← Product Code [NFO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO) · K250227

# JOVS Electric Stimulation Beauty Device (JE2) (K250227)

_Shenzhen Qianyu Technology Co., Ltd. · NFO · Oct 17, 2025 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K250227

## Device Facts

- **Applicant:** Shenzhen Qianyu Technology Co., Ltd.
- **Product Code:** [NFO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO.md)
- **Decision Date:** Oct 17, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

JOVS Electric Stimulation Beauty Device is indicated for over-the-counter aesthetic use including facial and neck stimulation.

## Device Story

Portable, reusable, battery-powered device for home aesthetic use. Features spring electrode head for transcutaneous electrical stimulation of facial and neck skin. User operates device via controls to select one of two output modes; device delivers pulsed biphasic square wave electrical stimulation. Includes automatic shut-off and patient override control. Healthcare provider does not interact with device; user self-administers treatment. Output parameters (voltage, current, frequency) are regulated to provide aesthetic stimulation. Benefits include non-invasive facial and neck skin stimulation.

## Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-2, -1-11, -2-10), battery safety (IEC 62133-2), and waveform output verification per FDA guidance for powered muscle stimulators.

## Technological Characteristics

Materials: PC, ABS, PVC, stainless steel. Energy: Internal 3.7V 1000mAh Li-ion battery, rechargeable via Type-C. Output: Pulsed biphasic square wave, 2.5kHz frequency, 110μs pulse width. Connectivity: Standalone. Form factor: Portable handheld unit (202x33x66mm). Software: Microprocessor-controlled with 2 output modes, automatic shut-off, and patient override. Sterilization: Non-sterile.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Beagank 4T Plus ([K233010](/device/K233010.md))
- BMR Face ([K103031](/device/K103031.md))
- Rejuvenique System Model: RJV-10 ([K011935](/device/K011935.md))

## Reference Devices

- MyoLight Microcurrent Handpiece ([K243063](/device/K243063.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 17, 2025

Shenzhen Qianyu Technology Co., Ltd.
% Tangyao Dai
RA Specialist
Feiying Drug &amp; Medical Consulting Technical Service Group
Rm 2401 Zhenye International Business Center
No. 3101-90, Qianhai Road
Shenzhen, Guangdong 518000
China

Re: K250227
Trade/Device Name: JOVS Electric Stimulation Beauty Device (JE2)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: January 24, 2025
Received: January 27, 2025

Dear Tangyao Dai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250227 - Tangyao Dai
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K250227 - Tangyao Dai
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120

Expiration Date: 07/31/2026

See PRA Statement below.

Submission Number (if known)

K250227

Device Name

JOVS Electric Stimulation Beauty Device (JE2)

Indications for Use (Describe)

JOVS Electric Stimulation Beauty Device is indicated for over-the-counter aesthetic use including facial and neck stimulation.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

Shenzhen Qianyu Technology Co., Ltd. 510(k)s -510(k) Summary

# 510(k) Summary

# K250227

"510(k) Summary" as required by 21 CFR Part 807.92.

# I. Submitter

Company name: Shenzhen Qianyu Technology Co., Ltd.

Address: Room 601, Han's Technology Center, No.9988, Shennan Avenue, Maling

Community, Yuehai Street, Nanshan District

Contact person: Guoyang Li

Title: Management Representative

Tel: +86-15099998872

E-mail: liguoyang@jovs-beauty.com

Date: 2025-10-8

# II. Subject Device

Name of Device: JOVS Electric Stimulation Beauty Device

Model(s): JE2

Common or Usual Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes

Regulation Name: Transcutaneous electrical nerve stimulator for pain relief

Regulatory Class: II

Product Code: NFO

Regulation Number: 21 CFR 882.5890

# III. Predicate and Reference Device

&gt; Predicate devices

|  No. | Manufacturer | Device name | Product code | 510(k) Number | Cleared Date  |
| --- | --- | --- | --- | --- | --- |
|  1. | Belega Inc. | Beagank 4T Plus | NFO | K233010 | 2023.11.21  |
|  2. | BIO-MEDICAL RESEARCH, LTD. | BMR Face | NFO | K103031 | 2011.11.10  |
|  3. | SALTON, INC. | Rejuvenique System Model: RJV-10 | NFO, GYB | K011935 | 2021.08.08  |

&gt; Reference device

{5}

Shenzhen Qianyu Technology Co., Ltd. 510(k)s -510(k) Summary

|  No. | Manufacturer | Device name | Product code | 510(k) Number | Cleared Date  |
| --- | --- | --- | --- | --- | --- |
|  1 | Raja Trading Company, Inc. | MyoLight Microcurrent Handpiece | NFO | K243063 | 2025.02.10  |

# IV. Device Description

The JOVS Electric Stimulation Beauty Device is a portable, non-sterile, reusable device designed to achieve the aesthetic effect. It mainly consists of main unit, charging and disinfection base, and power cable. The device is supplied by internal rechargeable lithium battery, which can be recharged by external charger through the Type-C charging cable. The device is unusable when charging.

The device is only home environment use, which has the spring electrode head (electric stimulation) to provide the following function:

The JOVS Electric Stimulation Beauty Device is indicated for over-the-counter aesthetic use including facial and neck stimulation.

# V. Indications for Use

JOVS Electric Stimulation Beauty Device is indicated for over-the-counter aesthetic use including facial and neck stimulation.

# VI. Comparison of Technological Characteristics With the Predicate Device

The JOVS Electric Stimulation Beauty Device has the same intended use as the predicates. The technological characteristics, features, specifications, materials are similar to the predicate devices and reference device. Any minor differences between the subject device and the listed predicate devices and reference device do not raise any issues of safety or effectiveness. Performance data supports that the device is safe and as effective as the predicate devices and reference device for its intended use.

Therefore, the JOVS Electric Stimulation Beauty Device may be found substantially equivalent to its predicate devices and reference device.

{6}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  Device Name and Model | JOVS Electric Stimulation Beauty Device Model: JE2 | BEAGANK 4T PLUS | BMR Face | Rejuvenique System Model: RJV-10 | MyoLight Microcurrent Handpiece | /  |
|  510(k) Number | K250227 | K233010 | K103031 | K011935 | K243063 | /  |
|  Regulation number | 882.5890 | 882.5890 | 882.5890 | 882.5890, 882.1275 | 882.5890 | Same  |
|  Product code | NFO | NFO | NFO | NFO, GYB | NFO | Same  |
|  Device classification | Class II | Class II | Class II | Class II | Class II | Same  |
|  OTC or prescription | OTC | OTC | OTC | OTC | Prescription | Same  |
|  Indications for use | JOVS Electric Stimulation Beauty Device is indicated for over-the-counter aesthetic use including facial and neck stimulation. | The BEAGANK 4T PLUS is a handheld portable device for over-the-counter aesthetic use including facial and neck stimulation or body skin stimulation. | BMR Face is intended for facial stimulation and is indicated for Over the Counter Cosmetic Use. | The Rejuvenique System is indicated for cosmetic use. | The device is indicated for aesthetic use including facial and neck or body skin stimulation. | Different. Compared to the predicate devices, the subject device is limited to use on the facial and neck areas. Both the subject and predicate devices use the same fundamental electrical stimulation technology and are intended for  |

Page 3 of 15

{7}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  |  |  | overlapping treatment areas and patient populations. Therefore, the difference in the indications for use does not impact the device’s safety or effectiveness.  |
|  Dimensions | 202×33×66mm | Not publicly available | 6.0x8.0x2.1(cm) | 4-1/2”×3”×1-1/4 (inch) | Handpiece #1: 62mm x 147 mm
Handpiece #2: 39mm x 139.5 mm | Different. The dimensions and weight of the subject device are different from the predicate devices, however this difference is insignificant and do not impact its safety and effectiveness.  |
|  Weight | 634g | 136g | 63g | 80 grams | Handpiece #1: 1100 g
Handpiece #2: 720 g |   |
|  Treatment area | Face and neck | Face, neck and body | Facial skin | Facial skin | Face, neck and body | Same  |
|  Power source | Adapter input: 5V=1A
Battery voltage: DC3.7V, 1000mAh | Internal rechargeable Lithium-ion battery | 3.6V rechargeable batteries | Single 9V battery | Mains Power 110-120, 220 - 240 VAC, 50- 60 Hz | Different. Although the power source is different from the predicate devices, the lithium battery of the subject device  |

Page 4 of 15

{8}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  |  |  | complies with the IEC 62133-2 standard, and the device has been tested for electrical safety as per IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 requirements, so this difference does not impact its safety and effectiveness.  |
|  Number of output modes | 2 | 4 | 3 | 1 | 1 | Different. Although the number of output modes is different from that of the predicate devices, the output parameters of each mode have been tested, and the subject device has passed IEC 60601-1 and IEC 60601-2-10, so this difference does not impact its safety and effectiveness.  |
|  Number of output channel | 1 | 1 | 2 | 1 | Not publicly available | Same  |
|  Regulated Current | Regulated voltage | Regulated voltage | Not publicly available | Regulated voltage | Regulated voltage | Same  |

Page 5 of 15

{9}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  or Regulated Voltage? |  |  |  |  |  |   |
|  Software/firmware/microprocessor control | YES | YES | YES | YES | YES | Same  |
|  Automatic Overload Trip? | NO | Not required due to circuit design | Not publicly available | NO | Not publicly available | Same  |
|  Automatic No-Load Trip? | NO | Not required due to circuit design | Not publicly available | NO | Not publicly available | Same  |
|  Automatic Shut Off? | YES | YES | Not publicly available | YES | YES | Same  |
|  Patient Override Control? | YES | YES | Not publicly available | YES | Not publicly available | Same  |
|  Indicator Display:
- On/Off Status?
- Low Battery?
- Voltage/ Current Level? | YES | YES | YES | YES | Not publicly available | Same  |

Page 6 of 15

{10}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  Timer Range | 5 minutes | Max 5 minutes | 10/15/20minutes | 16 minutes | 20 Minutes
Recommended
Treatment Time | Similar  |
|  Compliance with Voluntary Standards? | ISO 10993-5
ISO 10993-10
ISO 10993-23
IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 60601-2-10 | IEC 60601-1
IEC 60601-1-2 | ISO 10993-5
ISO 10993-10
IEC 60601-1
IEC 60601-1-2
IEC 60601-2-10 | Not publicly available | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-10
ISO 14971
IEC 62366 | Similar  |
|  Compliance with 21 CFR 898? | YES | YES | YES | YES | YES | Same  |
|  Housing materials | PC+ABS+PVC+Stainless steel | ABS | Plastic case built of abs, electrodes of skin conductive through adhesive hydrogel layer. | Not publicly available | Molded PC-ABS
plastic material with screw assembly construction. | Different. Although the materials of the subject device are not entirely the same as the predicate devices, the subject device has been tested for  |

Page 7 of 15

{11}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s –510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  |  |  | biocompatibility as per ISO 10993 series standards, so this difference does not impact its safety and effectiveness.  |
|  Output specifications |   |   |   |   |  |   |
|  Waveform Type | Pulsed biphasic, square wave | Rectangle, biphasic asymmetric | Symmetric, pulsed, biphasic | Pulsed biphasic, Rectangular (+phase), Spike (-phase) | Pulsed Biphasic rectangular modulated by trapezoidal | Same  |
|  Maximum Output Voltage | (+/-10%)
27V@500Ω
35.5V@2kΩ
36.7V@10kΩ | (+/-15%)
Mode 1:
0.12V@500Ω
0.51V@2KΩ
1.83V@10KΩ

Mode 2:
0.11V@500Ω | 15 V±10% @ 500 Ω
60 V±10% @ 2K Ω
32 V±10% @ 10K Ω | (+/-10%)
18.8V@ 500Ω
24.8V@ 2kΩ
28.0V@ 10kΩ | (+/-20%)
30V @ 500 Ω
33V @ 1,000 Ω
35V @ 10,000 Ω | Different. The “Maximum Output Voltage” and “Maximum Output Current” of the subject device are very similar to that of the predicate device 2 and basically within the range of the predicate device 2 and reference device  |

Page 8 of 15

{12}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s –510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  | 0.50V@2KΩ
1.87V@10KΩ

Mode 3:
16.6V@500Ω
21.0V@2KΩ
22.5V@10KΩ

Mode 4:
15.3V@500Ω
21.2V@2KΩ
22.9V@10KΩ |  |  |  | 1. And we have tested the product waveform parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not impact its safety and effectiveness.  |
|  Maximum Output Current | (+/-10%)
54mA@500Ω
17.75mA@2KΩ
3.67mA@10KΩ | Mode 1:
0.19mA@500Ω
0.18mA@2KΩ
0.16mA@10KΩ | 30mA±10%@500Ω
30mA±10%@2KΩ
3.2mA±10%@10KΩ | (+/-10%)
37.6mA @ 500Ω
12.4mA @ 2kΩ
2.8mA@ 10kΩ | 60mA @ 500 Ω
33mA @ 1000 Ω
3.5mA @ 10,000 Ω |   |

Page 9 of 15

{13}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s –510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  | Mode 2:
0.45mA@500Ω
0.39mA@2KΩ
0.24mA@10KΩ

Mode 3:
30.8mA@500Ω 8.52mA@2KΩ
2.26mA@10KΩ

Mode 4:
31.4mA@500Ω 8.60mA@2KΩ
2.15mA@10KΩ |  |  |  |   |
|  Pulse Width | 110μs±20% | Mode 1: 265μs
Mode 2: 265μs
Mode 3: 274μs
Mode 4: 610μs | 160-200μs | 300ms fixed | 128μs, 256μs, 384μs | Different. The pulse width of the subject device is similar to that of the predicate device 1 and reference device 1. And we have tested  |

Page 10 of 15

{14}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  |  |  | the product waveform parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not impact its safety and effectiveness.  |
|  Output Frequency | 2.5kHz±20% | Mode 1: 3.80kHz
Mode 2: 3.80kHz
Mode 3: 3.65kHz
Mode 4: 1.64kHz | 70-80Hz | 8 Hz fixed | 1 ~ 63 Hz | Different. The output frequency of the subject device is within the range of that of the primary predicate device. And we have tested the product waveform parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do  |

Page 11 of 15

{15}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s -510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|   |  |  |  |  |  | not impact its safety and effectiveness.  |
|  Net charge | 0μC@500Ω | Mode 1: 0.025μC@500Ω
Mode 2: 0.025μC@500Ω
Mode 3: -1.3μC@500Ω
Mode 4: 0μC@500Ω | 0μC@500Ω | 0μC@500Ω | 0μC (Pulse is biphasic symmetrical) | Same  |
|  Maximum phase charge | 7.19μC@500Ω | Mode1: 0.059μC@500Ω
Mode 2: 0.025μC@500Ω
Mode 3: 2.2μC@500Ω
Mode 4: 0.29μC@500Ω | Not publicly available | 11.3mC@500Ω | 23μC | Different. The maximum phase charge and maximum current density of the subject device are within the range of that of the primary predicate device and predicate device 2. The maximum power density of the subject device are within the range of that of the predicate device 1 and predicate device 2. And we have tested the product waveform  |
|  Maximum Current Density | 12.56mA/cm²@500Ω | Mode 1: 0.15~1.36mA/cm²@500Ω
Mode 2: 0.10~3.21mA/cm²@500Ω
Mode 3: 4.26~220mA/cm²@500Ω
Mode 4: 3.13~224.3mA/cm²@500Ω | 1mA/cm²@500Ω | 46.4mA/cm²@500Ω | 82μA/mm²  |   |

Page 12 of 15

{16}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s –510(k) Summary

|  Elements of Comparison | Subject Device | Primary Predicate Device | Predicate Device 1 | Predicate Device 2 | Reference Device 1 | Remark  |
| --- | --- | --- | --- | --- | --- | --- |
|  Maximum Power Density | 0.178W/cm²@500Ω | Mode 1: 8.51~11.39μW/cm²@500Ω
Mode 2: 17.46~33.87μW/cm²@500Ω
Mode 3: 1.124~106.6mW/cm²@500Ω
Mode 4: 0.56~51.78mW/cm²@500Ω | 3.91W/cm²@500Ω | 2.31mW/cm² @500Ω | 0.012W/ cm² | parameters according to FDA guidance documents and tested the device according to the requirements of IEC 60601-2-10, the tests are all passed, so these differences do not impact its safety and effectiveness.  |

Page 13 of 15

{17}

Shenzhen Qianyu Technology Co., Ltd. 510(k)s –510(k) Summary

## VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

### 1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the JOVS Electric Stimulation Beauty Device was conducted in accordance with the “Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020”, as recognized by FDA. The following testing was performed to, and passed, including:

- ISO 10993-5: 2009, Biological evaluation of medical devices –Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices – Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices – Part 23: Tests for irritation

### 2) Electrical Safety

- IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests
- IEC 60601-1-11 Medical Electrical Equipment – Part 1-11: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60601-2-10 Medical electrical equipment – Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- IEC 62133-2, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems

### 3) Software Verification and Validation

Software documentation consistent with *Basic Documentation Level* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

### 4) Performance test

The waveform test report has been conducted to verify the output specifications of the device according to Guidance for Powered Muscle Stimulator 510(k)s.

Page 14 of 15

{18}

Shenzhen Qianyu Technology Co., Ltd.
510(k)s – 510(k) Summary

## VIII. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.

Page 15 of 15

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K250227](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K250227)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
