Biosonic Technologies Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.
Device Story
Microprocessor-controlled facial toning device; delivers microcurrent electrical energy to facial tissues for aesthetic purposes. Desktop unit; operated by licensed practitioners. Input: user-selected intensity (Normal, Moderate, High), interval delay (1-6s), and sequence duration (10-30s). Output: biphasic, rectangular waveforms (3 Hz to 300 Hz) via handheld probes. Safety features: automatic overload trip, no-load trip, and shutoff; low battery/voltage/current indicators. Powered by 6V battery; no line power operation. Benefits: provides precise microcurrent delivery for facial toning; enhanced safety via integrated monitoring compared to non-microprocessor predicates.
Clinical Evidence
Bench testing only. Testing included product safety and electromagnetic compatibility per IEC 60601-1 and IEC 60601-1-2, and biocompatibility per ISO 10993.
Indicated for prescription cosmetic use for facial stimulation in patients requiring aesthetic tissue stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
# Section 005 - Traditional 510(k) Summary
As required in 21 CFR807.92, we hereby submit this Traditional 510(k) Summary.
Traditional 510(k) owners name, address, phone, fax, contact person & preparation date
The Traditional 510(k) owner is Biosonic Technologies, LLC.
- Address: 333 Las Olas Way, Suite 1906
Fort Lauderdale, FL 33311
Phone: (954) 300 - 2511
(866) 830 - 8759 Fax:
The contact person is Mr. Ray Baker, President, Biosonic Technologies, LLC.
Traditional 510(k) preparation date: 19 August, 2011
# SUBJECT DEVICE:
Name of the device, trade name, proprietary name, and classification name:
Trade name: Beautiful Image Model 900 Facial Toning Device
Common name: Facial Toning Device
Classification: Stimulator, Transcutaneous Electrical, Aesthetic Purposes.
CFR Reference: 21 CFR 882.5890
Product Code: NFO
# Device Description:
Beautiful Image Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.
{1}------------------------------------------------
To enhance safety, the device is operated only from battery voltage and will not operate while connected to line power. A proper recharge should be performed prior to using the machine.
The Model 900 is microprocessor controlled and housed in a white plastic case.
| Dimensions & Appearance: | |
|-------------------------------|--------------------------------------------------------------------------------------|
| Height (rear) | 5.5 inches (13.97 cm) |
| Height (front) | 3.3 inches (8.38 cm) |
| Width | 15.3 inches (38.86 cm) |
| Depth | 11.3 inches (28.7 cm) |
| Weight | 10.0 pounds (4.5 kgs) |
| Color | White case with blue front panel |
| Electrical Specifications | |
| Device power | Model 900 will operate only on battery power; it will<br>not operate on line voltage |
| Device Output | 800µA maximum, from constant current source |
| Complex wave output frequency | 3 Hz to 300 Hz |
| Maximum voltage | 30Vpp on any output with respect to common; 5Vpp<br>typical |
| Battery Charger | 9vdc @ 1 amp. Use of the charger disables the<br>patient circuits |
| Battery | 6 volt, 7 amp hours |
| Power Requirements CAC | |
| Adapter | |
| Input Voltage | 100 to 120 VAC, 60 Hz |
| Input Current | 0.15 amps |
| Configuration | Standard U.S. 2-Prong Plug |
| Device output Voltage | 9v DC |
| Output Current | 1 amp |
| Output Plug | De coas: 5.5mm OD, 2.1mm pin receptacle |
| Cable length | 6 ft (240 cm) |
| Environmental Specifications | |
| Operating Temperature | + 50 to 104° F (+10 to +40°C) |
| Storage Temperature | - 29 to + 167° F (-34 to +76° C) |
| Humidity | 0 to 95% - non-condensing |
Specifications:
5-2
{2}------------------------------------------------
# FUNCTIONAL DESCRIPTION
# OPERATING CYCLE:
The Model 900 operating cycle is initiated by momentarily pressing the POWER button located in the POWER section of the front panel. All indicator lights (LEDs) on the panel will illuminate for a short period of time except for the CHARGING and the BATTERY LOW or BATTERY FULL indicators (depending on the charge level of the battery).
The INTERVAL PACE lights (located in the WAVEFORM section of the panel) will step from left to right providing a visual indication of the approximate delay time remaining prior to the application of energy to the probes. The INTERVAL DELAY adjustment sets the amount of time for this delay.
The SEQUENCING lights (also located in the WAVEFORM section of the panel) indicate the application of each of the different waveform outputs from the equipment, seven in total. The SEQUENCE DURATION adjustment sets the amount of time for the total sequenced set.
# OPERATING RANGES:
Adjustment of the output energy intensity is set by the INTENSITY ADJUSTMENT button located in the PROCEDURE section of the panel. Each press of the button will advance the output energy from NORMAL, to MODERATE, to HIGH, and then the sequence will repeat.
The INTERVAL DELAY provides an adjustment range of approximately 1 to 6 seconds.
The SEQUENCE DURATION provides an adjustment range of approximately 10 to 30 seconds.
Waveform frequencies: automatically sequencing from 3 Hz to 300 Hz
# OPERATING MODE:
An indicator shows that the wand set is in use. The indicator will be illuminated when the wand is connected to the machine.
# CALIBRATION:
The Model 900 does not require calibration.
{3}------------------------------------------------
# INDICATION FOR USE:
The Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes. INTENDED USE:
Biosonic Technologies Model 900 Facial Toning Device is intended for facial stimulation and is indicated for prescription cosmetic use. The anatomical site for application of the Model 900 is the face.
# COMPARISON WITH THE PREDICATE DEVICE:
The predicate device for this Traditional 510(k) is the NūFace Facial Toning Device. The subject device is Biosonic Technologies Beautiful Image Model 900 Facial Toning Device, that is substantially equivalent to the predicate device, differing only in that the unit is larger and designed to support the handheld operation of the contacts from a desktop device by means of cables, and is intended for the use of licensed practitioners rather than the public.
# COMPARISON OF THE SUBJECT DEVICE AND PREDICATE DEVICE PARAMETERS:
| Parameter | Your Device | Predicate<br>Device |
|-------------------------------------------------------------------------------------------------|----------------------------------|-----------------------|
| 510(k) Number | K130065 | K072260 |
| Device Name and Model | Model 900 | NuFace<br>Original |
| Manufacturer | Biosonic<br>Technologies,<br>LLC | Carol Cole<br>Company |
| Power Source(s) † | One 6V battery | One 9V battery |
| - Method of Line Current Isolation | N/A | N/A |
| - Patient Leakage Current†† | N/A | N/A |
| - Normal Condition (μΑ) | N/A | N/A |
| - Single Fault Condition (μΑ) | N/A | N/A |
| Average DC current through electrodes when<br>device is on but no pulses are being applied (µA) | None | None |
| Number of Output Modes ††† | One | One |
| Number of Output Channels†††† | One | One |
| Synchronous or Alternating? | N/A | N/A |
{4}------------------------------------------------
| Parameter | Your Device | Predicate Device |
|-------------------------------------------|---------------|------------------|
| Method of Channel Isolation | N/A | N/A |
| Regulated Current or Regulated Voltage? | Both | Both |
| Software/Firmware/Microprocessor Control? | Yes | No |
| Automatic Overload Trip? | Yes | No |
| Automatic No-Load Trip? | Yes | No |
| Automatic Shut Off? | Yes | No |
| User Override Control? | Yes | Yes |
| Indicator Display Status | | |
| On/Off Display Status? | Yes | Yes |
| Low Battery? | Yes | No |
| Voltage/Current Level? | Yes | No |
| Timer Range (minutes) | None | None |
| Compliance with Voluntary Standards? | IEC 60601-1 | IEC 60601-1 |
| Compliance with 21 CFR 898? | Yes | Yes |
| Weight (lbs., oz.) | 10lbs | 0.5lbs |
| Dimensions (in.) [W x H x D] | 5.5x15.3x11.3 | 2.25x7x0.75 |
| Housing Materials and Construction | Thermoplastic | Thermoplastic |
† For AC line-powered devices, we recommend that you specify the line voltage and frequency, the method of line current isolation, and the measured patient leakage current. For battery-powered devices, we recommend that you specify the number, size and type of batteries.
++ We recommend that you follow IEC 60601-1 "Medical Electrical Equipment - Part 1: General Requirements for Safety" or an equivalent method to show that the levels of patient leakage current, measured under both normal and single fault conditions, are acceptable.
ttt For devices with more than one output mode, we recommend that you provide the information in Output Waveforms (see Section 7A above) and Output Specifications (see Section 7C below) for each output mode.
tttt For devices with more than one output channel, we recommend that you describe whether the outputs are delivered in a synchronous and/or alternating fashion and the method of achieving channel isolation.
| .<br>Parameter | - March Made de<br>문<br>. | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>Jevice | Predicate |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | | | Device |
| . B | of the first and a program of the company and any come and a | | |
| Parameter | | Your Device | Predicate<br>Device |
| Mode or Program Name | | Beautiful<br>Image | NuFace |
| | Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Pulsed<br>Monophasic |
| | Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular |
| | Maximum Output Voltage (volts) (+/- 5%) | 0.347 @500Ω | 0.16 @500Ω |
| | | 1.242 @ 2 kΩ | 0.78 @ 2 kΩ |
| | | 5.780 @10 kΩ | 2.6 @10 kΩ |
| | Maximum Output Current (mA) (+/- 5%) | 0.647 @500Ω | 0.324 @500Ω |
| | | 0.625 @ 2 kΩ | 0.358 @ 2 kΩ |
| | | 0.584 @10 kΩ | 0.263 @10 kΩ |
| Duration of primary (depolarizing) phase† (msec) See<br>Note 3 | | 0.648 - 322 | --- |
| Pulse Duration† (msec) | | 3.24 - 1610 | ---- |
| Frequency† (Hz) [or Rate† (pps)] | | 0.621 - 308.6 | ---- |
| For multiphasic | Symmetrical phases? | Yes | Yes / No |
| waveforms only: | Phase Duration† (msec),<br>(state range, if applicable),<br>(both phases, if asymmetrical | 0.324-161 | pulsed |
| Net Charge (micro coulombs (µC) per pulse) (If zero,<br>state method of achieving zero net charge.) | | 0μC @500Ω | N/A @500Ω |
| Maximum Phase Charge, (µC) | | 190 @500Ω | 18.13 @500Ω |
| Maximum Current Density, †† (mA/cm², r.m.s.) | | 1.486 @500Ω | 0.341 @500Ω |
| Maximum Average Current (average absolute value),<br>mA | | 0.493 @500Ω | @500Ω |
| Maximum Average Power Density, †† (W/cm²), (using<br>smallest electrode conductive surface area) | | 366E-6 @500Ω | 3.02μW@500Ω |
| Burst Mode | (a) Pulses per burst | N/A | 2.1 |
| (i.e., pulse trains): | (b) Bursts per second | N/A | 9.1 |
| | (c) Burst duration (seconds) | N/A | 2.3 |
| | (d) Duty Cycle: Line (b) x Line<br>(c) | N/A | 20.9 |
| ON Time (seconds) | | 10-30 | Constant |
| OFF Time (seconds) | | 1-6 | None |
| Additional Features (specify, if applicable) | | None | None |
{5}------------------------------------------------
† For continuously variable parameters, we recommend that you specify the full range; for parameters with discrete settings, we recommend that you specify all available selections.
tt We recommend that you calculate the maximum current density and maximum average power density values by using the conductive surface area of the smallest electrodes intended for use with the unit, and include sample calculations in your 510(k). We also recommend that you calculate the maximum power density by using the maximum duty cycle and by averaging over output duration of one second. The maximum average power density should be less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the risk of thermal burns.
{6}------------------------------------------------
# DISCUSSION OF WHY THE DIFFERENCES BETWEEN THE SUBJECT DEVICE AND THE PREDICATE DEVICE DO NOT ADVERSELY AFFECT SAFETY AND EFFECTIVENESS:
The predicate device has a 9v battery, while the subject device has a 6v battery. Generally, a lower voltage battery represents a lower risk factor, and this advantage counts in favor of the Model 900 while providing the device ample power to perform in its intended use. Furthermore, the predicate device does not have any software/firmware/microprocessor control, automatic overload trip, automatic no-load trip, automatic shutoff, low battery indicator display status, nor voltage/current level indicator display status. Each of these safety features exists within the subject device, and therefore provides a safer device than the predicate device. Additionally, the predicate device is much smaller and lighter than the robust Model 900, which will not as easily move about and is designed in this fashion in direct response to user design inputs. Regarding the current, the maximum current density of the Model 900 measures 1.486 mA/cm2 r.m.s. at 500 Ohms of resistance, versus the reported 0.341 mA/cm2 r.m.s. at 500 Ohms resistance of the NūFace, and although this appears to represent a difference within the overall current, it reflects simply a more precise surface contact area that applies the microcurrent as compared to the predicate. The overall current delivered from the Model 900 is unchanged. Finally, the maximum average power density is 366E-6 W at 500 Ohms versus 3.02 µW at 500 Ohms, which represents a larger energy output than the predicate, given the more precise surface contact area, while remaining as safe as the predicate, and more effective at the point of contact than the predicate. Thus, given the foregoing, the Model 900 is at least as safe and as effective as the predicate NūFace device.
All other parameters including the intended use and safety profile remain unchanged.
# NONCLINICAL TEST DATA:
Product Safety testing and electromagnetic compatibility testing conforming to IEC 60601-1 and IEC 60601-1-2, and biocompatibility testing conforming to ISO 10993 were performed in support of the overall safety and effectiveness characteristics of the device, and have confirmed that the Model 900 is safe, effective and substantially equivalent regarding the outputs of the predicate device.
# SUMMARY:
No new issues of safety or effectiveness are raised regarding the Model 900 Facial Toning Device.
End of Traditional 510(k) Summary Section
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 30, 2014
Biosonic Technologies, LLC c/o Shepard G. Bentley Bentley Biomedical Consulting, LLC 28241 Crown Valley Parkway, Suite 510(K) Laguna Niguel, CA 92677
Re: K130065
Trade/Device Name: Beautiful Image Model 900 Facial Toning Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: March 30, 2014 Received: April 1, 2014
Dear Mr. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your do rise to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a dolly and regulations administered by other Federal agencies. You must of any I vith all the Act's requirements, including, but not limited to: registration and listing (21
{8}------------------------------------------------
Page 2 - Mr. Shepard G. Bentley
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K130065
### Device Name
Beautiful Image Model 900 Facial Toning Device
## Indications for Use (Describe)
The Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aestheric purposes.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
.
# ം Date: 2014.04.30
-0 16:54:05 -04'00'
{10}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.