Postmarket surveillance data (adverse event reports from Europe and Asia); Peer-reviewed clinical literature
The sponsor used postmarket adverse event data from international markets to support the safety of OTC use and cited peer-reviewed literature to support the effectiveness of electrical stimulation for the proposed indications.
Postmarket surveillance; Adverse event reporting; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
International Postmarket Surveillance; Postmarket surveillance / observational
Users of the device in Europe and Asia; Number of Sites: Not applicable
Not applicable for this study
Safety (absence of reported adverse events)
Literature Review; Literature review
General population receiving electrical stimulation; Number of Sites: Not applicable
Not applicable for this study
Muscle performance improvement and pain reduction
Indications for Use
TENS: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS: It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Pinook Stimulator (BH-18) is a portable, battery-powered (3.7VDC) device combining TENS and Powered Muscle Stimulator (PMS) functions. Device delivers programmed electrical pulses via two channels through adhesive electrode pads placed on the body. Features 6 operation modes. Used by patients in home settings without medical supervision. Output parameters (voltage, current, frequency, pulse width) are controlled to ensure safety for OTC use. Healthcare providers or patients use the device to manage muscle pain or enhance muscle performance. Benefits include non-pharmacological pain relief and muscle conditioning.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, compliance with voluntary safety standards (IEC 60601-1, 60601-1-2, 60601-2-10, 60601-1-11, ISO 14971), and software verification. Effectiveness is supported by literature references regarding electrical stimulation for pain reduction and muscle performance.
Indicated for individuals seeking temporary relief of muscle pain (shoulder, waist, back, neck, arms, legs) due to exercise or household strain, and for stimulation of healthy muscles to improve/facilitate muscle performance. OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
# K132563
# APR 2 4 2014
# Summary Traditional 510(K)
# SUBMITTER INFORMATION
| Submitter's Name: | Pinook USA |
|-----------------------------|---------------------------------------------------------------|
| Submitter's Address: | 901 Central Florida Pkw<br>Suite A6<br>Orlando, Florida 32824 |
| Contact Person: | Mr. Dvir Lev-Ran |
| Email: | dvirkush@yahoo.com<br>dvir@pinookusa.com |
| Phone: | 855.606.4303 |
| Fax: | 888.505.0681 |
| Date of Summary Submission: | July 29, 2013 |
| Resubmitting on: | March 24th, 2014 |
| 510(K) Number: | K132563 |
## NEW DEVICE FOR WHICH SUBMITTING:
| Device Trade Name: | Pinook Stimulator |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Model: | BH-18 |
| Device Common Name: | Transcutaneous Electrical Nerve Stimulator and Powered<br>Muscle Stimulator |
| Classification Name: | Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>Stimulator, Muscle, Powered, For Muscle Conditioning |
| Device's Classification Panel: | Neurology<br>Physical Medicine |
| Regulatory Class: | Class II |
| Product Code:<br>Regulation Number: | NUH, NGX<br>882.5890 |
{1}------------------------------------------------
### MANUFACTURER INFORMATION:
| Name: | JOHARI DIGITAL HEALTHCARE LTD. | | |
|--------------------------------|-------------------------------------------------------------------------------------------------------------|--|--|
| Address and Registration: | G-582 - 583, EPIP, Boranada, Jodhpur 342008 | | |
| FDA Registration: | 8040537 | | |
| Predicate Device: | | | |
| Device Trade Name: | Hi-Dow | | |
| Model: | JQ-5C | | |
| Classification Name: | Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>Stimulator, Muscle, Powered, For Muscle Conditioning | | |
| 510(K) Number: | K102598 | | |
| Device's Classification Panel: | Neurology (As Per 21 CFR Section 882.5890)<br>Physical Medicine (As Per 21 CFR Sections 890.5850) | | |
| Regulatory Class: | Class II | | |
| Product Code: | NUH, NGX | | |
| Regulation Number: | 882.5890 | | |
| Manufacturer: | Hi-Dow International, Inc | | |
| Address: | 2071 Congressional Drive, Saint Louis, MO 61346 | | |
## DESCRIPTION OF THE NEW DEVICE
#### BH - 18:
The BH - 18 is a portable; battery powered (3.7VDC) multi-function device offering both Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) qualities in one device.
Two channels effectively transfer your desired choice of programmed electrical pulses directly through electrode adhesive pads to the suggested area of the body where the electrodes are placed, there are 6 modes of operation.
{2}------------------------------------------------
#### INTENDED USE OF DEVICE
#### TENS:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
### PMS:
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
### SUMMARY OF SUBSTANTIAL EQUIVALENCE
Comparison of BH - 18 and the predicate JQ-5C.
| S.No | Description | Pinook Stimulator | Hi-Dow |
|------|----------------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|
| 1. | Max Output Voltage over 10k, V | 84 V @ 10KΩ<br>79 V @ 2.2KΩ<br>61 V @ 500Ω | 84 V @ 10KΩ<br>79.2 V @ 2.2KΩ<br>62.4 V @ 500Ω |
| 2. | Max- Current over 10k, mA | 8.4 mA @ 10KΩ<br>39.5 mA @ 2.2KΩ<br>122 mA @ 500Ω | 8.4 mA @ 10KΩ<br>39.6 mA @ 2.2KΩ<br>124.8 mA @ 500Ω |
| 3. | Pulse Width, micro seconds | 100 | 100 |
| 4. | Pulse Period, msec | 16.3-833Ms | 16.3-833mS |
| 5. | Max. Pulse Frequency, Hz | 62 | 61.3 |
| 6. | Net Charge µC per pulse | 0 | 0 |
| 7. | Max Phase Charge over 500Ω, µC | 12.2 | 12.48 |
| 8. | Max Current Density over 500Ω, mA/cm2 | 11.77 | 12.04 |
| 9. | Max Power Density over 500 Ω, W/cm² | 0.718 W/cm² | 0.747 W/cm² |
| 10. | For multiphasic waveforms only:<br>-Symmetrical phases?<br>- Phase Duration† (include units) | YES<br>100μS | YES<br>100μS |
| 11. | ON Time (seconds) | 5 Seconds (M2, M3 & M4) | 5 Seconds (M2, M3 & M4) |
| 12. | OFF Time (seconds) | 3 Sec. (M2),<br>2 Sec. (M3 & M4) | 3 Sec. (M2),<br>2 Sec. (M3 & M4) |
### SUBSTANTIAL EQUIVALENCE
The electrical stimulation provided by the BH-18 is substantially equivalent to that commonly employed by muscle stimulators and TENS devices that have been cleared for
{3}------------------------------------------------
marketing without prescription labelling: i.e. for OTC sale. The pulses in the waveform combinations are restricted in amplitude and duration and is consistent with the other device quoted above.
The differences that exist between the devices are insignificant in the terms of safety or effectiveness.
The BH -18 has modes that offer substantially equivalent technical specifications. features and effective results as the predicate listed.
#### NON-CLINICAL TESTING PERFORMED
Compliance to applicable voluntary standards includes IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10,IEC 60601-1-11 and ISO 14971.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA guidance for the content of premarket submissions for software contained in medical devices.
#### CONCLUSION
The electrical stimulation provided by the BH-18 is similar to the commonly employed muscle stimulators and TENS devices that have been cleared for marketing without prescription labelling.
The BH-18 has the same intended uses and the similar technological characteristics as its OTC predicate. Moreover, verification and validation tests contained in this submission demonstrate that the differences in BH-18 still maintain the same safety and effectiveness as that of the cleared device.
In other words, the engineering differences do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
Concerns of safe and proper use of electrodes and electrode pad placement have been fully addressed by making the user conscious of the proper placement of the electrodes and proper operations of the device through detail in the User's Instruction Manual.
There are no new safety or effectiveness issues concerning the new device.
{4}------------------------------------------------
The safety of the device, to be used for the proposed indications without medical prescriptions or supervision, is established by the fact that no adverse events have been reported for units sold without a prescription in Europe and Asia. This also proves that its specific technical, safety measures and features are safe and effective when used without medical supervision.
The effectiveness of the device for the proposed indications is supported by a number of articles in peer-reviewed publications, which demonstrates that electrical stimulation does improve muscle performance as well as pain reduction.
Technological characteristics, features, specifications, materials and intended uses of the BH-18 are substantially equivalent to the quoted predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 24, 2014
Pinook USA, LLC c/o Dvir Lev-Ran 901 Central Florida Parkway, Suite A6 Orlando, FL 32824
Re: K132563
Trade/Device Name: Pinook Stimulator, Model BH-18 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH, NGX Dated: March 24, 2014 Received: March 27, 2014
Dear Mr. Lev-Ran:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Mr. Dvir Lev-Ran
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña. Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K132563
#### Device Name
Pinook Stimulator, Model BH-18
#### Indications for Use (Describe)
TENS: To be used for temporary relief of pain associated with sore and acting muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities
PMS: To be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Type of Use (Select one or both, as applicable)
[ ] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY 1 31, 97 : , "+ . :-.
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Carlos EBena -S
* . …
. :
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed announg to and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.