K024359 · Vista Scientific, LLC · MYU · Aug 8, 2003 · Neurology
Device Facts
Record ID
K024359
Device Name
VISTA SCIENTIFIC BARIUM SULFATE, MODEL 3001-1
Applicant
Vista Scientific, LLC
Product Code
MYU · Neurology
Decision Date
Aug 8, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.5300
Device Class
Class 2
Indications for Use
Vista Scientific Barium Sulfate™ is intended for use as an additive to Codman Cranioplastic™ (Type 1-Slow set) to provide radiopacity for imaging purposes.
Device Story
Vista Scientific Barium Sulfate is a sterile, non-pyrogenic dry powder additive. It is mixed with Codman Cranioplastic™ (Type 1-Slow set) bone cement during cranioplasty procedures. The primary function is to provide radiopacity, allowing the bone cement to be visualized via medical imaging. The device is used in a clinical setting by surgeons or operating room staff during the preparation of bone cement. By rendering the cement radiopaque, the device assists clinicians in post-operative monitoring and assessment of the cranioplasty site.
Clinical Evidence
No clinical data. Performance, biocompatibility, and microbiological bench testing were conducted to verify that the device performs as designed and is suitable for its intended use.
Technological Characteristics
Dry powder Barium Sulfate USP. Supplied in 6g quantities within a 120 mL polymer container, double sterile peel pouch. Non-pyrogenic. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as a radiopacifying additive to Codman Cranioplastic™ (Type 1-Slow set) bone cement to enable imaging of the cranioplasty site.
Regulatory Classification
Identification
Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.
Predicate Devices
Biotrace® Bone Cement Radio-Opacifier (Bryan Corporation) (K002063)
Submission Summary (Full Text)
{0}------------------------------------------------
AUG - 8 2003
#### 5 510(k) Summary
This summary of the 510(k) premarket notification for the Vista Scientific Barium Sulfate is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Manufacturer
Vista Scientific, LLC 7960 Niwot Road, Unit B-7 Niwot, CO 80503 Telephone: (303) 652-1501 Registration #: FDA forms 2891 and 2892 submitted. FDA letter acknowledging receipt of forms received 6 May 2003.
### Contact Person
Jonathan Modine General Manager
# Date Prepared
August 1, 2003
# Device Classification Name
Accessory, Barium Sulfate, Methyl Methacrylate For Cranioplasty, Product Code MYU (Class II)
# Trade Name
Vista Scientific AxioView Barium Sulfate
# Generic/Common Name
Barium Sulfate USP
# Predicate Device
Biotrace® Bone Cement Radio-Opacifier (Bryan Corporation) (K002063)
# Action taken to comply with Section 514 of the Act
No applicable mandatory performance standards or special controls exist for this device.
### Intended Use
Vista Scientific Barium Sulfate" is intended for use as an additive to Codman Cranioplastic™ (Type 1-Slow set) to provide radiopacity for imaging purposes.
{1}------------------------------------------------
# Product Description
Vista Scientific Barium Sulfate is a dry powder supplied sterile and non-pyrogenic for use as a radiopacifying additive to Codman Cranioplastic™ (Type 1-Slow set) bone cement. The Barium Sulfate is supplied in a quantity of 6 grams, packaged in a closed 120 mL polymer container, and sealed within a double sterile peel pouch package configuration.
# Substantial Equivalence
Vista Scientific Barium Sulfate is intended for use as an additive to Codman Cranioplastic™ (Type 1-Slow set) to provide radiopacity for imaging purposes.
It is substantially equivalent to other devices currently on the market for use as a radiopacifying additive for imaging purposes when used as an additive to Codman Cranioplastic™ (Type 1-Slow set).
Vista Scientific Barium Sulfate is equivalent to Biotrace® Bone Cement Radio-Opacifier (Bryan Corporation, K002063). Vista Scientific Barium Sulfate is substantially equivalent to the predicate device with regards to device design, intended use, patient population and anatomical site. Any differences in technological characteristics between the Vista Scientific Barium Sulfate and the predicate device do not raise any new issues of safety or effectiveness.
# Testing in Support of Substantial Equivalence
Performance, biocompatibility and microbiological testing have been conducted, and the results of the testing verify that the Vista Scientific Barium Sulfate performs as designed and is suitable for its intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 8 2003
Mr. Jonathan Modine General Manager Vista Scientific, LLC 7960 Niwot Road Unit B-7 Niwot, CO 80503
Re: K024359
Trade Name: Vista Scientific Barium Sulfate TM Regulation Number: 882.5300 Regulation Name: Methyl methacrylate for cranioplasty Regulatory Class: Class II Product Code: MYU Dated: June 3, 2003 Received: June 25, 2003
Dear Mr. Modine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 – Mr. Jonathan Modine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
fal Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Statement of Indications for Use 4
## INDICATIONS FOR USE
510(k) Number (if known):
Ko24359
Device Name: Vista Scientific Barium Sulfate™
Indications for Use:
Vista Scientific Barium Sulfate™ is intended for use as an additive to Codman Cranioplastic™ (Type 1-Slow set) to provide radiopacity for imaging purposes.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The Counter Use (Optional Format 1-2-96)
Miriam C. Provost
Division Sign-Off Division of General, Restorative and Neurological Devices
510(k) Number K624359