To be used in adults only for temporary relief of pain associated with sore and aching muscles in the upper extremities (arms), lower extremities (legs), and lower back due to strain from exercise or normal household and work activities.
Device Story
Tanyx is a disposable, portable, battery-operated transcutaneous electrical nerve stimulator (TENS) for OTC use. Device adheres directly to skin; emits controlled monophasic, rectangular electrical signals via integrated electrode grid to reduce pain. User operates via power and intensity buttons. No medical supervision required. Output modes include fixed or modulated signals; burst mode available. Device provides temporary pain relief for sore/aching muscles in arms, legs, and lower back. Benefits include ease of use, portability, and non-invasive pain management.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993-1, 5, 10) passed. Device safety testing (IEC 60601-1, IEC 60601-2-10) passed. Hazard analysis conducted per ISO 14971:2007. Usability study confirmed users could operate the device correctly and safely.
Technological Characteristics
Materials: ABS and Polypropylene. Sensing/Actuation: Microprocessor-controlled TENS. Energy: Battery-powered. Form factor: 153mm x 51mm x 8.2mm, 30g. Output: Monophasic, rectangular waveform, 55 Hz frequency, 6 output modes, 1 channel. Connectivity: Standalone. Sterilization: Not specified.
Indications for Use
Indicated for adults for temporary relief of pain associated with sore/aching muscles in arms, legs, and lower back due to exercise or normal household/work strain. Contraindications not specified.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
K23866
#### 510(k) Summary
Date of Submission: November 28, 2012
Submitter: Medecell US, Inc. 4901 Boulder Lane Hoffman Estates, IL 60010-5842 +1-847-323-5599 Phone: Robert Da Rocha, International Director Contact Person:
#### Subject Device:
Tanyx® Proprietary Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Classification Name: Regulation: 882.5899 Product Code: nuh
SEP 1.3 2013
#### Predicate Devices:
| Name of device | Model PM3030 |
|----------------|-------------------|
| Manufacturer: | Omron Health Care |
| K number: | K110068 |
| Product Code: | NUH |
#### Device Description:
TANYX® is based on TENS medical technology. The product is to reduce or eliminate pain through the emission of controlled and focused electrical signals. It is a disposable device, designed to be sold directly to consumers for the temporary relief of pain. TANYX® adheres to a person's skin and emits either a fixed or modulated electrical signal through a grid of integrated electrodes. It is a product which is characterized as portable, inexpensive, disposable, and (compared to other TENS-based products) easy to handle and operate. No medical supervision is required.
#### Intended Uses:
"To be used in adults only for temporary relief of pain associated with sore and aching muscles in the upper extremities (arms), lower extremities (legs), and lower back due to strain from exercise or normal household and work activities."
{2}------------------------------------------------
### Summary of Technological Characteristics:
| | Model PM3030 | TANYX® |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturer: Omron Healthcare, Inc. | Manufacturer: Medecell do Brasil |
| | | K110068 |
| Indications for Use | This device is intended for the relief of pain<br>associated with sore or aching muscles of the<br>lower back, arms, or legs due to strain from<br>exercise or normal household and work<br>activities. | To be used for temporary relief of pain<br>associated with sore and aching muscles in the<br>upper extremities (arms), lower extremities<br>(legs), and lower back due to strain from exercise<br>or normal household and work activities. |
| Patient Population | Adults | Adults |
| Prescriptive or OTC | OTC | OTC |
| Environment Of use | Clinics, hospital and home environments | Clinics, hospital and home environments |
| Number Of Output modes | 3 | 6 |
| Number of output channels | 1 | 1 |
| Waveform | Biphasic | Monophasic |
| Shape | Rectangular | Rectangular |
| Maximum Output Voltage<br>(max) | | |
| 500 ohm | 35.4 V | 47.6 Vpp |
| 2k ohm | 46.7 V | 64.0 Vpp |
| 10k ohm | 50.8 V | 72.8 Vpp |
| Frequency (Hz) | 0-100 Hz | 55 Hz |
| Maximum Output Current<br>(max) | | |
| 500 ohm | 4.4 mA | 95.2 mApp |
| 2k ohm | 1.7 mA | 29.1 mApp |
| 10k ohmn | 0.4 mA | 7.9 mApp |
| Maximum Average Current | 3.5mA | 6.02mA |
| Maximum Current Density | 0.095mA/Cm² | 0.12mA/Cm² |
| Maximum Average Power<br>Density | 89mW/Cm² | 98mW/Cm² |
| Maximum Phase charge (500<br>ohm) | 1.33 µC | 0.19 µC |
| Burst Mode | none | yes |
| Timer range (min) | 15 minutes | No |
| Dimensions | 55 mm x 95 mm x 19mm | 153mm x 51mm x 8.2mm |
| Weight | 60 grams | 30 grams |
| Material | Acrylonitrile Butadiene Styrene (ABS) | Acrylonitrile Butadiene Styrene (ABS) and<br>Polypropylene (PP) |
| Microprocessor control | Yes | yes |
| Automatic Overload trip | Yes | No |
| Automatic no-load trip | Yes | Yes |
| Automatic shut-off | Yes | No |
| User control | Power On/Off button | Power On/Off and Intensity buttons |
| Electrode cable | Yes | No |
. .
{3}------------------------------------------------
#### Differences between Tanyx and Predicate Device
Tanyx is viewed as substantially equivalent to the predicate device because the stimulation parameters are all in the same range for the intended use. The device includes a wide variation in terms of Current, Voltage Amplitude and Waveform, the basic parameters, without sacrificing biological effects. Maximum Output Current differences between Tanyx and the predicate device are due to different measurement methods. Tanyx was measured using a peak-to-peak value and predicate device uses RMS value, but both devices are in the same range of safety and efficacy. Other measurement differences are not significant from the predicated device. Operation and mechanical design don't show significant differences.
#### Non-Clinical Performance Data:
| Characteristic | Predicate | Subject |
|-------------------|---------------|---------|
| | | |
| Biocompatibility | | |
| ISO 10993-1 | Passed | Passed |
| ISO 10993-5:2009 | Not specified | Passed |
| ISO 10993-10:2010 | Not specified | Passed |
| | | |
| Device Safety | | |
| IEC 60601-1 | Passed | Passed |
| IEC 60601-2-10 | Passed | Passed |
| | | |
| Hazard Analysis | | |
| ISO 14971:2007 | Not Specified | Passed |
#### Risk Analysis Method:
The hazard analysis for the device was conducted according to EN ISO 14971:2007. The device was found to adequately minimize the risks for shock and burn as identified by FDA.
#### Usability Study:
A usability study was conducted and showed that users were able to use the device correctly and safely.
#### Conclusion:
Tanyx® is substantially equivalent to a legally marketed device. Although it has some technological characteristics that differ from that of predicate, TENS current state of art support that stimulation parameters of both devices are in the same range for the intended biological effects and practical use and raises no new questions of safety and effectiveness.
{4}------------------------------------------------
## Summary of substantial equivalence
.
| | Model PM3030 | TANYX® |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturer: Omron Healthcare, Inc. | Manufacturer: Medecell do Brasil |
| | K110068 | |
| Indications for Use | This device is intended for the relief of<br>pain associated with sore or aching<br>muscles of the lower back, arms, or<br>legs due to strain from exercise or<br>normal household and work activities. | To be used for temporary relief of pain<br>associated with sore and aching muscles<br>in the upper extremities (arms), lower<br>extremities (legs), and lower back due to<br>strain from exercise or normal household<br>and work activities. |
| Patient Population | Adults | Adults |
| Prescriptive or OTC | OTC | OTC |
| Environment Of use | Clinics, hospital and home<br>environments | Clinics, hospital and home environments |
| Number Of Output modes | 3 | 6 |
| Number of output channels | 1 | 1 |
| Waveform | Biphasic | Monophasic |
| Shape | Rectangular | Rectangular |
| Maximum Output Voltage<br>(max) | | |
| 500 ohm | 35.4 V | 47.6 Vpp |
| 2k ohm | 46.7 V | 64.0 Vpp |
| 10k ohm | 50.8 V | 72.8 Vpp |
| Frequency (Hz) | 0-100 Hz | 55 Hz |
| Maximum Output Current<br>(max) | | |
| 500 ohm | 4.4 mA | 95.2 mApp |
| 2k ohm | 1.7 mA | 29.1 mApp |
| 10k ohmn | 0.4 mA | 7.9 mApp |
| Maximum Average Current | 3.5mA | 6.02mA |
| Maximum Current Density | 0.095mA/Cm² | 0.12mA/Cm² |
| Maximum Average Power<br>Density | 89mW/Cm² | 98mW/Cm² |
| Maximum Phase charge<br>(500 ohm) | 1.33 µC | 0.19 μC |
| Burst Mode | none | yes |
| Timer range (min) | 15 minutes | No |
| Dimensions | 55 mm x 95 mm x 19mm | 153mm x 51mm x 8.2mm |
| Weight | 60 grams | 30 grams |
| Material | Acrylonitrile Butadiene Styrene (ABS) | Acrylonitrile Butadiene Styrene (ABS) and<br>Polypropylene (PP) |
| Microprocessor control | Yes | yes |
| Automatic Overload trip | Yes | No |
| Automatic no-load trip | Yes | Yes |
| Automatic shut-off | Yes | No |
| User control | Power On/Off button | Power On/Off and Intensity buttons |
| Electrode cable | Yes | No |
·
.
.
・
:
:
{5}------------------------------------------------
Qur product Tanyx was tested by the company "No Risk" from Brazil resulting the report R120331 - Medecell - Stimulator - Tanyx - parameters - 12042710 . The numbers collected from this document demonstrate that Tanyx is in accordance to the limits established to the safety of the differences between Tanyx and the other device, we believe do not raise any new question regarding safety and effectiveness because both are under the established limits.
In reference to the MAXIMUM OUTPUT VOLTAGE, our product have numbers under established limits within industry standards and the differences between the devices do not raise any new question of safety and effectiveness.
Regarding the MAXIMUM OUTPUT CURRENT, Tanyx again have our numbers under the established limits and the differences between the devices do not raise any new question of safety and effectiveness because both products are in accordance to industry standards.
Referent to the MAXIMUM PHASE CHARGE, the number for Tanyx is over the established limit and again, the difference between the devices do not raise any question regarding safety and effectiveness because both are under limits.
Regarding the MAXIMUM CURRENT DENSITY of Tanyx, the product is under the maximum limits and because of this do not raise any new question of safety and effectiveness. The differences between the devices do not raise any new question too, because both are under the maximum limits established.
In reference to MAXIMUM AVERAGE CURRENT, Tanyx is under the established maximum limits. The differences between the products do not raise new questions of safety and effectiveness because, as Tanyx, the other device is under the maximum limits.
Regarding MAXIMLUMN AVERAGE POWER DENSITY, the product is under the maximum limits and because of this do not raise any new question of safety and effectiveness. The differences between the devices do not raise any new question too, because both are under the maximum limits established.
In reference to Frequency (Hz) Tanyx has a 55 Hz (this number may vary ± 3%) and this frequency do not raise any new question of effectiveness and safety because it is tested and the differences between the devices to not raise any question two because they are under the limits of patient safety.
Tanyx is viewed as substantially equivalent to the predicate device because the stimulation parameters are all in the same range for the intended use. The device includes a wide variation in terms of Current, Voltage Amplitude and Waveform, the basic parameters, without sacrificing biological effects. Maximum Output Current differences between Tanyx and the predicate device are due to different methods. Tanyx was measured using a peak-to-peak value and predicate device uses RMS value, but both devices are
{6}------------------------------------------------
in the same range of safety and efficacy. Other measurement differences are not significant from the predicated device. Operation and mechanical design don´t show significant differences.
.
.
.
.
. . . .
. .
.
:
.
and the country of the country of the country
Page 8
.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The image is in black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### September 13, 2013
Medecell c/o Rhonda Alexander, M.S., M.P.A. Registrar Corp 144 Research Drive Hampton. VA 23666
Re: K123866
Trade/Device Name: Tanyx® Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: July 23, 2013 Received: July 24, 2013
Dear Ms. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Ms. Rhonda Alexander
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Joyce M. Whang -S
Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use
510(k) Number (if known): K123866
Device Name: Tanyx®
Indications For Use:
To be used in adults only for temporary relief of pain associated with sore and aching muscles in the upper extremities (arms), lower extremities (legs), and lower back due to strain from exercise or normal household and work activities.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ·
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.