Peña, PhD, MS Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
#### Device Name
Maxpower relief (PM3032) Indications for Use (Describe)
> This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder or foot due to strain from exercise or normal household and work activities.
Environments of Use: Clinics, hospital and home environments
Patient Population: Adult
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
#### 510(k) Summary Page 1 of 9 1/30/2015
| Omron Healthcare, Inc.<br>1925 West Field Court<br>Lake Forest, IL 60045 USA | Tel - 847-247-5626<br>Fax- 847-680-6269 |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Official Contact: | Renee Thornborough - Director QA/RA |
| Proprietary or Trade Name: | PM3032 |
| Common/Usual Name: | Transcutaneous electrical nerve stimulator for pain<br>relief. |
| Classification Name/Code: | NUH - stimulator, nerve, transcutaneous,<br>over-the-counter<br>21CFR 882.5890<br>Class II |
| Device Name: | Maxpower relief |
| Predicate Devices: | K110068 - Omron - PM3030<br>K121757 - Healthmate International - Pro8AB |
Device Description:
The Maxpower relief (Model PM3032) is a small, battery operated TENS device for pain relief intended for OTC use. The device complies with AAMI ANSI ES60601-1, IEC 60601-1-2 and IEC 60601-2-10.
The output modes are intended for application to the following areas: lower back, arms, legs, shoulder or foot. The specifications of each mode will be discussed in greater detail later in this section.
Software controls all controls and indicators. Software controls waveform characteristics.
This device is intended for the relief of pain associated with sore or aching, muscles of the shoulder, lower back, arms, or legs due to strain from exercise or normal household and work activities.
The accessories include an electrode cord / cable and electrodes pads (Long Life) which are placed on the specific body part.
As above the device is battery powered there is no connection to AC mains supply nor connection to any other device.
{5}------------------------------------------------
### 510(k) Summary Page 2 of 9 1/30/2015
We have provided the information suggested in Draft Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use from April 5, 2010. A checklist in accordance with this guidance can be found in Section 18 of this submission
### Intended User OTC
# Patient Population
Adults
## Indications for Use:
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder or foot due to strain from exercise or normal household and work activities.
## Environment of Use:
Clinics, hospital and home environments
## Contraindications:
- Use with a pacemaker or other implanted metallic or electronic device. ●
- Use with a life-supporting medical electronic device such as an artificial heart or . lung or respirator
- Use in the presence of monitoring equipment (e.g. cardiac monitors, ECG alarms) ●
- Simultaneous connection of a patient to a high frequency surgical unit ●
- Operation in close proximity (e.g. 1m) to a shortwave or microwave therapy . equipment
- . Use with another TENS device
## Predicate Device Comparison:
The Maxpower relief (PM3032) was compared to the predicates PM3030- K110068 and Healthmate Pro-8B - K121757 in the device comparison Table 5.1 below. Detailed electrical characteristics are provided in Table 5.2.
{6}------------------------------------------------
#### Table 5.1 Device Comparison
:
| Model Name<br>510(k) Number | New Device<br>Maxpower relief (PM3032) | Predicate Device<br>PM3030<br>K110068 | Predicate Device<br>Pro-8AB<br>K121757 | Comment |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Manufacturer | OMRON HEALTHCARE | OMRON HEALTHCARE | Healthmate International | |
| Indications for use | This device is intended for the<br>relief of pain associated with<br>sore or aching muscles of the<br>lower back, arms, legs,<br>shoulder or foot due to strain<br>from exercise or normal<br>household and work activities. | This device is intended for<br>the relief of pain<br>associated with sore or<br>aching, muscles of the<br>lower back, arms, or legs<br>due to strain from exercise<br>or normal household and<br>work activities. | To be used for temporary<br>relief of pain with sore<br>and aching muscles in the<br>shoulder, waist, back,<br>neck upper<br>extremities(arm), and<br>lower extremities(leg).<br>due to strain from exercise<br>or normal household work<br>activities. | Substantially<br>equivalent |
| Prescriptive | No, OTC | No, OTC | No, OTC | Identical |
| Power Source(s) | two AAA alkaline batteries | two AAA alkaline batteries | Rechargeable battery | PM3030 and<br>PM3032 identical |
| - Method of Line Current Isolation | N.A. not line powered | N.A. not line powered | N.A. not line powered | Identical all battery<br>powered |
| Patient Leakage Current - Normal<br>Condition (uA) | 1 | <0.1 | Not specified | All three device<br>comply with<br>60601-1 leakage<br>current<br>requirements |
| Patient Leakage Current - Single Fault<br>Condition (uA) | 8.9 max with mains<br>(250vAC/60Hz) | N.A | Not specified | All three device<br>comply with<br>60601-1 leakage<br>current<br>requirements |
{7}------------------------------------------------
| 510(k) Summary<br>Page 4 of 9<br>1/30/2015 | | | | | |
|---------------------------------------------------------------------------------------------------|----------------------------------------|---------------------------------------|----------------------------------------|--------------------------------------------------------------------------------------------------|--|
| Model Name<br>510(k) Number | New Device<br>Maxpower relief (PM3032) | Predicate Device<br>PM3030<br>K110068 | Predicate Device<br>Pro-8AB<br>K121757 | Comment | |
| Average DC current through electrodes<br>when device is on but no pulse are<br>being applied (uA) | 0 (uA) | 0 (uA) | 0 (uA) | Identical | |
| Number of output Modes | 9 | 3 | 8 | | |
| Number of output<br>channels | 1 ch | 1 ch | 2 ch | PM3030 and<br>PM3032 identical | |
| Method of Channel<br>Isolation | None, single channel | None, single channel | Not specified | PM3030 and<br>PM3032 identical | |
| Regulated Current or Regulated Voltage | Regulated Current | Regulated Current | Not specified | PM3030 and<br>PM3032 identical | |
| Software/Firmware/Microprocessor<br>Control | Microprocessor | Microprocessor | Not specified | PM3030 and<br>PM3032 identical | |
| Automatic Overload Trip | No | No | Not specified | PM3030 and<br>PM3032 identical | |
| Automatic No-Load Trip | Yes | No | No | PM3032 contains<br>circuit which will<br>stop stimulation if<br>pads are not<br>properly adhered | |
| Automatic shut Off | Yes | Yes | Yes | Identical | |
| User over ride control | User activated On/Off | User activated On/Off | User activated On/Off | Identical | |
,
{8}------------------------------------------------
| Model Name<br>510(k) Number | New Device<br>Maxpower relief (PM3032) | Predicate Device<br>PM3030<br>K110068 | Predicate Device<br>Pro-8AB<br>K121757 | Comment | |
|-------------------------------------|----------------------------------------|------------------------------------------------------------------|----------------------------------------|---------------------------------------------------|-----------------------------------------------------------------------|
| Indication display | ON/Off status | Yes | Yes | Yes | Identical |
| | Low Battery | Yes | No | Yes | PM3032 and<br>Pro-8AB identical |
| | Voltage/Current<br>Level | Yes | Yes | Yes | Identical |
| Timer Range (minutes) | | 15 | 15 | 10-60 | |
| Compliance with Voluntary standards | | ES 60601-1, IEC 60601-1-2,<br>IEC 60601-2-10, IEC 60601-<br>1-11 | IEC 60601-1, IEC 60601-<br>1-2 | IEC 60601-1, IEC<br>60601-1-2, IEC 60601-<br>2-10 | PM3032 complies<br>with currently<br>recognized version of<br>60601-1 |
| Compliance with 21 CFR 898 | | Yes (by compliance with<br>8.5.2.3 of ES 60601-1) | Yes | Not specified | PM3032 and PM3030<br>identical |
| Weight | | Approx. 100g (incl. batteries) | Approx. 60g (incl.<br>batteries) | Approx. 54g (incl.<br>batteries) | Similar |
| Dimensions (W x H x D) | | 52(W)x112(H)x25(D)mm | 55(W)x95(H)x18(D) | 50(W)x93(H)x10(D) | Similar |
| Patient contact | | Long Life Pads | Long Life Pads | Not specified | Identical electrode<br>pads. The Long Life<br>Pads (K120516) |
{9}------------------------------------------------
# Table 5.2 Detailed Electrical Comparisons Maxpower relief (PM303) Compared on Predicate Pro-8AB K121757
Press Presses Previous Max Previous Max power
| Parameter | Maxpower relief (P13032) | | | | | | | | Predicate Pro-8AB K121757 | | | | | | | | | |
|--------------------------------------------------|------------------------------------|-------------|-------------|-------------|-------------|-------------|-------------|-----------------|---------------------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|-------------|------|
| Mode of Program Name | TAP | KNEAD | RUD | ARM | LBACK | LEG | FOOT | JOINT | SHLDR | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum<br>Output<br>Voltage [V] | @500Ω<br>66.3 | 51.1 | 42.3 | 66.3 | 66.3 | 66.3 | 66.3 | 66.3 | 66.3 | 43.9 | 54.0 | 85.8 | 41.9 | 42.0 | 83.9 | 44.0 | 85.9 | |
| | @2kΩ<br>87.6 | 67.9 | 55.9 | 87.9 | 87.9 | 87.9 | 87.9 | 87.9 | 87.9 | 81.9 | 102.0 | 145.9 | 77.8 | 79.9 | 145.6 | 83.9 | 145.4 | |
| | @10kΩ<br>95.9 | 73.9 | 59.9 | 95.9 | 95.9 | 95.9 | 95.9 | 95.9 | 95.9 | 146.4 | 159.9 | 159.8 | 137.0 | 141.7 | 158.1 | 147.8 | 159.9 | |
| Maximum<br>Output<br>Current [mA] | @500Ω<br>132.6 | 102.2 | 84.6 | 132.6 | 132.6 | 132.6 | 132.6 | 132.6 | 132.6 | 87.8 | 108.0 | 171.6 | 83.8 | 84.0 | 167.8 | 88.0 | 171.8 | |
| | @2kΩ<br>43.8 | 34.0 | 28.0 | 44.0 | 44.0 | 44.0 | 44.0 | 44.0 | 44.0 | 41 | 51.0 | 73.0 | 38.9 | 40.0 | 72.8 | 42.0 | 72.7 | |
| | @10kΩ<br>9.6 | 7.4 | 6.0 | 9.6 | 9.6 | 9.6 | 9.6 | 9.6 | 9.6 | 14.6 | 16.0 | 16.0 | 13.7 | 14.2 | 15.8 | 14.8 | 16.0 | |
| Duration of primary phase [usec] | 100 | 150 | 150 | 100 | 150 | 100 | 100 | 100 | 100 | 90 | 90 | 90 | 90 | 90 | 90 | 90 | 90 | |
| Pulse Duration [usec] | 100 | 150 | 150 | 100 | 150 | 100 | 100 | 100 | 100 | 90 | 90 | 90 | 90 | 90 | 90 | 90 | 90 | |
| Frequency [Hz] | 1-15.43 | 25.28-79.22 | 84.86-132 | 1-132 | 1-237.6 | 1-19.16 | 1-19.16 | 42.43-<br>237.6 | 1-19.16 | 59 | 12-47 | 1.7 | 95 | 94 | 1.7-59 | 59 | 1.7-94 | |
| For<br>multiphasic<br>waveforms<br>only: | Symmetric<br>al phases | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| | Phase<br>Duration | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| Net Charge(μC per pulse)<br>(@500Ω) [μC] | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 2 | 3 | 4 | 5 | 6 | |
| Maximum Phase Charge<br>(@500Ω) [μC] | 13.26 | 15.33 | 12.69 | 13.26 | 19.89 | 13.26 | 13.26 | 13.26 | 13.26 | 7.9 | 9.72 | 15.44 | 7.54 | 7.56 | 15.10 | 7.92 | 15.46 | |
| Maximum Current Density<br>(@500Ω) [mA/cm²] | 3.03 | 2.33 | 1.93 | 3.03 | 3.03 | 3.03 | 3.03 | 3.03 | 3.03 | 3.38 | 4.16 | 6.62 | 3.23 | 3.24 | 6.47 | 3.39 | 6.62 | |
| Maximum Average Current<br>(@500Ω) [mA] | 0.17 | 0.37 | 0.71 | 0.83 | 0.38 | 0.43 | 0.17 | 0.22 | 0.88 | 0.47 | 0.46 | 0.03 | 0.72 | 0.71 | 0.89 | 0.47 | 1.45 | |
| Maximum Average Power<br>Density (@500Ω) [W/cm²] | 0.00017 | 0.00020 | 0.00039 | 0.00060 | 0.00015 | 0.00040 | 0.00016 | 0.00015 | 0.00082 | 0.00079 | 0.00095 | 0.00004 | 0.00116 | 0.00115 | 0.002 | 0.00079 | 0.00142 | |
| Burst<br>Mode | (a) Pulses per<br>burst | 46 | 71 | 118 | 23 | 1425 | 57 | 57 | 1425 | 76 | 159 | 940 | N/A*** | 66 | 1786 | 940 | 295 | 1767 |
| | (b) Bursts per sec | 0.33 | 0.59 | 0.59 | 1.00 | 0.14 | 0.25 | 0.25 | 0.14 | 0.20 | 0.24 | 0.05 | N/A*** | 0.87 | 0.04 | 0.05 | 0.15 | 0.05 |
| | **(c) Burst<br>duration | 3 | 8.5 | 17 | 8 | 14 | 8 | 8 | 14 | 10 | 4.2 | 21.1 | N/A*** | 1.15 | 23 | 21.17 | 6.6 | 20.5 |
| | (d) Duty cycle:<br>Line(b)xLine(c) | 1 | 5 | 10 | 8 | 2 | 2 | 2 | 2 | 2 | 1 | 1 | N/A*** | 1 | 1 | 1 | 1 | 1 |
| ON Time (seconds) | 3 | 0.9 | 0.9 | 0.3 | 6 | 3 | 3 | 6 | 4 | 2.7 | 20 | N/A*** | 0.7 | 19 | 20 | 5 | 18.8 | |
| OFF Time (seconds) | 0 | 0.8 | 0.8 | 0.7 | 1 | 1 | 1 | 1 | 1 | 1.5 | 11 | N/A*** | 0.45 | 4 | 1.17 | 1.6 | 1.7 | |
| Additional Features | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
{10}------------------------------------------------
## Output Specifications Maxpower relief (PM3032) Compared to Predicate PM3030 K110068
| Parameter | Maxpower relief (PM3032) | | | | | | | Predicate PM3030 K110068 | | | | | |
|--------------------------------------------------|------------------------------------|-------------|-------------|-------------|-------------|-------------|-------------|--------------------------|-------------|-------------|-----------------|---------------|---------|
| Mode of Program Name | TAP | KNEAD | RUB | ARM | LBACK | LEG | FOOT | JOINT | SHLDR | Arm Mode | Lower Back Mode | Leg/foot Mode | |
| Waveform | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic | |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular | |
| Maximum Output<br>Voltage [V] | @500Ω | 66.3 | 51.1 | 42.3 | 66.3 | 66.3 | 66.3 | 66.3 | 66.3 | 66.3 | 34.4 | 33.0 | 34.2 |
| | @2kΩ | 87.6 | 67.9 | 55.9 | 87.9 | 87.9 | 87.9 | 87.9 | 87.9 | 87.9 | 46.0 | 43.2 | 45.2 |
| | @10kΩ | 95.9 | 73.9 | 59.9 | 95.9 | 95.9 | 95.9 | 95.9 | 95.9 | 95.9 | 49.6 | 48.0 | 50.4 |
| Maximum Output<br>Current [mA] | @500Ω | 132.6 | 102.2 | 84.6 | 132.6 | 132.6 | 132.6 | 132.6 | 132.6 | 132.6 | 68.8 | 66.0 | 68.4 |
| | @2kΩ | 43.8 | 34.0 | 28.0 | 44.0 | 44.0 | 44.0 | 44.0 | 44.0 | 44.0 | 23.0 | 21.6 | 22.6 |
| | @10kΩ | 9.6 | 7.4 | 6.0 | 9.6 | 9.6 | 9.6 | 9.6 | 9.6 | 9.6 | 5.0 | 4.8 | 5.0 |
| Duration of primary phase [μsec] | | 100 | 150 | 150 | 100 | 150 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| Pulse Duration [μsec] | | 100 | 150 | 150 | 100 | 150 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| Frequency [Hz] | | 1 - 15.43 | 25.28-79.22 | 84.86-132 | 1 - 132 | 1 - 237.6 | 1 - 19.16 | 1 - 19.16 | 42.43-237.6 | 1 - 19.16 | 2 - 51.65 | 2 - 108 | 2 - 11 |
| For multiphasic<br>waveforms only | Symmetrical phases | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| | Phase Duration | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| Net Charge(μC per pulse) (@ 500Ω) [μC] | | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Maximum Phase Charge (@500Ω) [μC] | | 13.26 | 15.33 | 12.69 | 13.26 | 19.89 | 13.26 | 13.26 | 13.26 | 13.26 | 6.88 | 6.60 | 6.84 |
| Maximum Current Density (@500Ω) [mA/cm²] | | 3.03 | 2.33 | 1.93 | 3.03 | 3.03 | 3.03 | 3.03 | 3.03 | 3.03 | 1.57 | 1.51 | 1.56 |
| Maximum Average Current (@500Ω) [mA] | | 0.17 | 0.37 | 0.71 | 0.83 | 0.38 | 0.43 | 0.17 | 0.22 | 0.88 | 0.29 | 0.58 | 0.07 |
| Maximum Average Power Density (@500Ω)<br>[W/cm²] | | 0.00017 | 0.00020 | 0.00039 | 0.00060 | 0.00015 | 0.00040 | 0.00016 | 0.00015 | 0.00082 | 0.00018 | 0.00035 | 0.00005 |
| Burst Mode | (a) Pulses per burst | 46 | 71 | 118 | 23 | 1425 | 57 | 57 | 1425 | 76 | 180 | 540 | 33 |
| | (b) Bursts per second | 0.33 | 0.59 | 0.59 | 1.00 | 0.14 | 0.25 | 0.25 | 0.14 | 0.20 | 0.11 | 0.10 | 0.17 |
| | **(c) Burst duration | 3 | 8.5 | 17 | 8 | 14 | 8 | 8 | 14 | 10 | 18 | 20 | 6 |
| | (d) Duty cycle:<br>Line(b)xLine(c) | 1 | 5 | 10 | 8 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 1 |
| ON Time (seconds) | | 3 | 0.9 | 0.9 | 0.3 | 6 | 3 | 3 | 6 | 4 | 8 | 10 | 6 |
| OFF Time (seconds) | | 0 | 0.8 | 0.8 | 0.7 | 1 | 1 | 1 | 1 | 1 | 1 | 0 | 0 |
| Additional Features | | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
(**): Burst duration = pulse train duration
(***): Burst duration = pulse train duration
(***): Endurande model Automode I doesn't have a burst mode
(*) burn duration = pulse train duration
(***): Endurmeds model Automode I doesn't have a built like those.
{11}------------------------------------------------
### 510(k) Summary Page 8 of 9 1/30/2015
### Differences Between Other Legally Marketed Predicate Devices:
The Omron Maxpower relief (PM3032) is viewed as substantially equivalent to the predicate devices because: The Maxpower relief (PM3032) uses the exact same technology and has substantially equivalent indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness.
### Indications -
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, legs, shoulder or foot due to strain from exercise or normal household and work activities. These indications are virtually identical to the predicates.
Prescriptive - The Maxpower relief (PM3032) and predicates are all OTC.
Design and Technology - The Maxpower relief (PM3032) has equivalent design and features when compared to the predicates and has the identical technology to the predicate PM3030.
Performance and Specifications - The Maxpower relief (PM3032) has equivalent specifications of performance when compared to the predicates.
Compliance with standards - The predicate devices declare compliance with IEC 60601-1 and IEC 60601-1-2. The PM3032 complies with AAMI ANSI ES6060-1 (which replaced IEC 60601-1) and IEC 60601-1-2. The Pro-8AB and PM3032 comply with IEC 60601-2-10. Additionally the PM3032 complies with IEC 60601-1-11 for home healthcare.
### Materials -
The patient contacting materials of the Maxpower relief (PM3032) are the Long Life Pads. The Long Life Pads and were tested and the data presented in K 120516.
- Irritation and delayed-type hypersensitivity Extract in accordance with ISO 10993-10 .
- Closed-patch test for delayed hypersensitivity Rabbits in accordance with ISO . 10993-10
- Cytotoxicity in accordance with ISO 10993-5
### Patient Population -
The Maxpower relief (PM3032) and predicates are indicated for adults
Environment of Use - Clinics, hospital and home environments. Identical to the predicate K110068
### Non-Clinical Testing Summary:
The device has been tested to insure that all requirements have been met, this includes:
- . Testing of all controls
{12}------------------------------------------------
### 510(k) Summary Page 9 of 9 1/30/2015
- Testing of all indicators .
- . Testing of battery state indicators
- Testing of waveforms ●
The device has also been tested to the requirements of the following standards:
- AAMI / ANSI ES60601-1:2005 + A1: 2012 Medical electrical equipment part 1: . general requirements for basic safety and essential performance
- . IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility -Requirements and Tests
- . IEC 60601-1-11: 2010, medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-10: 2012 Medical electrical equipment part 2-10: particular . requirements for the basic safety and essential performance of nerve and muscle stimulators
The device has also been tested for drop, vibration and environmental temperature and humidity
Usability testing has also been performed.
### Clinical Testing Summary:
No clinical testing was performed
### Substantial Equivalence Conclusion
Omron maintains that the Maxpower relief (PM3032)is substantially equivalent to the predicate device in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.