K021594 · Center For Orthotic and Prosthetic Care,Llc · MVA · Nov 13, 2003 · Neurology
Device Facts
Record ID
K021594
Device Name
COPC BAND
Applicant
Center For Orthotic and Prosthetic Care,Llc
Product Code
MVA · Neurology
Decision Date
Nov 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5970
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The COPC Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial symmetry and/or shape in infants initiating treatment from three to 14 months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephalic, brachycephalic, and scaphocephalic patterned head shapes.
Device Story
COPC Band is a custom-fabricated cranial orthosis for infants with positional plagiocephaly. Device input consists of physical measurements and a plaster mold of the infant's head. Fabrication involves plaster modification techniques to create a custom shell. The orthosis features a thermoformable plastic outer shell, hypoallergenic foam lining, and securing straps. It functions by applying static pressure to prominent cranial regions while providing voids for flattened areas to guide skull growth. Used in clinical settings under practitioner supervision; prescribed by physicians. The device includes a natural corrugated stop in the clamshell design to prevent overtightening. Clinical practitioners monitor fit and alignment. The device benefits patients by passively correcting cranial asymmetry through controlled growth guidance.
Clinical Evidence
No clinical trials were performed. Safety and effectiveness were established through bench testing, including biocompatibility assessments of materials and pressure analysis. Pressure sensor testing on a cranial model demonstrated that the device maintains contact with cranial prominences while avoiding pressure on cranial depressions, confirming performance analogous to the predicate device.
Technological Characteristics
Custom-fabricated cranial orthosis. Materials: thermoformable modified polyethylene outer shell, hypoallergenic foam (aliplast) lining. Energy source: none (passive). Form factor: custom-molded clamshell with two-strap closure system. Includes integrated corrugated stop for pressure regulation. Sterilization: not specified.
Indications for Use
Indicated for infants aged 3 to 14 months with nonsynostotic positional plagiocephaly, including plagiocephalic, brachycephalic, and scaphocephalic head shapes, to improve cranial symmetry and shape.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{1}------------------------------------------------
# NOV 1 3 2003
The following summary is provided in pursuant to Section 513(I)(3)(A) of the Federal Food. Drug, and Cosmetic Act.
#### Applicant Information A.
- Submitter: Center for Orthotic and Prosthetic Care of KY, LLC. 1010 Clarks Lane, Louisville, KY 40217. Phone: (812) 941-0966, Fax: (812) 941-0958, Email: copc@bluegrass.net
- Contact: Keith Senn, Bob Havens, Center for Orthotic and Prosthetic Care of K Y, LLC 1010 Clarks Lane, Louisville, KY 40217. Email: copc(@bluegrass.net, Phone: (812) 941-0966
- . Summary Date: October 15, 2003
#### Device Name and Classification (K021594) B.
- Proprietary Name: COPC Band
- Common Name: Cranial Orthosis ●
- Classification Name: Cranial Orthosis ●
- Predicate Device: DOC™ Band, Cranial Orthosis, K964992, classified ● under 21 CFR § 882.5970
. . . : "
#### C. Device Description
The COPC Band is a cranial orthosis used to treat abnormally shaped craniums in infants three to 14 months of age. This condition is clinically known as positional or Deformational Plagiocephaly. The orthosis contains the protruding aspects of the cranium in a static equilibrium while guiding the growth of the flattened areas of the skull into the created spaces. The COPC Band is only available through a prescription issued by a physician.
The orthosis is custom designed for each patient from a modified model and measurements of the infant's head. The mold is prepared for fabrication using advanced plaster modification techniques. Each orthosis is composed of an outer shell of thermoformable plastic, an inner lining of hypoallergenic foam, and a strap for securing Optimum fit and alignment is insured and monitored by the clinical the orthosis. practitioner.
{2}------------------------------------------------
#### Intended Use D.
The COPC Band is intended for medical purposes to passively hold prominent cranial regions of an infant's skull in order to improve cranial symmetry and/or shape in infants initiating treatment from three to 14 months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephalic, brachycephalic, and scaphocephalic patterned head shapes.
#### Comparison to Predicate Device E.
The COPC Band and the predicate device are very similar with respect to production, instructions for use, materials, safety and effectiveness, and special controls. The COPC Band combines an outer shell of thermo-formable polymer and hypoallergenic foam lining in an analogous fashion as the predicate device fits the child in a similar fashion as the predicate device, with trim lines terminating over the same points. The primary difference between the proposed and predicate devices is the side closure strap of the orthosis, where the COPC Band incorporates two straps and the DOC Band utilizes a single strap. The COPC Band utilizes modified polyethylene and aliplast as the composite materials, and the DOC Band utilizes copolymer and aliplast as its composite materials. MSDS sheets for these materials confirm that they are safe, non-hazardous, and non-reactive with human skin under normal contact conditions and environments. The material is handled in an identical manner to the polymer used in the predicate device, incorporating all of the safety and standards of practice. The proposed indications of use are analogous to those presented by the predicate device, and biocompatibility, function, and effectiveness further parallel those of the predicate device.
#### F. Performance Data
The safety of the COPC Band is established under standard biocompatibility assessments for each material used. These assessments reveal that the device and the materials used are not expected to adversely affect the infants under the intended conditions of wear, which are analogous to those for the DOC Band. The materials are not reported to cause skin irritation or any toxic effects. Further, the product is designed to avoid improper migration or harmful levels of pressure. The interior of the device is smooth and poses no significant threat to the child during application within the normal scope of its intended use. The COPC Band has incorporated within its clamshell design a natural corrugated stop, which prevents overtightening and harmful pressure upon the infant's skull. This safety feature heightens the COPC Band's safety above that of the predicate device in that the predicate device has no such inherent safeguard. The COPC Band's performance has been demonstrated and compared to the predicate device through an analysis of pressure exerted by the device upon the child's skull. Pressure sensors applied to a cranial model with a COPC Band applied, demonstrates quantifiably that intimate contact exists between the regions of cranial prominences, while no pressure exists in the regions of the cranial depressions as they are positioned over the corrective voids. This data and device performance is directly analogous with the regions of contact between the predicate device and the patient using the device.
{3}------------------------------------------------
#### G. Summary
The safety and effectiveness data submitted to the FDA establishes that the COPC Band is safe and effective for its intended use and is substantially equivalent to applicable predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 3 2003
Center for Orthotic and Prosthetic Care, LLC C/O Joseph Terpenning, CO 2525 North Tenth St., #804 Arlington, Virginia 22201
Re: K021594
Trade/Device Name: COPC Band Regulation Number: 21 CFR 882.5970 Regulation Name: Cranial Orthosis Regulatory Class: Class II Product Code: MV A Dated: undated Received: October 31, 2003
Dear Mr. Terpenning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Joseph Terpenning, CO
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark M Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known);
Device Name: COPC Band
Indications for Use:
## Indications for Use Statement:
The COPC Band is intended for medical purposes to passively hold prominent cranial regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants initiating treatment from three to 14 months of age, with nonsynostotic positional plagiocephaly, including infants with plagiocephalic, brachycephalic, and scaphocephalic patterned head shapes.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) (Optional Format 1-2-96)
OR
Over the Counter Use
R. Mark A. Milken
(Division Sign-Off) Division of General. Restorative and Neurological Devices
510(k) Number K021544
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.