← Product Code [NUH](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH) · K133789

# DOLPHIN NEUROSTIM OTC (K133789)

_Acumed Medical , Ltd. · NUH · Mar 4, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K133789

## Device Facts

- **Applicant:** Acumed Medical , Ltd.
- **Product Code:** [NUH](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH.md)
- **Decision Date:** Mar 4, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the Dolphin Neurostim OTC is to deliver transcutaneous electrical nerve stimulation (TENS).

## Device Story

Dolphin Neurostim OTC is a handheld, 9V battery-powered TENS device for temporary muscular pain relief. It features a metal probe tip (cathode or anode) and an ancillary wrist strap ground. Operation includes two modes: a search mode to locate points of low resistance in a predetermined area and a treatment mode that applies low-frequency (2.4-3.0 Hz) Direct Current to the skin. The device is intended for over-the-counter use by patients. By delivering electrical stimulation to the skin, the device aims to provide pain relief for sore or aching muscles. It is a non-programmable device providing a pulsed biphasic waveform.

## Clinical Evidence

Bench testing only. Factory tests were conducted to verify signal accuracy and maintenance of calibrated energy patterns over the device's useful life. No clinical data provided.

## Technological Characteristics

Handheld, 9V battery-powered TENS device. Materials include a plastic case with metal electrodes. Operates via point stimulation using a metal probe tip and wrist strap ground. Delivers a low-frequency (2.4-3.0 Hz) Direct Current in treatment mode and a pulsed biphasic waveform. Maximum average power density 0.116 W/sq cm; maximum current density 0.21 A/sq cm; maximum charge per pulse 0.00044 coulombs.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Painmaster MCT Patch ([K130114](/device/K130114.md))
- TENS Pro 900 ([K023726](/device/K023726.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them. The profiles are stacked on top of each other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 4, 2015

Acumed Medical, LTD. c/o Mr. Richard Keen Compliance Consultants 1151 Hope Street Stamford. CT 06907-1659

Re: K133789

Trade/Device Name: Dolphin Neurostim OTC Regulation Number: 21 CFR 890.5890 Regulation Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Regulatory Class: Class II Product Code: NUH Dated: January 31, 2015 Received: February 4, 2015

Dear Mr. Keen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

## Felipe Aguel -S

Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K133789

Device Name Dolphin Neurostim™ OT C

Indications for Use (Describe)

The Dolphin Neurostim™ OT C is indicated for temporary relief of pain associated with sore and aching muscles in the back, arms, and legs due to strain from exercise or normal household and work activities.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 5. 510(k) Summary

(per 21 CFR 807.92)

#### 3 March 2015 Sponsor

Mr. Bruce Hocking President Acumed Medical Ltd. 3679 Lakeshore Blvd. W. Toronto, ON M8W 1P7 800-567-7246 416-253-6060 416-253-1911 Fax acumedmedical(@yahoo.com

#### Consultant

Mr. Richard Keen Compliance Consultants 1151 Hope Street Stamford, CT 06907-1659 203 329 2700 F 203 329 2345 rkeen@fda-complianceconsultants.com

| Proprietary Name:           |
|-----------------------------|
| Common Name:                |
| Device Classification Name: |
| Classification Number:      |

Product Code Reviewing Group Device Classification Establishment registration No. Predicate Device

Dolphin Neurostim tm OTC TENS device Transcutaneous Electrical Nerve Stimulator for Pain Relief 21 CFR 882.5890 NUH Neurological Therapeutic Devices Class II new submission Painmaster MCT Patch K130114 and TENS Pro 900 K023726

Trademark Notice: All Trademarks used other than those of ACUMED Medical are registered to their respective owners.

#### Device Description

The Dolphin Neurostim OTC is designed to perform transcutaneous electrical nerve stimulation (TENS) and as such is a Class II device, having Regulation Number: 882.5890 and Classification Number: NUH. The Dolphin Neurostim™ OTC (DNS) is a handheld, battery powered (9V) Transcutaneous Electrical Nerve Stimulators (TENS), specifically designed for the temporary relief of muscular pain. It combines the electrical characteristics of TENS with point stimulation delivered through a metal probe tip (cathode or anode). There is a metal ground on the device and an ancillary wrist strap that plugs into the device (cathode or anode). The device has two modes of operation. In search mode it locates points of low resistance in a predetermined area. In treatment mode it applies a low-frequency (2.4-3.0 Hz) Direct Current to the skin.

#### Indications For Use

The Dolphin Neurostim™ OTC is indicated for temporary relief of pain associated with sore and aching muscles in the back, arms, and legs due to strain from exercise or normal household and work activities.

### Intended Use

The intended use of the Dolphin Neurostim OTC is to deliver transcutaneous electrical nerve stimulation (TENS).

{4}------------------------------------------------

# 5. 510(k) Summary

(per 21 CFR 807.92)

#### Substantial Equivalence

ACUMED Medical has determined that the Dolphin Neurostim, OTC is substantially equivalent to the performance of a predicate Device. The differences between these systems are incidental and not significant. Both devices use similar technological characteristics and principles.

- Both devices deliver TENS waveform to patients.
- both devices are sold over the counter.

The Dolphin Neurostim tm is an over the counter version to its prescription device counterpart that has been sold in the US for 20 years. The Dolphin Neurostim has been sold over the counter in other countries for over 20 years for a total of 60,000 units.

| Comparable Parameter                                          | Dolphin Neurostim                                                                                                                                                                                             | Painmaster MCT Patch,                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                               | K133789                                                                                                                                                                                                       | K 130114                                                                                                                                                                                                                                                                                                                                                                                              |
| Transcutaneous Electrical Nerve<br>Stimulator for Pain Relief | Yes                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                   |
| Intended Use                                                  | This device is intended for the relief of pain<br>associated with sore or aching muscles of the<br>lower back, arms or legs due to strain from<br>normal exercise or normal household and<br>work activities. | The Painmaster MCT Patch is indicated for<br>temporary relief of pain associated with sore<br>and aching muscles in the lower back due to<br>strain from exercise or normal household and<br>work activities. This intended use is a subset<br>of the intended use for the device when it was<br>cleared for prescription use, and is also<br>identical to that of other OTC cleared TENS<br>devices. |
| technological characteristics                                 | Stimulation current delivered to electrodes<br>applied to skin.                                                                                                                                               | Stimulation current delivered to electrodes<br>applied to skin.                                                                                                                                                                                                                                                                                                                                       |
| Material                                                      | Plastic case with electrodes constructed of<br>metal.                                                                                                                                                         | Plastic case with electrodes constructed of<br>metal.                                                                                                                                                                                                                                                                                                                                                 |
| output waveform design                                        | single non-programmable milliamp<br>mode, delivering a pulsed biphasic waveform<br>that provides electrical stimulation<br>to the body to relieve pain.                                                       | single non-programmable microcurrent<br>mode, delivering a pulsed monophasic<br>waveform that provides electrical stimulation<br>to the body to relieve pain.                                                                                                                                                                                                                                         |
| maximum average power density<br>watts/sq cm                  | 0.116                                                                                                                                                                                                         | Below safety limit                                                                                                                                                                                                                                                                                                                                                                                    |
| maximum current density amps/sq cm                            | 0.21                                                                                                                                                                                                          | Below safety limit                                                                                                                                                                                                                                                                                                                                                                                    |
| maximum charge per pulse coulombs                             | 0.00044                                                                                                                                                                                                       | Below safety limit                                                                                                                                                                                                                                                                                                                                                                                    |

### Testing

The Dolphin Neurostim™, OTC has benefited from design, development, testing and production procedures that conform to Quality Systems. Testing has confirmed this device meets its product specification. A series of factory tests are conducted to verify the intended signals are accurate and can maintain a calibrated energy pattern over its useful life. Testing has established this device is substantially equivalent to the predicates.

### Conclusion

There are no substantial differences between the Dolphin Neurostim OTC defined in this 510(k) submission and the stated predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K133789](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NUH/K133789)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
