← Product Code [NFO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO) · K232172

# BTL-785BNF Handpiece (K232172)

_BTL Industries, Inc. · NFO · Sep 18, 2023 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K232172

## Device Facts

- **Applicant:** BTL Industries, Inc.
- **Product Code:** [NFO](/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO.md)
- **Decision Date:** Sep 18, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The device is indicated for aesthetic use including facial and neck or body skin stimulation.

## Device Story

BTL-785BNF Handpiece is an electrostimulation device for aesthetic facial, neck, and body skin stimulation. It operates via a main control unit with a touch screen interface that guides the user through therapy procedures. The device generates an electromagnetic field to interact with human tissues for neuro-muscle stimulation; it also incorporates a monopolar radiofrequency (RF) generator to provide low-degree warming for patient comfort. The system includes the handpiece, single-use conductive hydrogel electrodes (silver ink), a holding arm, connection cables, and a therapy discomfort button. It is used in a clinical setting, operated by a healthcare provider who monitors therapy parameters, remaining time, and therapy type on the control unit display. The device aims to provide aesthetic benefits through controlled stimulation and thermal comfort.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), software life cycle (IEC 62304), and biocompatibility (ISO 10993-1, -5, -10).

## Technological Characteristics

Electrostimulation device with monopolar RF warming. Output: Biphasic rectangular pulse modulated by trapezoidal. Mains power (100-240V). Electrodes: Silver conductive ink with hydrogel. Connectivity: Main control unit with touch screen. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62304, ISO 14971, ISO 10993. Software: IEC 62304 compliant.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI-LIFT Device ([K191951](/device/K191951.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.

September 18, 2023

BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752

Re: K232172

Trade/Device Name: BTL-785BNF Handpiece Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: July 3, 2023 Received: July 21, 2023

Dear David Chmel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K232172

Device Name BTL-785BNF Handpiece

Indications for Use (Describe)

The BTL-785BNF Handpiece device has the following indications for use:

The device is indicated for aesthetic use including facial and neck stimulation or body skin stimulation.

| Type of Use (Select one or both, as applicable)                                                               |                                                                                                              |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## General Information

| Sponsor:                     | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant:                   | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person:              | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com                                                         |
| Summary Preparation<br>Date: | June 23, 2023                                                                                                   |

#### Device

| Trade/Proprietary Name:      | BTL-785BNF Handpiece                                       |
|------------------------------|------------------------------------------------------------|
| Primary Classification Name: | Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Classification Regulation:   | 21 CFR 882.5890, Class II                                  |
| Classification Product Code: | NFO                                                        |

{4}------------------------------------------------

#### Legally Marketed Predicate Device

The BTL-785BNF Handpiece is a state-of-the-art device with accessories intended for electrostimulation therapy, and is substantially equivalent to the following products that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:

- Avazzia OTC TENS for Aesthetics, model BEST-AV1: EZZI-LIFT Device ( K191951) ●
#### Product Description

The BTL-785BNF Handpiece is a state-of-the-art electrostimulation device that enables the application of therapy by an electromagnetic field.

The subjected device is intended to be used together with a main control unit that is equipped with a large color touch screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device displays information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

The BTL-785BNF Handpiece is a specially developed device that can be used with the applicators and the main control unit. The generated electromagnetic field is interact with the tissues of the human body in order to achieve neuro-muscle stimulation. The radiofrequency is providing a low degree of warming to enhance the patient's comfort during the treatment.

The BTL-785BNF Handpiece consists of the following main components:

- BTL-785BNF Handpiece
- · Single-use application electrodes
- Holding arm
- · Connection cables
- Therapy discomfort button

#### Indications for Use

The BTL-785BNF Handpiece device has the following indications for use:

The device is indicated for aesthetic use including facial and neck or body skin stimulation.

#### Non-clinical Testing (Performance Data)

The BTL-785BNF Handpiece device has been thoroughly evaluated for electrical safety. The device has been found to comply with applicable medical device safety standards:

{5}------------------------------------------------

| IEC 60601-1    | Medical electrical equipment - Part 1: General requirements for basic safety and<br>essential performance                                                                                   |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2  | Medical electrical equipment – Part 1-2: General requirements for basic safety<br>and essential performance - Collateral standard: Electromagnetic disturbances<br>- Requirements and tests |
| IEC 60601-2-10 | Medical electrical equipment – Part 2-10: Particular requirements for the basic<br>safety and essential performance of nerve and muscle stimulators                                         |
| IEC 60601-1-6  | Medical electrical equipment - Part 1-6: General requirements for basic safety<br>and essential performance – Collateral standard: Usability                                                |
| IEC 62304      | Medical device software - Software life cycle processes                                                                                                                                     |
| ISO 14971      | Medical devices – Application of risk management to medical devices                                                                                                                         |
| ISO 10993-1    | Biological evaluation of medical devices – Part 1: Evaluation and testing within<br>a risk management process                                                                               |
| ISO 10993-5    | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity                                                                                                          |
| ISO 10993-10   | Biological evaluation of medical devices - Part 10: Tests for irritation and skin<br>sensitization                                                                                          |

# Clinical performance data

A clinical study was not conducted.

## Comparison with the Predicate Device

| 510(k) number                  | K232172                                                                                           | K191951                                                                                           |
|--------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Device name<br>Company name    | BTL-785BNF Handpiece<br>BTL Industries, Inc.                                                      | Avazzia OTC TENS for Aesthetics,<br>model BEST-AV1: EZZI-LIFT<br>Device Avazzia, Inc              |
| Type                           | Subject device                                                                                    | Predicate device                                                                                  |
| Product Code<br>and Regulation | Neurology<br>21 CFR 882.5890<br>NFO– Stimulator, Transcutaneous<br>Electrical, Aesthetic Purposes | Neurology<br>21 CFR 882.5890<br>NFO– Stimulator, Transcutaneous<br>Electrical, Aesthetic Purposes |

{6}------------------------------------------------

| Indications for Use                                                            | The device is indicated for aesthetic use including facial and neck or body skin stimulation.                                       | Device is indicated for over-the counter aesthetic use including facial and neck stimulation or body skin stimulation.         |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Anatomic Sites                                                                 | Face, neck, and body                                                                                                                | Face, neck, and body                                                                                                           |
| Number of Output Modes                                                         | 1                                                                                                                                   | 4                                                                                                                              |
| Low Battery Indicator                                                          | N/A                                                                                                                                 | Yes                                                                                                                            |
| Automatic shut off                                                             | N/A                                                                                                                                 | 60 min                                                                                                                         |
| Compliance with Voluntary Standards                                            | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 14971<br>IEC 62366                                                            | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 14971<br>IEC 62366                                                       |
| Compliance with 21 CFR 898                                                     | Yes                                                                                                                                 | Yes                                                                                                                            |
| Weight                                                                         | 2500 g                                                                                                                              | 150 g                                                                                                                          |
| Dimensions [W x H x D] inches                                                  | 38.98" x 53.94" x 34.65"                                                                                                            | 2.6" X 4.7" X 1.35"                                                                                                            |
| Housing Materials and Construction                                             | PCBs and leads inside plastic, or foam case housing                                                                                 | PCBs inside plastic case housing                                                                                               |
| User Interface Display                                                         | Yes on the main controll unit.                                                                                                      | LEDs and switches                                                                                                              |
| Energy type                                                                    | Electric stimulation                                                                                                                | Electric stimulation                                                                                                           |
| Additional energy                                                              | RF warming                                                                                                                          | N/A                                                                                                                            |
| Number, Size, and Type of Batteries                                            | Mains power<br>100-240V, 50-60 Hz                                                                                                   | Two 1.5 V AA batteries                                                                                                         |
| Waveform Shape                                                                 | Biphasic rectangular pulse modulated by trapezoidal                                                                                 | positive square wave followed by a damped sinusoidal waveform of variable duration depending on damping and body loading       |
| Max output voltage [V] (+/-20%)<br>- at 500 Ω<br>- at 2,000 Ω<br>- at 10,000 Ω | BTL-785-2, -8, -9<br>Max. 48,5V @ 10 - 373 Ω<br><br>48,5V @ 500 Ω<br>48,5V @ 2000 Ω<br>0 @ 10 000 Ω                                 | -42 @ 500 Ω<br>-122 @ 2000 Ω<br>-348 @ 10 000 Ω                                                                                |
|                                                                                | BTL-785-1, -7<br>Max. 48,5V @ 10 - 757 Ω<br><br>48,5V @ 500 Ω<br>48,5V @ 2000 Ω<br>0 @ 10 000 Ω                                     |                                                                                                                                |
|                                                                                | BTL-785-2, -8, -9<br>Max. 130 mA @ 10 - 373 Ohm                                                                                     |                                                                                                                                |
|                                                                                | Max output current<br>(+/-20%)<br>- at 500 Ω<br>- at 2,000 Ω<br>- at 10,000 Ω                                                       | 97 mA @ 500 Ω<br>24 mA @ 2000 Ω<br>0 mA @ 10000 Ω                                                                              |
|                                                                                |                                                                                                                                     | BTL-785-1, -7<br>Max. 64 mA @ 10 - 757 Ω<br><br>64 mA @ 500 Ω<br>24 mA @ 2000 Ω<br>0 mA @ 10000 Ω                              |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| Max current density at<br>500 Ω [μA/mm²]                                       | BTL-785-2 - 243 μA/mm²<br>BTL-785-8 - 335 μA/mm²<br>BTL-785-9 - 277 μA/mm²<br>BTL-785-1 - 168 μA/mm²<br>BTL-785-7 - 337 μA/mm²      | Built-in, Y,<br>Brush: 800 μA/mm²<br>Pencil: 19,000 μA/mm²                                                                     |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| Max average power<br>density at 500 Ω<br>[μW/mm²]                              | BTL-785-2 - 0,047 W/cm²<br>BTL-785-8 - 0,065 W/cm²<br>BTL-785-9 - 0,054 W/cm²<br>BTL-785-1 - 0,033 W/cm²<br>BTL-785-7 - 0,065 W/cm² | Built-in, Y,<br>Brush: 500 μW/mm²<br>Pencil: 3,500 μW/mm²                                                                      |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| Net Charge per pulse                                                           | Zero (pulse is biphasic symmetrical)                                                                                                | 4 μC                                                                                                                           |
| Max phase charge per<br>pulse at 500 Ω [μC]                                    | BTL-785-2, -8, -9<br>15 μC<br><br>BTL-785-1, -7<br>10 μC                                                                            | 10 μC                                                                                                                          |
|                                                                                | 10 μC                                                                                                                               |                                                                                                                                |
| Duration of primary<br>[μs]<br>(depolarizing phase)                            | 80 μs                                                                                                                               | 500 μs                                                                                                                         |
| Pulse Duration [μs]                                                            | 160 μs                                                                                                                              | 1 100 μs                                                                                                                       |
| Frequency [Hz]                                                                 | 250                                                                                                                                 | 15 to 121                                                                                                                      |
| Electrodes<br>(a) materials                                                    | Conductive ink (Silver)                                                                                                             | Stainless steel 316                                                                                                            |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| (b) electroconductive<br>media                                                 | Conductive hydrogel                                                                                                                 | n/a                                                                                                                            |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| (c) electrode-to-skin<br>impedance range                                       | 10-5000 Ohm                                                                                                                         | less than 15                                                                                                                   |
|                                                                                |                                                                                                                                     |                                                                                                                                |
| (d) max duration of<br>use (same as device<br>shut off)                        | 20 min                                                                                                                              | 60 min                                                                                                                         |
| (e) conductive<br>surface area                                                 | BTL-785-1 - 3.8 cm²<br>BTL-785-2 - 4 cm²<br>BTL-785-7 - 1.9 cm²<br>BTL-785-8 - 2.9 cm²<br>BTL-785-9 - 3.5 cm²                       | built in rectangular - 1.94 cm²<br>spherical - 5.60 cm²<br>brush/comb - 0.56 cm²<br>small circular,<br>pencil-like - 0.023 cm² |

{7}------------------------------------------------

{8}------------------------------------------------

#### Substantial Equivalence

The BTL-785BNF Handpiece device has equivalent technological characteristics and similar intended use compared to the predicate device.

The predicate device utilizes electric stimulation in order to achieve therapeutic effect. The subjected device BTL-785BNF Handpiece also utilizes electromagnetic stimulation for its main therapeutic effect; however, the device is also equipped with a monopolar radiofrequency generator. The RF warms the patient at the application site to increase patient convenience during therapy.

Any differences between the predicate devices and BTL-785BNF Handpiece have no significant influence on safety or effectiveness of the BTL-785BNF Handpiece.

Therefore, the BTL-785BNF Handpiece device is substantially equivalent to the predicate device.

#### Conclusion

Based upon the intended use, comparison of technical characteristics and performance testing provided in this premarket notification, the BTL-785BNF Handpiece device has been shown to be substantially equivalent to the currently cleared predicate device and secondary predicate device for requested intended use.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K232172](https://fda.innolitics.com/submissions/NE/subpart-f%E2%80%94neurological-therapeutic-devices/NFO/K232172)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
